AWG5200 Installation and Safety Manual
This document provides the AWG5200 safety and compliance information, powering the oscilloscope, and introduces the instrument
controls and connections.
Documentation
Review the following user documents before installing and using your instrument. These documents provide important operating
information.
Product documentation
The following table lists the primary product specific documentation available for your product. These and other user documents are
available for download from www.tek.com. Other information, such as demonstration guides, technical briefs, and application notes, can
also be found at www.tek.com.
Document Content
Installation and Safety Instructions Safety, compliance, and basic introductory information for hardware
products.
Help In-depth operating information for the product. Available from the
Help button in the product UI and as a downloadable PDF on
www.tek.com/downloads.
User Manual Basic operating information for the product.
Specifications and Performance Verification Technical Reference Instrument specifications and performance verification instructions
for testing instrument performance.
Programmer Manual Commands for remotely controlling the instrument.
Declassification and Security Instructions Information about the location of memory in the instrument.
Instructions for declassifying and sanitizing the instrument.
Service Manual Replaceable parts list, theory of operations, and repair and replace
procedures for servicing an instrument.
Rackmount Kit Instructions Installation information for assembling and mounting an instrument
using a specific rackmount.
How to find your product documentation and software
1. Go to www.tek.com.
2. Click Download in the green sidebar on the right side of the screen.
3. Select Manuals or Software as the Download Type, enter your product model, and click Search.
4. View and download your product files. You can also click the Product Support Center and Learning Center links on the page for
more documentation.
Important safety information
This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe
condition.
To safely perform service on this product, see the Service safety summary that follows the General safety summary.
Copyright © Tektronix
tek.com
*P071381000*
071-3810-00
General safety summary
Use the product only as specified. Review the following safety precautions to avoid injury and prevent damage to this product or any
products connected to it. Carefully read all instructions. Retain these instructions for future reference.
This product shall be used in accordance with local and national codes.
For correct and safe operation of the product, it is essential that you follow generally accepted safety procedures in addition to the safety
precautions specified in this manual.
The product is designed to be used by trained personnel only.
Only qualified personnel who are aware of the hazards involved should remove the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it is operating correctly.
This product is not intended for detection of hazardous voltages.
Use personal protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals
for warnings and cautions related to operating the system.
When incorporating this equipment into a system, the safety of that system is the responsibility of the assembler of the system.
To avoid fire or personal injury
Use proper power cord.
Use only the power cord specified for this product and certified for the country of use.
Ground the product.
This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly
grounded. Do not disable the power cord grounding connection.
Power disconnect.
The power cord disconnects the product from the power source. See instructions for the location. Do not position the equipment so that it is
difficult to operate the power cord; it must remain accessible to the user at all times to allow for quick disconnection if needed.
Observe all terminal ratings.
To avoid fire or shock hazard, observe all rating and markings on the product. Consult the product manual for further ratings information
before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do not operate without covers.
Do not operate this product with covers or panels removed, or with the case open. Hazardous voltage exposure is possible.
Avoid exposed circuitry.
Do not touch exposed connections and components when power is present.
Do not operate with suspected failures.
If you suspect that there is damage to this product, have it inspected by qualified service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or operates incorrectly. If in doubt about safety of the product,
turn it off and disconnect the power cord. Clearly mark the product to prevent its further operation.
Important safety information
2
Examine the exterior of the product before you use it. Look for cracks or missing pieces.
Use only specified replacement parts.
Do not operate in wet/damp conditions.
Be aware that condensation may occur if a unit is moved from a cold to a warm environment.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Remove the input signals before you clean the product.
Provide proper ventilation.
Refer to the installation instructions in the manual for details on installing the product so it has proper ventilation.
Slots and openings are provided for ventilation and should never be covered or otherwise obstructed. Do not push objects into any of the
openings.
Provide a safe working environment
Always place the product in a location convenient for viewing the display and indicators.
Avoid improper or prolonged use of keyboards, pointers, and button pads. Improper or prolonged keyboard or pointer use may result in
serious injury.
Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics professional to avoid stress injuries.
Use care when lifting and carrying the product. This product is provided with a handle or handles for lifting and carrying.
WARNING: The product is heavy. To reduce the risk of personal injury or damage to the device get help when lifting or carrying
the product.
WARNING: The product is heavy. Use a two-person lift or a mechanical aid.
Use only the Tektronix rackmount hardware specified for this product.
