APPA 103N ユーザーマニュアル

カテゴリー
測定、テスト
タイプ
ユーザーマニュアル
103N/105N/106
User Manual / 使用說明書 / 使用说明书
ユーザーマニュアル
Руководство пользователя
The Toughest Multimeters
堅固型萬用表
坚固型万用表
堅牢型マルチメーター
Мультиметр цифровой
RU
EN
TC
SC
JP
LIMITE D
WARRAN TY
YEAR S
103N / 105N / 106 EN
WARNING
1-1 Unpacking and Inspection
1-2 Meter Safety
INTRODUCTION
THESE SERVICING INSTRUCTIONS ARE FOR USE BY
QUALIFIED PERSONNEL ONLY.
TO AVOID ELECTRIC SHOCK, DO NOT PERFORM ANY
SERVICING OTHER THAN THAT
CONTAINED IN THE OPERATING INSTRUCTIONS
UNLESS YOU ARE QUALIFIED TO DO SO.
Upon removing your new Digital Multimeter from its packing,
you should have the following items:
1. Digital Multimeter.
2. Test lead set (one black, one red).
3. Instruction manual.
4. Protective holster.
5. K Type Temperature Sensor. (106 only)
6. Sensor Adaptor. (106 only)
Terms as Marked on Equipment.
Symbols in this Manual.
ATTENTION — Refer to manual.
DOUBLE INSULATION — Protection Class II.
DANGER — Risk of electric shock
This symbol indicates where cautionary or other
information is found in the manual.
FUSE
Battery
1
103N / 105N / 106 EN
1-3 Front Panel
Refer to Figure 1 and to the following numbered steps to
familiarize yourself with the meters front panel controls and
connectors.
1. Digital Display — The digital display has a 4000 counts
LCD readout with 82 segments analog bar graph, auto
polarity, decimal point, “ AC, DC, , , RANGE, H,
APO, REL, MAX, MIN, PMAX, PMIN and unit
annunciators. (PMAX, PMIN, REL are 105N / 106 only).
2. Rotary Switch — Select the Function and Range
desired.
3. COM Input Terminal — Ground input connector.
4. °C °F VΩ Hz Input Terminal — Positive input
connector for Volts, Ohms Capacitance, Frequency and
Temperature.
5. mA Input Terminal — Positive input connector for Amp
measurements (up to 400mA).
6. A Input Terminal — Positive input connector for Amp
measurements (up to 10A).
Push Switch Functions ( “MIN/MAX” and “Peak H ” and
“Rel “ switch are for 105N / 106 only)
Push switch functions control the special measurement
modes of the meter. Some special modes can be nested
in other special modes, while some push functions reset
all existing special mode. The following table summarizes
the push functions in each special mode.
HOLD
Active mode
Push Switch Functions
RANGE1 Hz2
BLUE
KEY1
MIN/
MAX
HOLD H PEAK H REL
REL
MIN/MAX HOLD
PEAK HOLD
HZ
2
103N / 105N / 106 EN
Note :
1. BLUE KEY switch and RANGE Switch resets all special
modes other than HZ mode .
2. HZ switch is only used to toggle between frequency
measurement mode and voltage or current measurement
mode, and thus can not be nested in any other special
modes.
3. The bar graph always shows the current value, regardless
of special mode.
7. Light switch — Press the switch to turn on or turn o the
back-light.
8. HZ — If “ HZ” switch is pushed in voltage or current
measurement mode, the meter enters frequency
counter mode with automatic range selection.
Therefore, pressing “RANGE” switch in HZ
mode does not change the frequency range.
However,“RANGE” switch changes the sensitivity
of frequency detection. If the input signal has a
small amplitude, the user shall increase the
sensitivity.
Note : Since HZ mode is treated as a temporary frequency
measurement in voltage or current mode, pressing
“RANGE” switch in HZ mode also changes the full
scale range of the original voltage or current mode
after VAHZ mode is cancelled.
9. PEAK H — The precision of PEAK hold mode
measurement can be enhanced by calibration. Calibration
for PEAK hold mode is invoked by pressing “PEAK H
switch for more than 2 seconds until the “caL” is displayed.
