Lovibond 00386460 - MD1x0/200 - Ozone T ユーザーマニュアル

タイプ
ユーザーマニュアル
Lovibond® Water Testing
Tintometer® Group
Manual of Methods
MD 100 • MD 110 • MD 200
Ozone
EN
Manual of Methods
Page 4
DE
Methodenhandbuch
Seite 16
ES
Manual de Métodos
Página 28
FR
Méthodes Manuel
Page 40
IT
Manuale dei Metodi
Pagina 52
PT
Métodos Manual
Página 64
NL
Handboek Methoden
Zijde 76
ZH
方法手册
Page 88
NL Handboek van Methoden 01/20EN Handbook of Methods 01/20
KS4.3 T
0.1 - 4 mmol/l KS4.3
Acid / Indicator
20
S:4.3
Instrument specific information
The test can be performed on the following devices. In addition, the required cuvette
and the absorption range of the photometer are indicated.
Instrument Type Cuvette λMeasuring Range
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l KS4.3
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
Alka-M-Photometer Tablet / 100 513210BT
Alka-M-Photometer Tablet / 250 513211BT
Application List
Waste Water Treatment
Drinking Water Treatment
Raw Water Treatment
Notes
1. The terms Alkalinity-m, m-Value, total alkalinity and Acid demand to KS4.3 are iden-
tical.
2. For accurate results, exactly 10 ml of water sample must be used for the test.
Display in the MD
100 / MD 110 /
MD 200
Bar code for the
detection of the
methods
Method name
Measuring range
Chemical Method
Method number
Language codes
ISO 639-1 Revision status
KS4.3 T / 20
How to use
ENMethod Reference Book4
EN
Implementation of the provision Acid capacity KS4.3 with Tablet
Select the method on the device
For this method, no ZERO measurements are to be carried out with the following de-
vices: XD 7000, XD 7500
10 ml
Fill 24 mm vial with 10 ml
sample.
Close vial(s).
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Dissolve tablet(s) by inver-
ting.
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Test
Press the TEST (XD:
START) button.
The result in Acid Capacity KS4.3 appears on the display.
Performing test
procedure
KS4.3 T / 20
NL Handboek van Methoden 01/20EN Handbook of Methods 01/20
How to use
ENMethod Reference Book 5
EN
Ozone T / M300
ENMethod Reference Book1.06
EN
Ozone T M300
0.02 - 2 mg/L O3O3
DPD / Glycine
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
DPD No.1 Tablet / 100 511050BT
DPD No. 1 Tablet / 250 511051BT
DPD No. 1 Tablet / 500 511052BT
DPD No. 3 Tablet / 100 511080BT
DPD No. 3 Tablet / 250 511081BT
DPD No. 3 Tablet / 500 511082BT
DPD No. 1 High Calcium e) Tablet / 100 515740BT
DPD No. 1 High Calcium e) Tablet / 250 515741BT
DPD No. 1 High Calcium e) Tablet / 500 515742BT
DPD No. 3 High Calcium e) Tablet / 100 515730BT
DPD No. 3 High Calcium e) Tablet / 250 515731BT
DPD No. 3 High Calcium e) Tablet / 500 515732BT
Glycine f) Tablet / 100 512170BT
Glycine f) Tablet / 250 512171BT
Set DPD No. 1/No. 3 100 Pc.#100 each 517711BT
Set DPD No. 1/No. 3 250 Pc.#250 each 517712BT
Set DPD No. 1/No. 3 High Calcium 100 Pc. #100 each 517781BT
Set DPD No. 1/No. 3 High Calcium 250 Pc. #250 each 517782BT
Set DPD No. 1/Glycine 100 Stck. #100 each 517731BT
Set DPD No. 1/Glycine 250 Stck. #250 each 517732BT
Ozone T / M300
ENMethod Reference Book1.0 7
EN
Preparation
1. Cleaning of vials:
As many household cleaners (e.g. dishwasher detergent) contain reducing
substances, the subsequent determination of oxidising agents (e.g. ozone and
chlorine) may show lower results. To avoid measurement errors, the glassware
used should be free of chlorine consumption. To achieve this, all glassware should
be placed in a sodium hypochlorite solution (0.1 g/L) for one hour and then rinsed
thoroughly with deionised water.
2. When preparing the sample, Ozone outgassing, e.g. through the pipette or
shaking, must be avoided. The analysis must take place immediately after taking
the sample.
3. Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7
before the analysis (use 0.5 mol/l Sulphuric acid or 1 mol/l Sodium hydroxide).
Ozone T / M300
ENMethod Reference Book1.08
EN
Determination of Ozone, in presence of Chlorine with tablet
Select the method on the device.
