Reference Manual Q3 Dosing Module
English
Referenzhandbuch Q3-Dosiermodul
Deutsch
Manual de referencia Módulo de dosificación Q3
Español
Manuel de référence Module de dosage Q3
Français
考手 Q3加样模块
简体中文
リファレンスマニュアル Q3分注モジュール
日本語
Table of Contents
1 Introduction 3
1.1 Further documents and information................................................................................ 3
1.2 Explanation of conventions and symbols used................................................................ 3
1.3 Acronyms and Abbreviations......................................................................................... 4
1.4 Compliance information................................................................................................ 4
2 Safety Information 6
2.1 Definitions of signal words and warning symbols............................................................ 6
2.2 Product-specific safety notes......................................................................................... 6
3 Design and Function 10
3.1 Function description..................................................................................................... 10
3.2 Overview..................................................................................................................... 10
3.3 Components description ............................................................................................... 11
3.4 Dosing heads and equipment........................................................................................ 13
3.5 Dosing head RFID tag .................................................................................................. 14
4 Installation and Putting into Operation 15
4.1 Selecting the location ................................................................................................... 15
4.2 Scope of delivery ......................................................................................................... 15
4.3 Setting up the powder dosing configuration..................................................................... 16
4.4 Setting up the liquid dosing configuration ....................................................................... 16
4.4.1 Overview...................................................................................................... 16
4.4.2 Interface of the pump..................................................................................... 16
4.4.3 Wiring the pump........................................................................................... 17
4.4.4 Connecting the tubes ..................................................................................... 17
5 Operation 22
5.1 Installing and removing the dosing head ........................................................................ 22
5.2 Installing the sealing insert............................................................................................ 23
5.3 Selecting and using an ErgoClip .................................................................................... 23
5.4 Powder dosing ............................................................................................................ 23
5.4.1 Using dosing heads without a vial .................................................................. 23
5.4.2 Filling the vial of a powder dosing head........................................................... 24
5.4.3 Storing powder dosing heads ......................................................................... 24
5.4.4 Avoiding electrostatic charges......................................................................... 24
5.5 Liquid dosing .............................................................................................................. 25
5.5.1 Attaching the dosing head to the bottle cap ...................................................... 25
5.5.2 Handling the bottle ........................................................................................ 26
5.5.3 Using the QL3 pump ..................................................................................... 27
5.6 Using other weighing methods ...................................................................................... 28
6 Maintenance 29
6.1 Cleaning..................................................................................................................... 29
6.1.1 Cleaning the housing..................................................................................... 29
6.1.2 Cleaning the sealing insert ............................................................................. 29
6.1.3 Cleaning the weighing chamber...................................................................... 30
6.1.4 Cleaning powder dosing heads....................................................................... 30
6.1.5 Cleaning the liquid dosing elements ................................................................ 30
6.2 Replacing the sealing ring and fastening nut on the liquid tube ......................................... 31
7 Troubleshooting 32
7.1 Error symptoms........................................................................................................... 32
8 Technical Data 34
8.1 General data ............................................................................................................... 34
Table of Contents 1Q3 Dosing Module
en
8.2 Model-specific data...................................................................................................... 35
8.3 Material specifications.................................................................................................. 36
8.3.1 Glossary of material acronyms ....................................................................... 36
8.3.2 Powder dosing head QH008 .......................................................................... 36
8.3.3 Powder dosing head QH012 .......................................................................... 37
8.3.4 Powder dosing head QH002-CNMW ............................................................... 38
8.3.5 Powder dosing head QH012-LNJW ................................................................ 39
8.3.6 Liquid dosing head QL001............................................................................. 40
8.3.7 Liquid dosing head QL003............................................................................. 41
8.3.8 QLL standard/advanced kit for bottles .............................................................. 41
8.4 Dimensions................................................................................................................. 42
9 Disposal 44
10 Accessories and Spare Parts 45
10.1 Accessories................................................................................................................. 45
10.2 Spare parts ................................................................................................................. 52
10.2.1 Q3 dosing module ........................................................................................ 52
10.2.2 QL3 pump.................................................................................................... 52
10.2.3 QLL spare parts kit for bottles ......................................................................... 53
Index 55
Table of Contents2 Q3 Dosing Module
Introduction 3Q3 Dosing Module
1 Introduction
Thank you for choosing a METTLER TOLEDO instrument. The instrument combines high performance with ease
of use.