Pressing the “Peak H ” switch to toggle between PMAX
and PMIN.Pressing “Peak H ” switch and held for longer
than one second to release the peak hold mode.
10. Range Switch, (Manual Range) — “Range” switch is
pressed to select manual ranging and to change ranges.
When “Range” switch is pressed on “RANGE” annunciator
on the LCD appears.
Press “RANGE” switch to select appropriate range to be
used.
Press “RANGE” switch and hold for 2 seconds to return
to Autoranging. In “ HZ” mode “RANGE” is not used
to change the full scale frequency range, but change the
sensitivity of frequency measurement.
11. Blue Switch — Press the switch to measure AC Voltage
/ Current or DC Voltage / Current in the voltage / Current
mode, or to measure Resistance or continuity or diode in
Ω / / mode or to measure Frequency or RPM in RPM
in Hz / PRM mode.
3
103N / 105N / 106 EN
12. MIN MAX — The meter displays the maximum or
minimum value of the input in MAX / MIN mode. When
“MIN MAX “ switch is pressed for the rst time, the meter
displays the maximum value. When “MIN MAX” switch
is pressed again, the meter displays the minimum value.
When “MIN MAX” switch is pressed for the third time,
the meter displays current input value and the “MAX”
“MIN” annunciator are blink. The meter returns to normal
operation if “MIN MAX” switch is pressed and held for
longer than one second. Pressing HOLD in MAX/MIN
mode makes the meter stop updating the maximum or
the minimum value.
13. REL In REL mode, the LCD panel displays
D(current) - D(relative), where D(relative) is the last
value before “REL ” switch is pressed, and D(current)
is the current value. If “REL ” switch is pressed again
in REL mode, the meter displays the reference value and
the “REL” annunciator is blink. The meter returns to
normal operation if “REL ” switch is pressed and held
for longer than one second. Pressing “HOLD H ” switch
in REL mode makes the meter stop updating the LCD
panel.REL with MIN MAX — MAX / MIN mode can be
nested in REL mode. The “MIN MAX’ meter displays the
maximum or the minimum value relative to the reference
when “MIN MAX” is pressed in REL mode.Before release
the REL function, the MAX/MIN function is needed to
release rst.
14. HOLD — HOLD mode makes the meter stop updating
the LCD panel. This mode can be nested in most of the
special modes. Enabling HOLD function in automatic
mode makes the meter switch to manual mode, but the
full scale range remains the same. HOLD function can
be released by changing the measurement mode,
pressing “RANGE” or “BLUE KEY” or push HOLD again.
When HOLD mode is nested in Peak H or MIN / MAX
or REL mode to release the Peak H or MIN / MAX or
REL function is needed to release HOLD function rst.
4
103N / 105N / 106 EN
2-1 General Specications
SPECIFICATIONS
Display : The Liquid Crystal Display (LCD) with a maximum
reading of 4000 and 82 segments bar graph.
Polarity Indication : Automatic, positive implied, negative
indicated.
Overrange Indication : “OL” or “-OL”.
Low Battery Indication : “ is displayed when the battery
voltage drops below operating voltage.
Sampling : 2 times/sec for digit. 12 times/sec for analog
bargraph.
Auto Power O : Approx 30 minutes.
Operating Ambient : 0°C to 30°C (≤ 80%R.H),
30°C to 40°C (≤ 75%R.H),
40°C to 50°C (≤ 45%R.H).
Storage Temperature : -20°C to 60°C , 0 to 80% R.H.
when battery removed from meter.
Temperature Coecient :
0.15 x (Spec.Acc’y) / °C , <18°C or > 28°C.
Power Requirements :
IEC LR03, AM4 or AAA size 1.5V x 2. (103N / 105N)
Standard 9V battery NEDA 1604, IEC6F22, JIS 006P.(106N)
Battery Life : Alkaline 500 hours. (103N / 105N),
Alkaline 300 hours. (106)
Dimensions (W x H x D) : 90mm x 200mm x 42mm ,
without holster. 100mm x 212mm x 55mm , with holster.
