Niigata seiki JLVC-30 ユーザーマニュアル

タイプ
ユーザーマニュアル

このマニュアルも適しています

PART IDENTIFICATION AND FUNCTION
SAFETY PRECAUTIONS Please Observe
To prevent harm to yourself or others, and to prevent damage
to property, always follow the procedures marked by the
following symbols.
Indicates risk of personal injury or property damage if not followed.
WARNING
Denotes a prohibition -
You MUST NOT do
Denotes a requirement -
You MUST do
Read the manual and follow the directions.Use of product other than as described in the manual
may cause accident.
Use only for measuring.Use for any purpose other than measuring may
damage or wear the instrument. Improper use may
also cause accident.
Use only in suitable environment.Do not use in locations that are humid, wet, very hot,
cold, or in direct sunlight.
Due to large size and weight, please use care and be
aware of your surroundings during use.
Handle with Care.Do not drop or subject instrument to excessive shock.
Do not place under heavy objects. Improper handling may
cause damage or poor accuracy.
Do not scratch instrument, for example by writing ID number.
Jaw tips are sharp Handle with Care.
Careless handling may cause injury.
Do not disassemble or modify.Do not remove the vernier slide, please do not attempt to
disassemble or modify as it may cause damage or poor
accuracy.
CALIBRATION
In order to maintain instrument accuracy, it is recommended that
accuracy is confirmed through calibration on a periodic basis.
Wear of measuring surfaces from repeated use may affect accuracy and
periodic accuracy checks should be performed.
Position measured object between these surfaces.
Measures external dimensions. Move Slide to open and close.
Measures the height of steps.
Measures internal dimensions. Move Slide to open and close.
Position surfaces inside object to be measured.
Locks the Vernier Slide.
Tighten to use Fine Adjustment Feed Wheel.
Indicates measured value in 1mm graduations.
Feed screw for fine adjust of Slide.
Scale for measuring with a resolution of 0.05mm.
Refer to following page for usage instructions.
Measured Value = Main Scale Reading + Vernier Scale Reading
Measured Value: 77.00mm
Measured Value: 77.30mm
Measured Value: 77.35mm
EXAMPLE 1
77
0.00
77.00
Reading
EXAMPLE 2
77
0.35
77.35
Reading
EXAMPLE 3
77
0.30
77.30
Reading
Main Scale
(1mm)
Vernier Scale
(0.05mm)
Main Scale
(1mm)
Vernier Scale
(0.05mm)
Main Scale
(1mm)
Vernier Scale
(0.05mm)
For Fine Adjustment, Loosen Lock Screw , and tighten Fine
Adjust Lock . Rotate Fine Adjust Feed wheel .
    
                    
       
PP
          
                     F P
PP
          
    
PP
          
      
          
      

      
          
JLVC-30
300mm
0.08mm
1.16kg
JLVC-60
600mm
0.10mm
3.29kg
0.05mm
Model No.
Max Meas. Length
Length Resolution
Maximum
permissible error
Weight
Outside Meas.
Surfaces
Outside Jaws
Step Meas.
Surfaces
Inside Meas. Surfaces
Slide (Vernier Scale)
Main Scale
Fine Adjust Lock
Lock Screw
Inside Jaws
Fine Adjust Feed
PREPARATION FOR USE
READING THE SCALE PREVENTING ERRORS
In order to prevent measurement error, please note the
following.
Material : Stainless Steel
SPECIFICATIONS
Wipe slide, measuring surfaces, and exterior with
a dry cloth to remove any dirt or oil.
For rust prevention, lightly apply anti-rust oil, or place in rust
preventive bag.
Store with Lock Screw loosened.
If stored with Lock Screw tight, expansion due to temperature
changes may cause excessive force which will cause inaccuracy.
Store in supplied case in a cool, dry location.
Store away from moisture or direct sunlight, and out of reach
of unauthorized personnel.
POSITIONING CALIPER
VIEWING THE SCALE
PREVENTING DEFLECTION
Measured
Object
Jaw
Check
alignment
Due to the large size and weight of the instrument, if not well
supported during use the body may deflect and introduce
measurement error.
We recommend use by more than one person during
measurement for proper support.
Niigata seiki Co., Ltd.
5-3-14, Tsukanome, Sanjo, Niigata, Japan, 955-0055
Tel. : +81-256-33-5522 Fax. : +81-256-33-5518
MAIL intl.sales@niigataseiki.co.jp
URL http://www.niigataseiki.co.jp
Thank you for purchasing Niigata Seiki Long Jaw Calipers.
This product is a precision measurement tool with a main scale and a vernier scale to indicate the measured value.
For safe and proper use, please read this instruction manual before use and follow the procedures described.
Please keep manual where it is accessible to user for future reference.
Keep this manual with the instrument if transferred or leased to a third party.
For inquiries about this product, please contact dealer or Niigata Seiki.
LONG JAW CALIPERS
PRECISION MEASURING TOOLS
JLVC
-
30/JLVC
-
60
Model No.
User Manual
Model JLVC-30 shown.
Before using the caliper, follow the steps below:
Do not apply excessive force to Slide.
Moving the Slide with excessive force may cause damage and measurement error.
Wipe off rust proofing and dirt from measuring surfaces and slide.
Contamination of surfaces may cause measurement error.
Allow the caliper and object to be measured time to reach the same temperature.
A temperature difference between the object to be measured and the
caliper may cause measurement error. Allow enough time for the
temperatures to equilibrate.
Performing an accuracy check.
Jaw Alignment
In closed position, make sure the measuring surfaces on the outside
jaws align properly and there is no gap (no light should leak past.)
Confirm 0 points on scale (see diagram at right)
With jaws closed as in step above, make sure the reading is zero and
the vernier scale graduations align at the 0 and 39 positions as shown.
Scale should be read from
directly above the calipers.
Due to change in height of
vernier scale relative to main
scale, if viewing direction is
not directly above the reading
may have parallax error.
Flat surfaces
Position the caliper
perpendicular to the surfaces
to be measured.Cylindrical surfaces
Position the caliper
perpendicular to the axis of
the cylinder.
Outside Meas. Surfaces
Outside Jaws
Step Meas. Surfaces
Inside Jaws
Inside Meas. Surfaces
Lock Screw
Fine Adjust Lock
Main Scale
Fine Adjust Feed
Slide (Vernier Scale)
AFTER USE CARE STORAGE NOTES
The measured value is determined by reading the main
scale, and then adding the Vernier Scale reading
determined by the position where the vernier and main
scale graduations align.
Measurements made on area of Measurement Surface
closest to the Main Scale will have the highest accuracy.
When taking measurements using the tips of the Jaws,
accuracy depends of measurement technique and may not
meet the stated accuracy specifications.
We provide calibration services.
Please contact agent in country of
purchase to make arrangements.
2312 I398-K1
URL http://www.niigataseiki.co.jp
            
PP
          
                     F P
PP
          
     
PP
          
      
          
      
          
      
          
.
JLVC
-
30/JLVC
-
60
(0256)33
-
5502
JLVC
-
30
300mm
0.08mm
1.16kg
JLVC
-
60
600mm
0.10mm
3.29kg
0.05mm
 
   
 
   
 
   
77.00
1
0.05
77.30
1
0.05
77.35
1
0.05
JLVC
-
30
  • Page 1 1
  • Page 2 2

Niigata seiki JLVC-30 ユーザーマニュアル

タイプ
ユーザーマニュアル
このマニュアルも適しています

他の言語で