APPA 170 Series ユーザーマニュアル

タイプ
ユーザーマニュアル
170 series
User Manual / 使用說明書 / 使用说明书
ユーザーマニュアル
Руководство пользователя
Clamp Meter
電流鉗表
电流钳表
クランプメーター
Клещи электроизмерительные
RU
EN
TC
SC
JP
LIMITE D
WARRAN TY
YEAR S
170 series EN
Read First
Safety Information
WARNING
Understand and follow operating instructions carefully. Use
the meter only as.
• If the equipment is used in a manner not specied by the
manufacturer, the protection provided by the equipment
may be impaired.
Always use proper terminals, switch position, and range for
measurements.
• To reduce the risk of re or electric shock, do not use this
product around explosive gas or in damp locations.
• Verify the Meter operation by measuring a known voltage.
If in doubt, have the Meter serviced.
• Do not apply more than the rated voltage, as marked on
Meter, between terminals or between any terminal and
earth ground.
• To avoid false readings that can lead to electric shock and
injury, replace the battery as soon as low battery indicator
blinks.
Avoid working alone so assistance can be rendered.
• Do not use the Tester if the Tester is not operating properly
or if it is wet.
• Individual protective device must be used if hazardous live
parts in the installation where the measurement is to be
carried out could be accessible.
• Disconnect the test leads from the test points before
changing the position of the function rotary switch.
• Never connect a source of voltage when the function rotary
switch is not in voltage position.
• When using test leads or probes, keep your ngers behind
the nger guards.
• Use caution with voltages above 30 Vac rms, 42 Vac peak,
or 60 Vdc. These voltages pose a shock hazard.
• Remove test lead from Meter before opening the battery
door or Meter case.
• DO NOT USE the test leads when the internal white
insulation layer is exposed.
• DO NOT USE the test leads above maximum ratings of
CAT. environment, voltage and current, that are indicated
on the probe and the probe tip guard cap.
• DO NOT USE the test leads without the probe tip guard
1
170 series EN
cap in CAT III and CAT IV environments.
• Probe assemblies to be used for MAINS measurements
shall be RATED as appropriate for MEASUREMENT
CATEGORY III or IV according to IEC 61010-031 and shall
have a voltage RATING of at least the voltage of the circuit
to be measured.
• Disconnect circuit power and discharge all high-voltage
capacitors before testing resistance, continuity, diodes, or
capacitance.
• De-energize the installation under test or wear suitable
protective clothing during tting and removal of the Flexible
Current Probe.
• Do not apply around or remove from UNINSULATED
HAZARDOUS LIVE conductors, which may render electric
shock, electric burn, or arc ash.
• Do not apply a current with frequency that is higher than
the frequency response range specied in the Electrical
specication section to APPA 172 .
Symbols as marked on the Meter
and Instruction card
Risk of electric shock
See instruction manual
DC measurement
AC measurement
Bluetooth
H Both direct and alternating current
Equipment protected by double or reinforced insulation
Low battery
Earth
Conforms to EU directives
Application around and removal from hazardous live conductors
is permitted
Do not discard this product or throw away.
Unsafe Voltage
To alert you to the presence of a potentially hazardous
voltage, when the Tester detects a voltage ≥ 30 V or a
voltage overload (OL) in V, mV, LoZ, PV. The symbol is
displayed.
2
170 series EN
Maintenance
Cleaning
Introduction
The Meter Description
Do not attempt to repair this Meter. It contains no user
serviceable parts. Repair or servicing should only be
performed by qualied personnel.
Periodically wipe the case with a dry cloth and detergent.
Do not use abrasives or solvents.
Front Panel Illustration
1. JAW
2. Volt seek Light.
3. Push-buttons.
4. Trigger.
5. Rotary switch for turn the Power On / O and select the
function.
6. 6,000 count digital display.
7. Input Terminal for Multi-function.
8. Common (Ground reference) Input Terminal.
1
2
3
5
6
7
4
3
8
JAW
Voltseek Light
Button
Rotary Switch
Display
"+" Input
Trigger
Button
Common input
Function button
3
170 series EN
Capacitance
Making Basic Measurements
Function
DCV / DC mV
172 173 179
ACV
Resistor/ Continuity
Capacitor/ Diode
Frequency
Volt Seek
HFR/ INRUSH
Backlight
Torch light
Data Logger
Bluetooth
AC 600A
DC 600A
DC/AC 1500A
DC/AC μA
LoZ
PV Voltage
Temperature
Flexible Current
MIN MAX
Preparation and Caution Before Measurement
Observe the rules of Warnings and Cautions.