1.1 Further documents and information
u www.mt.com/XPR-automatic
This document is available in other languages online.
u www.mt.com/Q3-RM
Instructions for cleaning a balance: "8 Steps to a Clean Balance"
u www.mt.com/lab-cleaning-guide
Search for software downloads u www.mt.com/labweighing-software-download
Search for documents u www.mt.com/library
For further questions, please contact your authorized METTLER TOLEDO dealer or service representative.
u www.mt.com/contact
1.2 Explanation of conventions and symbols used
Conventions and symbols
Key and/or button designations and display texts are shown in graphic or bold text, e.g., , Edit.
Note
For useful information about the product.
Refers to an external document.
Elements of instructions
In this manual, step-by-step instructions are presented as follows. The action steps are numbered and can
contain prerequisites, intermediate results and results, as shown in the example. Sequences with less than two
steps are not numbered.
Prerequisites that must be fulfilled before the individual steps can be executed.
1 Step 1
Intermediate result
2 Step 2
Result
en
Introduction4 Q3 Dosing Module
1.3 Acronyms and Abbreviations
Original term Explanation
EMC Electromagnetic Compatibility
FCC Federal Communications Commission
LPS Limited Power Source
POM Polyoxymethylene
RFID Radio-frequency identification
RM Reference Manual
sd Standard deviation
SELV Safety Extra Low Voltage
SOP Standard Operating Procedure
UM User Manual
USB Universal Serial Bus
1.4 Compliance information
National approval documents, e.g., the FCC Supplier Declaration of Conformity, are available online and/or
included in the packaging.
u http://www.mt.com/ComplianceSearch
Contact METTLER TOLEDO for questions about the country-specific compliance of your instrument.
u www.mt.com/contact
United States of America
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to
part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful inter-
ference when the equipment is operated in a commercial environment. This equipment generates, uses, and
can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications. Operation of this equipment in a residential area is
likely to cause harmful interference in which case the user will be required to correct the interference at his own
expense.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
1. This device may not cause harmful interference.
2. This device must accept any interference received, including interference that may cause undesired
operation.
Changes or modifications not expressly approved by the party responsible for compliance could void the user's
authority to operate the equipment.
Canada
This device contains licence-exempt transmitter(s)/receiver(s) that comply with Innovation, Science and
Economic Development Canada’s licence-exempt RSS(s). Operation is subject to the following two conditions:
1. This device may not cause interference.
2. This device must accept any interference, including interference that may cause undesired operation of the
device.
Singapore
Complies with
IMDA Standards
DA108390
Introduction 5Q3 Dosing Module
South Africa
TA 2019-2613
APPROVED
Thailand
en
Safety Information6 Q3 Dosing Module
2 Safety Information
Two documents named "User Manual" and "Reference Manual" are available for this instrument.
The User Manual is printed and delivered with the instrument.
The electronic Reference Manual contains a full description of the instrument and its use.
Keep both documents for future reference.
Include both documents if you transfer the instrument to other parties.
Only use the instrument according to the User Manual and the Reference Manual. If you do not use the
instrument according to these documents or if the instrument is modified, the safety of the instrument may be
impaired and Mettler-Toledo GmbH assumes no liability.
2.1 Definitions of signal words and warning symbols
Safety notes contain important information on safety issues. Ignoring the safety notes may lead to personal
injury, damage to the instrument, malfunctions and false results. Safety notes are marked with the following
signal words and warning symbols:
Signal words
DANGER
A hazardous situation with high risk, resulting in death or severe injury if not avoided.
WARNING
A hazardous situation with medium risk, possibly resulting in death or severe injury if
not avoided.
CAUTION
A hazardous situation with low risk, resulting in minor or moderate injury if not avoided.
NOTICE
A hazardous situation with low risk, resulting in damage to the instrument, other
material damage, malfunctions and erroneous results, or loss of data.
Warning symbols
General hazard Notice
2.2 Product-specific safety notes
Intended use
This dosing system is designed to be used in analytical laboratories by trained staff. The dosing system is
intended for weighing and dosing powder or liquid samples.
Any other type of use and operation beyond the limits of use stated by Mettler-Toledo GmbH without consent
from Mettler-Toledo GmbH is considered as not intended.
Responsibilities of the instrument owner
The instrument owner is the person holding the legal title to the instrument and who uses the instrument or
authorizes any person to use it, or the person who is deemed by law to be the operator of the instrument. The
instrument owner is responsible for the safety of all users of the instrument and third parties.
Mettler-Toledo GmbH assumes that the instrument owner trains users to safely use the instrument in their
workplace and deal with potential hazards. Mettler-Toledo GmbH assumes that the instrument owner provides
the necessary protective gear.
