MDO3014

Tektronix MDO3014, MDO3000 Series, MDO3012, MDO3022, MDO3024, MDO3032, MDO3034, MDO3052, MDO3054, MDO3102, MDO3104 Installation And Safety Instructions

  • MDO3000シリーズオシロスコープの設置および安全に関する手順書の内容を理解しました。このドキュメントでは、機器の設置、安全上の注意事項、操作方法について詳しく説明されています。オシロスコープの機能、使用方法、トラブルシューティングなど、ご質問にお答えしますので、お気軽にお尋ねください。
  • オシロスコープの電源を入れる手順は?
    TPP0250プローブを補正するにはどうすればよいですか?
    ファームウェアをアップデートするにはどうすればよいですか?
xx
MDO3000 Series
Oscilloscopes
Installation and Safety
ZZZ
Instructions
*P071324900*
071-3249-00
MDO3000 Series
Oscilloscopes
Installation and Safety
ZZZ
Instructions
xx
www.tektronix.com
071-3249-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries
or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverto
n, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tektronix.com to nd contacts in your area.
Table of Contents
Preface ............................................................................................................... 1
Warranties ...................................................................................................... 1
Accessories and Replaceable Parts .......................................................................... 1
Documentation ................................................................................................. 2
Important safety information ...................................................................................... 4
General safety summary ...................................................................................... 4
Service safety summary....................................................................................... 7
Terms in this manual .......................................................................................... 8
Symbols and terms on the product........................................................................... 8
Compliance information ........................................................................................... 9
EMC compliance .............................................................................................. 9
Safety compliance ............................................................................................ 10
Environmental considerations............................................................................... 13
Operating Requirements .......................................................................................... 14
Electrical Ratings ............................................................................................. 16
Input Ratings .................................................................................................. 17
Environmental Ratings....................................................................................... 17
Physical Specications....................................................................................... 18
Installation Procedure ............................................................................................. 19
A Tour of Your Instrument ....................................................................................... 20
Front-Panel Menus, Controls, and Connectors ............................................................ 20
Rear-Panel Connectors....................................................................................... 21
Power-On and Power-Off Procedure ............................................................................ 22
Functional Check ............................................................................................. 23
Compensating a TPP0250, TPP0500B or TPP1000 Passive Voltage Probe ......... .................. 24
Application Module Free Trial .............................................................................. 26
Upgrading Firmware ......................................................................................... 27
Connecting Your Oscilloscope to a Computer............................................................. 28
Getting Acquainted with the Oscilloscope............ .......................... .............................. .. 31
Using the Menu System...................................................................................... 31
Using the Menu Buttons ..................................................................................... 33
Using Spectral Analysis Controls........................................................................... 34
Using Other Controls......................................................................................... 35
Identifying Items in the Time Domain Display............................................................ 38
Identifying Items in the Frequency Domain Display ..................................................... 42
Identifying Items in the Arbitrary Function Generator Display ......................................... 43
Identifying Items in the Digital Voltmeter Display ....................................................... 44
まえがき .................................................................................................. 45
保証期間 ............................................................................................ 45
MDO3000 Installation and Safety Instructions i
Table of Contents
アクセサリおよび交換可能
.................................................................... 45
マニュアル ........................................................................................... 47
安全性に関する重要な情報............................................................................ 48
安全にご使用いただくために ...................................................................... 48
安全に保守点検していただくために .............................................................. 51
本マニュアル内の用語 ............................................................................. 52
本製品に使用される記号と用語 ................................................................... 52
適合性に関する情報.................................................................................... 53
EMC ......................................................................................... 53
安全性 ............................................................................................... 54
環境条件について.................................................................................. 57
動作の要件 .............................................................................................. 58
電気定格 ............................................................................................ 60
入力定格 ............................................................................................ 61
環境要件 ............................................................................................ 61
物理仕様 ............................................................................................ 62