4
170 series EN
CAUTION
CAUTION
When connecting the test leads to the DUT (Device Under
Test) connect the common test leads before connecting the
live test leads ; when removing the test leads, remove the
live test leads before removing the common test leads.
Do not use the LoZ mode to measure voltages in circuits
that could be damaged by this mode’s low impedance.
Dial the switch to select the measuring function.
LoZ mode : Display measurement result at AC or DC value,
it depends on whichever is greater.
Measuring Voltage
5
170 series EN
Measuring PV Voltage
PV String
View side
CAUTION
This function is only available with the dedicated PV test
probe (APPA ATL-PV).
Always select correct DC / AC mode to perform high voltage
measurement. This meter will ash symbol and the correct
mode symbol (AC / DC) if the input voltage is dierent and
dangerous.
Dial the switch and press the Function button to select
AC/DC mode.
6
170 series EN
Measuring Current
Measuring Resistance/Continuity/
Capacitance/Diode
Dial the switch and press the Function button to select the
measuring function.
Note : Torch will be on when jaw is opened.
Dial the switch and press the Function button to select the
measuring function
OK
I+(-I)=0
OK
I
OK
Iwith respect to
earth for the jaw.
Tactile Barrier for
hand guard.
Do not hold
the meter across
the Barrier
CAT.IV 600V
CAT.III 1000V
7
170 series EN
Measuring Temperature °C / °F
Measuring μA
Dial the switch and press the Function button to select
AC/DC mode.
Funtion
Button
Dial the switch and press the Function button to
select °C / °F mode.
8
170 series EN
Measuring Current with Flex Clamp Meter
Using the Function
Press the Function button to change the function on the
same switch position.
Keep the range of Flex Clamp meter (APPA sFlex-T) at
3000A/3V output ratio.
Position
Switch
PV AC
AC
Hz
Hz
Hz
V
A
A
μA
Ω
V
Function
DC
DC
mVV
A
V
A
A
Ω
°C °F
9
170 series EN
Measuring Frequency
Backlight
MIN/MAX
Button
Funtion
Press
> 2 sec
2 Sec
Press Function button for over 2 sec. to turn Backlight
on / o.
The MAX/MIN mode records the min and max input values.
When the input goes below the recorded min value or above
the recorded max value, the meter records the new value.
Press Hold button to pause the recording.
MIN
MAX
to exit the MINMAX mode
Press
MIN
Press
Button > 2 sec
MAX
MIN
MIN
MAX
Press
MAX
Press
MAX
MIN
Press
MIN
MAX
10
170 series EN
Volt Seek
Smart Hold
Press MIN/MAX button for over 2 sec. to enter / exit Volt
Seek mode. Press MIN/MAX button to switch high/low
sensitivity.
The meter will beep continuously and the display will ash if
the measured signal is larger than the display reading by 50
counts. (However, it can not detect across the AC and DC
Voltage / Current).
MIN
MAX
MIN
MAX
Volt seek
2 Sec
Volt seek
2 Sec
Do not hold across this
line when sensing the
hazardous voltage.
Sensor
position
The Volt Seek LED indicates the electric eld. If the Volt
Seek LED is not on, voltage could still be present.
WARNING
HOLD
ZERO
11
170 series EN
DCA ZERO
High Frequency Rejection (HFR)
Remove the jaw from the conductor before performing
DCA ZERO.
Press HOLD button for over 2sec. to compensate the
residual magnetism.
The High Frequency Rejection mode equip a low pass lter
in the AC measurements. The cut-o frequency (-3dB point)
of low pass lter is 800Hz.
HOLD
ZERO
> 2 sec
Press
HOLD
ZERO
> 2 sec
Press
HFR
HFR
Press
12
170 series EN
INRUSH
In inrush current mode, select the suitable measurement
range by pressing HFR/INRUSH button before triggering the
inrush current measurement.
The hazardous voltage may be present even if the LCD
reading is very low in HFR mode. Verify the voltage again
without HFR mode.