In addition, choose the test: in presence of Chlorine
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
Empty vial except for a few
drops.
24 mm
Add DPD No. 1 tablet . Add DPD No. 3 tablet . Crush tablet(s) by rotating
slightly.
Ozone T / M300
ENMethod Reference Book1.0 9
EN
Fill up vial with sample to
the 10 mL mark.
Close vial(s). Dissolve tablet(s) by
inverting.
Sample
Test
9
4
5
6
7
8
10
11 12
1
2
3
2 min
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
Wait for 2 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
Remove the vial from the
sample chamber.
Empty vial. Thoroughly clean the vial and
vial cap.
Ozone T / M300
ENMethod Reference Book1.010
EN
24 mm
10 mL
24 mm
Fill a second vial with
10 mL sample .
Add GLYCINE tablet. Crush tablet(s) by rotating
slightly.
Close vial(s). Dissolve tablet(s) by
inverting.
Add one DPD No. 1 tablet
and one DPD No. 3 tablet
straight from the foil into the
first cleaned cuvette
24 mm
Crush tablet(s) by rotating
slightly.
Fill prepared vial with
prepared glycine solution.
Close vial(s).
Ozone T / M300
ENMethod Reference Book1.0 11
EN
Sample
Test
Dissolve tablet(s) by
inverting.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
9
4
5
6
7
8
10
11 12
1
2
3
2 min
Wait for 2 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L Ozone; mg/l total chlorine appears on the display.
Determination of Ozone, in absence of chlorine with tablet
Select the method on the device.
In addition, choose the test: without Chlorine
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Ozone T / M300
ENMethod Reference Book1.012
EN
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
Empty vial except for a few
drops.
24 mm
Add DPD No. 1 tablet . Add DPD No. 3 tablet . Crush tablet(s) by rotating
slightly.
Fill up vial with sample to
the 10 mL mark.
Close vial(s). Dissolve tablet(s) by
inverting.
Ozone T / M300
ENMethod Reference Book1.0 13
EN
Sample
Test
9
4
5
6
7
8
10
11 12
1
2
3
2 min
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
Wait for 2 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L Ozone appears on the display.
Ozone T / M300
ENMethod Reference Book1.014
EN
Analyses
The following table identifies the output values can be converted into other citation forms.
Unit Cite form Scale Factor
mg/l O31
mg/l Cl21.4771
Chemical Method
DPD / Glycine
Appendix
Interferences
Persistant Interferences
1. All oxidising agents in the samples react like chlorine, which leads to higher results.
2. Concentrations above 6 mg/L Ozone can lead to results within the measuring
range of up to 0 mg/L. In this case, the water sample must be diluted. 10 ml of the
diluted sample should be mixed with the reagent and the measurement taken again
(plausibility test).
Bibliography
Colorimetric Chemical Analytical Methods, 9th Edition, Lovibond
Derived from
DIN 38408-3:2011-04
e) alternative reagent, used instead of DPD No.1/No.3 in case of turbidity in the water sample caused by high
concentration of calcium and/or high conductivity | f) additionally required for determination of bromine, chlorine dioxide
and ozone in the presence of chlorine | # including stirring rod, 10 cm
DE Methodenhandbuch 01/20
KS4.3 T
0,1 - 4 mmol/l KS4.3
Säure / Indikator
20
S:4.3
Instrumentspezifische Informationen
Der Test kann auf den folgenden Geräten durchgeführt werden. Zusätzlich sind die
benötigte Küvette und der Absorptionsbereich der Photometer angegeben.
Geräte Küvette λMessbereich
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0,1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0,1 - 4 mmol/l KS4.3
Material
Benötigtes Material (zum Teil optional):
Reagenzien Form/Menge Bestell-Nr.
Alka-M-Photometer Tablette / 100 513210BT
Alka-M-Photometer Tablette / 250 513211BT
Anwendungsbereich
Abwasserbehandlung
Trinkwasseraufbereitung
Rohwasserbehandlung
Anmerkungen
1. Die Begrie Alkalität-m, m-Wert, Gesamtalkalität und Säurekapazität KS4.3 sind
identisch.
2. Die exakte Einhaltung des Probevolumens von 10 ml ist für die Genauigkeit des
Analysenergebnisses entscheidend.
Displayanzeige
im MD 100 MD 110
/ MD 200
Barcode zur
Methodenerkennung
Methoden Name
Messbereich
Chemische
Methode
Methodennummer
Sprachkürzel nach
ISO 639-1 Revisionsstand
KS4.3 T / 20
Leitfaden für den Anwender
ENMethod Reference Book16
DE
Durchführung der Bestimmung Säurekapazität KS4.3 mit Tablette
Die Methode im Gerät auswählen.
