Prepare your treatment
1. To turn on the RecoveryAir Prime pump, press and hold the power button.
2. Access two preset program options when connected to the Therabody app:
Recover and Warm-up.
3. Once you have chosen your desired program, use the (<) and (>) buttons to
adjust the pressure setting.
4. Then, toggle through the time options using the Time button to set your
desired treatment time (20min, 40min, 60min, or continuous).
5. Once you have selected the treatment options that are right for you, press
start, relax, and you’ll be on your way to recovery in no time.
RecoveryAir Prime
Your RecoveryAir Prime pneumatic compression system comes with a pump, a set
of compression boots, a blocker plug, a DC power adapter, and a carrying pouch.
Advanced compression simplied for everybody.
See Warnings on pg. 5 for
important safety instructions.
1. To turn on the RecoveryAir Prime pump, press and hold the power button.
2. To get started, put your boots on first before attaching them to the pump. This allows you to
get into the most comfortable position and avoid any disruption in the air flow connection.
3. Next, plug in the connector from the boots into the pump. Make sure you hear a “click” so that
you know the RecoveryAir Prime is fully connected, is plugged in, and there is a proper seal.
4. Once connected and comfortable, you are ready to customize your treatment.
Getting started
2
EN
Feature Callouts
Buttons Indicators
Battery Indicator
3 LED lights to indicate battery life
Warning Indicator
Indicator lights up solid ON or blinking when a
device malfunction is detected
Bluetooth Indicator
Indicator turns ON only when connected to
the App
Start/Pause/Stop Button
Power Button
Time Button:
4 time intervals
(20, 40, 60, and continuous)
Pressure UP/Down Button:
Adjustable Pressure Range
(20-100 mmHg) in increments
of 10 mmHg
Adjustable pressure changes
via the Therabody app in
increments of 5 mmHg
3
EN
RecoveryAir Prime Programs
About negative gradient for all programs:
The RecoveryAir Prime’s unique true negative gradient of pressure sequentially travels up the limb from the
foot toward the heart in four internal overlapping chambers. The spiraling overlap of chambers safely maximizes
circulation.
The pressure can be the same in two consecutive chambers, but the pump won’t allow a back chamber to
have a higher pressure than a front chamber.
The minimum pressure level of each chamber is 20 mmHg.
Recover and Warm-Up preset programs
Access the Recover and Warm-Up preset programs via the Therabody app.
The Recover program is the perfect go-to when experiencing fatigue, tension, or soreness brought on by
your everyday activities or post workout.
The Warm-Up program is recommended for use before workout.
The only difference between the two programs is the default pressure level and treatment time.
Sequential inflation cycle
Both the Recovery and Warm-Up preset programs use the Sequential inflation cycle.
A directional massage is applied, starting at the base of the treated area, and progresses upwards towards
the torso and then releases. (see Figure 1)
Starting with Chamber 1, once the inflating chamber has reached its preset pressure level, the pressure
level for that chamber is held and the next chamber begins inflating.
Chamber 4
Chamber 3
Chamber 2
Chamber 1
Figure 1 - Sequential Cycle Mode
4
EN
Smart Features
The RecoveryAir Prime also connects to the Therabody
app via Bluetooth, giving you control of your RecoveryAir
Prime session from the convenience of your smartphone.
Not only can you easily access the Recover and Warm-up
treatment options, but you can also adjust the pressure
and time.
Scan the QR code
to get started.
Charging the RecoveryAir Prime
1. After turning off the pump, plug the power adapter into the charging port of the
RecoveryAir Prime pump.
2. You can continue to use the RecoveryAir Prime while charging.
3. Charging is complete when all 3 LED lights of the LED battery indicator are lit.
Traveling with your RecoveryAir Prime
Compression System
When traveling with your RecoveryAir Prime Compression System, we recommend:
Roll the hose tightly and slide it into the foot section of your garment. It is important to keep the
hose from creasing to protect the hose’s integrity and its ability to properly distribute airflow.
Once you secure the hose, simply fold your garment to fit into your travel bag or carry-on luggage.
Secure your pump in the provided carrying pouch.