5
103N / 105N / 106 EN
2-2 Environmental Conditions
2-3 Electrical Specications
(1) DC Volts
Indoor use.
Maximum Altitude : 2000 Meter.
Pollution Degree : 2
Installation Category : EN61010-1,EN61010-2-033, 1000V
Cat.II, 600V Cat.III.
Accuracy is ± (% reading + number of digits) at 23°C ± 5°C ,
less than 80% R.H.
Input Impedance : 10MΩ . (over 1000MΩ in 400mV range).
The circuits not connected to mains.
The circuits directly connected to Low-voltage installation.
The building installation.
The source of the Low-voltage installation.
Application FieldCAT
II
III
IV
I
Range Resolution
Accuracy
103N 105N / 106
±(0.3%reading +
2digits)
400mV 100μV
40V 10mV
4V 1mV
400V 100mV
1000V 1V
±(0.3%reading +
2digits)
±(0.1%reading +
2digits)
±(0.4%reading +
2digits)
±(0.25%reading +
2digits)
Over voltage
protection
1000V rms
(2) AC Volts
Range Resolution
Accuracy
103N 105N / 106
±(2.0%reading +
8digits)*
400mV 0.1mV
40V 10mV
4V 1mV
400V 100mV
750V 1V
±(2.0%reading +
8digits)*
***
±(1.5%reading +
5digits)
40Hz to 60Hz
±(1.0%reading +
5digits)
60Hz to 1KHz
±(1.3%reading +
5digits)**
±(1.5%reading +
5digits)
40Hz to 60Hz
±(1.3%reading +
5digits)
60Hz to 1KHz
Over voltage
protection
1000V rms
Accessories : Protective Holster , battery, instruction
manual, test leads.
6
103N / 105N / 106 EN
(3) DC Current
AC Conversion Type :
103N / 105N : Average sensing rms indication.
106 : AC conversions are ac-coupled, true rms responding,
calibrated to the rms value sine wave input.
The specied accuracy is for since wave at full scale and
non-sine wave at half scale below 500Hz with crest factor
up to 2.
Input impedance : 10MΩ // less than 100PF.
* Frequency Response : 50Hz to 60Hz.
** Frequency Response : 40Hz to 300Hz.
*** Frequency Response : 40Hz to 500Hz for 4V Range.
AC Conversion Type : AC conversions are average
sensing , calibrated to the rms value sine wave input.
Overload Protection :
1A , 600V IR 10KA fuse (Bussmann BBS-1 or equivaknt) for
mA input. (Size 10.3mm x 34.9mm)
15A , 600V IR 100KA fuse (Bussman KTK 15 or equivalent)
for A input. (Size 10.3mm x 38.1mm)
Range Resolution
Accuracy
103N 105N / 106
±(0.6%reading +
2digits)
40mA 10μA 300mV max
10A 10mA 3V max
400mA 0.1mA 3V max
±(0.4%reading +
2digits)
±(0.7%reading +
2digits)
±(1.2%reading +
3digits)
±(0.5%reading +
2digits)
±(1.0%reading +
3digits)
Voltage
Burden
(4) AC Current
Range Resolution
Accuracy
103N 105N / 106
±(1.5%reading +
5digits)
40mA 1μA 300mV max
10A 10mA 3V max
400mA 0.1mA 3V max
±(1.5%reading +
5digits)
±(2.5%reading +
5digits)
±(2.0%reading +
5digits)
Voltage
Burden
Frequency Response : 40Hz to 1KHz.
AC Conversion Type :
103N / 105N : Average sensing rms indication.
106 : AC conversions are ac-coupled, true rms responding,
calibrated to the rms value sine wave input.
The specied accuracy is for since wave at full scale and
non-sine wave at half scale below
500Hz with crest factor up to 2.
7
103N / 105N / 106 EN
Overload Protection : 1A , 600V IR 10KA fuse (Bussmann
BBS-1 or equivaknt) for mA input. (Size 10.3mm x 34.9mm)
15A , 600V IR 100KA fuse (Bussman KTK 15 or equivalent)
for A input. (Size 10.3mm x 38.1mm)
AC Conversion Type : AC conversions are average
sensing, calibrated to the rms value sine wave input.
