Vaisala HMD65 ユーザーマニュアル

タイプ
ユーザーマニュアル
EN
DE
FR
ES
PT
RU
JA
ZH
M212264EN-B
Quick Guide
HMD60 Series Humidity and Temperature
Transmitters for Ducts in HVAC
HMD65
PUBLISHED BY
Vaisala Oyj
Vanha Nurmijärventie 21, FI-01670 Vantaa, Finland
P.O. Box 26, FI-00421 Helsinki, Finland
+358 9 8949 1
Visit our Internet pages at www.vaisala.com.
No part of this document may be
reproduced, published or publicly
displayed in any form or by any means,
electronic or mechanical (including
photocopying), nor may its contents be
modified, translated, adapted, sold or
disclosed to a third party without prior
written permission of the copyright holder.
Translated documents and translated
portions of multilingual documents are
based on the original English versions. In
ambiguous cases, the English versions are
applicable, not the translations.
The contents of this document are subject
to change without prior notice.
Local rules and regulations may vary and
they shall take precedence over the
information contained in this document.
Vaisala makes no representations on this
document’s compliance with the local
rules and regulations applicable at any
given time, and hereby disclaims any and
all responsibilities related thereto.
This document does not create any legally
binding obligations for Vaisala towards
customers or end users. All legally binding
obligations and agreements are included
exclusively in the applicable supply
contract or the General Conditions of Sale
and General Conditions of Service of
Vaisala.
Table of Contents
English.............................................................................................................................................5
Deutsch........................................................................................................................................ 23
Français.........................................................................................................................................41
Español.........................................................................................................................................59
Português.................................................................................................................................... 77
Русский........................................................................................................................................95
日本語.............................................................................................................117
中文................................................................................................................ 135
3
4 M212264EN-B
Introduction to HMD60 Series
The duct mounted HMD60 HUMICAPâ Humidity and Temperature Transmitters are designed
for monitoring humidity and temperature in demanding HVAC and light industrial applications.
HMD60 series transmitters provide stable, reliable, and highly accurate (up to ±1.5 %RH and
±0.1 °C (0.18 °F)) measurements, and are resistant to chemicals and dust.
HMD60 series transmitter options include the HMD62 and TMD62 analog output transmitters
with loop powered 4 … 20 mA current output, and the analog and digital output transmitter
HMD65 with analog voltage output (0 … 10 V) and digital Modbus RTU and Bacnet output
(RS-485).
Thanks to easy access to electronics also when the transmitter is installed to a duct,
configuration and adjustment can be carried out quickly and conveniently. Available
configuration and adjustment interface options range from physical trimmers and DIP switches
on the transmitter's circuit board to Modbus, BACnet, and Vaisala Insight PC software for
Windowsâ.
HMD65 Basic Features and Options
Humidity and temperature measurement:
available humidity parameters: RH, T
d
, T
df
, A, X, T
w
, H
T measurement in °C or °F
Analog output: 2 analog 0 … 10 V output channels for humidity and temperature
measurements
Digital output (RS-485): Modbus RTU and BACnet MS/TP
Power supply input: 15 … 35 VDC / 16 … 24 VAC
Configuration and adjustment options:
RH and T measurement field adjustment with trimmers
Humidity output parameter selection and Modbus/BACnet serial setting configuration
with DIP switches
Configuration and adjustment with Vaisala Insight PC software
Configuration with Modbus and BACnet
Field adjustment with MI70 hand-held indicator
Output Parameter Scaling
Default temperature analog output scale: -20 … +80 °C (-4 … +176 °F)
Default scaling for humidity parameters: see Table 2 (page 15).
To change the default scaling of an analog output parameter, use Vaisala Insight PC
software. See the instructions in HMD65 User Guide.
More Information
For more detailed instructions for installing, configuring, and maintaining the HMD60 series
transmitters, see HMD62 and TMD62 User Guide in English M212016EN and HMD65 User Guide
in English M212243EN available at www.vaisala.com/HMD60.
