Vaisala HMT140 ユーザーマニュアル

  • Vaisala HMT140シリーズWi-Fiデータロガーのクイックガイドの内容を読み込みました。このデバイスの機能、設定方法、トラブルシューティングなど、ご質問にお答えします。例えば、バッテリー交換方法やネットワーク接続設定、データ送信方法などについて詳しい情報をご提供できます。
  • バッテリー交換方法は?
    ネットワークへの接続方法は?
    データ送信をトリガーするには?
EN
DE
FR
JA
ZH
M211484EN-G
Quick Guide
Vaisala Wi-Fi Data Logger
HMT140
PUBLISHED BY
Vaisala Oyj
Vanha Nurmijärventie 21, FI-01670 Vantaa, Finland
P.O. Box 26, FI-00421 Helsinki, Finland
+358 9 8949 1
Visit our Internet pages at www.vaisala.com.
© Vaisala Oyj 2019
No part of this document may be
reproduced, published or publicly
displayed in any form or by any means,
electronic or mechanical (including
photocopying), nor may its contents be
modified, translated, adapted, sold or
disclosed to a third party without prior
written permission of the copyright holder.
Translated documents and translated
portions of multilingual documents are
based on the original English versions. In
ambiguous cases, the English versions are
applicable, not the translations.
The contents of this document are subject
to change without prior notice.
Local rules and regulations may vary and
they shall take precedence over the
information contained in this document.
Vaisala makes no representations on this
document’s compliance with the local
rules and regulations applicable at any
given time, and hereby disclaims any and
all responsibilities related thereto.
This document does not create any legally
binding obligations for Vaisala towards
customers or end users. All legally binding
obligations and agreements are included
exclusively in the applicable supply
contract or the General Conditions of Sale
and General Conditions of Service of
Vaisala.
This product contains software developed
by Vaisala or third parties. Use of the
software is governed by license terms and
conditions included in the applicable
supply contract or, in the absence of
separate license terms and conditions, by
the General License Conditions of Vaisala
Group.
Table of Contents
English.............................................................................................................................................5
Deutsch........................................................................................................................................ 23
Français.........................................................................................................................................41
日本語...............................................................................................................59
中文.................................................................................................................. 77
Drilling Template........................................................................................................................95
3
4 M211484EN-G
Product Overview
The Vaisala Wi-Fi Data Logger HMT140 measures relative humidity and temperature using the
connected probe and analog signals – RTD, Voltage, Current Loops, and Boolean Contacts. It is
powered with 3 × 3.6 Volt batteries, outputs data via wireless transmitter, and comes with or
without an optional LCD display.
The HMT140 series of transmission devices supports several dierent signal measurements.
Available channel types are Relative Humidity (RH) and Temperature (T), RTD, Boolean
Contact, Voltage, and Current. The channel types are set at the factory during time of order.
Other wireless parameters can be changed afterwards using the configuration port and the
included software.
The HMT140 is available either with a fixed probe directly attached to the data logger housing
or a remote probe with dierent cable lengths (3/5/10 m (9.8/16.4/32.8 ft)).
The HMT140 is also available with an optional LCD display without backlight. It is typically
installed mounted on the wall with up to 4 screws (not included). Duct installation kits are
available as accessories. There is a drilling template inside the back cover of this guide to help
you position the screws correctly.
For more information on the HMT140 Data Logger, see the HMT140 User Guide (M211488EN),
available at www.vaisala.com/hmt140.
5
ENGLISH
Dimensions
145 [5.7]72 [2.8]
120 [4.7]
37 [1.5]
124 [4.9]
Ø 12 [0.47]
116 [4.6]
mm
[in]
Figure 1 HMT140 Data Logger Dimensions
6 M211484EN-G
46 [1.8]
39 [1.5]
4 × M4
91 [3.6]
92 [3.6]
78 [3.1]
mm
[in]
Figure 2 HMT140 Wall Assembly Dimensions
7
ENGLISH
Opening Data Logger Cover
1. If the device is not mounted already, hold it against a flat surface.
2. Push on the cover with your thumb, and pull the bottom part of the cover towards
yourself.
