DOC023.97.80614
CL17sc
08/2019, Edition 1
User Manual
Manuel de l'utilisateur
Manual del usuario
Manual do Usuário
用户手册
取扱説明書
사용
설명서
คู่มือผู้ใช้
English..............................................................................................................................3
Français......................................................................................................................... 27
Español.......................................................................................................................... 52
Português...................................................................................................................... 77
中文............................................................................................................................... 101
日本語........................................................................................................................... 124
한글............................................................................................................................... 148
ไทย.................................................................................................................................. 172
2
Table of contents
Specifications on page 3 Operation on page 18
Product overview on page 6 Calibration and adjustment on page 18
Product components on page 7 Maintenance on page 18
Installation on page 7 Troubleshooting on page 24
User navigation on page 17 Replacement parts and accessories on page 25
Configuration on page 17
Specifications
Specifications are subject to change without notice.
Table 1 General specifications
Specification Details
Dimensions (W x H x D) 32.9 x 34.2 x 17.7 cm (12.9 x 13.5 x 7.0 in.)
Enclosure IP66 per IEC 60529 with the door closed and latched
1
Shipping weight 4.1 kg (9 lb) without bottles; 5.1 kg (11.2 lb) with full bottles
Mounting Wall mount
Protection class III
Pollution degree 3
Electrical installation category I (indoor installation)
Power requirements 12 VDC, 400 mA maximum (supplied by the controller)
Operating temperature 5 to 40 °C (41 to 104 °F)
Operating humidity 0 to 90% non-condensing relative humidity
Storage temperature -40 to 60 °C (-40 to 140 °F)
Fittings Sample line: ¼-in. OD quick-connect fitting for plastic tubing
Drain lines: slip-on fitting for ½-in. ID soft plastic tubing
Indicator lights Analyzer status and measurement cycle
Certifications CE DoC, FCC SDoC, IC, RCM, KC
Warranty 1 year (EU: 2 years)
Table 2 Sample requirements
Specification Details
Pressure 0.31 to 5.17 bar (4.5 to 75 psig) supplied to Y-strainer; 0.1 to 0.34 bar (1.5 to 5 psig) supplied to
analyzer
Flow rate 60 to 200 mL/min through the instrument
Temperature 5 to 40 °C (41 to 104 °F)
Filtration Y-strainer with 40 mesh screen or higher
1
Drain holes closed with plugs for testing requirements.
English 3
Table 3 Measurement specifications
Specification Details
Light source LED, measurement at 510 nm; 1 cm light pathlength
Measurement range 0.04–10 mg/L free or total residual chlorine as Cl
2
Measurement interval 150 seconds
Event log and data log 2 weeks capacity minimum, all measurements
Service log 24 hours capacity minimum
Accuracy ±5% or ±0.04 mg/L from 0 to 5 mg/L (the larger value) as Cl
2
±10% from 5 to 10 mg/L as Cl
2
Precision ±5% or ±0.01 mg/L (the larger value) as Cl
2
Limit of detection 0.04 mg/L as Cl
2
Limit of quantitation 0.09 mg/L
Calibration Factory calibration
Optional: 2-point user calibration with calibration standards in Calibration Verification Kit
Reagent usage 0.5 L of buffer solution and 0.5 L of indicator solution in 31 days
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious
injury.
C A U T I O N
Indicates a potentially hazardous situation that may result in minor or moderate injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special
emphasis.
4 English
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument
could occur if not observed. A symbol on the instrument is referenced in the manual with a
precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public
disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to
the user.
Certification
C A U T I O N
This equipment is not intended for use in residential environments and may not provide adequate protection to
radio reception in such environments.
Canadian Radio Interference-Causing Equipment Regulation, IECS-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing
Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne
sur les équipements provoquant des interférences.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC
Rules. Operation is subject to the following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including interference that may cause
undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for
compliance could void the user's authority to operate the equipment. This equipment has been tested
and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference, in which case the user will be required to
correct the interference at their expense. The following techniques can be used to reduce
interference problems:
1. Disconnect the equipment from its power source to verify that it is or is not the source of the
interference.
2. If the equipment is connected to the same outlet as the device experiencing interference, connect
the equipment to a different outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the interference.
