Lovibond® Water Testing
Tintometer® Group
Manual of Methods
MD 100 • MD 110 • MD 200
Aluminium | Chloride | Copper | DEHA | Hydrazine | Iron |
Oxygen | Phosphate | Polyacrylate | Silica
EN
Manual of Methods
Page 4
DE
Methodenhandbuch
Seite 60
ES
Manual de Métodos
Página 116
FR
Méthodes Manuel
Page 172
IT
Manuale dei Metodi
Pagina 228
PT
Métodos Manual
Página 284
NL
Handboek Methoden
Zijde 340
ZH
方法手册
Page 396
NL Handboek van Methoden 01/20EN Handbook of Methods 01/20
KS4.3 T
0.1 - 4 mmol/l KS4.3
Acid / Indicator
20
S:4.3
Instrument specific information
The test can be performed on the following devices. In addition, the required cuvette
and the absorption range of the photometer are indicated.
Instrument Type Cuvette λMeasuring Range
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l KS4.3
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
Alka-M-Photometer Tablet / 100 513210BT
Alka-M-Photometer Tablet / 250 513211BT
Application List
Waste Water Treatment
Drinking Water Treatment
Raw Water Treatment
Notes
1. The terms Alkalinity-m, m-Value, total alkalinity and Acid demand to KS4.3 are iden-
tical.
2. For accurate results, exactly 10 ml of water sample must be used for the test.
Display in the MD
100 / MD 110 /
MD 200
Bar code for the
detection of the
methods
Method name
Measuring range
Chemical Method
Method number
Language codes
ISO 639-1 Revision status
KS4.3 T / 20
How to use
ENMethod Reference Book4
EN
Implementation of the provision Acid capacity KS4.3 with Tablet
Select the method on the device
For this method, no ZERO measurements are to be carried out with the following de-
vices: XD 7000, XD 7500
10 ml
Fill 24 mm vial with 10 ml
sample.
Close vial(s).
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Dissolve tablet(s) by inver-
ting.
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Test
Press the TEST (XD:
START) button.
The result in Acid Capacity KS4.3 appears on the display.
Performing test
procedure
KS4.3 T / 20
NL Handboek van Methoden 01/20EN Handbook of Methods 01/20
How to use
ENMethod Reference Book 5
EN
Aluminium PP / M50
ENMethod Reference Book1.06
EN
Aluminium PP M50
0.01 - 0.25 mg/L Al AL
Eriochrom Cyanine R
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
VARIO Aluminium Reagent, Set F20 1 Pieces 535000
Preparation
1. To get accurate results the sample temperature must be between 20 °C and 25 °C.
2. To avoid errors caused by contamination, rinse the vial and the accessories
with Hydrochloric acid (approx. 20%) before the analysis. Then rinse them with
deionised water.
Aluminium PP / M50
ENMethod Reference Book1.0 7
EN
Determination of Aluminium with Vario Powder Pack
Select the method on the device.
Blank
20 mL
Prepare two clean 24 mm
vials. Mark one as a blank.
Put 20 mL sample in
100 mL measuring beaker
Add Vario ALUMINIUM ECR
F20 powder pack.
9
4
5
6
7
8
10
11 12
1
2
3
30 sec
Dissolve the powder by
mixing.
Press the ENTER button. Wait for 30 second(s)
reaction time.
Blank
1
Add Vario HEXAMINE
F20 powder pack.
Dissolve the powder by
mixing.
Place 1 drops Vario
ALUMINIUM ECR Masking
Reagent in the blank.
Aluminium PP / M50
ENMethod Reference Book1.08
EN
24 mm
10 mL
Place 10 mL pre-treated
sample in each vial.
Close vial(s). Press the ENTER button.
9
4
5
6
7
8
10
11 12
1
2
3
5 min
Blank
Zero
Wait for 5 minute(s)
reaction time.
Place blank in the sample
chamber. Pay attention to
the positioning.
Press the ZERO button.
Sample
Test
Remove the vial from the
sample chamber.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
The result in mg/L Aluminium appears on the display.
Aluminium PP / M50
ENMethod Reference Book1.0 9
EN
Analyses
The following table identifies the output values can be converted into other citation
forms.
Unit Cite form Scale Factor
mg/l Al 1
mg/l Al2O31.8894
Chemical Method
Eriochrom Cyanine R
Appendix
Interferences
Removeable Interferences
A low test result may be given in the presence of Fluorides and Polyphosphates. The
effect of this is generally insignificant unless the water has fluoride added artificially.
In this case, the following table should be used to determine the actual concentration
of aluminium.
Fluoride Displayed value: Aluminium [mg/L]
[mg/L
F]
0.05 0.10 0.15 0.20 0.25 0.30
0.2 0.05 0.11 0.16 0.21 0.27 0.32
0.4 0.06 0.11 0.17 0.23 0.28 0.34
0.6 0.06 0.12 0.18 0.24 0.30 0.37
0.8 0.06 0.13 0.20 0.26 0.32 0.40
1.0 0.07 0.13 0.21 0.28 0.36 0.45
1.5 0.09 0.20 0.29 0.37 0.48 ---
Bibliography
Richter, F. Fresenius, Zeitschrift f. anal. Chemie (1943) 126: 426
According to
APHA Method 3500-Al B
Chloride L (B) / M92
ENMethod Reference Book1.010
EN
Chloride L (B) M92
0.5 - 20 mg/L Cl-CL-
Mercury Thiocyanate / Iron Nitrate
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
Chloride Reagent Set 1 Pieces 56R018490
Chloride L (B) / M92
ENMethod Reference Book1.0 11
EN
Determination of Chloride with liquid reagent
Select the method on the device.