設置手順................................................................................................. 63
ご使用機器のツアー .................................................................................... 64
前面パネルのメニュー、コントロール、コネクタ ................................................... 64
後部パネル・コネクタ ............................................................................... 65
電源投入、電源切断の手順............................................................................ 66
機能チェック......................................................................................... 67
TPP0250 型、TPP0500B 型、TPP1000 型受動電圧プローブの補正............................ 68
アプリケーション・モジュールの無料トライアル ................................................... 70
ファームウェアのアップグレード: .................................................................. 71
オシロスコープとコンピュータの接続 .............................................................. 72
オシロスコープの概要 .................................................................................. 75
メニュー・システムの使用........................................................................... 75
メニュー・ボタンの使用 ............................................................................. 77
スペクトラム解析コントロールの使 .............................................................. 78
他のコントロールの使用 ............................................................................ 79
時間領域表示の項目............................................................................... 83
周波数領域表示の項目 ............................................................................ 87
任意波形ファンクション・ゼネレータ表示の項目 ................................................. 88
デジタル電圧計表示の項目 ....................................................................... 89
前言 ...................................................................................................... 91
保修 .................................................................................................. 91
附件和可更换部件 ................................................................................. 91
文档 .................................................................................................. 92
重要安全信息 ........................................................................................... 94
常规安全概要....................................................................................... 94
ii MDO3000 Installation and Safety Instructions
Table of Contents
维修安全概要....................................................................................... 97
本手册中的术语 .................................................................................... 97
产品上的符号和术语 .............................................................................. 97
符合性信息 .............................................................................................. 98
EMC .......................................................................................... 98
安全符合性.......................................................................................... 99
环境注意事项...................................................................................... 102
操作要求................................................................................................ 103
电源额定值......................................................................................... 105
输入额定值......................................................................................... 106
环境额定值......................................................................................... 106
物理技术规格...................................................................................... 107
安装步骤................................................................................................ 108
仪器概览................................................................................................ 109
前面板菜单、控件和连接 ..................................................................... 109
后面板连接器...................................................................................... 110
开机和关机步骤 ....................................................................................... 111
功能检查 ........................................................................................... 112
补偿 TP
P0250、TPP0500B TPP1000 无源 ........................................ 113
应用模块免费试用 ................................................................................ 115
升级固件 ........................................................................................... 115
将示波器连接到计算机........................................................................... 116
熟悉示波器 ............................................................................................. 119
使用菜单系统...................................................................................... 119
使用
菜单按钮...................................................................................... 121
使用频谱分析控件 ................................................................................ 122
使用其它控件...................................................................................... 123
识别时域显示中的项 ............................................................................. 126
识别频域画面中的项 ............................................................................. 130
识别任意波形函数发生器画面中的项 .......................................................... 131
识别数字电压表显示中的 ..................................................................... 132
MDO3000 Installation and Safety Instructions iii
Table of Contents
iv MDO3000 Installation and Safety Instructions
Preface
This manual describes the installation and operation of the following
oscilloscopes:
MDO3104 MDO3054 MDO3034 MDO3024 MDO3014
MDO3102 MDO3052 MDO3032 MDO3022 MDO3012
Important safety precautions to avoid injury and prevent damage to this
product or any products connected to it
EMC (electromagnetic compliance), safety, and environmental standards with
which the product complies
Voltage, power, and environmental requirements to use the product
Installation procedure
Power-on and power-off procedure
Front- and rear-panel controls and connectors
Time, frequency, arbitrary function generator, and digital voltmeter displays
Wa rranties
Warranties
Warranty Description
MDO3000 oscilloscope: Three year
warranty
P6316, TP P0250, TPP0500B, and
TPP1000 probes: One year warranty
For details, refer to the warranties in the front
of the electronic (PDF) user manual
Accessories and Replaceable Parts
Optional Accessories
Tektronix part number Description
MDO3AERO MIL-STD-1553 Serial Triggering and Analysis
MDO3AUDIO Audio Serial Triggering and Analysis (I
2
S, LJ, RJ, TDM)
MDO3AUTO Automotive serial triggering and analysis (CAN and LIN)
MDO3COMP Computer triggering and analysis (RS-232, RS-422, RS-485 and UART)
MDO3EMBD Embedded serial triggering and analysis (I
2
C and SPI)
MDO3FLEX FlexRay Serial Triggering and Analysis
MDO3USB Universal Serial B us Triggering and Analysis (LS, FS, HS).