(5) Resistance
Range Resolution
Accuracy
103N 105N / 106
±(0.7%reading +
3digits)
±(0.7%reading +
3digits)
±(1.5%reading +
5digits)
±(0.6%reading +
3digits)
±(1.5%reading +
5digits)
400Ω 0.1Ω
40KΩ 10Ω
4KΩ 1Ω
400KΩ 100Ω
4MΩ
40MΩ
1KΩ
10KΩ
±(0.7%reading +
3digits)
±(0.6%reading
+3digits)
±(0.4%reading +
2digits)
Over voltage
protection
600V rms
Open circuit Voltage : -1.3V approx.
(6) Diode Check and Continuity
Range Resolution Accuracy Max. Test Current
Current
Max. Open Circuit
Circuit Voltage
±(1.5%reading +
5digits)*
1mV 1.5mA 3V
* For 0.4V — 0.8V.
Overload Protection : 600V rms max.
Continuity : The internal sounder operates when resistance
is less than 30Ω approximately.
(7) Frequency / RPM
Range Resolution Accuracy
Sensitivity
4.0KHz/40KRPM
40KHz/400KRPM
400KHz/4MRPM
4MHz/40MRPM 1KHz/30KRPM 300mV rms
40MHz/
400MRPM
10KHz/
300KRPM 1V rms
* 400MHz/
4000MRPM
100KHz/
3MRPM * *
1Hz/30RPM
10Hz/300RPM
100Hz/3KRPM
150mV rms ≥ 20Hz
1.5V rms ≤ 20Hz Frequency :
±(0.01%reading
+ 1digit)
RPM :
±(0.01%reading
+ 10digits)
600V rms
Over voltage
protection
8
103N / 105N / 106 EN
(8) Capacitance
(9) Temperature (°C )
Range
Temperature
Resolution Accuracy
Accuracy
Overload
Protection
Overload Protection
4nF
-20°C — 0°C
1pF
±(2%reading + 4°C )
40nF
1°C — 100°C
10pF
±(2%reading + 4°C )
400nF
1°C — 100°C
100pF
±(2%reading + 3°C )
4μF
1°C — 100°C
1nF
±(3%reading + 2°C )
40μF 10nF
400μF 100nF
*4mF 1μF
*40mF 10μF
±(3%reading +
10digits)
±(2%reading +
8digits)
600V rms
600V rms
** ±(5%reading +
20digits)
* In these two ranges the reading maybe rolling within
specication.
** Specify reading < half of full scale range .
For 106 only.
* For 105N / 106 only.
* * The spec is not guaranty.
9
103N / 105N / 106 EN
(11) PEAK HOLD
Function FunctionRange Range
Accuracy Accuracy
400mV Unspecied 40mA 3*
40mA 3*
400mA 3*
400mA 3*
10A 3*
10A 3*
Unspecied
±(1.5%reading
+300digits) 2*
±(3%reading
+60digits)
±(1.5%reading
+60digits)
±(3%reading
+60digits)
±(1.5%reading
+60digits)
±(1.5%reading
+300digits) 2*
±(1.5%reading
+60digits)
±(1.5%reading
+60digits)
400mV
40V
40V
400V
400V
1000V
750V
4V
4V
DCV DCV
ACV ACV
Note :
1. With zero calibrated before measurement.
2* 4V range species readings above 10% of full scale of range.
3* Amp ranges specify reading <90% of full scale of range.
4. In the noise generating eld, may aect intervals.
(10) Temperature (°F )
Temperature Accuracy Overload Protection
-4°F — 32°F ±(2%reading + 8°F)
33°F — 212°F ±(1%reading + 6°F)
213°F — 932°F ±(2%reading + 6°F)
933°F — 1472°F ±(3%reading + 4°F)
600V rms
For 106 only.