5. Try combinations of the above.
English
5
Product overview
D A N G E R
Chemical or biological hazards. If this instrument is used to monitor a treatment process and/or dialysis
feed water for which there are regulatory limits and monitoring requirements related to public health,
public safety, food or beverage manufacture or processing, it is the responsibility of the user of this
instrument to know and abide by any applicable regulation and to have sufficient and appropriate
mechanisms in place for compliance with applicable regulations in the event of malfunction of the
instrument.
The CL17sc analyzer measures the free chlorine or total chlorine concentration in water in the range
of 0.04 to 10 mg/L at 150 second intervals. Figure 1 gives an overview of the analyzer.
Connect the analyzer to an SC Controller for power, operation, data collection, data transmission and
diagnostics. Refer to the SC Controller manual for an overview of the controller.
Note: More than one analyzer can be connected to an SC Controller if the controller has more than one quick-
connect fitting.
Figure 1 CL17sc
1 Indicator bottle 5 Analyzer status light (Table 5 on page 7)
2 Measurement cycle lights (Table 4 on page 7) 6 Flow meter
3 Pump clamp 7 Buffer bottle
4 Tubing clip 8 Colorimetric cell
Flow meter
The analyzer has a flow meter that measures the sample flow through the analyzer. Refer to Figure 1
on page 6.
Sample flows through the analyzer only when the analyzer flushes the cell. At other times, there is no
sample flow and the flow rate shown is zero. Push the RIGHT arrow on the controller to show the
flow rate.
6
English
Measurement cycle lights
The measurement cycle lights show the step of the measurement cycle being done. Refer to Table 4.
Table 4 Measurement cycle lights
Lights on
Description The cell is flushed with water. The buffer solution and the indicator
solution are added to the cell.
The sample is measured.
Analyzer status light
The analyzer status light changes from green to yellow when there is a warning or to red when there
is an error. Refer to Table 5.
Table 5 Analyzer status light
Color Description
Green Normal operation
Yellow The system needs attention to prevent a failure in the future. Measurements continue.
To show the warnings, push menu and select DIAGNOSTICS > [select analyzer] > WARNING LIST.
Refer to Warnings—Yellow light on page 24.
Red The system needs immediate attention. Measurements have stopped.
To show the errors, push menu and select DIAGNOSTICS > [select analyzer] > ERROR LIST. Refer to
Errors—Red light on page 24.
Product components
Make sure that all components have been received. Refer to Figure 2. If any items are missing or
damaged, contact the manufacturer or a sales representative immediately.
Figure 2 Product components
1 CL17sc analyzer 3 Tubing Kit (tubing harness and stir bar)
2 Installation kit
Installation
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
English 7
Installation guidelines
The analyzer is supplied with a standpipe installation kit or a pressure regulator installation kit to
control the water pressure to the analyzer. The user selects the installation kit at the time of
purchase.
Install the analyzer indoors in an environment with minimum vibration. Refer to the operating
temperature and humidity specifications in Specifications on page 3.
Do not install the analyzer in direct sunlight or near a heat source.
For standpipe installations, install the analyzer in a location with 61 cm (2 ft) of free space above
the analyzer for the standpipe mounting.
Install the analyzer near an open drain. Refer to local regulatory agency instructions for disposal.
Step 1: Air purge (optional)
Air purge may be necessary if the analyzer is installed in a location with high humidity and/or fumes
that cause corrosion. The air purge keeps positive pressure in the instrument with dry and clean air.
Supply clean, dry instrument-quality air at 0.003 m
3
/minute (0.1 ft
3
/minute) at 20 psig maximum.
Refer to the illustrated steps that follow. The
3
/
8
-inch quick-connect fitting and tubing is supplied by
the user.
8 English
Step 2: Mount and plumb the analyzer
C A U T I O N
Only the tubing specified by Hach Company may be used with the instrument. Other tubing may
absorb chlorine that is in the sample water and have a negative effect on chlorine readings (decrease
the chlorine readings).
1. Attach the analyzer (item 9) to a wall with four screws. Make sure that the analyzer is level.
Note: The user supplies the mounting hardware.
2. Attach the SC Controller (item 7) to a wall, panel or pipe. Refer to the SC Controller
documentation for instructions.
3. Assemble the Y-strainer. Refer to the instructions on the packing list.
4. Use the 1-inch conduit clamp to mount the Y-strainer assembly (item 5 in Figure 3 or Figure 4).
5. Do step 6 or 7.
The analyzer is supplied with a standpipe installation kit (refer to Figure 3) or a pressure regulator
installation kit (refer to Figure 4) to control the water pressure to the analyzer. The user selects
the installation kit at the time of purchase.