Für diese Methode muss bei folgenden Geräten keine ZERO-Messung durchgeführt
werden: XD 7000, XD 7500
10 ml
24-mm-Küvette mit 10 ml
Probe füllen.
Küvette(n) verschließen.
Sample
Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Tablette(n) durch Um-
schwenken lösen.
Sample
Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Test
Taste TEST (XD: START)
drücken.
In der Anzeige erscheint das Ergebnis als Säurekapazität KS4.3.
Durchführung der
Messung
KS4.3 T / 20
DE Methodenhandbuch 01/20
Leitfaden für den Anwender
ENMethod Reference Book 17
DE
Ozon T / M300
DEMethod Reference Book1.018
DE
Ozon T M300
0,02 - 2 mg/L O3O3
DPD / Glycin
Material
Benötigtes Material (zum Teil optional):
Reagenzien Form/Menge Bestell-Nr.
DPD No.1 Tablette / 100 511050BT
DPD No. 1 Tablette / 250 511051BT
DPD No. 1 Tablette / 500 511052BT
DPD No. 3 Tablette / 100 511080BT
DPD No. 3 Tablette / 250 511081BT
DPD No. 3 Tablette / 500 511082BT
DPD No. 1 High Calcium e) Tablette / 100 515740BT
DPD No. 1 High Calcium e) Tablette / 250 515741BT
DPD No. 1 High Calcium e) Tablette / 500 515742BT
DPD No. 3 High Calcium e) Tablette / 100 515730BT
DPD No. 3 High Calcium e) Tablette / 250 515731BT
DPD No. 3 High Calcium e) Tablette / 500 515732BT
Glycine f) Tablette / 100 512170BT
Glycine f) Tablette / 250 512171BT
Set DPD No. 1/No. 3#je 100 517711BT
Set DPD No. 1/No. 3#je 250 517712BT
Set DPD No. 1/No. 3 High Calcium#je 100 517781BT
Set DPD No. 1/No. 3 High Calcium#je 250 517782BT
Set DPD No. 1/Glycine #je 100 517731BT
Set DPD No. 1/Glycine #je 250 517732BT
Ozon T / M300
DEMethod Reference Book1.0 19
DE
Vorbereitung
1. Reinigung der Küvetten:
Da viele Haushaltsreiniger (z.B. Geschirrspülmittel) reduzierende Stoffe enthalten,
kann es bei der nachfolgenden Bestimmung von Oxidationsmitteln (z.B. Ozon,
Chlor) zu Minderbefunden kommen. Um diesen Messfehler auszuschließen, sollten
die Glasgeräte chlorzehrungsfrei sein. Dazu weden die Glasgeräte für eine Stunde
unter Natriumhypochloritlösung (0,1 g/L) aufbewahrt und danach gründlich mit VE-
Wasser gespült.
2. Bei der Probenvorbereitung muss das Ausgasen von Ozon, z.B. durch Pipettieren
und Schütteln vermieden werden. Die Analyse muss unmittelbar nach der
Probennahme erfolgen.
3. Stark alkalische oder saure Wässer müssen vor der Analyse in einen pH-Bereich
zwischen 6 und 7 gebracht werden (mit 0,5 mol/l Schwefelsäure bzw. 1 mol/l
Natronlauge).
Ozon T / M300
DEMethod Reference Book1.020
DE
Durchführung der Bestimmung Ozon, neben Chlor mit Tablette
Die Methode im Gerät auswählen.
Wählen Sie zudem die Bestimmung: neben Chlor
10 mL
Sample
24-mm-Küvette mit 10 mL
Probe füllen.
Küvette(n) verschließen. Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Zero
Taste ZERO drücken. Küvette aus dem
Messschacht nehmen.
Die Küvette bis auf einige
Tropfen entleeren.
24 mm
Eine DPD No. 1 Tablette
zugeben.
Eine DPD No. 3 Tablette
zugeben.
Tablette(n) unter leichter
Drehung zerdrücken.
Ozon T / M300
DEMethod Reference Book1.0 21
DE
Küvette bis zur 10-mL-
Marke mit der Probe
auffüllen.
Küvette(n) verschließen. Tablette(n) durch
Umschwenken lösen.
Sample
Test
9
4
5
6
7
8
10
11 12
1
2
3
2 min
Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Taste TEST (XD: START)
drücken.
2 Minute(n) Reaktionszeit
abwarten.
Nach Ablauf der Reaktionszeit erfolgt automatisch die Messung.
Küvette aus dem
Messschacht nehmen.
Küvette entleeren. Die Küvette und den
Küvettendeckel gründlich
reinigen.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99

Lovibond 00386460 - MD1x0/200 - Ozone T ユーザーマニュアル

タイプ
ユーザーマニュアル