High speed is decode only; available on 1 GHz models only.
MDO3000 Installation and Safety Instructions 1
Preface
Optional Accessories, (cont.)
Tektronix p art number Description
MDO3LMT Limit/Mask Test
MDO3PWR
Power Measurement Analysis
TekVPI probes that work with MDO3000 Series
oscilloscopes
Visit the Oscilloscope Probe and Accessory Selector Tool on the Tektronix
website at www.tektronix.com/probes
TPA-BNC TekVPI to TekProbe II B NC Adapter
Optional Instrument Upgrades
Tektronix p art number Description
MDO3AFG Arbitrary function generator
MDO3MSO
16 digi
tal channels; includes P6316 digital probe
MDO3SA Increase spectrum analyzer input frequency to 3 GHz.
MDO3SEC
Add pa
ssword protected security to enable or disable all communication
ports and rmware upgrades to any MDO3000 Series oscilloscope.
Bandw
idth upgrades
Upgrade the analog bandwidth on MDO3000 Series products post-purchase.
Visit www.tektronix.com for information on available upgrade products.
Documentation
The
following table lists the documentation that is available for the product and
shows where you can nd it: in a printed manual, on the product documentation
CD-ROM, or on the Tektronix Web site at www.tektronix.com.
Table 1: Product documentation
Item Purpose Location
Installation and Safety Instructions (this
manual)
Provides safety and compliance
information along with hardware
installation instructions to present the
associated safety warnings. This manual
is available in English, Japanese, and
Simplied Chinese
Printed manual and also
available in electronic format at
www.tektronix.com/manuals
User Manual Provides operation and application
information. This manual is available
in English, French, Italian, German,
Spanish, Japanese, Portuguese,
Simplied Chinese, Traditional Chinese,
Korean, and Russian
Product Documentation CD and available
at www.tektronix.com/manuals
Specications and Performance
Verication Technical Reference
Specications and procedures for
checking instrument performance.
Product Documentation CD and available
at www.tektronix.com/manuals
2 MDO3000 Installation and Safety Instructions
Preface
Table 1: Product documentation, (cont.)
Item Purpose Location
Programmer Manual
Command reference for remotely
controlling the instrument.
Product Documentation CD and available
at www.tektronix.com/manuals
Service Manual Provides information about adjustments,
repair, and replaceable parts.
Available at www.tektronix.com/manuals
MDO3000 Installation and Safety Instructions 3
Important safety information
Important saf
ety information
This manual c
ontains information and warnings that must be followed by the user
for safe operation and to keep the product in a safe condition.
To safely perform service on this product, additional information is provided at
the end of this section.(See page 7, Se rvice s afety summary .)
General safety summary
Use the product only as specied. Review the following safety precautions to
avoid injury and prevent damage to this product or any products connected to it.
Carefully read all instructions. Retain these instructions for future reference.
Comply with local and national safety codes.
For correct and safe operation of the product, it is essential that you follow
generally accepted safety procedures in addition to the safety precautions specied
in this manual.
The product is designed to be used b y trained personnel only.
Only qualied personnel who are aware of the hazards involved should remove
the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it is
operating correctly.
This product is not intended for detection of hazardous voltages.
Use personal protective equipment to prevent shock and arc blast injury where
hazardous live conductors are exposed.
While using this product, you may need to access other parts of a larger system.
Read the safety sections of the other component manuals for warnings and
cautions related to operating the system.
When incorporating this equipment into a system, the safety of that system is the
responsibility of the assembler of the system.