10
103N / 105N / 106 EN
(12) Auto Power O (APO)
(13) Disable Auto Power O
(14) RS232
The APO sign on the LCD panel indicates the meter is
working in the Auto Power O mode. If the meter idles
for more than 30 minutes, the meter automatically turns
the power o. When this happens, the state (non-logic
measurement) of the meter is saved, the meter can be
turned back on by pushing any key switch, except back
light switch, or changing the rotary switch. If the meter
is Re-Powered on by pushing key switch, the LCD
displays the saved state when meter auto power o
itself, pushing Hold key switch to disable the hold state.
The meter will give periodic alarm in 15 seconds before
automatically turns power o by itself, any key press or
rotary change reset Auto-Power-O.
In order to disable auto power o function, power on
the meter with pressing anyone of switches, except
HOLD switch, back light switch or BLUE switch (for
function change) is pressed down.
Sent a single command string “H” ( Hex code “48” )
to the multimeter or shine the ashlight on the RS232
socket of the meter to enable or disable the RS232.
When enable the RS232 the LCD display shows the
RS232 annunciator.
The serial data is send from RS232 cable twice every
A/D conversion cycle. The data format complies with
JIS 7BIT transmission code with a baud rate of 2400.
The user can use RS232 interface to read the data.
A single dada packet includes a start bit (always 0), 7
data bits, an odd parity check bit, and a stop bit (always
1). The following gure shows
01
LSB M SB
-12V
12V
11
103N / 105N / 106 EN
One data block consists of 11 packets, or 110 bits. The
following gure shows the format of a data block. The
range packet indicates the full scale range of the meter.
Digit 3 through digit 0 is just the digits on the LCD panel.
The function packet indicates the measurement mode of
the meter. Status, option 1 and option 2 gives the status of
the meter. CR and LF are delimiters used to separate the
blocks.
The meter always outputs the current input value to the
serial port. Each block is repeated twice in one conversion
cycle.
The detailed data format of each packet is listed below.
This packet indicates the measurement mode of the meter.
The following table summaizes the transmitted code for
each mode.
01 01 01
01 01 01
01 01 01
01 01
CR LF
status option1o ption2
ranged igit3d igit2
digit1 digit0 function
14-1 FUNCTION
Code Measurement Mode
0111011
0111101
0111001
0111111
0110011
0110101
0110001
0110010
0110110
0110100
Voltage
μA Current
mA Current
A Current
Ω
Continuity
Diode
Frequency / RPM1
Capacitance
Temperature2
12
103N / 105N / 106 EN
Note :
1. The Judge bit in the Status packet determines whether it
is frequency mode or RPM mode.
2. The Judge bit in the Status packet determines whether
the unit is Celsius or Fahrenheit.
This packet indicates the full scale range of the meter.
When the meter operates in continuity mode, diode mode,
or current (A) mode, this packet is always 0110000 since the
full scale range in these modes are xed. The following table
lists the code for each range in each measurement mode.
Digit 3 is the most signicant digit on the LCD panel, and
digit 0 is the least signicant digit. When the LCD panel
shows OL, the serial port outputs 4000.
14-2 RANGE
14-3 DIGIT 3 — DIGIT 0
Digit
Code V mA ΩFrequency RPM Capacitance
Code
0
1
2
3
4
5
6
7
8
9
0110000
0110001
0110010
0110011
0110100
0110101
0110110
0110111
400.0mV
4.000V
40.00V
400.0V
4000V
4.000nF
40.00nF
400.0nF
4.000μF
40.00μF
400.0μF
4.000mF
40.00mF
400.0Ω
4.000KΩ
40.00KΩ
400.0KΩ
4.000MΩ
40.00MΩ
4.000KHz
40.00KHz
400.0KHz
4.000MHz
40.00MHz
400.0MHz
40.00KRPM
400.0KRPM
4.000MRPM
40.00MRPM
400.0MRPM
4000MRPM
40.00mA
400.0mA
0110000
0110001
0110010
0110011
0110100
0110101
0110110
0110111
0111000
0111001
13
103N / 105N / 106 EN
The format of this packet is shown below. The Judge eld is
meaningful only when the Function packet indicates
Frequency / RPM mode .
In Frequency / RPM mode, judge is 0 if the meter operates
in Frequency mode; otherwise, it is 1. Sign eld indicates
whether the minus sign on the LCD panel is on or o. BATT
eld is 1 if battery low condition is true. OL indicates input
overow.
