Hach sensION+ pH3 ユーザーマニュアル

タイプ
ユーザーマニュアル
DOC022.97.90248
sensION
+ pH3
04/2011, Edition 1
User Manual
Manuel d'utilisation
Manual del usuario
Manual do usuário
用户手册
取扱説明書
사용 설명서
คูมือผูใช
English..................................................................................................................................................................................................3
Français.............................................................................................................................................................................................14
Español..............................................................................................................................................................................................25
Português..........................................................................................................................................................................................36
中文......................................................................................................................................................................................................47
日本語.................................................................................................................................................................................................57
한글......................................................................................................................................................................................................68
ไทย.........................................................................................................................................................................................................79
2
Specifications
Specifications are subject to change without notice.
Specification Details
Dimensions 35 x 20 x 11 cm (13.78 x 7.87 x 4.33 in.)
Weight 1100 g (2.43 lb)
Meter enclosure IP42
Power requirements
(external)
100–240 V, 0.4 A, 47-63 Hz
Meter protection class Class II
Storage temperature –15 to +65 °C (5 to +149 °F)
Operating temperature 0 to 40 °C (41 to 104 °F)
Operating humidity < 80% (non-condensing)
Resolution pH: 0.1/0.01/0.001, ORP: 0.1/1 mV, temperature: 0.1
ºC (0.18 ºF)
Measuring error (± 1 digit) pH: ≤ 0.002, ORP: ≤ 0.2 mV, temperature: ≤ 0.2 ºC
(≤ 0.36 ºF)
Reproducibility (± 1 digit) pH: ± 0.001, ORP: ± 0.1 mV, temperature: ± 0.1 ºC (±
0.18 ºF)
Connections Combined or indicator probe: BNC connector (Imp.
>10
12
Ω); Reference electrode: banana connector;
A.T.C.
type Pt 1000: banana or telephonic connector;
magnetic stirrer: RCA connector
Temperature correction Manual, Pt 1000 temperature probe (A.T.C.), NTC 10
kΩ probe
Measurement display lock Continuous measurement, by stability
Display Liquid crystal, backlit, 128 x 64 dots
Keyboard PET with protective treatment
Certification CE
General information
Revised editions are found on the manufacturer’s website.
Safety information
Please read this entire manual before unpacking, setting up or operating
this
equipment. Pay attention to all danger and caution statements. Failure
to do so could result in serious injury to the operator or damage to the
equipment.
Make sure that the protection provided by this equipment is not impaired,
do not use or install this equipment in any manner other than that specified
in this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will
result in death or serious injury.
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided,
could result in death or serious injury.
C A U T I O N
Indicates a potentially hazardous situation that may result in minor or moderate
injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument.
Information that requires special emphasis.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed. A symbol, if noted
on the instrument, will be included with a danger or caution statement in
the manual.
English 3
This symbol, if noted on the instrument, references the instruction
manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of
in European public disposal systems after 12 August of 2005. In
conformity with European local and national regulations (EU Directive
2002/98/EC),
European electrical equipment users must now return old
or end-of-life equipment to the Producer for disposal at no charge to
the user.
Note: For return for recycling, please contact the equipment producer or supplier
for instructions on how to return end-of-life equipment, producer-supplied electrical
accessories, and all auxillary items for proper disposal.
Product overview
The sensION
+ meters are used with probes to measure various
parameters in water.
The sensION
+ PH3 meter measures pH, ORP (mV) or temperature.
Product components
Refer to Figure 1 to make sure that all components have been received.
If
any items are missing or damaged, contact the manufacturer or a sales
representative immediately.
Figure 1 Meter components
1 Electrolyte for the probe 5 Power supply
2 Buffer solutions (pH 4.01, pH 7.00
and pH 10.00)
6 Rod with o-ring
3 Calibration beakers (with magnetic
bar inside)
7 Probe (included with kits only)
4 Probe holder 8 Meter
Installation
Assemble the probe holder
Follow the numbered steps to assemble the probe holder and to connect
the magnetic stirrer.
4 English
1 2
3 4
English 5
Connect to AC power
D A N G E R
Electrocution hazard. If this equipment is used outdoors or in potentially
wet locations, a Ground Fault Circuit Interrupt (GFCI/GFI) device must
be used to connect the equipment to its main power source.
The meter can be powered by AC power with the universal power adapter.
1. Select the correct adapter plug for the power outlet from the adapter
kit.
2. Connect the universal power adapter to the meter (Figure 2).
3. Connect the universal power adapter to an AC receptacle (Figure 3).
4. Turn the meter on.
Figure 2 Connector panel
1 Magnetic stirrer connector 4 Temperature probe connector
2 Combined
pH electrode (or indicator)
connector
5 Separated temperature probe
connector
3 Reference electrode (separated
electrodes) connector
6 Power supply
Figure 3 AC power connection
6 English
User interface and navigation
User interface
Keypad description
1 CALIBRATION key: start a
calibration
, view calibration data and
change the calibration frequency
5 DOWN key: scroll to other options,
change a value
2 TEMPERATURE and STIRRER key:
change the temperature (°C, °F) or
the stirring speed
6 ON/OFF: turn on or turn off the meter
3 UP key: scroll to other options,
change a value
7 mV measurement key: start a ORP
(mV) measurement
4 pH measurement key: start a pH
measurement
Display description
The
meter display shows the concentration, units, temperature, calibration
status, date and time.