To avoid re or personal
injury
Use proper power cord. Use only the power cord specied for this product and
certied for the country of use.
Do not use the provided power cord for other products.
Ground the product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, make sure that the product is properly g rounded.
4 MDO3000 Installation and Safety Instructions
Important safety information
Do not disable t
he power cord g rounding connection.
Power disconnect. The power cord disconnects the product from the power
source. See instructions for the location. Do not position the equipment so that
it is difcult to operate the power cord; it must remain accessible to the user at
all times to allow for quick disconnection if needed.
Connect and disconnect properly. Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Use only ins ulated voltage probes, test leads, and adapters supplie d with the
product, or indicated by Tektronix to be suitable for the product.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product. Do not exceed the
Measurement Category (CAT) rating and voltage or current rating of the lowest
rated individual component of a p roduct, probe, or a ccessory. Use caution w hen
using 1:1 test leads because the probe tip voltage is directly transmitted to the
product.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Do not oat the common terminal above the r ated voltage for that terminal.
Do not operate without covers. Do not operate this product with covers or panels
removed, or with the case open. Hazardous voltage exposure is possible.
Avoid exposed circuitry. Do not touch exposed connections and components
when power is present.
Do not operate with suspected failures. If you suspect that there is damage to this
product, have it inspected by qualied service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged
or operates incorrectly. If in doubt about safety of the product, turn it off and
disconnect the power cord. Clearly mark the product to prevent its further
operation.
Before use, inspect voltage probes, test leads, and accessories for mechanical
damage and replace when damaged. Do not use probes or test leads if they are
damaged, if there is exposed metal, or if a wear indicator shows.
Examine the exterior of the product before you use it. Look for cracks or missing
pieces.
Use only specied replacement parts.
Use proper fuse. Use only the fuse type and rating specied for this product.
MDO3000 Installation and Safety Instructions 5
Important safety information
Do not operate i
n wet/damp conditions. Be aware that condensation may occur if
a unit is moved from a cold to a warm environment.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry. Remove the input signals before you clean
the product.
Provide proper ventilation. Refer to the installation instructions in the manual for
details on installing the product so it has proper ventilation.
Slots a nd openings are provided for ventilation and should never be covered or
otherwise obstructed. Do not push objects into any of the openings.
Provide a safe working environment. Always place the product in a location
convenient for viewing the display and indicators.
Avoid improper or prolonged use of keyboards, pointers, and button pads.
Improper or prolonged keyboard or pointer use may result in serious injury.
Be sure your work a rea meets applicable ergonomic st andards. Consult with an
ergonomics professional to avoid stress injuries.
Use only the Tektronix rackmount hardware specied for this product.
Probes and te st leads
Before connecting probes or test leads, connect the p ower cord from the power
connector to a properly grounded power outlet.
Keep ngers behind the nger guards on the probes.
Remove all probes, test leads and accessories that are not in use.
Use only correct Measurement Category (CAT), voltage, temperature, altitude,
and amperage rated probes, test le ads, and adapters for any measurement.
Beware of high voltages. Understand the voltage ratings for the probe you are
using and do not exceed those ratings. Two ratings are important to know and
understand:
The maximum measurement voltage from the probe tip to the probe reference
lead.
The maximum oating voltage from the probe reference lead to earth ground
These two voltage ratings depend on the probe and your application. Refer to the
Specications section o f the manual for more information.
WARNING. To prevent electrical shock, do not exceed the m axi mum measurement
or maximum oating voltage for the oscilloscope input BNC connector, probe
tip, or probe reference lead.
6 MDO3000 Installation and Safety Instructions
Important safety information
Connect and dis
connect properly. Connect the probe output to the measurement
product b efore connecting the probe to the circuit under test. Connect the
probe r eference lead to the circuit under test before connecting the probe input.
Disconnect the probe input and the probe reference lead from the circuit under test
before disconnecting the probe from the measurement product.
Connect and
disconnect properly. De-energize the circuit under test before
connecting or disconnecting the current probe.