5
EN
Superior, hygienic design
It’s important to take proper care of your RecoveryAir Prime compression boots because the longer they
last, the longer you will feel the benefits. RecoveryAir Prime compression boots have non-porous medical
grade material that help fight bacteria and provide a resistant surface. Our internal overlapping chambers
provide a smooth surface to clean, this means no wiping under chamber flaps where bacteria can build up.
This also means, no smells.
How to Clean
Here are simple steps to cleaning your RecoveryAir Prime compression boots:
1. Unzip the boots completely and lay them on a flat surface.
2. Take a disinfectant wipe or spray, and gently clean the inside of the boots. Avoid any oil based
cleaning product.
3. Once you have cleaned your garment, make sure to fully dry the surface by hanging them up or
wiping them down.
4. We also recommend you wipe down the exterior of the pump to reduce surface contaminants.
What it is and why it’s important:
We use medical-grade material covering the internal chamber to help reduce the potential for microbial
growth and moisture retention. Whether you are an individual, or a clinic, our boots are designed to help
limit the potential for bacterial transmission.
Important Safety Information
Read all instructions before using the RecoveryAir Prime Compression System for the first time
Warning
This system is intended for use by people in good health. This system is not recommended for people who have heart problems,
or vascular problems, have a condition requiring the use of any medical device, or have any condition that may affect their normal
well-being.
If you are, or may be, pregnant, consult with your physician before use.
Do not use this system over insensitive or numb areas, or in the presence of poor circulation. Do not use if you have been
6
EN
Important Safety Information
diagnosed with blood clots, deep vein thrombosis or phlebitis. This system should not be used over swollen or inflamed areas or
skin eruptions. Do not use in the presence of unexplained calf pain.
Consult your physician prior to use.
Safety Warnings and Precautions
Warning
Explosion Hazard: Do not use the RecoveryAir Prime Compression System in the presence of flammable gases, including
flammable anesthetics.
Electric Shock Hazard: Do not allow liquid to enter any part of the RecoveryAir Prime Compression System pump. Do not
immerse in water or liquid. To clean, follow instructions found on page 5.
Electric Shock Hazard: To prevent electric shock, do not open the pump. Do not attempt to service the pump yourself.
All repairs should be performed only by Therabody trained and authorized service personnel. Service by unauthorized
personnel will void any warranty.
Stop using the system if there is any change in RecoveryAir Prime Compression System’s performance.
Do not modify any part of the RecoveryAir Prime Compression System.
The RecoveryAir pneumatic compression system includes small parts that could cause choking in children.
Keep away from children and pets.
Do not inflate the garment without wearing it over the intended body area or with open zippers. Doing so can damage the
garment.
Do not apply excessive force to the garment straps. Do not use the straps for any purpose other than intended by the
manufacturer. Use of excessive force and/or misuse shall void the manufacturer’s warranty.
Do not stand or walk while wearing the garment.
Disconnect the pump from the electrical outlet before cleaning and let it dry completely before reconnecting it to the
electrical outlet.
Caution
Do not operate the RecoveryAir Prime Compression System while operating a vehicle.
Do not store or transport the RecoveryAir Prime Compression System beyond the specified temperature, humidity, and
atmospheric pressure range.
Do not use the RecoveryAir Prime Compression System beyond the specified temperature range: 10 to 30°C (50 to 86°F).
Do not use the RecoveryAir Prime Compression System beyond the specified humidity range: 30%rH–75%rH, non-
condensing.
Only use the RecoveryAir Prime Compression System up to 3000m above mean sea level.
To prevent any damage to the RecoveryAir Prime Compression System, keep it away from dust, lint, and dirt. Keep away from
sources of heat or moisture.
To prevent any damage to equipment, use only accessories, detachable parts, and materials described in this User Manual.
7
EN
While in use, place the pump on a horizontal firm surface only. Do not place the pump on a bed, blanket, mattress, pillow, or
soft furniture. Do not cover the pump.
The equipment is to be installed and put into service according to the EMC information provided in Chapter 11 - EMC
Manufacturer Declarations.
Portable and mobile RF communication equipment might affect the equipment.