Terms in this manual
These terms may appear in this manual:
WARNING: Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION: Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
Terms in this manual
3
Symbols on the product
When this symbol is marked on the product, be sure to consult the manual to find out the nature of the potential
hazards and any actions which have to be taken to avoid them. (This symbol may also be used to refer the user to
ratings in the manual.)
The following symbols(s) may appear on the product.
Compliance information
This section lists the safety and environmental standards with which the instrument complies. This product is intended for use by
professionals and trained personnel only; it is not designed for use in households or by children.
Compliance questions may be directed to the following address:
Tektronix, Inc.
PO Box 500, MS 19-045
Beaverton, OR 97077, USA
tek.com
Safety compliance
This section lists the safety compliance information.
Equipment type
Test and measuring equipment.
Safety class
Class 1 – grounded product.
Pollution degree description
A measure of the contaminants that could occur in the environment around and within a product. Typically the internal environment inside a
product is considered to be the same as the external. Products should be used only in the environment for which they are rated.
Pollution Degree 1. No pollution or only dry, nonconductive pollution occurs. Products in this category are generally encapsulated,
hermetically sealed, or located in clean rooms.
Pollution Degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is caused by
condensation must be expected. This location is a typical office/home environment. Temporary condensation occurs only when the
product is out of service.
Pollution Degree 3. Conductive pollution, or dry, nonconductive pollution that becomes conductive due to condensation. These are
sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct sunshine, rain, or direct
wind.
Pollution Degree 4. Pollution that generates persistent conductivity through conductive dust, rain, or snow. Typical outdoor locations.
Symbols on the product
4
Pollution degree rating
Pollution Degree 2 (as defined in IEC 61010-1). Note: Rated for indoor, dry location use only.
IP rating
IP20 (as defined in IEC 60529).
Measurement and overvoltage category descriptions
Measurement terminals on this product may be rated for measuring mains voltages from one or more of the following categories (see
specific ratings marked on the product and in the manual).
Measurement Category II. For measurements performed on circuits directly connected to the low-voltage installation.
Measurement Category III. For measurements performed in the building installation.
Measurement Category IV. For measurements performed at the source of low-voltage installation.
Note: Only mains power supply circuits have an overvoltage category rating. Only measurement circuits have a measurement
category rating. Other circuits within the product do not have either rating.
Mains overvoltage category rating
Overvoltage Category II (as defined in IEC 61010-1)
Environmental compliance
This section provides information about the environmental impact of the product.
Product end-of-life handling
Observe the following guidelines when recycling an instrument or component:
Equipment recycling Production of this equipment required the extraction and use of natural resources. The equipment may
contain substances that could be harmful to the environment or human health if improperly handled at the
product’s end of life. To avoid release of such substances into the environment and to reduce the use of
natural resources, we encourage you to recycle this product in an appropriate system that will ensure that
most of the materials are reused or recycled appropriately.
This symbol indicates that this product complies with the applicable European Union requirements according to Directives
2012/19/EU and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For information about
recycling options, check the Tektronix Web site (www.tek.com/productrecycling).
Perchlorate materials This product contains one or more type CR lithium batteries. According to the state of California, CR
lithium batteries are classified as perchlorate materials and require special handling. See www.dtsc.ca.gov/
hazardouswaste/perchlorate for additional information.
Operating requirements
Place the instrument on a cart or bench, observing clearance requirements:
Top and bottom: 0 cm (0 in)
Left and right side: 5.08 cm (2 in)
Rear: 0 cm (0 in)
CAUTION: To ensure proper cooling, keep sides of the instrument clear of obstructions.
Operating requirements
5
Power supply requirements
The power supply requirements for your instrument are listed in the following table.
WARNING: To reduce the risk of fire and shock, ensure that the mains supply voltage fluctuations do not exceed 10% of the
operating voltage range.
Source Voltage and Frequency Power Consumption
100 VAC to 240 VAC, 50/60 Hz 750 W
Environmental requirements
The environmental requirements for your instrument are listed in the following table. For instrument accuracy, ensure that the instrument
has warmed up for 20 minutes and meets the environmental requirements listed in the following table.
Requirement Description
Temperature (operating) 0 °C to 50 °C (+32 °F to +122 °F)
Humidity (operating) 5% to 90% relative humidity at up to 30 °C (86 °F)
5% to 45% relative humidity above 30 °C (86 °F) up to +50 °C (122
°F) noncondensing
Altitude (operating) Up to 3,000 m (9,843 feet)
Install the instrument
Unpack the instrument and check that you received all items listed as Standard Accessories. Check the Tektronix Web site
(www.tektronix.com) for the most current information.