This packet contains information on special measurement
modes. The format of this packet is shown below. The three
non-constant elds is set to one when the meter operates in
the corresponding special modes.
This packet contains information on the operation mode of
the meter. The format is shown below. The DC eld indicates
that the meter operates in DC measurement mode, either
voltage or current. The AC eld indicates that the meter
operates in AC measurement mode. The AUTO eld is set
to 1 if the meter operates in automatic mode, and is set to
0 when the meter operates in manual mode. The APO eld
indicates whether auto power o function is enabled or not.
Carriage return. The transmitted code is 0001101.
Line feed. The transmitted code is 0001010.
14-4 STATUS
14-5 OPTION 1
14-6 OPTION 2
14-7 CR
14-8 LF
0
0
0
Bit 6
Bit 6
Bit 6
1
1
1
Bit 5
Bit 5
Bit 5
1
1
1
Bit 4
Bit 4
Bit 4
Judge
PMAX
DC
Bit 3
Bit 3
Bit 3
Sign
PMIX
AC
Bit 2
Bit 2
Bit 2
BATT
0
AUTO
Bit 1
Bit 1
Bit 1
OL
VAHZ
APO
Bit 0
Bit 0
Bit 0
14
103N / 105N / 106 EN
The sensitivity in the Hz mode is 1/10 of full scale range.
The accuracy is same as Frequency mode. The measuring
frequency is from 40Hz up to 1KHz.
This instrument has been designed and tested in
accordance with IEC Publication 1010, Safety Requirements
for Electronic Measuring Apparatus and has been supplied
in a safe condition. This instruction manual contains some
Information and warnings which have to be followed by the
user to ensure safe operation and to retain the instrument in
safe condition.
1. Before measurement, warm up for at least 60 seconds.
2. When the rotary function selector is changed during
measurement, be sure do so only after removing the test
leads from
the equipment.
3. If the equipment is used near noise generating equipment,
be aware that may become unstable or indicate large errors.
4. Maximum rated voltage to earth for voltage and current
measurements terminals is 1000V CAT.II, 600V CAT.III .
1. Connect the red test lead to the “ VΩHz” input terminal
and the other (black) test lead to the “COM” terminal.
2. Set the rotary function to the VA position.
3. Measurement of AC voltage can be performed by pushing
the “BLUE” key switch.
TO AVIOD ELECTRICAL SHOCK, HAZARD OR
DAMAGE TO METER, DO NOT A ATTEMPT TO
MEASURE VOLTAGE THAT MIGHT EXCEED 1000V
ms. DO NOT APPLY MORE THAN 1000V rms
BETWEEN THE COMMON INPUT TERMINAL AND
EARTH GROUND.
(15) Hz Sensitivity
OPERATION
3-1 Preparation and Caution before Measurement
3-2 Voltage Measurements
WARNING
15
103N / 105N / 106 EN
1. Connect the red test lead to “mA” terminal and the other
(black) test lead to “COM” terminal, or use the “A” and
“COM” terminal in the 10A range.
2. Set function selector rotary switch to “ mA ” or “ A ” .
3. Measurement of AC current can be performed by pushing
the “BLUE” switch.
4. Connect the test leads to the circuit to be measured.
1. Connect the red test lead to the “ VΩHz” terminal and
the other (black) test lead to the “COM” terminal.
2. Set the rotary function selector to “Ω ” position to
measure the resistance.
3. For correct reading, ensure that the device being tested
contains no voltage.
4.Connect the test leads across the resistor to be measured.
In order to ensure the best accuracy in measurement of
low resistance, short the test leads before measurement
and memorize the test probe resistance in mind. This is
necessary to subtract for the resistance of the test leads
themselves.
1. Connect the red test lead to the “ VΩHz” terminal and
the other (black) test lead to the “COM” terminal.
2. Set the rotary function selector to “Ω ” position.
3. Connect the test leads to the circuit to be measured. The
internal sounder operates if the resistance of the circuit
measured is lower than 30Ω approximately.