5
ENGLISH
Transmitter Parts
1
2
3
4
5
6
7
8
9
Figure 1 HMD65 Transmitter Parts Overview
1
Captive screw (2 pcs, cross-head) for attaching the lid of the transmitter.
2 Screw (2 pcs) for mounting the transmitter on the installation surface.
3 Transmitter lid. Open the captive screws of the lid to access input and output electronics.
4 Transmitter base. Contains the input and output connectors on the transmitter board: see
Transmitter Board (page 10).
5 Cable gland (M16 x 1.5 lead-through) for leading wires into the transmitter. See HMD65
User Guide for cable gland and conduit options.
6 Alternative lead-through (M20 x 1.5) for wiring.
7 Probe body. Long (shown) and short probe options available: see Transmitter Dimensions
(page 7).
8 Probe filter (default option: AISI 316L stainless steel). See HMD65 User Guide for filter
options.
9 HUMICAPâ sensor inside the probe filter.
Do not touch the sensor element.CAUTION!
6 M212264EN-B
Installation
Transmitter Dimensions
The dimensions are given in millimeters and [inches].
299 [11.77]
49 [1.92]
250 [9.84]
101 [3.97]
mm
[in]
101 [3.97]
82[3.23]
82 [3.23]
Ø 12 [0.47]
19.5 [0.77]
149 [5.87]
49 [1.92]
100 [3.94]
101 [3.97]
Ø 12 [0.47]
19.5 [0.77]
Figure 2 Dimensions with Long and Short Probe
7
ENGLISH
Duct Mounting Overview
90°
1
4
3
2
5
6
7
Figure 3 Duct Installation Overview
1
Make sure there is a minimum clearance of 5 m (16.5 ft) between the probe body and any
possible humidifier.
2 When installing the transmitter, drill a second hole approximately 30 cm (12 in) from the
installation hole, towards the direction of the air flow, and plug it with a removable seal.
This second hole is intended for later use in reference measurement with another device
when calibrating or adjusting the transmitter.
3 Check that the duct diameter is suitable for the probe body (see Transmitter Dimensions
(page 7)). Ideally, the sensor (probe head) should be installed in the middle of the duct.
4 Maximum air flow speed: 50 m/s (with sintered filter).
5 Avoid installing the transmitter in dead legs. Supersaturation can occur in areas where
there is no air flow.
6 Do not install the probe in a downward angle. Condensation can travel to the sensor along
the probe body if the probe points down.
7 Install the probe in a 90° angle so that the sensor is placed as close to the middle of the
duct as possible.
Avoid installing in a location where condensation can fall on the
sensor inside the duct.
CAUTION!
8 M212264EN-B
Installing into Duct
Medium size crosshead screwdriver for mounting screws and lid screws.
Small slotted screwdriver for screw terminals.
Drill with 3.5 mm (0.14 in) and 13 … 15 mm (0.51 … 0.59 in) bits for making the
installation holes.
Tools for cutting and stripping wires.
2 x Ø 3.5 [0.14]
Ø 13 ... 15 [0.51 ... 0.59]
Ø 12 [0.47]
82 [3.23]
82 [3.23]
mm
[in]
Ø 3.5 [0.14]
Ø 3.5 [0.14]
3.1 ... 3.4 Nm [2.3 ... 2.5 ft-lbs]
Figure 4 Drilling and Mounting Screws
1. Select an installation location for the transmitter on the duct surface and drill a
Ø 13 … 15 mm (0.51 … 0.59 in) hole for inserting the probe.
2. Push the probe through the hole on the duct until the transmitter body meets the duct.
3. Attach the transmitter body to the duct with 2 Ø 3.5 mm (0.14 in) screws.
Check that the insulation ring sits tightly over the installation hole. If the duct
has a negative pressure, external air can be drawn into the duct and aect
the measurement if the installation hole is not sealed tightly.