Protective equipment
Chemical-resistant gloves Goggles
Safety Information 7Q3 Dosing Module
Lab coat
Safety notes
WARNING
Death or serious injury due to electric shock
Contact with parts that carry a live current can lead to death or injury.
1 Only use the METTLER TOLEDO power cable and AC/DC adapter designed for your
instrument.
2 Connect the power cable to a grounded power outlet.
3 Keep all electrical cables and connections away from liquids and moisture.
4 Check the cables and the power plug for damage and replace them if damaged.
WARNING
Injury and/or damage due to hazardous substances
Chemical, biological or radioactive hazards can be associated with the substances
processed by the instrument. During dosing procedures, small amounts of the dosed
substance may become airborne and penetrate the instrument or contaminate its
surroundings.
The substance characteristics and related hazards is the full responsibility of the instrument
owner.
1 Be aware of possible hazards associated with the substance and take adequate safety
measures, e.g., those stated on the safety data sheet provided by the manufacturer.
2 Make sure that every instrument part in contact with the substance will not get altered or
damaged by the substance.
WARNING
Injury or damage due to powder handling
Powders may get compacted in the dosing head and block it. Too much force applied by
the dosing head mechanism may lead to a break and potentially hazardous substances
may get airborne.
1 Handle the dosing heads with care.
2 If the dosing head seems blocked, avoid further dosing. Remove the head from the
instrument and turn it upside down to loosen the powder.
3 Cease work immediately in the event of leak or break.
WARNING
Injury and/or damage due to reacting, flammable, or explosive substances
During the dosing procedure, substances could be combined and cause an exothermic
reaction or explosion. This includes powders, liquids, and gases.
The sample characteristics and related hazards is the full responsibility of the instrument
owner.
1 Be aware of possible hazards associated with reacting, flammable, or explosive
substances.
2 Ensure a working temperature low enough to prevent the formation of flames or an
explosion.
en
Safety Information8 Q3 Dosing Module
WARNING
Injury or death due to toxic, explosive, or flammable substances
If you use toxic, explosive, or flammable liquids with the pump, the exhaust air will be
contaminated.
Connect a tube to the exhaust air outlet to collect the contaminated air.
WARNING
Injury and/or damage due to reacting substances
When pressure is released from the bottle, the air/gas in the bottle moves back towards the
pump. The air/gas coming from the coupled outlets mixes in the pump. Molecules of the
substances in the various bottles can get in contact through this contaminated air/gas.
1 Do not connect bottles with incompatible liquids to the same pump simultaneously.
2 Before connecting a second, incompatible liquid to the pump, disconnect the first bottle
and purge the pump cavity with clean air/gas.
WARNING
Injury and damage to the pump or the bottle due to high pressure
High pressure from external gas can damage the pump or the bottle.
1 Use a regulator on the external gas line.
2 Ensure that the pressure of the external gas does not exceed 0.5bar (7.2psi).
CAUTION
Injury due to splashing liquids
If the pressure in the bottle is not released, liquid might splash when removing the micro
dosing valve, opening the bottle, or removing the liquid tube.
Always release pressure before removing the micro dosing valve, opening the bottle, or
removing the liquid tube.
CAUTION
Injury due to leaking liquids
Wrongly cut tubing can result in leaking connections.
Cut the tubes with a tube cutter or a sharp knife.
CAUTION
Injury due to moving parts
Do not reach into the working area while parts of the instrument are moving.
CAUTION
Injury due to sharp objects or broken glass
Instrument components, e.g., glass, can break and lead to injuries.
Always proceed with focus and care.
Safety Information 9Q3 Dosing Module
NOTICE
Damage to the instrument or malfunction due to the use of unsuitable parts
Only use parts from METTLER TOLEDO that are intended to be used with your
instrument.
NOTICE
Damage to the instrument
The instrument contains no user-serviceable parts.
1 Do not open the instrument.
2 In the event of problems, please contact a METTLER TOLEDO representative.
NOTICE
Damage to the instrument due to inappropriate cleaning methods
If liquid enters the housing, it can damage the instrument. The surface of the instrument can
be damaged by certain cleaning agents, solvents, or abrasives.
1 Do not spray or pour liquid on the instrument.
2 Only use the cleaning agents specified in the Reference Manual (RM) of the instrument
or the guide "8 Steps to a Clean Balance".