Power on the instrument
Procedure
1. Connect the AC power cord to the rear of the instrument.
2. Use the front-panel power button to switch the instrument on.
Install the instrument
6
The power button indicates four instrument power states:
No light – no power applied
Yellow – standby mode
Green – powered on
Flashing Red – over heat condition (instrument shuts down and cannot restart until internal temperature returns to a safe level)
Power off the instrument
Procedure
1. Press the front-panel power button to shut down the instrument.
The shutdown process takes approximately 30 seconds to complete, placing the instrument in standby mode. Alternatively, use the
Windows Shutdown menu.
Note: You can force an immediate shutdown by pressing and holding the power button for four seconds. Unsaved data is lost.
2. To completely remove power to the instrument, perform the shutdown just described, and then remove the power cord from the
instrument.
Install the instrument
7
Connecting to the instrument
Connecting to a network
You can connect your instrument to a network for file sharing, printing, Internet access, and other functions. Consult your network
administrator and use the standard Windows utilities to configure the instrument for your network.
Connecting peripheral devices
You can connect peripheral devices to your instrument, such as a keyboard and mouse (provided). A mouse and keyboard can substitute
for the touchscreen and are particularly helpful for opening and saving files.
Controlling the instrument using a remote PC
Use your PC to control the arbitrary waveform generator through a LAN using the Windows Remote Desktop function. If your PC has a
larger screen, it will be easier to see details such as zooming waveforms or making cursor measurements. You can also use a third party
software application (installed on your PC) to create a waveform and import it through a network.
Preventing instrument damage
Overheat protection
The instrument is protected against overheating damage by continuously monitoring the internal temperature. If the internal temperature
exceeds the maximum rated operating range, two actions occur.
The instrument shuts down.
The Power button flashes red.
Note: An indication that the internal temperature is escalating is continual calibration warnings due to temperature change.
If an overheat condition was detected, the power button will continue to flash red, even after the instrument cools (unless power is
disconnected). This is done to indicate that an overheat condition has occurred, regardless of how much time has passed.
Restarting the instrument (or removing and reapplying power) will stop the power button from flashing red. But if the overheat condition still
remains while attempting to restart the instrument, the power button may immediately (or in a short time) start flashing red again and the
instrument will shut down.
Common causes of overheating include:
The ambient temperature requirement is not being met.
The required cooling clearance is not being met.
One or more instrument fans are not working properly.
Connectors
The arbitrary waveform generator has both output and input connectors. Do not apply external voltage to any output connector and ensure
proper restrictions are met for any input connector.
CAUTION: Always turn off the signal outputs when you connect or disconnect cables to/from the signal output connectors. If
you connect a (Device Under Test) DUT while the instrument signal outputs are in the On state, it may cause damage to the
instrument or to the DUT.
Install the instrument
8
External device connections
For many applications, powered external devices may need to be used on the output of the AWG. These may include Bias-Ts, Amplifiers,
transformers etc. It is important to guarantee that these components are adaptable for the specific AWG and that they are configured as
required by the device manufacturer.
Note: The term Device means external powered devices such as bias-t, whereas Device Under Test (DUT) refers to the circuit
being tested.
It is critical that there is minimal inductive kickback into the AWG output when the device is connected or disconnected. Inductive kickback
can occur if the external device can hold a charge and then discharge when a ground path becomes available such a connection to the
output termination of the AWG channel output. To minimize this inductive kickback care should be taken before connecting the device to
the AWG output.
Some simple guidelines to follow for device connection are:
1. Always use a grounded wrist strap when connecting cables.
2. Make sure the power supply to the device is turned off or unplugged.
3. Establish ground connection between the device and AWG test system.
4. Make sure the DUT's power supply is turned off or set at 0 volts.
5. Discharge cables to ground before connecting to the AWG.
6. Engage connector between device and AWG output.
7. Power up device power supply.
8. Set device voltage power supply (bias level voltage for bias-t) to desired voltage.
9. Power up DUT power supply.
Enhancements for your instrument
Upgrades and plug-ins purchased with your instrument are pre-installed. You can view these by going to Utilities > About my AWG. If you
purchase an upgrade or plug-in after you've received your instrument, you may need to install a license key to activate the feature. Use
the Install Licenses dialog box to enable the upgrades that you purchased from Tektronix for your instrument. For the most current list of
upgrades, go to www.tektronix.com or contact your local Tektronix representative.