4. Optional: Drill a second hole for reference measurements approximately 30 cm (12 in)
from the transmitter installation hole. See Figure 3 (page 8).
5. Open the 2 captive screws on the transmitter body and remove the lid.
6. Attach the input/output wiring to the screw terminals on the transmitter component
board. See Wiring (page 11). Tighten cable glands firmly after wiring.
7. Check that the DIP switches and trimmers are in the correct position. See Transmitter
Board (page 10) for more information on DIP switches and trimmers.
8. Close the transmitter lid and switch on the transmitter's power supply input.
9
ENGLISH
Transmitter Board
Figure 5  HMD65 Transmitter Board: Service Port, DIP switches, Trimmers, and Screw Terminals
1
RS-485 termination (120 Ω resistor) ON/OFF switch.
2 RS-485 (Modbus/BACnet) screw terminals.
3 Power supply input (15 … 35 VDC or 16 … 24 VAC) screw terminals.
4 Service port for MI70 hand-held indicator and Insight PC software cable connection.
5 DIP switches for setting the HMD65 Modbus RTU or BACnet MS/TP MAC address.
6 DIP switches for selecting Modbus/BACnet communication bit rate and parity (Modbus
only).
7 DIP switches for humidity output parameter selection.
8 DIP switch for selecting either Modbus or BACnet mode.
9 Trimmer for humidity measurement adjustment.
10 Screw terminals for humidity measurement output.
11 Screw terminals for temperature measurement output.
12 Trimmer for temperature measurement adjustment.
13 Indicator LEDs: flash when there is RS-485 transmit (TX) or receive (RX) activity.
10 M212264EN-B
Wiring
0 ... 10 V ANALOG OUTPUT WIRING
DIGITAL (RS-485) COMMUNICATION WIRING
CONTROLLER
RS-485 +
RS-485 -
RSGND
HMD65
HMD65
+
-
V
V
POWER SUPPLY
(16 ... 24 VAC)
15 ...35 VDC
1)
1)
1) HMD65 GROUNDING SCREW TERMINALS ARE INTERNALLY
CONNECTED TO HMD65 CHASSIS GROUNDING.
CONNECT GROUNDING AS APPLICABLE
TO INSTALLATION SITE WIRING IMPLEMENTATION.
1)
CONTROLLER
+
-
POWER SUPPLY
(16 ... 24 VAC)
15 ...35 VDC
ANALOG INPUT
ANALOG INPUT GROUND
ANALOG INPUT
ANALOG INPUT GROUND
Figure 6 HMD65 Wiring Diagrams (Analog and Digital Output Options)
Make sure that you prepare or connect only de-energized wires.WARNING!
11
ENGLISH
Inputs and Outputs
Table 1 HMD65 Inputs and Outputs
Property Specification
Analog output
1 x RH output
1)
, 0 … 10 V
1 x T output, 0 … 10 V
Load resistance: 10 kΩ min.
Digital output (RS-485) Isolated, supports Modbus RTU and BACnet
MS/TP protocols
BACnet MS/TP Address range: 0 … 127 (master mode only)
Modbus RTU Address range: 1 … 247
Power supply input
Using a power supply with
overload protection is
recommended for electrical
safety.
15 … 35 VDC
16 … 24 VAC
Power consumption 1.0 W (typical, for both AC and DC)
Service port connector M8 4-pin male connector for MI70 hand-held
indicator (requires cable accessory 219980SP)
or Vaisala Insight PC software cable connection
(requires USB cable accessory 219690)
2)
Cable lead-throughs M16 x 1.5 lead-through, options available from
Vaisala:
Cable Gland M16x1.5 (Vaisala order code:
254280SP). This is the default option
delivered with the transmitter.
Conduit Fitting M16x1.5, ½”NPT (Vaisala
order code: 210675SP)
Alternative M20 x 1.5 lead-through
Screw terminal wire size
0.5 ... 2.5 mm
2
1) Available calculated parameters for HMD65 include T
d
, T
df
, A, X, T
w
, and H.