Connect th
e probe reference lead to earth ground only.
Do not connect a current probe to any wire that carries voltages above the current
probe vol
tage rating.
Inspect the probe and accessories. Before each use, inspect probe and accessories
for damage (cuts, tears, or defects in the probe body, accessories, or cable jacket).
Do not use if damaged.
Ground
-referenced oscilloscope use. Do not oat the reference lead of this probe
when using with ground-referenced oscilloscopes. The reference lead must be
connected to earth potential (0 V).
Servicesafetysummary
The Service safety summary section contains additional information required to
safely perform service on the product. Only qualied personnel should perform
service procedures. Read this Service safety summary and the General safety
summary before performing any service procedures.
To avoid electric shock. Do not touch exposed connections.
Do not service alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering rst aid and resuscitation is
present.
Disconnect power. To avoid electric shock, switch off the product power and
disconnect the power cord from the mains pow er before removing any covers or
panels, or opening the case for servicing.
Use care when servicing with power on. Dangerous voltages or currents may exist
in this product. Disconnect power, remove battery (if applicable), and disconnect
test leads before removing protective panels, soldering, or replacing components.
Verify safety after r epair. Always recheck ground continuity and mains dielectric
strength after performing a repair.
MDO3000 Installation and Safety Instructions 7
Important safety information
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Symbols and terms on the product
These ter
ms may appear on the product:
DANGER indicates an injury hazard immediately ac cessible as you read
the mark
ing.
WARNING indicates an injury hazard not immediately accessible as you
read th
e marking.
CAUTION indicates a hazard to property including the product.
When this symbol is marked on the product, be sure to consult the manual
to nd out the nature of the potential hazards and any actions which have to
betakentoavoidthem. (Thissymbolmayalsobeusedtorefertheuserto
ratings in the manual.)
The following symbol(s) may appear on the product:
8 MDO3000 Installation and Safety Instructions
Compliance information
Compliance in
formation
This section
lists the EMC (electromagnetic compliance), safety, and
environmental standards with which the instrument complies.
EMC compliance
EC Declaration of
Conformity EMC
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility.
Compliance was demonstrated to the following specications as listed in the
Ofcial Journal of the European Communiti
es:
EN 61326-1:2006, EN 61326-2-1:2006. EMC requirements for electrical equipment
for measurement, control, and laboratory use.
123
CISPR 11:2003. Radiated and conducted emissions, Group 1, Class A
IEC 61000-4-2:2001. Electrostatic discharge immunity
IEC 61000-4-3:2002. RF electromagnetic eld immunity
4
IEC 61000-4-4:2004. Electrical fast transient/burst immunity
IEC 61000-4-5:2001. Power line surge immunity
IEC 61000-4-6:2003. Conducted RF immunity
5
IEC 61000-4-11:2004. Voltage dips and interruptions immunity
6
EN 61000-3-2:2006. AC powe r line h armonic emissions
EN 61000-3-3:1995. Voltage changes, uctuations, and icker
European contact.
Tektronix UK, Ltd.
Western Pe ninsula
Western Road
Bracknell, RG12 1RF
United Kingdom
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic
interference.
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a
test object.
3
For compliance with the EMC standards listed here, high quality shielded interface cables should be used.
4
Oscilloscope: 3.0 division waveform displacement and 6.0 division increase in peak-to-peak noise. RF:
Residual spurious signals in the RF section can typically increase to –50 dBm when the instrument is subjected
to electromagnetic interference per the IEC 61000-4-3 test for frequencies up to 1 GHz, and to –35 dBm for
frequencies above 1 GHz.
5
Oscilloscope: 1.0 division waveform displacement and 2.0 division increase in peak-to-peak noise. RF:
Residual spurious signals in the RF section can typically increase to –85 dBm when the instrument is subjected
to electromagnetic interference per the IEC 61000-4-6 test.