Your instrument can be enhanced by several different methods:
Software enhancements: Enhancements ordered at the time of your purchase are pre-installed. These can also be purchased post
sales and may require the installation of software in addition to installing a license to activate.
Hardware enhancements: Features that require/enable hardware on the instrument. These can be ordered with the purchase of the
instrument or as a post-purchase addition.
Plug-ins: Applications that enhance a host application. Plug-ins designed to operate with an AWG5200 series instrument are also able
to operate with the SourceXpress Waveform Creation software. Plug-ins with a floating license can be moved between instruments or
SourceXpress.
Introduction to the instrument
Connectors and controls are identified and described in the following images and text.
Introduction to the instrument
9
Front-panel connectors
Table 1: Front-panel connectors
Connector Description
Analog outputs (+ and –)
AWG5202 – Two channels
AWG5204 – Four channels
AWG5208 – Eight channels
These SMA type connectors supply the complimentary (+) and (-) analog output signals.
The channel LEDs light to indicate when the channel is enabled and the output is electrically
connected. The LED color matches the user defined waveform color.
The channel (+) and (-) connectors are electrically disconnected when the All Outputs Off
control is activated.
AC outputs (+) The (+) connector of each channel can supply a single-ended analog signal when an AC output
mode is activated for the channel. The AC output provides for additional amplification and
attenuation of the output signal.
The (-) connector of the channel is electrically disconnected. For best EMI reduction, install a
50 Ω termination to the (-) connector when using the AC output mode.
USB Two USB2 connectors.
Removable hard disk drive (HDD) The HDD contains the operating system, product software and all user data. By removing the
HDD, user information such as setup files and waveform data is removed from the instrument.
Chassis ground Banana type ground connection.
CAUTION: Always turn off the signal outputs when you connect or disconnect cables to/from the signal output connectors. Use
the All Outputs Off button (either the front-panel button or the screen button) to quickly disable the Analog and Marker outputs.
(Marker outputs are located on the rear panel.) When the All Outputs Off is enabled, the output connectors are electrically
disconnected from the instrument.
Do not connect a DUT to the front-panel signal output connectors when the instrument signal outputs are on.
Do not power on or off the DUT when the generator signal outputs are on.
Introduction to the instrument
10
Front-panel controls
The following illustration and table describe the front panel controls.
Buttons/Keys Description
Play/Stop The Play/Stop button starts or stops playing the waveform.
The Play/Stop button displays the following lights:
No light - no waveform playing
Green - playing a waveform
Flashing green - preparing to play a waveform
Amber - play out temporarily inhibited due to a settings change
Red - Error preventing play out
When a waveform is playing, it is only present at the output connectors if the following conditions are met:
The channel is enabled.
The All Outputs Off is not active (outputs are connected).
General purpose knob The general purpose knob is used to increment or decrement values when a setting is enabled (selected)
for change.
Note: The general purpose knob operation mimics the actions of the up and down arrow keys on
a keyboard as defined by the Windows operating system. Because of this, rotating the knob when
a desired control is not selected may result in seemingly odd behavior of the control or accidental
changes to some other control.
Numeric keypad The numeric keypad is used to directly enter a numeric value into a selected control setting.
Units prefix buttons (T/p, G/n, M/μ, and k/m) are used to complete an input with the numeric keypad. You
can complete your entry by pushing one of these prefix buttons (without pressing the Enter key).
If you push the units prefix buttons for frequency, the units are interpreted as T (tera-), G (giga-), M (mega-),
or k (kilo-). If you push the buttons for time or amplitude, the units are interpreted as p (pico-), n (nano-), μ
(micro-), or m (milli-).
Left and Right Arrow
buttons
Use the arrow buttons to change (select) the focus of the cursor in the Frequency control box when and
IQ waveform is assigned to the channel. The Digital Up Converter (DIGUP) must be licensed to assign IQ
waveforms to a channel..
Force Trigger (A or B) The A or B Force Trigger buttons generate a trigger event. This is only effective when the Run mode is set
to Triggered or Triggered Continuous.
Table continued…
Introduction to the instrument
11
Buttons/Keys Description
All Outputs Off The All Outputs Off button provides a quick disconnect of the Analog, Marker, and Flag outputs, whether
those outputs are enabled or not. (All Outputs Off overrides the channel output enable controls.)
When activated, the button lights, the outputs are electrically disconnected, and the channel output
front-panel lights are turned off.
When the All Outputs Off is deactivated, the outputs return to their previously defined state.