2) Vaisala Insight software for Windows available at www.vaisala.com/insight.
Do not modify the unit or use it in ways not described in the
documentation. Improper modification may lead to safety hazards, equipment
damage, failure to perform according to specification, or decreased equipment
lifetime.
CAUTION!
12 M212264EN-B
Configuration Options
Vaisala Insight Software
Vaisala Insight software is a configuration software for Vaisala Indigo-compatible probes and
transmitters. The supported operating systems are Windows 7 (64-bit), Windows 8.1 (64-bit),
and Windows 10 (64-bit).
To ensure support for your HMD60 series transmitter, download the latest version
of Insight at www.vaisala.com/insight.
With the Insight software, you can:
See real-time measurements, device information and status.
Configure outputs and scaling.
Calibrate and adjust the device.
HMD60 can be connected to Insight using a Vaisala USB cable (order code 219690).
Connecting to Insight Software
Figure 7 Connecting Transmitter to Insight
1. Open the Insight software.
2. Connect the USB cable to a free USB port on the PC.
3. Connect the USB cable to the service port of the transmitter.
4. Wait for Insight software to detect the transmitter.
13
ENGLISH
Trimmers
1
Figure 8 Component Board Adjustment
Trimmer
1 Use a Phillips head screwdriver to
rotate the RH or T adjustment trimmer.
To increase the measurement output
value, rotate the trimmer clockwise. To
decrease, rotate counterclockwise.
Note that there is a slight delay before
the measurement output changes
after rotating the trimmer.
You can adjust the transmitter's RH or T measurement output with the trimmers on the
component board. During trimmer adjustment, the output of the transmitter is corrected using
the trimmers until the output matches the known value of a reference.
In order to make an adjustment with the trimmers, you need a reference measurement source.
You can either insert a reference instrument into the environment that HMD65 is installed in
and compare the readings of the instruments, or remove HMD65 from the installation
environment and use a calibration and adjustment tool (for example, Vaisala Humidity
Calibrator HMK15) to generate an environment with a known value.
±0 %RH
−1 %RH
−2 %RH
−3 %RH
−4 %RH
−5 %RH
+1 %RH
+2 %RH
+3 %RH
+4 %RH
+5 %RH
±0 °C
−0.06 °C
−0.12 °C
−0.18 °C
−0.24 °C
−0.3 °C
+0.06 °C
+0.12 °C
+0.18 °C
+0.24 °C
+0.3 °C
RH: −5 %RH ... +5 %RH T: −0.3 °C ... +0.3 °C
RH T
Figure 9 RH and T Trimmer Adjustment Ranges (Indicative)
You can only calibrate the relative humidity measurement (RH) and temperature
measurement (T). Other parameters are calculated internally based on RH and T.
Check that the output selection DIP switch is set to RH when making adjustments
with the physical trimmer; when using the Insight PC software, set all DIP
switches to the OFF position. For further information on using the adjustment
trimmers, see HMD65 User Guide.
If you use the Insight PC software to adjust the measurement or to
restore the factory settings, always return the physical trimmer to the middle
position before starting. When you make an adjustment with Insight, the
position in which the trimmer is at that point is set as the ±0 point.
CAUTION!
14 M212264EN-B
DIP Switch Humidity Output Selection
Figure 10 HMD65 DIP Switch Example: T
df
Output Selected
RH Relative humidity
Td Dew point temperature
Tdf Dew point/frost point temperature
A Absolute humidity
X Mixing ratio
Tw Wet-bulb temperature
H Enthalpy
You can change the humidity parameter that is output on the RH channel of HMD65 with the
DIP switches on the component board. Select the parameter you want the transmitter to
output by sliding the parameter's DIP switch to the right (ON). In the example in Figure 10
(page 15), the selected output parameter is dew point/frost point temperature (T
df
). Keep the
other DIP switches in the OFF position (left).