8 M211484EN-G
Probe Assembly with Duct Installation Kit
1 2 3
1 Probe (HMP110)
2 Duct installation kit
3 Probe cable
1. Pass the probe cable through the plastic pipe of the duct installation kit.
2. Connect the probe cable to the HMP110.
42
1
2
3
Ø24
Ø3.2
1 Mounting screw
2 Tension screw
3 HMP110 assembled in duct installation
kit plastic pipe
To attach the probe assembly to the duct:
1. Drill the holes for the duct installation kit as follows:
Use a 24-mm (0.9-in) drill bit to drill a hole in the wall.
Use a 3.2-mm (0.1-in) drill bit to make 4 holes for the installation screws (four
ST4.2×16-C-Z DIN 7981 screws). The holes should be arranged in a square around the
24-mm (0.9-in) hole, at a distance of 42 mm (2 in) from each other.
2. Mount the probe holder using the screws.
3. Adjust the depth of the plastic pipe and tighten the screw to lock the probe in place.
9
ENGLISH
29
[1.1]
59 [2.3]
20
[0.79]
Ø 15 [0.59]
205 [8.1]
42 [1.7]
53 [2.1]
Ø 22
[0.87]
[0.24]
[0.47]
266 [10]
mm
[in]
42 [1.7]
42 [1.7]
Figure 3 Dimensions of HMP110 Probe with Duct Installation Kit
10 M211484EN-G
Connections and Wiring
1
2
3
3
4
5
6
7
9
8
Figure 4 HMT140 Components Diagram
1
CONFIG header
2 SERVICE button
3 Infrared sensor
4 Status LED
5 LCD
6 Power switch for battery
7 Cable grommet (rubber plug)
8 Cable gland
9 Channel wire terminals
11
ENGLISH
Figure 5 RTD Wiring Diagram
Figure 6 RTD and Boolean Wiring Diagram
Figure 7 Boolean Wiring Diagram
12 M211484EN-G
Figure 8 Voltage Wiring Diagram
Figure 9 Current Wiring Diagram
13
ENGLISH
Installing HMT140 Utility Software
Install the Vaisala HMT140 Utility software to configure the HMT140 for use on your network.
The software is compatible with Windows 10 (64-bit), Windows 8.1 (64-bit), Windows 7 (32-bit
and 64-bit), Windows XP (32-bit), Windows Server 2008, and Windows Server 2012.
To install the HMT140 Utility software:
1. Disconnect the HMT140 Configuration Cable if it is already connected to your PC.
2. Insert the CD that came with the HMT140 Configuration Cable, or download the software
from www.vaisala.com/hmt140.
3. Run the HMT140 Utility installation program (HMT140UtilitySetup.exe).
4. Accept the license agreement terms, then select Next to accept the installation default
settings. The installation of the driver may take several minutes.
5. After the software is installed, connect the HMT140 Configuration Cable to a USB port on
your PC to complete the installation setup.
6. When the installation is complete, disconnect the cable from your PC. You are now ready
to set up and configure your device.
If you wish to remove the Vaisala HMT140 Utility software at a later date, use the uninstall
function in the Windows Control Panel.
14 M211484EN-G
Setting up HMT140 Hardware
To set up the HMT140 for the first time:
1. Make sure the HMT140 Configuration Cable is disconnected from your PC.
2. Open the HMT140 cover (see Opening Data Logger Cover (page 8)).
3. Make sure the power switch on the device is in the OFF position.
4. Install 3 × 3.6 V lithium batteries as indicated in Figure 4 (page 11).
Use only batteries recommended by Vaisala (see HMT140 User Guide,
M211488EN).
5. Turn the power switch on the device to the
ON position, and wait for the setup process to
complete (5 seconds).
15
ENGLISH
Connecting HMT140 to Your Network
After you install the HMT140 Utility and set up the HMT140, use the Configuration Cable and
HMT140 Utility software to connect and configure the HMT140 for your network.
