Hill-Rom MicroTymp 4 Portable Tympanometer リファレンスガイド

タイプ
リファレンスガイド
Welch Allyn MicroTymp® 4 Quick Guide
4341 State Street Road, Skaneateles Falls, NY 13153 www.welchallyn.com
GSI D-0122380
© 2019, Welch Allyn, Inc.
DIR No. 80025202 Ver. B, Revision Date 2019-07
WELCH ALLYN® MICROTYMP® 4
HAND HELD PORTABLE TYMPANOMETER
QUICK GUIDES
Welch Allyn MicroTymp® 4 Quick Guide
4341 State Street Road, Skaneateles Falls, NY 13153 www.welchallyn.com
GSI D-0122380
© 2019, Welch Allyn, Inc.
DIR No. 80025202 Ver. B, Revision Date 2019-07
Contents:
Welch Allyn MicroTymp® 4 Tympanometer Quick Guide English
Welch Allyn MicroTymp® 4 手持ち型携帯用ティンパノメー クイックガイド- 日本語
Welch Allyn MicroTymp® 4 Quick Guide
4341 State Street Road, Skaneateles Falls, NY 13153 www.welchallyn.com
GSI D-0122380
© 2019, Welch Allyn, Inc.
DIR No. 80025202 Ver. B, Revision Date 2019-07
The operating manual contains information pertinent to the use of the MicroTymp 4 system, including
safety information, as well as maintenance and cleaning recommendations. Read the manual in its
entirety prior to use of the MicroTymp 4 device on a patient.
Handset
Press the On/Off key to turn the MicroTymp 4 on (refer to the diagram above). No warm-up time is required,
although a short diagnostic routine will run for a few seconds. During this time the internal pump will operate. To
switch off, again press and hold the On/Off key for a few seconds.
Press the up and down navigation keys to scroll through the menus or set values
Press the right navigation key to accept a menu choice or go to the next step.
Press the left navigation key to cancel an operation or go back to the previous step.
The function of the left and right keys is usually shown on the bottom line of the display.
Cradle LED Indicators
The LED indicators on the instrument cradle show the status of the AC mains connection and the
battery charging.
LED displays green when power is applied to the cradle; otherwise it will be off.
LED shows green when the handset is in the cradle and its internal battery pack is charging; it will be off
when the handset is removed.
Probe
LED
Indicators
LCD Display
Navigation
Keys
Welch Allyn MicroTymp® 4 Quick Guide
4341 State Street Road, Skaneateles Falls, NY 13153 www.welchallyn.com
GSI D-0122380
© 2019, Welch Allyn, Inc.
DIR No. 80025202 Ver. B, Revision Date 2019-07
Prior to Testing and Testing Environment
A qualified health care professional should perform a thorough otoscopic examination to establish that the condition
of the ear is suitable for the test and that no contraindications are present.
Tympanometric and reflex testing should always be performed in a quiet room or in an acoustic booth. No specific
action is required by the patient during the automatic test. However, the patient must be advised to remain still and
avoid speaking or swallowing while the probe is applied to the ear.
The probe tip must be fitted with a new ear tip before it is presented to a patient’s ear canal. The ear
tip must be fitted completely to the probe tip and must not occlude any of the four holes in the probe
tip. The ear tip size is chosen to suit the patient’s ear and provide a comfortable pressure seal.
Performing a test
From the MAIN MENU select NEW TEST.
Select the ear(s) to be tested.
Place the ear tip into the ear canal to obtain a seal.
Once an adequate seal is detected, a tympanogram measurement is made.
When the tympanogram is complete the instrument will perform the reflex test(s), if selected.
When the measurement is complete withdraw the probe and the
tympanogram will be displayed.
Review the tympanogram to ensure that the peak admittance point selected
by the MicroTymp 4 is suitable. If required, it is possible to select an
alternative peak using the and keys.
To repeat the test, press .
When satisfied with the tympanogram press .
If the reflex test was carried out the results will be displayed.
Use the and keys to view the results for each of the reflex tone levels
used. If the reflex test was performed at more than one frequency use the
and keys to view the results for the other frequencies. If the MicroTymp 4
was set to test for a reflex at all levels of the stimulus press to view a
summary of the levels and frequencies at which a reflex was detected.
Press to return and view the tympanogram, reflex results or to repeat the test. When satisfied with the results
press .
If both ears were chosen for test the entire sequence will now be repeated for the other ear.
When the selected ears have been tested and the results saved, the PROCESS RESULTS menu will be displayed:
PRINT (Print the results)
SAVE RESULTS (Save the results in the internal database)
VIEW TEST (Review the results as described above)
MAIN MENU (Return to the main menu)
Results of the last test will be erased as soon as a new test is started. Test results should be saved to the internal
database or printed to ensure that data is not lost.