WARNING
Press
> 2 sec
Inrush
2 Sec
HFR
2 Sec
Inrush
2
1
The current is > trigger level
.
Press
> 2 sec
HFR
Detecting
13
170 series EN
Auto Power O
Wake up the meter by dialing the switch or pressing any
button.
Press the function button and turn the meter on. Then press
the function button to select the time. The time can be 5
minutes, 10 minutes, 20 minutes, and disabled (OFF).
After Specified Time
Without Operation
Time Setting of Auto Power O
Funtion
Button
1
2
Funtion
Button
14
170 series EN
Testing LCD Monitor
To turn on the meter after keeping HOLD button down.
Pressing Bluetooth button while powering-up to select the –
Logger mode, Manual Saving mode and Clear memory.
Function of LOG Button
2
1
HOLD
Press
12
3
Press
Press
Press
15
170 series EN
Data Logger
Manual Saving Mode
The meter can store up to 4000 data in memory.
Press Bluetooth button for more than 2 seconds to activate
Data logger mode. The meter will enter Time interval setting
mode. Press Bluetooth button again to select time interval.
The interval can be 1 second, 5 seconds, 10 seconds, 30
seconds, 60 seconds.
Press
Start to record auto
matically after 2sec
without operaction.
LOG
2 Sec
>2Sec
LOG
2 Sec
Press
Press Press
CAUTION
All stored data will be cleared next startup. Download the
stored data by App rst if needed.
LOG
2 Sec
To save a reading
Show stored reading
amount for 2 seconds
HOLD
Flash
Press
LOG
2 Sec
>2Sec
16
170 series EN
Bluetooth
CAUTION
All stored data are saved until switching to data logger mode
or executing the clear function.
The meter uses Bluetooth low energy (BLE) V4.0 wireless
technology to transfer the real-time reading and the stored
data. The open-air communication range is up to 10m.
Download “APPA Connect” App via the following QR Code.
Turn on Bluetooth function of the meter and open “APPA
Connect” to connect the DMM. The Bluetooth icon of the
meter will freeze on LCD after the connection establishes
successfully.
Press
APPA Connect on Google Play APPA Connect on iTune Store
17
170 series EN
Default Temperature Units Setting
Turn on the meter after keeping MIN/MAX button down+
MAX
1
MIN
MIN
MAX
2
OR
Low Battery and Battery Replacement
Replace the battery as soon as the low battery indicator
appears, to avoid false reading.
Refer to the following gure to replace the batteries
1
3
2
Remove test leads from Meter before opening the battery
cover or Meter case.
WARNING
18
170 series EN
Specications
General Specications
Display : 6000 counts.
Overrange Indication : ”OL” or “-OL”
Measure : Samples 3 times per second .
Max Conductor Size of JAW :
37mm diameter for 172/173
42mm diameter for 179
Dimensions (W x H x D) :
62mm x 240mm x 41mm for 172/173
62mm x 254mm x 41mm for 179
Weight :approx. 430g (including battery) for 172/173
approx. 480g (including battery) for 179
Low Batteries Indication :
Voltage drops below operating voltage will ash.
Power Requirement :
AA Size Battery x 2 (R6, LR6, 15D, 15A)
Battery Life : 200 hours ALKALINE Battery for 173/179
300 hours ALKALINE Battery for 172
(without Backlight)
Operating Temperature : -10 ~10°C
10°C ~ 30°C ( ≤ 80% RH),
30°C ~ 40°C ( ≤ 75% RH),
40°C ~ 50°C ( ≤ 45%RH)
Storage Temperature :
-20°C to 60°C , 0 to 80% R.H. (batteries not tted)
Altitude : 6561.7 ft (2000m)
Safety : EN 61010-1, EN 61010-2-032, EN 61010-2-033
for CAT III 1000V, CAT IV 600V, EN 61326-1
Drop Protection : 4 feet drop to hardwood on concrete
oor Vibration : Random Vibration per MIL-PRF-28800F
Class 2
Pollution degree : 2
Indoor Use
The circuits directly connected to Low-voltage installation.
The building installation.
The source of the Low-voltage installation.
Application FieldCAT
III
IV
II
19
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127

APPA 170 Series ユーザーマニュアル

タイプ
ユーザーマニュアル