The selected parameter uses the default scaling shown in Table 2 (page 15).
Table 2 HMD65 Default Parameter Scaling
Parameter Default Scaling for 0 … 10 V Output Range
RH 0 … 100 %RH
T
d
-40 ... +80 °C (-40 ... +176 °F)
T
df
-40 ... +80 °C (-40 ... +176 °F)
A
0 … 300 g/m
3
(0 … 131.1 gr/ft
3
)
X 0 … 600 g/kg (0 … 4200 gr/lb)
T
w
-40 ... +80 °C (-40 ... +176 °F)
H -40 … 1600 kJ/kg (-9.5 ... 695.6 Btu/lb)
If you need to change the default scaling of a parameter, configure the output
with Vaisala Insight PC software. See the instructions in HMD65 User Guide.
If you use the Insight software to further configure the output, note
that the DIP switch selections override the Insight configuration. When using
Insight to configure the output, set all humidity parameter DIP switches to the
OFF position (left) to ensure they do not cause a conflict with the Insight
settings.
CAUTION!
15
ENGLISH
Modbus and BACnet Communication (RS-485)
3
2
1
7
4 5 6
Figure 11 Modbus and BACnet DIP Switches and Screw Terminals
1 DIP switch for setting RS-485 termination (120 Ω resistor) ON/OFF
2 Screw terminals for RS-485 (Modbus/BACnet) communication
3 Screw terminals for power supply input wiring (15 … 35 VDC / 16 … 24 VAC)
4 DIP switches for setting the device MAC address: see Figure 12 (page 16)
5 DIP switches for setting the communication bitrate (4800 … 115200 bps) and parity (N/E)
6 DIP switch for selecting either Modbus RTU or BACnet MS/TP mode
7 LED indicators for RS-485 transmit/receive activity
Setting Device MAC Address with DIP Switches
Figure 12 MAC Address DIP Switch Example
DIP
switches 32,
8, and 1 set
to ON
The MAC address is
encoded in eight bit binary
form, with each numbered
switch representing a single
bit. This example shows
address 41 selected: DIP
switches 32, 8, and 1
(decimal: 41, binary:
00101001) are set to ON.
Modbus and BACnet MAC Address Ranges
The BACnet MS/TP MAC address range for the transmitter is 0 … 127 (master mode only).
16 M212264EN-B
The Modbus RTU MAC address range for the transmitter is 1 … 247.
Configuring an address above the range maximum results in the address defaulting back to
the maximum address (127 or 247). Addresses below the range minimum default to the
minimum address (0 or 1).
Bit Rate and Parity Options
The bit rate 4800 is used only for Modbus RTU (use 9600 and above for BACnet MS/TP).
If the bit rate DIP switches are all set to OFF (left), the following defaults are used:
Modbus RTU: 19200
BACnet MS/TP: 38400
The parity selection (N/E) only has an eect on Modbus RTU communication.
Additional Configuration Options and Further Information
For HMD65 Modbus registers, see Modbus Registers (page 18).
For a description of the HMD65 BACnet protocol implementation and further information on
configuring BACnet, see the BACnet reference documentation in HMD65 User Guide available
at www.vaisala.com/hmd60.
To configure Modbus and BACnet settings beyond the communication settings available with
DIP switch selections, use Vaisala Insight PC software (see Connecting to Insight Software
(page 13) and instructions in HMD65 User Guide).
17
ENGLISH
Modbus Registers
The Modbus registers available for HMD65 include measurement output registers in metric and
non-metric units, pressure compensation setpoint configuration, status registers, and
communication test registers. For status registers, see HMD65 User Guide.
The Modbus communication settings are configured using the DIP switches on HMD65
component board: see Modbus and BACnet Communication (RS-485) (page 16).
Registers are numbered in decimal, starting from one. Register
addresses in actual Modbus messages (Modbus Protocol Data Unit (PDU)) start
from zero.
Please check the reference documentation of your Modbus host (PLC) for the
notation of Modbus register addresses.