For indoor use only.
Use only the DC power adapter provided with the pump.
The massage sensation should be pleasant and comfortable. If you experience pain or discomfort during or after the
massage or if there is onset of bruising or irritation during or after the massage, discontinue use and consult your physician.
Do not hand or machine wash. Surface wipe only.
Do not allow liquid to get into the air inlets.
Do not use bleach.
Do not dry clean.
Do not wring, iron, tumble, or force heat dry.
Important Safety Information
Label Description Location
Level of protection type BF
applied part On pump base
Class III equipment
(only in the EU per IEC60335-
2-32)
Equipment relying on
limitation of voltage to ex-
tra-low-voltage (ELV) values
as provision for basic protec-
tion and with no provision for
fault protection.
On pump base
(only in the EU)
Labels
The following labels and symbols appear on the pump, garments and/or packaging.
Label Description Location
IP 22 Degree of protection against
ingress of water On pump base
Read instructions before use On pump base
Read instructions before use (for
China only) On pump base
Read instructions before use On garment label
8
EN
Label Description Location
Serial number of the
console
On console
base
Fragile, handle with
care On package
Keep dry
On package
and console
base.
This side Up On package
Keep away from
sunlight On package
Transportation &
storage humidity
limitation
On package
Transportation &
storage atmospheric
pressure limitation
On package
Transportation & stor-
age temperature
limitation
On package
Label Description Location
Double insulation On DC power
adapter
Direct current On pump base
Therabody, Inc.
6100 Wilshire Blvd. Suite 200
Los Angeles, CA 90048
On garment label
and on pump base
Date of Manufacture On pump base
UDI :
(01) 00810036057288
(10)
(21)
2143
00001
Unique Device Identication
(UDI)
On pump base
and package, and
garment bag and
package
Separate collection for waste
electrical and electronic
equipment
On pump base
In accordance with Directive
2014/35/EU electrical equip-
ment designed for use within
certain voltage limits, and
Directive 2014/30/EU elec-
tromagnetic compatibility
On pump
9
EN
Label Description Location
Do not wash On garment tag
Do not dry-clean On garment tag
Do not tumble dry On garment tag
Label Description Location
Do not bleach On garment tag
Do not iron On garment tag
Indications for Use
RecoveryAir is indicated for the temporary relief of minor muscle aches and pains, and for the temporary increase in
circulation to the treated areas in people who are in good health. RecoveryAir simulates kneading and stroking of tissues by
using an inflatable garment.
Note
The pump is “multi-voltage” and can be used for travel.
An appropriate socket adapter must be used in countries with incompatible wall outlets.
Use only RecoveryAir garments with the RecoveryAir pump.
When immediate garment deflation is necessary, disconnect the garment hose bundle from the pump to
immediately deflate the air pressure in the garment.
10
EN
Note Possible Cause Action
The pump is not working. No electricity Inspect the electrical wall outlet.
DC power adapter Verify that the DC power adapter cable is connected to
the DC adapter socket on the console, and the DC pow-
er adapter is connected to the 100-240 Volt wall outlet.
DC power adapter cable Examine the cable for any defects.
The Status Indicator is On in yellow Malfunction Contact Therabody.
The pump starts working and stops immediately. The air cannot move through the garment
hose.
Examine hoses for kinks, twists and folds.
One garment inflates but the second one does not. The second garment does not receive air. Examine its hoses for kinks, twists and folds.
The pump stops working, the Status Indicator turns
On in yellow.
Hose is not connected properly to garment
or pump, or prong plug is not inserted into
unused air outlet.
Examine and fasten all air connections. When treating
only one limb, always plug the unused air outlet with the
prong plug provided with the pump. If all air connections
are OK and the problem persists, contact Therabody.
The pump works at a very low pressure, regardless
of the pressure set by the user.
Defective garment Replace garment and check again.
Warranty
For full warranty information, please visit www.therabody.com/warranty. To request a copy of the warranty by mail, you may send a
request to the following address:
Therabody - WarrantyAttn: Customer Service
6100 Wilshire Blvd.,
Los Angeles, CA 90048
Please note, this is not a return address or a retail location. No products or packages will be accepted at this location.