3 Only use a lightly moistened, lint-free cloth or a tissue to clean the instrument.
4 Wipe off any spills immediately.
en
Design and Function10 Q3 Dosing Module
3 Design and Function
3.1 Function description
The Q3 dosing module can be added to any XPR analytical balance to perform automatic dosing of powders
and/or liquids. The powder or liquid dosing head is attached to the dosing module and stays outside of the
weighing chamber during the whole dosing procedure. The top door of the balance is replaced by a top panel
with an opening through which the powder or liquid is dosed. The top panel moves down with the dosing
module and the dosing head until the distance to the sample vessel is appropriate for dosing. The weighing
chamber is therefore closed during the whole dosing procedure, providing the best possible dosing perfor-
mances.
The Q3 dosing module is delivered with three ErgoClips: the ErgoClip adapter holder, the ErgoClip adapter
stabilizer, and the ErgoClip height extension. The different designs can accommodate vessels of different
sizes and shapes. All three ErgoClips are compatible with XPR analytical balances with tall draft shield.
When liquid is dosed, the QL3 pump and QLL kit for bottle are used together with the Q3 dosing module. The
QL3 pump is used to build pressure in the bottle. As soon as the pressure is sufficiently high, the micro
dispensing valve in the liquid dosing head opens and liquid ascends the liquid tube. Several pumps can be
connected to the balance simultaneously and each pump can accommodate up to three bottles.
3.2 Overview
1
2
3
4
5
6
7
8
9
8
1
Dosing lift
6
Powder dosing head
2
Dosing module
7
Sealing insert
3
Top panel
8
Dosing head holder
4
Vial adapter
9
Dosing head release button
5
ErgoClip adapter holder
Design and Function 11Q3 Dosing Module
3.3 Components description
Dosing module
The dosing module is designed to hold the dosing head
and can automatically move up and down during the
dosing procedure.
Dosing lift
The dosing lift is fixed to the back of the weighing
chamber. It contains the mechanism that moves the
dosing module up and down during the dosing procedure
and is controlled through the terminal.
Top panel
The Q3 dosing module includes a top panel especially
designed to minimize the influence of the environment on
the dosing process. An opening in the top panel allows the
tip of the dosing head to penetrate the weighing chamber
while dosing. The top panel is attached to the bottom part
of the dosing module, moving up and down together with
it.
While the dosing module is at its uppermost position,
other weighing methods and accessories can be used
normally.
Dosing head holder
The dosing head holder consists of four pins that hold the
dosing head in place. Once installed, the dosing head is
secured and can only be released by pressing one of the
release buttons on either side of the holder, while the
dosing module is at its uppermost position. The holder
includes an RFID reader to identify the attached dosing
head.
en
Design and Function12 Q3 Dosing Module
Optical sensor for HeightDetect
The dosing module is equipped with an optical sensor to
detect the height of the sample vessel opening (Height-
Detect). This allows automatic vertical positioning of the
dosing module for fast and safe dosing.
ErgoClip adapter holder and adapters
The Q3 dosing module includes the ErgoClip adapter
holder, which can be installed on the SmartGrid weighing
pan. It can hold adapters of various sizes allowing perfect
positioning of different sample vessels. Additional adapters
are available as accessories. See [Accessories
Page45].
The Q3 dosing module is delivered with a variety of
ErgoClips for various usage. See [Selecting and using an
ErgoClip
Page23]
ErgoClip adapter stabilizer
The Q3 dosing module includes the ErgoClip adapter
stabilizer, which can be installed on the SmartGrid
weighing pan. The ErgoClip adapter stabilizer is especially
designed to provide more stability to the adapter. It can
therefore be used for tall and narrow vessels.
ErgoClip height extension
The Q3 dosing module includes the ErgoClip height
extension, which can be installed on the SmartGrid
weighing pan. This ErgoClip has a flat surface and can
therefore accommodate any sample vessel. It is especially
useful for short sample vessels, such as weighing boats.
Design and Function 13Q3 Dosing Module
3.4 Dosing heads and equipment
Powder dosing head
Powder dosing heads are used for automated powder
dosing. If a small quantity of powder is required, the
dosing head can be used without a vial attached to it. In
this case, the opening can then be closed using the cap
included in the delivery. The dosing heads can be stored in
appropriate storage containers. Vials and storage
containers need to be purchased separately.
If a printer is connected, you can use a pre-defined
template to print a label with the dosing-head data from
the RFID tag and affix this label to the dosing head.