Rear-panel connectors
Table 2: Rear-panel connectors
Connector Description
Aux Outputs
AWG5202 - Four
AWG5204 - Four
AWG5208 - Eight
SMB connectors to supply output flags to mark the state of sequences.
These outputs are not affected by the All Outputs Off state.
Chassis ground Banana type ground connection.
Trigger Inputs A and B SMA type input connectors for external trigger signals.
Streaming ID RJ-45 connector for future enhancement.
Sync Clock Out SMA type output connector used to synchronize the outputs of multiple AWG5200 series generators.
This output is not affected by the All Outputs Off state.
Sync to Hub Connector for future enhancement.
eSATA eSATA port to connect external SATA devices to the instrument.
Pattern Jump In 15-pin DSUB connector to provide a pattern jump event for Sequencing. (Requires SEQ license.)
Table continued…
Introduction to the instrument
12
Connector Description
VGA VGA video port to connect an external monitor to view a larger copy of the instrument display (duplicate) or
to extend the desktop display. To connect a DVI monitor to the VGA connector, use a DVI-to-VGA adapter.
USB Device USB Device connector (type B) interfaces with the TEK-USB-488 GPIB to USB adapter and provides
connectivity with GPIB based control systems.
USB Host Four USB3 Host connectors (type A) to connect devices such as a mouse, keyboard, or other USB devices.
Tektronix does not provide support or device drivers for USB devices other than the optional mouse and
keyboard.
LAN RJ-45 connector to connect the instrument to a network.
Power Power cord input.
Marker outputs SMA type output connectors for marker signals. Four per channel.
These outputs are affected by the All Outputs Off state.
Sync In SMA type connector to use a synchronization signal from another AWG5200 series instrument.
Sync Out Connector for future enhancement.
Clock Out SMA type connector to provide a high speed clock that is related to the sample rate.
This output is not affected by the All Outputs Off state.
Clock In SMA type connector to provide an external clock signal.
Ref In SMA type input connector to provide a reference timing signal (variable or fixed).
10 MHz Ref Out SMA type output connector to provide a 10 MHz reference timing signal.
This output is not affected by the All Outputs Off state.
Cleaning the instrument
Inspect the arbitrary waveform generator as often as operating conditions require. Follow these steps to clean the exterior surface.
WARNING: To avoid personal injury, power off the instrument and disconnect it from line voltage before performing any of the
