APPA 607 Insulation Multimeter ユーザーマニュアル

カテゴリー
マルチメータ
タイプ
ユーザーマニュアル
LIMITED
WARRANTY
YEARS
User Manual / 使用說明書 / 使用说明书
ニュアル
Руководство пользователя
607
EN Insulation Multimeter
絕緣萬用電表
绝缘万用电表
絶縁抵抗計
Измеритель сопротивления изоляции
TC
SC
JP
RU
Read First
!
!Safety Information
Understand and follow operating instructions carefully.
Use the meter only as specified in this manual; otherwise,
the protection provided by the meter may be impaired.
Identifies hazardous conditions and actions that could cause
BODILY HARM or DEATH
• When using test leads or probes, keep your fingers behind
the finger guards.
• Remove test lead from Meter before opening the battery door
or Meter case.
• Use the Meter only as specified in this manual or the
protection by the Meter might be impaired.
• Always use proper terminals, switch position, and range for
measurements.
• Verify the Meter’s operation by measuring a known voltage.
If in doubt, have the Meter serviced.
• Do not apply more than the rated voltage, as marked on
Meter, between terminals or between any terminal and earth
ground.
• Only replace the blown fuse with the proper rating as
specified in this manual.
• Use caution with voltages above 30 Vac rms, 42 Vac peak, or
60 Vdc. These voltages pose a shock hazard.
• To avoid false readings that can lead to electric shock and
injury, replace battery as soon as low battery indicator blinks/
appears.
• Disconnect circuit power and discharge all high-voltage
apacitors before testing resistance, continuity, diodes, or
capacitance.
• Do not use Meter around explosive gas or vapor.
• To reduce the risk of fire or electric shock do not expose this
product to rain or moisture.
• Probe assemblies to be used for MAINS measurements shall
be RATED as appropriate for MEASUREMENT CATEGORY
III or IV according to IEC 61010-031 and shall have a voltage
RATING of at least the voltage of the circuit to be measured.
Warning
!
EN
607
1
• Never attempt a voltage measurement with the test lead
inserted into the INSULATION input terminal.
Avoid working alone so assistance can be rendered.
• Do not use the Tester if the Tester is not operating properly or
if it is wet.
• Individual protective device must be used if hazardous live
parts in the installation where the measurement is to be
carried out could be accessible.
• Do not attempt a current measurement when the open
voltage is above the fuse protection rating. Suspected open
voltage can be checked with voltage function.
• Never attempt a voltage measurement with the test lead
inserted into the A input terminal.
!CAUTION
• Disconnect the test leads from the test points before
changing the position of the function rotary switch.
• Never connect a source of voltage with the function rotary
switch in Ω,,°C ,mA, INSULATION position.
• Do not expose Meter to extremes in temperature or high
humidity.
• Never set the meter in Ω,,°C ,mA, INSULATION function to
measure the voltage of a power supply circuit in equipment
that could result in damage the meter and the equipment
under test.
EN
607
2
Symbols as marked on the Meter and
Instruction manual
!
Risk of electric shock
See instruction card
DC measurement
Equipment protected by double or reinforced insulation
Battery
Earth
AC measurement
Conforms to EU directives
Application around and removal from hazardous live
conductors is permitted
Do not discard this product or throw away.
Both direct and alternating current
Fuse
Unsafe Voltage
To alert you to the presence of a potentially hazardous voltage,
when the Tester detects a voltage 30 V or a voltage overload
(OL) in V, mV, insulation function. The ”“ symbol is displayed
and High voltage indicator is turned on.
!WARNING
To avoid electric shock and injury, never use standard test
probes and INSULATION test probe into the input terminal at
the same time.
EN
607
3
The Meter Description
Front Panel Illustration
1. LCD display : 10000 counts .
2. Push-buttons for features.
3. Rotary switch for turn the Power On / Off and select the
function.
4. Input Terminal for Insulation function.
5. Input Terminal for V, Ω, , Hz, °C functions.
6. Common (Ground reference) Input Terminal for all functions
expect Insulation function.
7. Input Terminal for mA or Common Input Terminal for
Insulation function.
EN
607
1
2
Display
3Funtions &
Power On/Off
4Insulation
Features
6Common
Input
6mA &
Insulation
5V, Ω, , Hz, °C
Input
4
Making Basic Measurements
When connecting the test leads to the DUT (Device Under Test)
connect the common (mA) test lead before connecting the
live lead ; when removing the test leads removing the test
live lead before removing the common test lead.
The figures on the following pages show how to make basic
measurements.
Measuring AC / DC Voltage
EN
607
Preparation and Caution Before Measurement
Observe the rules Warnings and Cautions.
!! !
5
Measuring Resistance / Continuity / Diode
Press the Blue button to select the measuring function.
Testing Diode
EN
607
6
Testing Continuity
The buzzer allows you to quick continuity tests without having
to watch the display. The buzzer sounds when a short (< 30Ω)
is defected.
Measuring Capacitance
EN
607
7
Measuring Temperature °C / °F
Measuring AC / DC Current
EN
607
8
Measuring Insulation Resistance
Press the Range button to select the test voltage
(50V / 100V / 250V / 500V / 1000V)
Lock test voltage
Press the Hold button to lock the test voltage. Press the button
again to cancel the lock mode.
Make the reading stably
Press the Blue button to make the reading stably, the “Smooth”
appears on the display. Press the Blue button again to cancel
this mode.
EN
607
Select test voltage
9
Measuring Insulation Resistance
Insulation tests should only be performed on dead circuits.
Check the fuse before testing, see page 19 for how to check
the fuse.
To measure insulation resistance, follow the steps below.
1. Insert test probes in the “ + “and ““ input terminals.
2. Turn the rotary switch to Insulation position.
3. Press the Range button to select the test voltage.