CAUTION!
16-bit integers have a maximum value of +32767. Certain measurement
parameters can exceed this value when x100 scaling is used (see measurement
registers 0100
hex
… 0107
hex
and 0180
hex
… 0187
hex
). Whenever possible, it is
recommended to use 32-bit float values.
Measurement Data Registers
Table 3 Modbus Measurement Data Registers (Read-Only)
Register
Number
(Decimal)
Address
(Hexadecimal)
Register Description Data Format Unit
Floating Point Values (Metric)
1 0000
hex
Relative humidity 32-bit float %RH
0001
hex
3 0002
hex
Temperature 32-bit float °C
0003
hex
5 0004
hex
Dew point temperature 32-bit float °C
0005
hex
7 0006
hex
Dew/frost point
temperature
32-bit float °C
0007
hex
9 0008
hex
Absolute humidity 32-bit float
g/m
3
0009
hex
11 000A
hex
Mixing ratio 32-bit float g/kg
000B
hex
18 M212264EN-B
Register
Number
(Decimal)
Address
(Hexadecimal)
Register Description Data Format Unit
Floating Point Values (Metric)
13 000C
hex
Wet-bulb temperature 32-bit float °C
000D
hex
15 000E
hex
Enthalpy 32-bit float kJ/kg
000F
hex
Floating Point Values (Non-metric)
129 0080
hex
Relative humidity 32-bit float %RH
0081
hex
131 0082
hex
Temperature 32-bit float °F
0083
hex
133 0084
hex
Dew point temperature 32-bit float °F
0085
hex
135 0086
hex
Dew/frost point
temperature
32-bit float °F
0087
hex
137 0088
hex
Absolute humidity 32-bit float
gr/ft
3
0089
hex
139 008A
hex
Mixing ratio 32-bit float gr/lb
008B
hex
141 008C
hex
Wet-bulb temperature 32-bit float °F
008D
hex
143 008E
hex
Enthalpy 32-bit float Btu/lb
008F
hex
Integer Values (x100, Metric)
1)
257 0100
hex
Relative humidity 16-bit signed
integer
%RH
258 0101
hex
Temperature 16-bit signed
integer
°C
259 0102
hex
Dew point temperature 16-bit signed
integer
°C
260 0103
hex
Dew/frost point
temperature
16-bit signed
integer
°C
19
ENGLISH
Integer Values (x100, Metric)
1)
261 0104
hex
Absolute humidity 16-bit signed
integer
g/m
3
262 0105
hex
Mixing ratio 16-bit signed
integer
g/kg
263 0106
hex
Wet-bulb temperature 16-bit signed
integer
°C
264 0107
hex
Enthalpy 16-bit signed
integer
kJ/kg
Integer Values (x100, Non-metric)
1)
385 0180
hex
Relative humidity 16-bit signed
integer
%RH
386 0181
hex
Temperature 16-bit signed
integer
°F
387 0182
hex
Dew point temperature 16-bit signed
integer
°F
388 0183
hex
Dew/frost point
temperature
16-bit signed
integer
°F
389 0184
hex
Absolute humidity 16-bit signed
integer
gr/ft
3
390 0185
hex
Mixing ratio 16-bit signed
integer
gr/lb
391 0186
hex
Wet-bulb temperature 16-bit signed
integer
°F
392 0187
hex
Enthalpy 16-bit signed
integer
Btu/lb
1) NOTE: 16-bit integers have a maximum value of +32767. Certain measurement parameters
(for example, mixing ratio and enthalpy) can exceed this value in x100 scaling. In such cases,
the value of the parameter is cut o at +32767, and measurements above this value are not
reported. Verify that the measurement ranges in your application are suitable for the 16-bit
integer format with x100 scaling; whenever possible, it is recommended to use 32-bit float
values.
20 M212264EN-B
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154

Vaisala HMD65 ユーザーマニュアル

タイプ
ユーザーマニュアル