Medtronic Guardian 4 ユーザーガイド

タイプ
ユーザーガイド
Sensor UserGuide
Guía del usuario del sensor
头用户指
- 1 -
English
Introduction
The Guardian 4 sensor is part of the
Continuous Glucose Monitoring (CGM)
system. The sensor converts small
amounts of glucose from the interstitial
fluid under the skin into an electronic
signal. The system then uses these
signals to provide sensor glucose
values.
Sensor assembly
Indications for use
The Guardian 4 sensor (MMT-7040) is intended to monitor glucose
levels for the management of diabetes for persons ages seven
years and older. It is indicated for use as an adjunctive device to
complement, not replace, information obtained from the standard
blood glucose monitoring devices. The sensor is intended for
single use and requires a prescription. The Guardian 4 sensor is
indicated for up to seven days of continuous use. Refer to the
system user guide for treatment decisions.
Contraindications
Refer to the system guide for contraindications associated with
Guardian 4 sensor use.
User safety
Warnings
Read this entire user guide before attempting to insert the
Guardian 4 sensor. The one-press serter (MMT-7512) is the only
serter approved for use with the sensor. Failure to follow directions,
or the use of a different insertion device, may result in improper
insertion, pain, or injury.
Do not attempt to connect a transmitter or recorder that is not
compatible with the sensor. The sensor is designed to work with
approved transmitters only. Connecting the sensor to a transmitter
or recorder that is not approved for use with the sensor may
damage the components. Refer to the system user guide for a list
of compatible products.
Do not use continuous glucose monitoring if hydroxyurea, also
known as hydroxycarbamide, is taken. Hydroxyurea is used to
treat certain diseases, such as cancer and sickle cell anemia.
Hydroxyurea use results in higher sensor glucose readings
compared to blood glucose readings. Taking hydroxyurea while
using continuous glucose monitoring can result in substantially
- 2 -
higher sensor glucose readings in reports than actual blood
glucose readings. For pump users, taking hydroxyurea while using
continuous glucose monitoring can result in hypoglycemia caused
by over-delivery of insulin.
Always check the label of any medication being taken to confirm
if hydroxyurea or hydroxycarbamide is an active ingredient. If
hydroxyurea is taken, consult a healthcare professional. Use
additional blood glucose meter readings to verify glucose levels.
Taking medications that contain acetaminophen or paracetamol,
including, but not limited to fever reducers and cold medicine, while
wearing the sensor, may falsely raise sensor glucose readings.
The level of inaccuracy depends on the amount of acetaminophen
or paracetamol active in the body and may be different for each
person. Always check the label of any medications to confirm
whether acetaminophen or paracetamol is an active ingredient.
Do not expose the sensor to MRI equipment, diathermy devices,
or other devices that generate strong magnetic fields. The
performance of the sensor has not been evaluated under those
conditions and may be unsafe. If the sensor is exposed to a strong
magnetic field, discontinue use and contact 24-hour Technical
support for further assistance.
Always inspect the packaging for damage prior to use. Sensors are
sterile and non-pyrogenic, unless the package has been opened or
damaged. If the sensor packaging is open or damaged, discard the
sensor directly into a sharps container. Use of a non-sterile sensor
may result in infection at the insertion site.
Do not allow children to put small parts in their mouth. This product
poses a choking hazard for young children.
Do not use the Guardian 4 sensor if you are pregnant or critically
ill. Since the sensor has not been studied in these populations,
the impact of medications common to these conditions on sensor
performance is unknown and the sensor may be inaccurate in
these populations.
Healthcare professionals and caregivers:
Always wear gloves to insert the sensor. A retractable needle is
attached to the sensor. Minimal bleeding may occur.
Cover the sensor with sterile gauze to remove the needle
housing from the sensor.
Place the needle housing directly into a sharps container after
sensor insertion to prevent accidental needlestick injury.
Watch for bleeding at the insertion site (under, around, or on top of
the sensor).