Welch Allyn MicroTymp® 4 Quick Guide
4341 State Street Road, Skaneateles Falls, NY 13153 www.welchallyn.com
GSI D-0122380
© 2019, Welch Allyn, Inc.
DIR No. 80025202 Ver. B, Revision Date 2019-07
取扱説明書には、安全情報、メンテナンス、クリーニングについての推奨事項などの、MicroTymp 4
ステムの使用に関する情報が記載されています。 MicroTymp 4デバイスを患者に使用する前に、取扱説
明書全体に目を通してください。
ハンドセット
オン/オフキーを押して、MicroTymp 4 をオンにします (上図参照)。ウォーミングアップ時間は不要ですが、数秒間の
短い診断ルーチンが実行されます。この間、内部ポンプが作動します。スイッチをオフにするには、再度オン/オフ
キーを数秒間長押しします。
メニューや設定値を表示するには、上▼のナビゲーションキーを押して画面をスクロールします。
選択したメニューの確定、または次のステップに進むには、右ナビゲーションキーを押します。
操作のキャンセル、または前のステップに戻るには、左ナビゲーションキーを押します
左右キーの機能は、通常画面最下部に表示されます。
クレードルLEDインジケー
装置クレードルLEDインジケータは、AC電源接続とバッテリーの充電状態を示します。
クレードルの電源がオンになると、LEDディスプレイが緑に点灯します。電源が入っていない場合は、
点灯しません。
LEDライトは、クレードルにハンドセットが入っている時と内蔵バッテリーパックを充電している時緑
に点灯し、ハンドセットが入っていない時は点灯しません。
プローブ
LEDインジ
ケータ
LCDディス
プレイ
ナビゲーシ
ョンキー
Welch Allyn MicroTymp® 4 Quick Guide
4341 State Street Road, Skaneateles Falls, NY 13153 www.welchallyn.com
GSI D-0122380
© 2019, Welch Allyn, Inc.
DIR No. 80025202 Ver. B, Revision Date 2019-07
テスト前、およびテスト環境
耳の状態が検査に適し、禁忌がないことを確認するために、資格を保持する医療専門家による入念な耳鏡検査が必要
となります。
ティンパノメトリーと反射テストは常に静かな部屋か音響ブースで実施してください。 自動検査中は、患者による
特別な操作は必要ありません。ただし、プローブを耳に当てている間は、静止したまま、話したり飲み込んだりしな
いよう患者に指示してください。
プローブチップを患者の外耳道に挿入する前に、プローブチップに新しいイヤーチップを取り付けてく
ださい。 イヤーチップをプローブチップに完全にはめ込み、プローブチップの4つの穴がどれも塞がれ
ていないようにします。イヤーチップのサイズは、患者の耳に合うものを選択し、快適な圧力で耳穴を
塞ぎます。
検査の実施
メインメニューから新規検査を選択します。
検査する耳を選択します。
イヤーチップを外耳道に挿入して耳穴を塞ぎます
適切に穴が塞がれたことが検知されると、ティンパノグラム測定が行われます。
ティンパノグラム完了後に、装置は反射テストを実行します (選択されている場合)
測定を完了してプローブを抜くと、ティンパノグラムが表示されます。
MicroTymp 4よって選択されたピークアドミタンス点が適切であることを確
認するために、ティンパノグラムを再確認してください。 必要に応じて、▲
と▼キーを使用して代替のピークを選択することが可能です。
検査を繰り返すには、◄を押します。
ティンパノグラムの結果に満足したら►を押します。
反射テストを実施した場合はその結果が表示されます。
使用された各反射トーンレベルの結果を表示するには、▲キーと▼キーを使用
します。反射テストを複数の周波数で実行した場合は、▲キーと▼キーを使用
して他の周波数の結果を表示します。 刺激の全レベルで反射テストを行うよ
うにMicroTymp 4を設定した場合、►を押すと反射が検出されたレベルと周波数
の概要を表示されます。
キーを押して、ティンパノグラムと反射結果に戻って確認するか、検査を繰り返します。 検査結果に満足した
ら、►を押します。
両耳の検査を選択した場合、もう一方の耳にもすべてのシーケンスが繰り返されます。
選択した耳が検査されて結果が保存されると、結果処理メニューが表示されます:
印刷(結果を印刷)
結果を保存(結果を内部データベースに保存)
検査の表示(上記のように結果を確認)
メインメニュー (メインメニューに戻る)
最後の検査結果は、新しい検査が開始されるとすぐに消去されます。 検査結果は、内部データベースに保存する
か、データが失われないように印刷する必要があります。