4. Connect the probes to the circuit.
5. Push and hold the Test button to start the test.
The “Test” and ”” appear on the display. The secondary
display shows the test voltage applied to the circuit under
test. The primary display shows the resistance.
6. Keep the probes on the test points and release the Test
button. The resistance reading appears on the primary
display until a new test is started or a different function or
range is selected or > 30 V is detected.
=
EN
607
10
Measuring Frequency for ACV, ACmV and
ACmA
The meter measures the frequency of a voltage or current
signal by counting the number of times the signal crosses a
thresholdlevel each second. To measure frequency, follow the
steps below.
1. Turn the rotary switch to V, mV or mA position.
2. Press the Hz button to measure the frequency.
3. Press the Hz button or change the rotary switch position
to exit this function.
Display Hold
In the Hold mode, the meter holds the reading. To use the Hold
mode, follow the steps below.
1. Press the Hold button to activate Hold mode. The “ Hold ”
appears on the display.
2. Press the Hold button, Blue button, Range button or change
the rotary switch position to exit this function.
EN
607
11
Display MAX / MIN / AVG
The “MIN/MAX/AVG“ mode records minimum and maximum
input values. When the inputs go below the recorded minimum
value or above the recorded maximum value, the meter beeps
and records the new value. MIN/MAX/AVG mode can also
calculate an average of maximum and minimum. To use the
MIN/MAX/AVG mode, follow the steps below.
1. Press the “MIN/MAX“ button to activate MIN/MAX/AVG
mode. The “MIN MAX” blinks on the display.
2. Press the “MIN/MAX“ button to step through the present
readings, maximum, minimum and average (AVG).
3. Press the “MIN/MAX“ button for 2 seconds, Blue button,
Range button or change the rotary switch position to exit
this function.
EN
607
12
Relative mode for Ω /
In the Relative mode, the meter records the present reading as
reference and the later reading will subtract it. To use the
Relative mode, follow the steps below.
1. Turn the rotary switch to Ω or position.
2. Press the Hz button to activate Relative mode. The “ 0 “
blinks on the display.
3. Press the Hz button, Blue button, Range button or change
the rotary switch position to exit this function.
EN
607
13
Auto Test and Manual Test
The meter has both “AutoTest “ mode and Manual Test mode.
• In the Auto Test mode, the meter compares the reading of
AC and DC, and the bigger reading appears on the display.
The meter beeps when the AC/DC mode has change.
• In the Manual Test mode, you override “AutoTest “ and select
the AC/DC mode yourself.
When you turn the rotary switch to V, mV or mA position, it
defaults to Auto Test mode and the “AutoTest ” appears on the
display.
1. To enter the Manual Test mode, press the Blue button.
2. In the Manual Test mode, press the Blue button to change
the AC/DC mode.
3. Press the Blue button for 2 seconds to activate Auto Test
mode.
High Frequency Reject mode
When the rotary switch in V position, the HFR mode can be
used. To use HFR mode, press the Blue button in the Manual
Test mode.
EN
607
14
Auto Range and Manual Range
EN
607
The meter has both Auto Range mode and Manual Range
mode.
• In the Auto Range mode, the meter selects the range with
the best resolution.
• In the Manual Range mode, you select the range yourself.
When you turn the meter on, it defaults to Auto Range mode
and the “Range” appears on the display.
1. To enter the Manual Range mode, press the Range button.
The “Range” disappears on the display.
2. In the Manual Range mode, press the Range button to
increment the range. After the highest range, the meter
returns to the lowest range.
3. Press the Range button for 2 seconds to activate Auto
Range mode.
15
Store and Recall
EN
607
You can store the reading on the display, and recall the on the
display after.
• To store the reading press the Store button.
The data amount and “MEM” blink on the secondary display.
• Each function has a separate memory space. Each memory
space has the maximum 100 amounts.
To recall the reading on the display, press the Store button
for 2 seconds to activate Recall mode. The data amount
shows on the secondary display. In the Recall mode, you
can make the following operation.
Press the ← button or → button to select the data amount.
Press the ← button or → button for 2 seconds to search
data quickly.
• Press the Blue button to clear all stored data in this function.
• Press the Store button for 2 seconds or change the rotary
switch position to exit this function.
16
Auto Power Off
Wake-up the meter by switching rotor or pressing any button.
Auto Backlight
The backlight is automatically turned on at dark environment.
BUZZER
The meter beeps once for every valid key-press, and beeps
twice for every invalid key-press.
Power On Options
Press button while turning the meter on from OFF position.
Blue button : Disable APO
MAX/MIN button : Disable auto backlight
Store button : Clear all stored data
Test button : Display LCD test frame
Range button : Default °C / °F reading
EN
607
17
Battery and Fuse Replacement
Refer to the following figure to replace fuse and the batteries :
!CAUTION
•Use only a fuse with the amperage, interrupt, voltage, and
speed rating specified.
•Fuse rating : 440mA,1000V AC/DC, IR 10kA, fast blow fuse.
•Replace the batteries as soon as the low batteries indicator
" " appears, to avoid false reading.
•1.5V x 4 alkaline batteries.
Maintenance
Do not attempt to repair this Meter. It contains no user
serviceable parts. Repair or servicing should only be
performed by qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and detergent.
Do not use abrasives or solvents.
EN
607
18
Testing the fuse
Test the fuse as described below.
1. Open the battery door and remove the fuse.
2. Use the continuity function to check the fuse.
EN
607
19
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130

APPA 607 Insulation Multimeter ユーザーマニュアル

カテゴリー
マルチメータ
タイプ
ユーザーマニュアル