following procedures.
CAUTION: To avoid damage to the surface of the instrument, do not use any abrasive or chemical cleaning agents.
Use extreme care when cleaning the surface of the display. The display is easily scratched if excessive force is used.
Procedure
1. Remove loose dust on the outside the instrument with a lint-free cloth. Use care to avoid scratching the front-panel display.
2. Use a soft cloth dampened with water to clean the instrument. If needed, use a 75% isopropyl alcohol solution as a cleaner. Do not
spay liquids directly on the instrument.
Cleaning the instrument
13
AWG5200 シリーズ 任意波形ジェネレータインストールおよび安全
に関する取扱説明書
このマニュアルでは、AWG5200 シリーズ機器の設置と基本的な操作方法について説明します。詳細な操作方法
については、ご使用の機器に搭載されているヘルプを参照してください。このマニュアルは次の機器を対象として
います。
AWG5202 型任意波形ゼネレータ(2チャンネル)
AWG5204 型任意波形ゼネレータ(4チャンネル)
AWG5208 型任意波形ゼネレータ(8チャンネル)
ニュアル
本機に関する各種情報の参照先は以下のとおりです。
AWG5200 シリーズのインストールおよび安全に関する取扱説明書 14
参照する項目 参照するマニュアル
インストールと操作(概要) インストールおよび安全に関するマニュアル(本書)
操作方法およびユーザ・インタフェース 本機のヘルプ機能(Help(ヘルプ)メニューから呼び出
し)
ルプ・システムの PDF 版も当社 Web サイトでご利用
になれます。当社部品番号 077-1446-xx
プログラマ・コマンド プログラマ・マニュアル。当社部品番号 077-1452-xx
仕様および性能検査手順 仕様および性能検査のテクニカル・リファレンス。当社
部品番号 077-1453-xx
安全性に関する重要な情報
このマニュアルには、操作を行うユーザの安全を確保し、製品を安全な状態に保つために順守しなければならな
い情報および警告が記載されています。
安全にご使用いただくために
製品は指定された方法でのみご使用ください。人体への損傷を避け、本製品や本製品に接続されている製品の
破損を防止するために、安全性に関する次の注意事項をよくお読みください。すべての指示事項を注意深くお読
みください。必要なときに参照できるように、説明書を安全な場所に保管しておいてください。
該当する地域および国の安全基準に従ってご使用ください。
本製品を正しく安全にご使用になるには、このマニュアルに記載された注意事項に従うだけでなく、一般に認めら
れている安全対策を徹底しておく必要があります。
本製品は訓練を受けた専門知識のあるユーザによる使用を想定しています。
製品のカバーを取り外して修理や保守、または調整を実施できるのは、あらゆる危険性を認識した専門的知識の
ある適格者のみに限定する必要があります。
使用前に、既知の情報源と十分に照らし合わせて、製品が正しく動作していることを常にチェックしてください。
本製品は危険電圧の検出用にはご利用になれません。
危険な通電導体が露出している部分では、感電やアーク・フラッシュによってけがをするおそれがありますので、
保護具を使用してください。
本製品をご使用の際に、より大きな他のシステムにアクセスしなければならない場合があります。他のシステム
の操作に関する警告や注意事項については、その製品コンポーネントのマニュアルにある安全に関するセクショ
ンをお読みください。
本機器をシステムの一部としてご使用になる場合には、そのシステムの構築者が安全性に関する責任を果たさ
なければなりません。
火災や人体への損傷を避けるには
適切な電源コードを使用してください: 本製品用に指定され、使用される国で認定された電源コードのみを使用し
てください。
本製品を接地してください: 本製品は、電源コードのグランド線を使用して接地します。感電を避けるため、グラン
ド線をアースに接続する必要があります。本製品の入出力端子に接続する前に、製品が正しく接地されているこ
とを確認してください。電源コードのグランド接続を無効にしないでください。
電源を切断してください: 電源コードの取り外しによって主電源が遮断されます。スイッチの位置については、使
用説明書を参照してください。電源コードの取り扱いが困難な場所には設置しないでください。必要に応じてすぐ
に電源を遮断できるように、ユーザが常にアクセスできる状態にしておく必要があります。
すべての端子の定格に従ってください: 発火や感電の危険を避けるために、本製品のすべての定格とマーキン
グに従ってください。本製品に電源を接続する前に、定格の詳細について、製品マニュアルを参照してください。
コモン端子を含むいかなる端子にも、その端子の最大定格を超える電圧をかけないでください。
カバーを外した状態では使用しないでください: カバーやパネルを外した状態やケースを開いたまま動作させない