MDO3000 Installation and Safety Instructions 9
Compliance information
6
Performance Cr
iterion C applied at the 70%/25 cycle Voltage-Dip and the 0%/250 cycle Voltage-Interruption test
levels (IEC 61000-4-11).
Australia / New Zealand
Declaration o
f
Conformity EMC
Complies with the EMC provision of the Radiocommunications Act p er the
following st
andard, in accordance with ACMA:
CISPR 11:2003. Radiated and Conducted Emissions, Group 1, Class A, in
accordance
with EN 61326-1:2006 and EN 61326-2-1:2006.
Australia / New Zealand contact.
Baker & McKenzie
Level 27, AMP Centre
50 Bridge Street
Sydney NSW 2000, Australia
Russian Federation
This product is approved by the Russian government to carry the GOST mark.
Safety compliance
This section lists the safety standards with which the product complies and other
safety compliance information.
EU declaration of
conformity l ow voltage
Compliance was demonstrated to the following specication a s listed in the
Ofcial Journal of the European Union:
Low Voltage Directive 2006/95/EC.
EN 61010-1. Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use Part 1: General Requirements.
EN 61010-2-030. Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use Part 2-030: Particular
requirements for testing and measuring circuits.
U.S. nationally recognized
testing laboratory listing
UL 61010-1. Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use Part 1: General Requirements.
UL 61010-2-030. Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use Part 2-030: Particular
requirements for testing and measuring circuits.
10 MDO3000 Installation and Safety Instructions
Compliance information
Canadian certication
CAN/CSA-C22.2
No. 61010-1. Safety Requirements for Electrical
Equipment for Measurement, Control, and Laboratory Use Part 1: General
Requirements.
CAN/CSA-C22.2 No. 61010-2-030. Safety Requirements for Electrical
Equipment for Measurement, Control, and Laboratory Use Part 2-030:
Particular requirements for testing and measuring circuits.
Additional compliances
IEC 61010-1. Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use Part 1: General Requirements.
IEC 61010-2-030. Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use Part 2-030: Particular
requirements for testing and measuring circuits.
Equipment type
Test and measuring equipment.
Safety class
Class1–groundedproduct.
Pollution degree
descriptions
A measure of the contaminants that could occur in the environment around
and within a product. Typically the internal environment inside a product is
considered to be the same as the external. Products should be used only in the
environment for which they are rated.
Pollution degree 1. No pollution or only dry, nonconductive pollution occurs.
Products in t his category are generally encapsulated, hermetically sealed, or
located in clean rooms.
Pollution degree 2. Normally only dry, nonconductive pollution occurs.
Occasionally a temporary conductivity that is caused by condensation must
be expected. This location is a typical ofce/home environment. Temporary
condensation occurs only when the product is out of service.
Pollution degree 3. Conductive pollution, or dry, nonconductive pollution
that becomes conductive due to condensation. These are sheltered locations
where neither temperature nor humidity is controlled. The area is protected
from direct sunshine, rain, or direct wind.
Pollution d egree 4. Pollution that generates persistent conductivity through
conductive dust, rain, or snow. Typical outdoor locations.
P
ollution degree rating
Pollution degree 2 (as dened in IEC 61010-1). Rated for indoor, dry location
use only.
IP rating
IP20 (as dened in IEC 60529).
MDO3000 Installation and Safety Instructions 11
Compliance information
Measurement and
overvoltage category
descriptions
Measurement te
rminals on this product may be rated for measuring mains voltages
from one or more of the following categories (see specic ratings marked on
the product and in the manual).
Category II. Circuits directly connected to the building wiring at utilization
points (socket outlets and similar points).
Category III. In the building wiring and distribution system.
Category IV
. At the source of the electrical supply to the building.
NOTE. Only mains power supply circuits have an overvoltage category rating.
Only measurement circuits have a measurement category rating. Other circuits
within the product do not have either rating.
Mains overvoltage
category rating
Overvoltage category II (as dened in IEC 61010-1).
12 MDO3000 Installation and Safety Instructions
/