Troubleshooting
11
EN
Storage environment
The pump can be transported or stored for short periods of time within:
Temperature range of -4 - 158°F (-20- 70°C)
Humidity range of 10-93% RH non-condensing
Atmospheric pressure range of 190 - 1060hPa
Allow the pump to reach a reasonable room temperature of 50 - 86°F (10 - 30°C) before operating.
When the system has been stored in extreme temperature conditions of -20°C (-4°F) or 70°C (158°F) between uses, wait
for two (2) hours before using again.
Radiation Exposure statement
This equipment complies with FCC/IC RSS-102 radiation exposure limits set forth for an uncontrolled environment.
12
TW
準備進行治療
1. 要開啟RecoveryAirPrime幫浦請按住電源按鈕
2. 連線到Therabody應用程式時可使用兩個預設療程選項Recover和Warm-up
3. 一旦選擇了所需的療程之後請使用(+)和(-)(<)和(>)按鈕來調整按壓壓力設定
4. 接著使用Time按鈕來切換時間選項設置所需的治療時間20分鐘40分鐘60
分鐘或連續
5. 一旦選擇了適合的治療選項按下開始放輕鬆您將很快就會恢復狀態
RecoveryAir Prime
您的RecoveryAirPrime氣動壓縮系統附有一個幫浦一雙壓縮筒靴一個插頭塊一個DC電
源轉接器以及一個隨身收納袋
為每個人簡化了進階壓縮療程
請參閱第16頁的《警告》
了解重要的安全性說明
1. 要開啟RecoveryAirPrime幫浦請按住電源按鈕
2. 要開始使用請先穿上筒靴然後將它們連接到幫浦這可以讓您找到最舒適的位置並且避免任何氣流
連接的中斷
3. 接下來將靴子的連接器插入幫浦中請確保聽到喀嚓」好讓您知道RecoveryAirPrime已完全連接
插入並且有著適當的密封
4. 一旦連接並找到舒適的位置即可自訂您的治療
開始使用
13
TW
按钮
功能特性標註
指示燈
電池指示燈
以3個全亮的燈條來指示電池壽命
警告指示燈
檢測到設備故障時指示燈會常亮」或閃爍
藍牙指示燈
指示燈僅會在連線到應用程式時亮起
Start開始/Pause暫停/
Stop停止按鈕
Power Button(電源鍵
Time 按鈕
4組時間間隔204060和連續)
Pressure 按鈕
可調整的壓力範圍20-100
mmHg以10mmHg遞增
通過Therabody應用程式
調整的壓力變化能以5mmHg
遞增
14
TW
RecoveryAirPrime的獨特真負壓梯度在四個内部重疊加壓艙中依序從足部往上向心臟方向移動加壓艙的螺
旋重疊能安全地最大化循環
兩個連續加壓艙中的壓力可以相同但幫浦不允許後艙的壓力高於前艙
各個加壓艙的最低壓力等級為20mmHg
Recover Warm-Up 預設療程
通過Therabody應用程式使用Recover和Warm-Up
Recover療程是在日常活動或健身後感到疲勞緊張或痠痛時的完美選擇
建議在健身前使用Warm-Up療程
兩個療程間的唯一區別是預設的壓力等級和治療時間
依序充氣循環
Recovery和Warm-Up預設療程都使用Sequential充氣循環
進行定向按摩從治療區域的底部開始並往上
朝著軀幹方向進行接著釋放壓力見圖1)
從加壓艙1開始一旦加壓艙達到其預設壓力等級
加壓艙的壓力等級將會維持不變並且下
一個加壓艙則開始充氣
RecoveryAir Prime 療程
關於所有療程的副梯度
加壓艙 4
加壓艙 3
加壓艙 2
加壓艙 1
圖 1 Sequential 循環模式
智慧功能特性
RecoveryAirPrime能同時通過藍牙連線到Therabody
應用程式讓您通過智慧型手機便利地控制您的
RecoveryAirPrime療程
不僅可以便利地使用Recover和Warm-up治療選項
可以調整按壓壓力和時間
Scan the QR code
to get started.