でください。危険性の高い電圧に接触してしまう可能性があります。
露出した回路への接触は避けてください: 電源が投入されているときに、露出した接続部分やコンポーネントに触
れないでください。
AWG5200 シリーズのインストールおよび安全に関する取扱説明書 15
故障の疑いがあるときは使用しないでください: 本製品に故障の疑いがある場合には、資格のあるサービス担当
者に検査を依頼してください。
製品が故障している場合には、使用を停止してください。製品が故障している場合や正常に動作していない場合
には、製品を使用しないでください。安全上の問題が疑われる場合には、電源を切って電源コードを取り外してく
ださい。誤って使用されることがないように、問題のある製品を区別しておいてください。
使用する前に、製品の外観に変化がないかよく注意してください。ひび割れや欠落した部品がないことを確認して
ください。
指定された交換部品のみを使用するようにしてください。
湿気の多いところでは動作させないでください: 機器を寒い場所から暖かい場所に移動する際には、結露にご注
意ください。
爆発性のガスがある場所では使用しないでください:
製品の表面を清潔で乾燥した状態に保ってください: 製品の清掃を開始する前に、入力信号を取り外してくださ
い。
適切に通気してください: 適切な通気が得られるように製品を設置できるように、マニュアルの設置手順を参照し
てください。
製品には通気用のスロットや開口部があります。その部分を覆ったり、通気が妨げられたりすることがないように
してください。開口部には異物を入れないでください。
安全な作業環境を確保してください: 製品は常にディスプレイやインジケータがよく見える場所に設置してくださ
い。
キーボードやポインタ、ボタン・パッドを不適切に使用したり、長く押しすぎたりしないでください。キーボードやポイ
ンタの使用を誤ると、大けがにつながる可能性があります。
作業場が該当する人間工学規格を満たしていることを確認してください。ストレスに由来するけががないように、
人間工学の専門家に助言を求めてください。
製品を持ち上げたり運んだりする作業は慎重に行ってください。本製品には持ち運び用のハンドルが取り付けら
れています
本製品には指定された当社のラック取り付け金具のみを使用してください。
安全性に関する重要な情報
16 AWG5200 シリーズのインストールおよび安全に関する取扱説明書
警告:
本機はかなりの重量があります。怪我をしたり装置が損傷することがないように、製品を持ち運ぶときには
誰かの手を借りてください。
本マニュアル内の用語
本マニュアルでは以下の用語を使用しています。
警告:
人体や生命に危害をおよぼすおそれのある状態や行為を示します。
注意:
本機やその他の接続機器に損害を与えるおそれのある状態や行為を示します。
本機に関する用語
本機では次の用語を使用します。
危険: ただちに人体や生命に危険をおよぼす可能性があることを示します。
警告: 人体や生命に危険をおよぼす可能性があることを示します。
注意: 本製品を含む周辺機器に損傷を与える可能性があることを示します。
本製品に関する記号と用語:
製品にこの記号が表記されているときは、マニュアルを参照して、想定される危険性とそれらを回避す
るために必要な行動について確認してください(マニュアルでは、この記号はユーザに定格を示すため
に使用される場合があります。)
本製品では、次の記号を使用します。
安全性に関する重要な情報
AWG5200 シリーズのインストールおよび安全に関する取扱説明書 17
適合性に関する情報
このセクションでは、本製品が適合している EMC 基準、安全基準、および環境基準について説明します。
家庭での使用や子供による使用に対応して設計されていません。
以下の適合性に関するご質問は、以下の住所宛に、直接お問い合わせいただくこともできます:
Tektronix, Inc.PO Box 500, MS 19045
Beaverton, OR 97077, USA
www.tek.com
安全性
このセクションでは、製品が適合している安全規格およびその他の基準について説明します。
汚染度について
製品内部およびその周辺で発生する可能性がある汚染度の尺度です。通常、製品の内部環境は外部環境と同
じ規定が適用されるものとみなされます。製品は、その製品に指定されている環境でのみ使用してください。
汚染度 1:汚染なし、または乾燥した非伝導性の汚染のみが発生します。このカテゴリの製品は、通常、被包
性、密封性のあるものか、クリーン・ルームでの使用を想定したものです。
汚染度 2:通常、乾燥した非導電性の汚染のみが発生します。ただし、結露によって一時的な導電性が発生
することもまれにあります。これは、標準的なオフィスや家庭内の環境に相当します。一時的な結露は製品非
動作時のみ発生します。
汚染度 3:伝導性のある汚染、または結露のために伝導性のある汚染となる乾燥した非伝導性の汚染。これ
らは、温度、湿度のいずれも管理されていない屋内環境に相当します。日光や雨、風に対する直接の曝露か
らは保護されている領域です。
汚染度 4:伝導性のある塵、雨、または雪により持続的に伝導性が生じている汚染。これは一般的な屋外環
境に相当します。
汚染度
汚染度 2IEC 61010-1 の定義による)注:乾燥した屋内でのみ使用できます。
IP 定格
IP20IEC 60529 で定義)。
測定および過電圧カテゴリについて
本製品の測定端子は、測定する電源電圧について次の 1つまたは複数のカテゴリに評価されます(製品やマニ
ュアルへの特定の評価を参照)。
測定カテゴリ II:低電圧インストレーションに直接接続された回路で実施する測定用。
測定カテゴリ III:建築物の屋内配線で実施する測定
測定カテゴリ IV:低電圧電源を使用して実施する測定
NOTE.