- 3 -
If bleeding occurs, do the following:
1. Apply steady pressure, using sterile gauze or a clean cloth
placed on top of the sensor, for up to three minutes. The use of
unsterile gauze may cause site infection.
2. If bleeding stops, connect the transmitter (or recorder) to the
sensor.
If bleeding does not stop, do not connect the transmitter to the
sensor because blood may get into the transmitter connector,
and may damage the device.
If bleeding continues, causes excessive
pain or discomfort, or is significantly visible
in the plastic base of the sensor, do the
following:
Plastic base
1. Remove the sensor and continue to apply
steady pressure until the bleeding stops.
Discard the sensor in a sharps container.
2. Check the site for redness, bleeding, irritation, pain,
tenderness, or inflammation. Treat based on instructions from
a healthcare professional.
3. Insert a new sensor in a different location.
For questions or concerns related to sensor use, contact
24-Hour Technical Support for assistance.
For medical questions or concerns, contact a healthcare
professional.
Precautions
Wash hands with soap and water before inserting the Guardian 4
sensor to help prevent site infection.
Do not insert the sensor through tape. Inserting the sensor through
tape may cause improper sensor insertion and function.
Only use alcohol to prepare the insertion site. Using alcohol to
prepare the insertion site makes sure that residue is not left on the
skin.
Rotate the sensor insertion site so that sites do not become
overused.
Do not clean, resterilize, or try to extract the needle from the
needle housing. An accidental needlestick or puncture may occur.
Do not reuse sensors. Reuse of a sensor may cause damage to
the sensor surface and lead to inaccurate glucose values, site
irritation, or infection.
- 4 -
Risks and side effects
Risks related to sensor use include
Skin irritation or other reactions
Bruising
Discomfort
Redness
Bleeding
Pain
Rash
Infection
Raised bump
Appearance of a small “freckle-like” dot where the sensor was
inserted
Allergic reaction
Fainting secondary to anxiety or fear of needle insertion
Soreness or tenderness
Swelling at insertion site
Sensor fracture, breakage or damage
Minimal blood splatter associated with sensor needle removal
Residual redness associated with adhesive or tapes or both
Scarring
Reagents
The Guardian 4 sensor contains two biological reagents: glucose
oxidase, and human serum albumin (HSA). Glucose oxidase is
derived from Aspergillus niger and manufactured to meet industry
requirements for the extraction and purification of enzymes for use
in diagnostic, immunodiagnostic, and biotechnical applications.
The HSA used on the sensor consists of purified and dried albumin
fraction V derived from pasteurized human serum, which is cross-
linked via glutaraldehyde. Approximately 3 μg of glucose oxidase
and approximately 10 μg of HSA are used to manufacture each
sensor. HSA is approved for IV infusion in humans at quantities
much larger than in the sensor.
Removing the sensor
To change the Guardian 4 sensor, disconnect the transmitter from
the sensor as described in the Guardian 4 transmitter user guide.
Gently pull the sensor from the body to remove it. Discard the
sensor in a sharps container.
- 5 -
Components
BA
One-press serter
A. bump on both buttons
B. thumbprint marking
C
D
A
B
Glucose sensor assembly
A. pedestal
B. needle housing
C. sensor
D. clear liner
A
B
Sensor base
A. sensor connector
B. sensor snaps
Transmitter
BB
A
First piece of oval tape
A. liner 1
B. liner 2
BB
A
Second piece of oval tape
A. liner 1
B. liner 2
- 6 -
Where to insert the sensor
Choose an insertion site for the applicable age group and make
sure that the site has an adequate amount of subcutaneous fat.
Ages 7-17 years
Back of the upper arm
Caution: Guardian 4 sensor is indicated for arm use only.
Do not use Guardian 4 sensor in other sites, including the
abdomen or buttocks, due to a difference in performance
that could result in hypoglycemia or hyperglycemia
Ages 18 years and older
Back of the upper arm
Note:
Assistance will likely be needed for sensor insertion into the
back of the upper arm. Some users find it difficult to insert the
sensor into their arm by themselves.