6. If a standpipe installation kit was selected at the time of purchase, install the standpipe as follows:
a. Assemble the standpipe. Refer to the instructions on the packing list.
b. Use the smaller conduit clamp to mount the standpipe assembly a minimum of 61 cm (24 in.)
above the top of the analyzer. Refer to item 10 in Figure 3.
c. With the black tubing, install the sample bypass tubing (item 8). Push the tubing behind and
then into the standpipe. Put the tubing 10–13 cm (4–5 in.) in the top of the standpipe.
d. With the clear tubing, plumb the bottom of the standpipe to an open drain (item 11). Refer to
Drain line guidelines on page 12.
7. If a pressure regulator installation kit was selected at the time of purchase, install the pressure
regulator (item 10 in Figure 4).
8. With the black tubing, connect the Y-strainer inlet (item 5) to the sample source.
Refer to Sample line guidelines on page 12 and Sample requirements on page 12.
9. With the black tubing, connect the Y-strainer outlet (or the outlet of the pressure regulator) to the
sample inlet fitting (item 1) on the bottom of the analyzer. Make the tubing as short as possible
and a maximum of 2.5 m (8 ft).
10. If a pressure regulator was installed, install the sample bypass tubing (item 8 in Figure 4).
11. Assemble and plumb the grab sample assembly (item 4) as follows:
a. Assemble the grab sample assembly.
b. Plumb the tee fitting between the Y-strainer and the analyzer. Refer to Figure 3 or Figure 4.
12. With the clear tubing, connect the sample drain (item 2) and case drain (item 3) to an open drain.
Refer to Drain line guidelines on page 12.
English
9
Figure 3 Installation with a standpipe
1 Sample inlet tubing 7 SC Controller
2 Sample drain tubing 8 Sample bypass tubing
3 Case drain tubing (optional) 9 Analyzer
4 Grab sample assembly 10 Standpipe
5 Y-strainer assembly 11 Standpipe drain tubing
6 Shut-off valve at Y-strainer inlet
10 English
Figure 4 Installation with a pressure regulator
1 Sample inlet tubing 6 Shut-off valve at Y-strainer bypass
2 Sample drain tubing 7 SC Controller
3 Case drain tubing (optional) 8 Sample bypass tubing
4 Grab sample assembly 9 Analyzer
5 Y-strainer assembly 10 Pressure regulator (not adjustable)
English 11
Sample line guidelines
Select a good, representative sampling point for the best instrument performance. The sample must
be representative of the entire system.
To prevent erratic readings:
Collect samples from locations that are sufficiently distant from points of chemical additions to the
process stream.
Make sure that the samples are sufficiently mixed.
Make sure that all chemical reactions are complete.
Install the analyzer as near the sample source as possible (4.6 m (15 ft) maximum).
Sample requirements
The water from the sample source(s) must agree with the specifications in Specifications on page 3.
Keep the sample flow rate and operating temperature as constant as possible for best performance.
Drain line guidelines
N O T I C E
Incorrect installation of the drain lines can cause liquid to go back into the instrument and cause damage.
Make the drain lines as short as possible.
Make sure that the drain lines have a constant slope down.
Make sure that the drain lines do not have sharp bends and are not pinched.
Make sure that the drain lines are open to air and are at zero back pressure.
12 English
Step 3: Install the stir bar and tubing harness
Do the illustrated steps that follow.
Step 4: Install the reagent bottles
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
Items to collect:
Personal protective equipment (refer to MSDS/SDS)
Indicator bottle
Buffer bottle
DPD compound bottle
1. Put on the personal protective equipment identified in the safety data sheets (MSDS/SDS).
2. Install the buffer bottle on the right side and the indicator bottle on the left side. Refer to the
illustrated steps that follow.
Note: (Optional) Use the top of the analyzer as a shelf.
English
13
14 English
Step 5: Set the sample flow to on
1. Slowly turn the shut-off valve(s) to the open position ¼ turn.
2. Make sure that there are no leaks at the plumbing connections. If there is a leak, push the tube
farther into the fitting or tighten the connection with a wrench.
3. For standpipe installations, do the steps that follow:
a. Fully open the shut-off valve at the Y-strainer outlet.
b. Turn the shut-off valve at the Y-strainer inlet (item 6 in Figure 3 on page 10) until a small
stream of water comes out of the standpipe drain tubing.