To configure the HMT140 for your network with standard settings:
1. Make sure the HMT140 cover is open and the correct batteries are installed.
2. Turn on the HMT140 and wait 5 seconds to make sure the startup process completes.
3. Plug the USB connector on the HMT140 Configuration Cable to the computer, if it is not
already connected.
4. From the Start menu, open the HMT140 Utility.
5. Connect the end of the HMT140 Configuration Cable labeled USB Wi-Fi Programmer to
the 4-pin HMT140 CONFIG header.
6. On the HMT140 Utility Setup tab, select Retrieve.
7. If the IP address is static, uncheck the IP Address Dynamically Assigned option and enter
the network parameters. Otherwise leave this option checked.
8. Enter the parameters for the WLAN (wireless LAN router settings).
9. Enter the Destination IP address/name and port number of the host (viewLinc server
settings).
If you do not know the information required, request these parameters from
your network administrator.
10. Select the Sensor tab and select a Transmit Period Time and number of Tries, or accept
the default settings.
11. Select Update. The HMT140 Utility will configure the HMT140.
12. Select the Status tab to check that the settings appear as configured. Select Get if the
settings do not appear automatically.
13. Disconnect the HMT140 Configuration Cable from the HMT140 and wait for 8 seconds.
14. Press the SERVICE button on the HMT140. Optional: Enter a name for the saved settings
on the Setup tab in the Saved Settings field, and select the disk icon to save.
To configure custom settings for the HMT140, see HMT140 User Guide, M211488EN.
Refer to the Vaisala viewLinc product guides for more information about
using HMT140 devices with the Vaisala viewLinc continuous monitoring
system.
16 M211484EN-G
Quick Reference Information
Remove batteries before shipping.WARNING!
To replace the batteries and reset the battery meter of the HMT140:
1. Open the HMT140 cover.
2. Turn the power switch o.
3. Remove the batteries by pulling them up firmly. If the battery does not come out easily,
one of the battery tabs can be pulled out gently while pulling up.
4. After replacing the batteries (use only recommended batteries), press the SERVICE
button for about 10 seconds and turn the power switch ON. The LED indicator flashes and
the unit beeps several times.
5. Release the SERVICE button.
To trigger data packet transmission using the IR Sensor:
1. Make sure the HMT140 Configuration Cable is disconnected from the HMT140. This will set
the HMT140 to Sensor mode.
2. Turn on the HMT140 and wait for the display to turn on, then o.
3. Place your hand over the IR Sensor and leave it covered for 5 seconds. The display will
turn o and then on again, and will then beep twice to confirm transmission. The HMT140
will then attempt to connect and transmit data to the host server.
17
ENGLISH
Technical Data
Table 1 HUMICAP Humidity and Temperature Probe HMP110
Property Description/Value
Relative Humidity
Measurement range 0 ... 100 %RH
Accuracy (incl. Non-Linearity, Hysteresis, and Repeatability):
At temperature range 0 ... +40 °C (+32 ... +104 °F):
0 ... 90 %RH ±1.5 %RH
90 ... 100 %RH ±2.5 %RH
At temperature range -40 ... 0 °C, +40 ... +80 °C (-40 ... +32 °F, +104 ... +176 °F):
0 ... 90 %RH ±3.0 %RH