For best sensor glucose performance, and to prevent accidental
sensor removal:
Do not insert the sensor into muscle, tough skin, or scar tissue.
Avoid areas that are constrained by clothing or accessories.
Avoid areas subjected to vigorous movement during exercise.
- 7 -
Inserting the sensor
WARNING: Always wear gloves when inserting the sensor into
another person to avoid contact with patient blood. Minimal
bleeding may occur. Contact with patient blood may cause
infection.
11. Wash hands thoroughly with soap
and water.
22. Choose an insertion site that has
a sufficient amount of fat.
33. Clean the insertion site with
alcohol. Let the area air dry.
44. Open the sensor package.
- 8 -
55. Hold the pedestal and remove the
glucose sensor assembly from the
package. Place the pedestal on a
clean, flat surface such as a table.
6
Correct
tucked tab
Incorrect
6. Confirm that the
adhesive tab of the
sensor is tucked
under the sensor
connector and
sensor snaps.
7
thumb on thumbprint marking
Correct Incorrect
7. Using either hand,
place a thumb on
the thumbprint
marking to hold the
serter. Fingers must
not touch the serter
buttons.
88. Push the serter
down onto the
pedestal until the
base of the serter
sits flat on the table
and there is a click.
- 9 -
99. With either hand,
place two fingers
on the base of the
pedestal. With the
other hand, grip the
serter and pull the
serter upwards.
arrow
Note: The arrow on the side of the
serter aligns with the needle inside
the serter.
WARNING: Never point a loaded serter toward any body part
where insertion is not desired. An accidental button-push may
cause the needle to inject the sensor in an undesired location,
causing minor injury.
10 10. Place the serter on top
of the prepared insertion
site.
11 11. Press and release both
serter buttons at the
same time. Continue to
hold the serter on top
of the insertion site for
five seconds or more to
let the adhesive stick to
the skin.
- 10 -
12 12. Lift the serter away
from the insertion site.
Fingers must not press
the buttons while lifting
the serter.
B
A
AC
D
E
Sensor base
A. sensor snaps
B. sensor connector
C. adhesive tab
D. adhesive liner
E. adhesive pad
If the sensor is inserted without assistance, complete step 13a. If a
healthcare professional or caregiver assisted with sensor insertion,
complete step 13b.
13a Patient:
13. a. Hold the sensor base
against the skin at the sensor
connector and at the opposite
end of the sensor base. Hold
the needle housing at the
top and pull away from the
sensor.
OR
13b
Healthcare professional or
caregiver:
13. b. Wrap sterile gauze around
the sensor. Hold the sensor
base against the skin at the
sensor connector and the
opposite end of the sensor
base. Hold the needle
housing at the top and pull
away from the sensor.
- 11 -
WARNING: Always watch for bleeding at the insertion site. If
bleeding occurs under, around, or on top of the sensor, apply
steady pressure using sterile gauze or a clean cloth placed
on top of the sensor for up to three minutes. The use of
unsterile gauze may cause an infection. If bleeding does not
stop, remove the sensor and apply steady pressure until the
bleeding stops.
Note: After insertion, use of adhesive products, such as
Skin Tac, in addition to the oval tape is optional. If optional
adhesive products are used, apply them to the skin under the
adhesive pad prior to removing the liner. Adhesive products
may also be applied to the adhesive pad or the skin around the
sensor base. Allow the product to dry before continuing.
14 14. Remove the adhesive liner from under
the adhesive pad. Pull the liner away
from the sensor, staying close to the
skin. Do not pull on the sensor when
you remove the liner.
Note: Do not remove the adhesive
liner from the rectangular adhesive
tab. This tab will be used to secure
the transmitter in a later step.
Note: If the sensor base moves,
hold the sensor base down.
15 15. Firmly press the adhesive pad against
the insertion site to confirm that the
sensor base remains on the skin.
- 12 -
16 16. Untuck the adhesive tape from under
the sensor connector.
17 17. Straighten the sensor adhesive tab so
that it lies flat against the skin.