4. For pressure regulator installations, do the steps that follow:
a. Fully open the shut-off valves at the Y-strainer inlet and outlet.
b. Turn the shut-off valve at the Y-strainer bypass (item 6 in Figure 4 on page 11) until a small
stream of water comes out of the Y-strainer bypass tubing.
Step 6: Electrical installation
Connect the analyzer to the controller
W A R N I N G
Electrical shock hazard. Externally connected equipment must have an applicable country safety
standard assessment.
Connect the analyzer cable to the quick-connect fitting of the SC controller. Refer to Figure 5.
Keep the connector cap to seal the connector opening in case the cable must be removed.
Note: Extension cables are available. Refer to Replacement parts and accessories on page 25. The maximum
cable length is 15 m (49 ft).
English
15
Figure 5 Connect the cable to a quick-connect fitting
Connect the controller to power
Connect the controller to line power by hard-wiring in conduit or wiring to a power cord. Refer to the
controller documentation for instructions.
Connect external devices to the controller
Connect the controller relays, analog outputs and digital inputs to external devices as necessary.
Refer to the controller documentation for instructions.
Step 7: Set up and prime the analyzer
1. If the controller does not recognize the analyzer is connected, push menu, then select
TEST/MAINT > SCAN SENSORS.
2. Push menu, then select SENSOR SETUP > [select analyzer] > PRIME.
Step 8: Set the flow rate
1. Push the RIGHT arrow on the controller to show the flow rate. Sample flows through the analyzer
only when the analyzer flushes the cell. At other times, there is no sample flow and the flow rate
shown is zero. Refer to Table 4 on page 7 for descriptions of the measurement cycle steps.
2. Use the shut-off valves to set the flow rate between 60 and 200 mL/min when the analyzer
flushes the cell with water.
Step 9: Install the latest software
Make sure that the SC Controller has the latest software installed. Use an SD card to install the latest
software on the SC Controller.
1. Go to the product page for the applicable SC Controller on http://hach.com.
2. Click the "Downloads" tab.
3. Scroll down to "Software/Firmware".
4. Click the link for the SC Controller software.
5. Save the files to an SD card.
6. Install the files on the SC Controller. Refer to the software installation instructions supplied with
the software files.
16
English
Configuration
Step 10: Configure the analyzer
Set the analyzer name, signal average, type of chlorine measured, bubble reject and chlorine alarm
setpoints.
1. Push menu, then select SENSOR SETUP > [select analyzer] > CONFIGURE.
2. Select and configure each option.
Option Description
EDIT NAME Sets the name of the analyzer. The name of the analyzer shows on the controller
display and in the log files.
SIGNAL AVERAGE Sets the number of measurements used to calculate the average measurement
shown on the display. Options: 1 (default), 2, 3 or IRREGULAR VALUE.
When SIGNAL AVERAGE is set to 1, signal averaging is disabled.
When SIGNAL AVERAGE is set to 2 or 3, an average reading shows on the display.
For example, the measurement on the display is equal to the last and previous
measurement divided by two when SIGNAL AVERAGE is set to 2.
When SIGNAL AVERAGE is set to IRREGULAR VALUE, the analyzer rejects a
reading that is unusually higher or lower than the latest readings. When a reading is
rejected, the last good reading shows on the display and is saved to the data log. No
more than three consecutive readings are rejected.
Signal averaging corrects for reading fluctuations that are caused by bubbles and/or
larger particles in the sample.
MEASUREMENT Sets the type of chlorine measured. If the buffer and indicator bottles have "Free
Chlorine" on their labels, select FREE CHLORINE (default).
If the buffer and indicator bottles have "Total Chlorine" on their labels, select TOTAL
CHLORINE.
BUBBLE REJECT Sets the bubble reject to on or off (default). Set BUBBLE REJECT to ON to decrease
noise caused by bubbles in the sample. Air bubbles in the sample can cause the
readings to not be stable.
HIGH CL ALARMS Sets the chlorine concentration setpoint for the high chlorine alarm—0.00 to
10.00 mg/L (default: 4.00 mg/L).
LOW CL ALARMS Sets the chlorine concentration setpoint for the low chlorine alarm—0.00 to
10.00 mg/L (default: 0.20 mg/L).