90 ... 100 %RH ±4.0 %RH
Factory calibration uncertainty at +20 °C
(+68 °F)
±1.5 %RH
Humidity sensor Vaisala HUMICAP 180R
Stability ±2 %RH over 2 years
Temperature
Measurement range -40 ... +80 °C (-40 ... +176 °F)
Accuracy over Temperature Range:
+15 ... +25 °C (+59 ... +77 °F) ±0.2 °C (±0.36 °F)
0 ... +15 °C, +25 ... +40 °C (+32 ... 59 °F,
+77 ... +104 °F)
±0.25 °C (±0.45 °F)
-40 ... 0 °C, +40 ... +80 °C (-40 ... +32 °F,
+104 ... +176 °F)
±0.4 °C (±0.72 °F)
Temperature sensor Pt1000 RTD 1/3 Class B IEC 751
HMP110 probe -40 ... +80 °C (-40 ... +176 °F)
Storage temperature -50 ... +70 °C (-58 ...+158 °F)
Table 2 Temperature Probes
Property Description/Value
Resistive Temperature Input Signals
Temperature sensor Pt100 RTD / 4 wire, Class A IEC 751
Input impedance 5.1K Ω
Measurement range -196 ... +90 °C (-320.8 ... +194 °F)
Accuracy over Temperature Range:
18 M211484EN-G
Property Description/Value
-196 ... -90 °C (-320.8 ... -130 °F)
-90 ... -30 °C (-130 ... -22 °F)
-30 ... 0 °C (-22 ... +32 °F)
0 ... +50 °C (+32 ... +122 °F)
+50 ... +90 °C (+122 ... +194 °F)
±2.5 °C (±4.5 °F)
±0.75 °C (±1.35 °F)
±0.5 °C (±0.9 °F)
±0.25 °C (±0.45 °F)
±0.75 °C (±1.35 °F)
Table 3 Analog Inputs
Property Description/Value
Current Input Signals 0 ... 22 mA
Resolution 0.67 μA
Accuracy ±0.15 % F.S. at +25 °C (+77 °F)
Input impedance 62 Ω
Overload protection 40 mA
Voltage Input Signals 0 ... 5 V, 0 ... 10 V
Resolution 0.0034 % F.S.
Accuracy ±0.15 % F.S. at +25 °C (+77 °F)
Input impedance 37K Ω
Overload protection 50 V max.
Isolation One common per logger
Contact Inputs Open/Closed with magnetic reed switch cable
connections (Dry Contact)
Table 4 Measurement Variants
Model Measurement Channels
HMT141 1 Temperature and 1 Humidity Channel
HMT143 2 RTD Temperature Channels
HMT144 2 Voltage Inputs (0 ... 5 VDC)
HMT145 2 Voltage Inputs (0 ... 10 VDC)
HMT146 2 Current Inputs (0 ... 20 mA)
HMT147 2 Door Contacts
HMT148 1 RTD Temperature and 1 Door Contact
HMT14D 1 RTD Temperature and 1 Current Input (0 ... 20 mA)
HMT14E 1 RTD Temperature and 1 Voltage Input (0 ... 5 VDC)
HMT14F 1 RTD Temperature and 1 Voltage Input (0 ... 10 VDC)
19
ENGLISH
Table 5 Operating Environment
Property Description/Value
Operating Temperature
Data logger body, no display -40 ... +60 °C (-40 ... +140 °F)
Data logger body, with display -20 ... +60 °C (-4 ... +140 °F)
IP rating IP65
Maximum operating altitude 2000 m (6 500 ft)
Table 6 General Specifications
Property Description/Value
Memory 3,060 samples
Sample rate User selectable (interval: 2 … 60 minutes)
Display (optional) 128 × 64 resolution full graphics
B&W display without backlight
Table 7 Mechanical Specifications
Property Description/Value
Weight (with battery/without probe) 300 g (10.6 oz)
Screw terminals 26 AWG ... 20 AWG
HMP110 probe interface 4-pin M8 female panel connector
HMP110 probe cable lengths 3 m, 5 m, and 10 m (9.8 ft, 16.4 ft, and 32.8 ft)
Hermetic Door Switch Sensor cable length 7.6 m (24.9 ft)
Material
Data logger housing PBT plastic
Display window PC plastic
HMP110 probe body Stainless steel (AISI 316)
HMP110 probe grid filter Chrome coated ABS plastic
RTD Temperature Sensor
Sensor tip material Stainless steel (AISI 316)
Sensor tip length 50.8 mm (2 in)
Sensor tip diameter 4.76 mm (0.19 in)
Cable length 5 m (16.4 ft)
20 M211484EN-G
/