Applying oval tape
11. Remove the liner marked 1.
22. Apply the tape as shown and
press down firmly.
wide part of tape covers half
of sensor base
- 13 -
33. Remove liners marked 2 from
each side.
44. Smooth the tape.
55. Connect the transmitter to the
sensor.
Note: Wait for the green light on
the transmitter to flash. If the green
light does not flash, refer to the
Troubleshooting section of the
Guardian 4 transmitter user guide.
66. Cover the transmitter with the
adhesive tab.
Note: Do not pull the tab too tightly.
- 14 -
77. To apply a second tape, remove
liner marked 1.
88. Apply the second tape in
the opposite direction to the
first tape and place it on the
transmitter. Press down firmly.
99. Remove liners marked 2 from
each side.
10 10. Smooth the tape.
Note: For details on how to enter sensor settings into a
compatible display device, refer to the system user guide.
wide part of tape covers end
of transmitter and skin
- 15 -
Maintenance
Storage
CAUTION: Do not freeze the sensor, or store it in direct
sunlight, extreme temperatures, or humidity. These conditions
may damage the sensor.
Only store sensors at room temperature between 36 °F to 80 °F
(2 °C to 27 °C). Discard sensor after the “Use-by date” indicated
on the label, if the package is damaged, or if the seal is broken.
Disposal
Dispose of the Guardian 4 sensor into a sharps container.
Assistance
Department Telephone Number
24-Hour Technical Support
(calls within the United States) +1 800 646 4633
24-Hour Technical Support
(calls outside the United States) +1 818 576 5555
Website www.medtronicdiabetes.com
Technical specifications
Approximate dimensions
1.50 x 2.60 x 2.00 inches (3.80 x 6.70 x 5.20 centimeters)
Approximate weight
0.09 ounces (2.80 grams)
Sensor life of use
The Guardian 4 sensor can be used one time and has a maximum
life of up to 170 hours (seven days). The 170-hour life span of the
sensor begins when the sensor is connected to the transmitter.
Icon table
Use-by date
Medical device
Do not re-use
Caution: consult instructions for use for important
warnings or precautions not found on the label.
(1x)
One sensor per container/package
(5x)
Five sensors per container/package
- 16 -
(2x)
Two tapes per package
(10x)
Ten tapes per package
Consult instructions for use
Catalogue or model number
Batch code
Sterilized using irradiation
Do not use if package is damaged
Single sterile barrier system
XX°C
XX°F
XX°C
XX°F
Storage temperature limit
Open here
Manufacturer
Date of manufacture
Do not resterilize
Fragile, handle with care
Keep dry
Recyclable, contains recycled content.
Authorized representative in the European Community.
Conformité Européenne (European Conformity). This
symbol means that the device fully complies with
applicable European Union Acts.
Magnetic Resonance (MR) unsafe
Non-pyrogenic
Requires prescription in the USA
Icon glossary
For definitions of the symbols on the device and package labels,
see www.medtronicdiabetes.com/symbols-glossary.
©2023 Medtronic. All rights reserved. Medtronic, Medtronic logo, and Engineering the extraordinary are
trademarks of Medtronic. ™* Third-party brands are trademarks of their respective owners. All other
brands are trademarks of a Medtronic company.
- 17 -
Español
Introducción
El sensor Guardian 4 forma parte del
sistema de monitorización continua de
glucosa (MCG). El sensor convierte
pequeñas cantidades de glucosa del
líquido intersticial que se encuentra
bajo la piel en una señal eléctrica. A
continuación, el sistema utiliza estas
señales para proporcionar valores de
glucosa del sensor. Conjunto de sensor
Indicaciones de uso
El sensor Guardian 4 (MMT-7040) está indicado para controlar los
niveles de glucosa en el tratamiento de la diabetes en personas
mayores de siete años. Está indicado para utilizarse como
dispositivo adicional para complementar la información obtenida
mediante los dispositivos de monitorización de glucosa en sangre
habituales, y no para sustituirla. El sensor está diseñado para un
solo uso y requiere prescripción facultativa. El sensor Guardian 4
está diseñado para utilizarse durante siete días como máximo de
manera continuada. Consulte la guía del usuario del sistema para
tomar decisiones acerca de la terapia.