SENSOR INFO Shows the analyzer serial number, software version, boot version and driver version.
DEFAULT SETTINGS Select YES to change the configuration settings back to the factory default values.
SERVICE For service use only
System configuration
Refer to the controller documentation for system configuration, general controller settings and setup
for outputs and communications.
User navigation
Refer to the controller documentation for keypad description and navigation information.
Push the RIGHT arrow key on the controller multiple times to show more information on the home
screen and to show a graphical display.
English
17
Operation
Data, event and service logs
The controller provides a data log, event log and service log for each connected instrument. The
CL17sc measurements are saved to the data log at 150 second intervals. The event log shows the
events that have occurred. The event log and data log capacity are 2 weeks minimum. The service
log capacity is 24 hours minimum.
Refer to the controller documentation to download the data log, event log and/or service log. The
data log is an XML file that can be changed to CSV or Excel format. The event log and service log
are files in CSV format.
Measure a GRAB SAMPLE IN
Use the GRAB SAMPLE IN menu to add a water sample of known chlorine concentration to the cell
for measurement. When necessary, measure a grab sample for verification or to measure a water
sample collected from another location.
1. Push menu, then select SENSOR SETUP > [select analyzer] > GRAB SAMPLE IN.
2. Complete the steps that show on the display. When prompted, remove the cell lid and add 50 mL
of the grab sample to the cell. Some of the sample will flow to the sample drain.
Modbus registers
A list of Modbus registers is available for network communication. Refer to the manufacturer's
website for more information.
Calibration and adjustment
The calibration curve of the analyzer is set at the factory for performance to specifications.
Calibration can be done for performance verification.
No user adjustments to the factory calibration curve are recommended unless required by a
regulatory agency for compliance reporting purposes, or a large repair of the analyzer is done.
For more information, refer to the instructions in the Calibration Verification Kit or contact technical
support.
Maintenance
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
N O T I C E
Do not disassemble the instrument for maintenance. If the internal components must be cleaned or repaired,
contact the manufacturer.
18 English
Maintenance schedule
Table 6 shows the recommended schedule of maintenance tasks. Facility requirements and
operating conditions may increase the frequency of some tasks.
Table 6 Maintenance schedule
Task 1 month 6 months As necessary
Clean the cell on page 19 X
2
Replace the reagent bottles on page 20 X
Replace the stir bar and tubing harness
3
X
Clean the screen in the Y-strainer on page 21 X
Clean the cell
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
Clean the cell at 1-month intervals or more frequently if necessary.
Items to collect:
Cotton swabs
4
Sulfuric Acid Solution
4
1. Put on the personal protective equipment identified in the safety data sheets (MSDS/SDS).
2. Push menu, then select SENSOR SETUP > [select analyzer] > TASKS > CLEAN CELL.
3. Push enter to stop measurements.
4. Select an option.
Option Description
HOLD The controller outputs are held at the last measured value.
TRANSFER The controller outputs change to the transfer value.
5. When the status light flashes, do the illustrated steps that follow. Make sure that there is no
unwanted material in the cell. When done, push enter.
2
Clean the cell more or less frequently as necessary.
3
Refer to the instructions supplied with the Tubing Kit.
4
Refer to Replacement parts and accessories on page 25.
English 19
6. When "TASK COMPLETE" shows on the display, push enter.
The analyzer starts a measurement cycle in approximately 30 seconds.
Replace the reagent bottles
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
Replace the reagents bottles at 1-month intervals.
1. Put on the personal protective equipment identified in the safety data sheets (MSDS/SDS).
2. Push menu, then select SENSOR SETUP > [select analyzer] > TASKS > CHANGE REAGENTS.
3. Push enter to stop measurements.
4. Select an option.
Option Description
HOLD The controller outputs are held at the last measured value.
TRANSFER The controller outputs change to the transfer value.
5. Wait for the status light to flash.
6. Replace the buffer bottle as follows:
Note: (Optional) Use the top of the analyzer as a shelf.
a. Remove the cap and seal from the new buffer bottle.
b. Remove the used buffer bottle from the analyzer.
c. Put the buffer tubing in the new buffer bottle on the right side of the analyzer. Tighten the cap.
7. Replace the indicator bottle as follows:
a. Remove the cap and seal from the indicator bottle and the brown DPD bottle.
b. Fill the brown DPD bottle approximately ¼ full with indicator solution.
20
English
/