Powder test head
The powder test head is used to test a powder dosing
system. The dosing head is filled with calcium carbonate
(CaCO
3
) and uses a predefined test program to check the
dosing functionality of the system. The powder test head is
programmed to perform 15 tests. Each test weighs 10
doses of a pre-defined amount of powder. If your instrument fails the tests, inform a METTLER TOLEDO service
technician.
Weight test head
The weight test head performs an automated repeatability
test by placing a small test weight on given tare weights.
The repeatability of the system is determined based on 10
consecutive measurements of the test weight. If your
instrument fails the tests, inform a METTLER TOLEDO
service technician.
Liquid dosing head
Liquid dosing heads are used for automated liquid dosing.
They are used together with a pump and bottle. The liquid
dosing head contains an RFID tag which stores infor-
mation about the dosing head and the used substance.
en
Design and Function14 Q3 Dosing Module
3.5 Dosing head RFID tag
1
1
Each powder and liquid dosing head is equipped with an
integrated RFID tag (1) that stores and exchanges data
with the instrument.
Various data is stored in the RFID tag of the dosing heads,
such as the substance name, lot ID, filling date, expiry
date, etc. It also includes customized data fields.
This data can be edited on the terminal and should be set
before using a new dosing head such that the data is
available for reports and labels.
For powder dosing heads, the counter for the remaining
powder dosing cycles is based on the Dose limit of the
dosing head, which can be defined by the customer. With
every dosing cycle started, the counter is decreased by 1.
If the counter drops to zero, consider replacing the dosing head. If the vial of the old dosing head still contains
a considerable amount of powder, you may remove the vial from the old dosing head and screw it onto the
new dosing head. Copy the user data and the powder content value from the old dosing head to the new
dosing head.
The quantity of powder added to the dosing head can be stored in the RFID tag during the filling procedure. The
RFID information is updated after each dosing operation such that the powder quantity in the dosing head is up
to date. If the remaining quantity of powder is insufficient for the next dosing cycle, a warning message will
appear on the terminal.
For further information, consult the Reference Manual (RM) of your XPR balance.
u www.mt.com/XPR-analytical-RM
Installation and Putting into Operation 15Q3 Dosing Module
4 Installation and Putting into Operation
This instrument must be installed by a METTLER TOLEDO service technician.
The Q3 dosing module is compatible with all XPR analytical balances.
4.1 Selecting the location
A balance is a sensitive precision instrument. The location where it is placed will have a profound effect on the
accuracy of the weighing results.
Requirements of the location
Place indoors on stable
table
Ensure sufficient spacing Level the instrument Provide adequate lighting
Avoid direct sunlight Avoid vibrations Avoid strong drafts Avoid temperature fluctu-
ations
Sufficient spacing for balances: > 15 cm all around the instrument
Take into account the environmental conditions. See "Technical Data".
Note
The powder delivery mechanism of the Q3 dosing module might cause the instrument to vibrate. Do not place
on the same working surface as other instruments that are sensitive to vibrations.
4.2 Scope of delivery
Q3 dosing module
Dosing module
Dosing lift
Dosing lift back cover with screws
Top panel dosing module
Sealing inserts, 5 pcs
Powder dosing head
ErgoClip adapter holder
ErgoClip adapter stabilizer
ErgoClip height extension
Various vial adapters, 4 pcs
Declaration of Conformity
User Manual
Recommended options
EasyHub USB Integrable antistatic kit
The scopes of delivery for the QL3 pump and the QLL kits for bottles are in the documentation delivered with
these products.
en
Installation and Putting into Operation16 Q3 Dosing Module
4.3 Setting up the powder dosing configuration
After the Q3 dosing module has been installed on your XPR analytical balance by a METTLER TOLEDO service
technician, simply install a powder dosing head and start a Automated dosing or Automated solution prep.
method.
4.4 Setting up the liquid dosing configuration
4.4.1 Overview
1
2
3
4
5
6
7
8
9
10
11
1
Dosing lift
7
ErgoClip adapter holder
2
Dosing module
8
QL3 pump (available separately)
3
Liquid dosing head
9
QLL kit for bottles (available separately)
4
Sealing insert
10
Air tube (included in QLL kit for bottles)
5
Top panel
11
Liquid tube (included in QLL kit for bottles)
6
Vial adapter
4.4.2 Interface of the pump
Rear
Front
11
22
3
3
4
4
5
5
6
6
1
Air outlets (to bottle)
4
Socket for AC/DC adapter
2
Pressure release button and indicator light
5
USB-B port (to host)
3
Air exhaust outlet
6
Air inlet
/