3-3 Current Measurements
3-4 Resistance Measurement
3-5 Continuity Check by Buzzer
NOTICE
UNSTABLE DISPLAY MAY OCCUR ESPECIALLY AT
400mV RANGE, EVEN THOUGH YOU DON’T PUT
TESTED LEADS INTO INPUT TERMINALS. IN THIS
CASE, IF AN ERRONEOUS READING IS SUSPECTED,
SHORT THE “ VΩHz” TERMINAL AND THE “COM”
TERMINAL, AND MAKE SURE THE ZERO DISPLAY.
16
103N / 105N / 106 EN
1. Set the rotary switch at “Ω ” position.
2. Connect black test lead to “COM” terminal and red lead to
VΩHz ” input terminal.
3. Connect test leads to the diode normally the forward
voltage drop of good silicon diode is shown between.
400V to 0.900V.
If the diode under test is defective. “0.000” (short circuit)
or “OL” (non-conductance) is displayed. Reverse check
of diode if the diode under test is defective “0.000” or
other value are to be displayed.
1. Connect the red test lead to the “ VΩHz ” terminal and
the other (black) test lead to the “COM” terminal.
2. Set the rotary function selector to “Hz RPM” position to
measure the frequency or RPM with the blue switch.
3. Connect the test leads to the circuit to be measured.
1. Connect the red test lead to the “ VΩHz” terminal and
the other (black) test lead to the “COM” terminal.
2. Set the rotary function selector to “ ” position to measure
capacitance.
3. Connect the test leads to the circuit to be measured.
4. In order to ensure the best accuracy in measurement of
low capacitance, open the test leads before measurement
and memorize the test probe capacitance in mind. This is
necessary to subtract for the capacitance of the test
probe themselves or use the relative mode for 105N.
3-6 Diode Check
3-7 Hz / RPM Measurements
3-8 Capacitance Measurement
17
103N / 105N / 106 EN
WARNING :
TO AVOID ELECTRICAL SHOCK REMOVE TEST
LEAD BEFORE OPENING THE COVER.
MAINTENANCE
1. Do not attempt to repair this Tester. It contains no
user-serviceable parts. Repair or serving should only be
performed by qualied personnel.
2. Periodically wipe the case with a dry cloth and detergent
do not use abrasives or solvents.
The meter is powered by 1.5V x 2 battery. Refer to Figure 2
and use the following procedure to replace the battery:
1. Disconnect the test leads and turn the meter o. Remove
the test leads from the front terminals.
2. Position the meter face down. Remove the screws from
the battery case bottom.
3. Lift the end of the battery case bottom until it gently
unsnaps from the case bottom.
4. Lift the batteries from the battery case top and carefully
disconnect the battery connector leads.
5. Snap the battery connector leads to the terminals of a
new battery and reinsert the battery into the case top.
Make sure that the battery leads do not become pinched
between the case bottom and case top.
6. Replace the case top and battery case.
Refer to Figure 3 and the following procedure to examine or
replace the meters fuse:
1. Perform steps 1 though 3 of the battery replacement
procedure.
2. Than remove the two screws from the case bottom and
lift the case bottom until it gently unsnaps from the case top.
3. Remove the defective fuse by gently prying one end of
the fuse loose and sliding the fuse out of the fuse holder.
4. Install a new fuse of same size and rating. Make sure the
new fuse is centered in the fuse holder.
5. Replace the case top and case bottom and battery case
bottom. Make sure that the battery leads do not be
come pinched between the case halves. Reinstall the
three screws.
4-1 Maintenance and Cleaning
4-2 Battery Installation or Replacement
4-3 Fuse Replacement
18
103N / 105N / 106 EN
Figure 2. Battery Replacement.
Battery
Cover
Battery Box
1.5V
Battery
Screw
Screw
Figure 3. Fuse Replacement
Top Case
Fuse 1
15A (600V)
1A (600V)
Screw
Screw
Screw
Battery Cover
Bottom
Case
19
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112

APPA 103N ユーザーマニュアル

カテゴリー
測定、テスト
タイプ
ユーザーマニュアル