15
TW
妥善保養您的RecoveryAirPrime壓縮筒靴很重要因為它們的使用壽命越長您能感受到好處的時間就越
RecoveryAirPrime壓縮筒靴採用無孔醫療級材料有助於防止細菌滋生並提供耐磨表面我們的內部重疊加
壓艙有著容易清潔的光滑表面這意味著不需要擦拭會累積細菌的加壓艙封蓋這也意味著沒有難聞氣味
如何清潔
以下是清潔RecoveryAirPrime壓縮筒靴的簡易步驟
1. 完全解開筒靴的拉鍊並將它們放置於平坦的表面上
2. 用消毒濕紙巾或噴劑輕輕地擦拭筒靴內部避免使用任何油性清潔產品
3. 清潔完裝備後確保通過懸掛或擦拭它們好讓表面完全乾燥
4. 我們還建議擦拭幫浦的外部以減少表面污染物
它是什麼和為什麼它很重要
我們使用醫療級材料覆蓋內部加壓艙有助於減少細菌滋生和濕氣滯留的可能性無論是個人還是診所使用
們筒靴設計的目的為協助限制細菌傳播的可能性
為 RecoveryAir Prime 充電
帶著 RecoveryAir Prime 壓縮系統去旅行
1. 關閉裝置電源後將電源轉接器插入RecoveryAirPrime幫浦的充電埠
2. 可以在充電時繼續使用RecoveryAirPRO
3. LED電池指示燈的所有3個LED燈點都亮起時即代表充電完成
當帶著RecoveryAirPrime壓縮系統去旅行時我們建議
緊緊地捲收軟管並將其收入裝備的足部區域避免軟管起皺很重要這可以保護軟管的完整性及其正確
分配氣流的能力
一旦固定軟管後只需將裝備折疊放入旅行包或隨身行李即可
將幫浦固定在隨附的隨身收納袋
卓越衛生的設計
16
TW
重要的安全性說明
首次使用RecoveryAirPrime壓縮系統之前請詳細閱讀所有說明
警告
本系統供身體健康的人使用對於有心臟問題或血管問題需要使用任何醫療設備或有可能影響其正常健康的任何情況的人不建議使用此系統
如果您已經懷孕或可能懷孕請在使用前諮詢您的醫生
請勿在不敏感或麻木身體部位或血液循環不良的情況下使用此系統如果您被診斷有血栓深部靜脈栓塞或靜脈炎請勿使用此系統不應使用於腫
脹或發炎區域或皮膚藥物過敏時不要在不明原因的小腿疼痛情況下使用
使用前請諮詢您的醫生
•
安全性警告和預防措施
警 告:
爆炸危險 請勿在存在易燃氣體(包括易燃麻醉劑的情況下使用RecoveryAirPrime壓縮系統.
電擊危險 請勿讓液體進入到RecoveryAirPrime壓縮系統幫浦的任何部位請勿浸入水中或液體中如要清潔請按照第15頁上的說明進行操作
電擊危險 為防止電擊請勿打開幫浦請勿嘗試自行維修幫浦所有的維修只能由Therabody培訓和授權的服務人員進行由未經授權的人員提供服務
將使得任何保固失效
如果RecoveryAirPrime壓縮系統的性能發生任何變化請停止使用該系統
請勿改動RecoveryAirPrime壓縮系統的任何部分
RecoveryAir氣動壓縮系統包含可能導致孩童窒息的小型零件遠離孩童和寵物
請勿在未將裝備穿在預期的身體部位或打開拉鍊的情況下為其充氣這樣做會損壞裝備
請勿對裝備繫帶施加過大的力量請勿將繫帶用於製造商規定之外的任何目的使用過大的力量和/或誤用將使製造商的保固失效
請勿在穿著裝備時站立或行走
清潔前請斷開幫浦與電源插座的連接並讓它完全乾燥後再將其重新連接到電源插座
謹慎
請勿在駕駛車輛時操作RecoveryAirPrime壓縮系統
請勿在超出指定溫度濕度和大氣壓力範圍的情況下存放或運輸RecoveryAirPrime壓縮系統
請勿在超出指定溫度範圍10至30°C50至86°F的情況下使用RecoveryAirPrime壓縮系統