過電圧カテゴリ定格に該当するのは主電源回路のみです。測定カテゴリ定格に該当するのは測定回路の
みです。製品内部のその他の回路にはいずれの定格も該当しません。
主電源過電圧カテゴリ定格
過電圧カテゴリ II(IEC 61010-1 の定義による)
環境基準に対する適合性
このセクションでは本製品が環境におよぼす影響について説明します。
使用済み製品の処理方法
機器またはコンポーネントをリサイクルする際には、次のガイドラインを順守してください。
機器のリサイクル. 本製品の製造には天然資源が使用されています。この製品には、環境または人体に有害とな
る可能性のある物質が含まれているため、製品を廃棄する際には適切に処理する必要があります。有害物質の
放出を防ぎ、天然資源の使用を減らすため、本製品の部材の再利用とリサイクルの徹底にご協力ください。
このマークは、本製品が WEEE(廃棄電気・電子機器)およびバッテリに関する指令 2012/19/EC およ
2006/66/EC に基づき、EU の諸要件に準拠していることを示しています。リサイクル方法について
は、当社の Web サイトのサービス・セクション(www.tek.com/productrecycling)を参照してください。
過塩素酸塩の取り扱い. 本製品には CR リチウム電池が搭載されていますCR リチウム電池はカリフォルニア州
法により過塩素酸塩材として規定され、特別な取り扱いが求められています。詳細についてはwww.dtsc.ca.gov/
hazardouswaste/perchlorate を参照してください。
適合性に関する情報
AWG5200 シリーズのインストールおよび安全に関する取扱説明書 18
環境要件
本機の環境要件を以下の表に示します。保証確度が得られるよう、本機を使用する前に 20 分間ウォーム・アップ
し、以下の表の環境要件が満たされていることを確認してください。
要件 説明
温度(動作時) 0~+ 50(+ 32F~+ 122F
湿度(動作時) 30以下で相対湿度 5%90%86F
3086F)超~50122F)で相対湿度 5%45%(結露
のない状態)
高度(動作時) 最高 3,000m9,843 フィート)
動作要件
次の設置条件を満たす空間を確保して、カートまたはベンチに設置します。
上部および底部:0cm0インチ)
左側および右側:5.08cm2インチ)
後部:0cm0インチ)
注意:
排気が確実に行われるように、機器の両側に障害物を置かないでください。
電源要件
本機の電源要件を以下の表に示します。
警告:
出火および感電のリスクを減らすため、主電源の電圧変動が動作電圧範囲の
10%
を超えていないことを
確認してください。
電源電圧と周波数 消費電力
100VAC240VAC50Hz/60Hz 750W
クリーニング
任意波形ゼネレータは使用状況に応じた頻度で点検してください。筐体表面のクリーニング手順は次のとおりで
す。
警告:
人体への損傷を避けるために、以下の手順を実行する前に、本機の電源をオフにして電源コードを取り外
してください。
注意:
研磨剤や化学洗剤は使用しないでください。本機の表面が損傷する可能性があります。
ディスプレイ表面のクリーニングには細心の注意が必要です。過重な力が加わると簡単に引っかき傷が付いてし
まいます。
1. 無塵布で機器の表面についた塵を落とします。前面パネルのディスプレイを傷つけないように注意してくださ
い。
2. 水で湿らせた柔らかい布を使用して機器を拭きます。必要であれば、75%イソプロピル・アルコール溶液を使
用してください。液体を本機に直接噴霧するのは避けてください
本機の設置
AWG5200 シリーズのインストールおよび安全に関する取扱説明書 19
機器の電源の投入
1. 本機リアパネルに AC 電源コードを接続します。
2. 前面パネルの電源ボタンを押して、本機の電源をオンにします。
電源ボタンは次の 4通りの電源状態を表します。
消灯:電源オフ
黄色:スタンバイ・モード
緑:電源オン
赤色(点滅):過熱状態(本機はシャットダウンし、内部温度が安全なレベルに下がるまで再起動すること
はできません)
機器の電源の切断
1. 前面パネルの電源ボタンを押して本機をシャットダウンします。
シャットダウン・プロセスが完了し、本機がスタンバイ・モードに移行するまでに約 30 秒かかります。Windows
Shutdown メニューを使用してシャットダウンすることもできます。
注:
本機を即時にシャットダウンするには、電源ボタンを
4
秒間長押ししてください。この場合、保存されてい
ないデータは失われます。
2. 機器の電源を完全にオフにするには、シャットダウン実行後、電源コードを取り外します。
本機の検査
本機には、機能確認用の診断ルーチンが次の 2種類用意されています。
パワーオン・セルフテスト(POST - 本機は電源投入時に毎回セルフテストを実行します。
System(システム)メニューの診断機能 - System(システム)メニューから内部診断機能を実行することができま
す。手順は次のとおりです。
1. ワークスペース・タブで Utilities(ユーティリティ)を選択し、続いて System(システム)を選択します。
2. Diagnostics & Calibration(診断と校正)をクリックします。
3. Diagnostics and Calibration(診断と校正)画面で Diagnostics(診断)をクリックします。
4. 実行する診断項目セットを選択します。
POST OnlyPOST のみ):パワーオン時に POST として自動的に実行されるテストのセットです。これらの
テストでは、内部のデバイス通信、システム・メモリ、リアルタイム・クロックなどの項目が診断されます。
本機の設置
AWG5200 シリーズのインストールおよび安全に関する取扱説明書 20
/