Contraindicaciones
Consulte la guía del sistema para conocer las contraindicaciones
asociadas al uso del sensor Guardian 4.
Seguridad del usuario
Advertencias
Lea esta guía del usuario en su totalidad antes de intentar insertar
el sensor Guardian 4. El dispositivo de inserción One-press
(MMT- 7512) es el único dispositivo de inserción aprobado para
utilizar con el sensor. Si no se siguen las indicaciones o se utiliza
un dispositivo de inserción diferente, pueden producirse una
inserción incorrecta, dolor o lesiones.
No intente conectar un transmisor o registrador que no sea
compatible con el sensor. El sensor se ha diseñado para funcionar
únicamente con los transmisores aprobados. Si conecta el sensor
a un transmisor o registrador que no esté aprobado para utilizarse
con él, pueden producirse daños en los componentes. Consulte
la guía del usuario del sistema para ver una lista de productos
compatibles.
No utilice la monitorización continua de glucosa si toma
hidroxiurea, también conocida como hidroxicarbamida. La
hidroxiurea se utiliza en el tratamiento de determinadas
- 18 -
enfermedades, como el cáncer y la anemia de células falciformes.
La toma de hidroxiurea da lugar a lecturas de glucosa del
sensor más altas que las lecturas de glucosa en sangre. Tomar
hidroxiurea mientras se utiliza la monitorización continua
de glucosa puede provocar lecturas de glucosa del sensor
sustancialmente más altas que las lecturas de glucosa en sangre
reales. Para los usuarios de la bomba, tomar hidroxiurea durante
el uso de la monitorización continua de glucosa puede provocar
hipoglucemia debido a una infusión excesiva de insulina.
Compruebe siempre el prospecto de los medicamentos que
toma para verificar si contienen hidroxiurea o hidroxicarbamida
como principio activo. Si está tomando hidroxiurea, consulte a un
profesional sanitario. Utilice lecturas adicionales del medidor de
glucosa en sangre para comprobar los niveles de glucosa.
Tomar medicamentos que contengan paracetamol (acetaminofén),
tales como antipiréticos y medicamentos contra el resfriado,
durante el uso del sensor puede elevar erróneamente las lecturas
de glucosa del sensor. El nivel de inexactitud depende de la
cantidad de paracetamol (acetaminofén) que esté activa en el
cuerpo y puede variar de una persona a otra. Compruebe siempre
el prospecto de los medicamentos para verificar si contienen
paracetamol (acetaminofén) como principio activo.
No exponga el sensor a equipos de RM, dispositivos de diatermia
u otros dispositivos que generen campos magnéticos potentes. No
se ha evaluado el funcionamiento del sensor en estas condiciones
y puede no ser seguro. Si el sensor se expone a un campo
magnético intenso, deje de utilizarlo y póngase en contacto con la
línea de asistencia técnica 24 horas para obtener ayuda.
Inspeccione siempre el embalaje por si tuviese daños antes de
utilizar este dispositivo. Los sensores son estériles y apirógenos,
a menos que el embalaje esté abierto o dañado. Si el embalaje
del sensor presenta daños o está abierto, deseche el sensor
directamente en un contenedor para objetos cortantes. El uso de
un sensor no estéril puede producir una infección en la zona de
inserción.
No permita que los niños se lleven a la boca piezas pequeñas. El
producto conlleva un peligro de asfixia para los niños pequeños.
No utilice el sensor Guardian 4 si tiene una enfermedad grave
o está embarazada. Puesto que el sensor no se ha estudiado
en estas poblaciones, se desconoce el efecto que pueden
tener los medicamentos habituales para estas situaciones en el
funcionamiento del sensor, que puede no ser preciso en estas
poblaciones.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56

Medtronic Guardian 4 ユーザーガイド

タイプ
ユーザーガイド