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
20
Hold cuvettes vertically
and add equal drops by
pressing slowly.
Add 20 drops
KS251 (Chloride Reagenz
A).
Close vial(s).
Chloride L (B) / M92
ENMethod Reference Book1.012
EN
20
Invert several times to mix
the contents.
Add 20 drops
KS253 (Chloride Reagenz
B).
Close vial(s).
Sample
Test
Invert several times to mix
the contents.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
9
4
5
6
7
8
10
11 12
1
2
3
5 min
Wait for 5 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L Chloride appears on the display.
Chloride L (B) / M92
ENMethod Reference Book1.0 13
EN
Analyses
The following table identifies the output values can be converted into other citation
forms.
Unit Cite form Scale Factor
mg/l Cl-1
mg/l NaCl 1.65
Chemical Method
Mercury Thiocyanate / Iron Nitrate
Appendix
Interferences
Persistant Interferences
1. Reducing substances such as sulfite and thiosulfate, that can reduce iron (III) to
iron (II) or mercury (II) to mercury (I) may interfere. Cyanide, Iodine and Bromide
give a positive intereference.
Derived from
DIN 15682-D31
DIN ISO 15923-1 D49
Copper T / M150
ENMethod Reference Book1.014
EN
Copper T M150
0.05 - 5 mg/L Cua) Cu
Biquinoline
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
Copper No. 1 Tablet / 100 513550BT
Copper No. 1 Tablet / 250 513551BT
Copper No. 2 Tablet / 100 513560BT
Copper No. 2 Tablet / 250 513561BT
Set Copper No. 1/No. 2 100 Pc.#100 each 517691BT
Set Copper No. 1/No. 2 250 Pc.#250 each 517692BT
Preparation
1. Strong alkaline or acidic water samples must be adjusted to pH 4 to 6 before
analysis.
Copper T / M150
ENMethod Reference Book1.0 15
EN
Determination of Copper, free with tablet
Select the method on the device.
In addition, choose the test: free
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
24 mm
Add COPPER No. 1 tablet
.
Crush tablet(s) by rotating
slightly.
Close vial(s).
Copper T / M150
ENMethod Reference Book1.016
EN
Sample
Test
Dissolve tablet(s) by
inverting.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
9
4
5
6
7
8
10
11 12
1
2
3
2 min
Wait for 2 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L free Copper appears on the display.
Determination of Copper, total with tablet
Select the method on the device.
In addition, choose the test: total
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Copper T / M150
ENMethod Reference Book1.0 17
EN
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
24 mm
Add COPPER No. 1 tablet
.
Crush tablet(s) by rotating
slightly and dissolve.
Add COPPER No. 2 tablet .
24 mm
Crush tablet(s) by rotating
slightly.
Close vial(s). Dissolve tablet(s) by
inverting.
Copper T / M150
ENMethod Reference Book1.018
EN
Sample
Test
9
4
5
6
7
8
10
11 12
1
2
3
2 min
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
Wait for 2 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L total Copper appears on the display.
Copper T / M150
ENMethod Reference Book1.0 19
EN
Chemical Method
Biquinoline
Appendix
Interferences
Persistant Interferences
1. Cyanide CN- and Silver Ag+ interfere with the test result.
Method Validation
Limit of Detection 0.05 mg/L
Limit of Quantification 0.15 mg/L
End of Measuring Range 5 mg/L
Sensitivity 3.8 mg/L / Abs
Confidence Intervall 0.026 mg/L
Standard Deviation 0.011 mg/L
Variation Coefficient 0.42 %
Bibliography
Photometrische Analyse, Lange/Vedjelek, Verlag Chemie 1980
a) determination of free, combined and total | # including stirring rod, 10 cm
DEHA PP / M167
ENMethod Reference Book1.020
EN
DEHA PP M167
0.02 - 0.5 mg/L DEHA DEHA
PPST
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
VARIO DEHA Reagent Set 1 Pieces 536000
Preparation
1. To avoid errors caused by iron deposits, rinse the glassware with Hydrochloric acid
(approx. 20%) before the analysis and then rinse with deionised water.
Remarks
1. Because the reaction depends on temperature, the temperature must be maintained
at 20 °C ± 2 °C.
2. Keep the sample vial in the dark or in the sample chamber during colour
development time. If the Reagent solution is exposed to UV-light (sunlight) it causes
high measurement results.
DEHA PP / M167
ENMethod Reference Book1.0 21
EN
Determination of DEHA (N,N-Diethylhydroxylamine) with Vario
Powder Pack and Fluid Reagent
Select the method on the device.
Blank
Blank
10 mL
Sample
10 mL
Prepare two clean 24 mm
vials. Mark one as a blank.
Put 10 mL deionised
water in the blank.
Put 10 mL sample in the
sample vial.
SampleBlank
Add a Vario OXYSCAV
1 Rgt powder pack in
each vial.
Close vial(s). Invert several times to mix
the contents.
SampleBlank
0.2 mL
Add 0.2 mL Vario DEHA
2 Rgt solution to each
vial.
Close vial(s). Invert several times to mix
the contents.
/