請勿在超出指定濕度範圍的情況下使用RecoveryAirPrime壓縮系統30%rH–75%rH非凝結
僅在平均海平面以上3000公尺的範圍內使用RecoveryAirPrime壓縮系統
為了防止RecoveryAirPrime壓縮系統受到任何損壞請使其遠離灰塵棉絨和污垢遠離熱源或濕氣
為了防止對設備造成任何損壞請只使用本用戶手冊中描述的配件可拆卸零件和材料
使用時只能將幫浦放置在水平穩固的表面上請勿將幫浦放置在床被毯床墊枕頭或軟家具上請勿蓋住幫浦
設備將根據第11章–《EMC製造商聲明中提供的EMC資訊進行安裝和使用
可攜式和移動式RF通訊設備可能會影響該設備
只能在室內使用
只能使用幫浦隨附的DC電源轉接器
按摩的感覺應該是愉快和舒適的如果在按摩期間或之後感到疼痛或不適或者在按摩期間或之後出現瘀血或刺激請停止使用並諮詢您的醫師
請勿手洗或機洗只能表面擦拭
請勿讓液體進入進氣口
請勿使用漂白劑
請勿乾洗
請勿擰乾熨燙滾筒烘乾或強加熱度烘乾
17
TW
標籤
以下標籤和符號出現在幫浦裝備和/或包裝上
標籤 說明 位置
Therabody, Inc.
6100 Wilshire Blvd.
Suite 200
Los Angeles, CA 90048
在裝備標籤和
幫浦底座上
製造日期 在幫浦底座上
UDI :
(01) 00810036057288
(10)
(21)
2143
00001
單一裝置識別
(Unique Device Identi󼴩-
cation, UDI)
在幫浦底座和
包裝與裝備收
納袋和包裝上
廢棄電器和電子設備分
開收集 在幫浦底座上
符合 2014/35/EU 電氣設
備指令設計用於在特定
電壓限制範圍內使用
及指令 2014/30/EU
磁兼容性
在幫浦上
控制台序號 在控制台底
部上
易碎品請小心輕放 在包裝上
標籤 說明 位置
IP 22 防進水等級 在幫浦底座上
使用前閱讀說明 在幫浦底座上
使用前閱讀說明(僅適用於中國) 在幫浦底座上
使用前閱讀說明 在裝備標籤上
防護等級 BF 類型的觸身部件 在幫浦底座上
Class III 設備
(根據 IEC60335-2-32僅適用
於歐盟)
依靠將電壓限制為超低電壓
(extra-low-voltage, ELV) 值作為
基本保護措施的設備並且其未提
供故障保護
在幫浦底座上
(僅適用於歐盟
雙重絕緣 在 DC 電源轉
接器上
直流電 在幫浦底座上
18
TW
標籤 說明 位置
保持乾燥 在包裝和控制台
底部上
此面朝上 在包裝上
避免日光照射 在包裝上
運輸和存放濕
度限制 在包裝上
運輸和存放大氣壓
力限制 在包裝上
運輸和存放溫
度限制 在包裝上
標籤 說明 位置
請勿清洗 在裝備標籤上
請勿乾洗 在裝備標籤上
請勿烘乾 在裝備標籤上
請勿漂白 在裝備標籤上
請勿熨燙 在裝備標籤上
使用說明
RecoveryAir適用於暫時緩解輕微的肌肉酸痛以及暫時增進身體健康人士其受治療區域的血液循環RecoveryAir使用充氣裝
備來模擬組織的揉捏和按摩
備註
幫浦為「多重電壓並且可以用於旅行
在牆上插座不相容的國家/地區必須使用合適的插座轉接器
只能將RecoveryAir裝備與RecoveryAir幫浦一起使用
當需要立即將裝備放氣時斷開裝備軟管束與幫浦的連接即可立即釋放裝備中的氣壓
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52

Therabody RecoveryAir Prime ユーザーマニュアル

タイプ
ユーザーマニュアル