Figure 4 Single screen display
1 Measurement unit and value (pH,
ORP (mV))
3 Sample temperature (ºC or ºF)
2 Measurement
mode or time and date 4 Visual measurement timer
Navigation
Use the calibrate key to calibrate the probe. Use the parameter key to
take a sample measurement. Use the TEMPERATURE and STIRRER key
to change the temperature (°C, °F) or the stirring speed. Use the arrow
keys
to scroll to other options or to change a value. Be sure to look
at the display during tasks as the screens change quickly. Refer to each
task for specific instructions.
Start-up
Turn the meter on and off
N O T I C E
Make sure that the probe is connected to the meter before the meter is turned on.
Push to turn on or turn off the meter. If the meter does not turn on, make
sure
that the AC power supply is properly connected to an electrical outlet.
Change the language
The display language is selected when the meter is powered on for the
first time.
Use the
or to select a language from a list. The meter returns
automatically to the measurement screen after 3 seconds.
Note: To change the language from the main menu, push .
English 7
Standard operations
Calibration
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
the personal protective equipment appropriate to the chemicals that are handled.
Refer to the current material safety data sheets (MSDS) for safety protocols.
Calibration procedure
This
procedure is for general use with liquid calibration solutions. Refer to
the documents that are included with each probe for additional information.
Note: Solutions should be stirred during calibration. For more information about the
stirring settings, refer to Change the stirring settings on page 9.
1. Pour the buffers or calibration solutions into the labeled calibration
beakers.
2. From the main menu push
to select pH or mV calibration.
3. Rinse the probe with deionized water and put the probe into the first
calibration beaker. Be sure that there are no air bubbles in the
membrane.
4. Push to measure the first calibration solution.
The next calibration solution is shown.
5. Rinse the probe with deionized water and put the probe into the first
calibration beaker. Be sure that there are no air bubbles in the
membrane.
6. Push to measure the second calibration solution.
The next calibration solution is shown.
7. Rinse the probe with deionized water and put the probe into the first
calibration beaker. Be sure that there are no air bubbles in the
membrane.
8. Push to measure the third calibration solution.
When the calibration is good, the display briefly shows 3rd Buffer OK
and then returns to the main menu.
View the calibration data
Data from the most recent calibration can be shown.
1. From the main menu push .
2. Use the to view the last calibration data. The meter returns
automatically to the measurement screen after 3 seconds.
Set the calibration reminder
The calibration reminder can be set between 0 to 23 hours or 1-7 days.
The display shows the remaining time to the new calibration.
1. From
the main menu push
and then push until the display shows
Cal. frequency.
2. Use the or to change the value.
The meter returns automatically to the measurement screen after 3
seconds.
Sample measurements
Each probe has specific preparation steps and procedures for taking
sample measurements. For step-by-step instructions, refer to the
documents that are included with the probe.
Note: Solutions should be stirred during measurement. For more information about
the stirring settings, refer to Change the stirring settings on page 9.
Note: The temperature is not shown during an ORP (mV) measurement when the
temperature probe is not connected.
Push pH or ORP (mV) to take a sample measurement. During
measurements, the parameter flashes and the timer shows the
stabilization time. To change the measurement parameter (if applicable),
push and hold pH or ORP (mV).
To measure the sample continuously, push pH or ORP (mV) to start a
measurement and push pH or ORP (mV) again during the stabilization.
The parameter flashes to indicate the continuous measurement mode.
Advanced operations
Change the date and time
The date and time can be changed from the Date and Time menu.
1. Use
the
to enter the Date and Time menu. The date and time format:
dd-mm-yyyy 24h.
8 English
The attribute to be changed will be automatically highlighted and will
advance to the next attribute if no or is pushed.
2. To change an attribute, push or when the highlighted attribute is
selected.
The current date and time will be shown on the display.
Adjust the display contrast
1. Push
the
and at the same time to enter the Display contrast menu.
2. Use
the
or to adjust the contrast of the display. The meter returns
automatically to the measurement screen after 3 seconds.
Change the stirring settings
The stirring speed can be changed during calibration and during a
measurement.
1. Push during a calibration or during a measurement to enter the
stirring menu.
2. Use the or to change the stirring speed in %.
Change the temperature units
The temperature units can be changed to Celsius or Fahrenheit.
1. From the main screen push .
2. Push to select between Celsius or Fahrenheit.
Maintenance
W A R N I N G
Multiple hazards. Do not disassemble the instrument for maintenance. If the
internal components must be cleaned or repaired, contact the manufacturer.
C A U T I O N
Personal injury hazard. Only qualified personnel should conduct the tasks
described in this section of the manual.
Clean the instrument
N O T I C E
Never use cleaning agents such as turpentine, acetone or similar products to clean
the instrument including the display and accessories.
Clean the exterior of the instrument with a moist cloth and a mild soap
solution.
Clean the probe
Clean the probe as needed. Refer to Troubleshooting on page 11 for
more information about cleaning. Refer to the probe documentation for
information about the probe maintenance.
Use the cleaning agents listed in Table 1 for contaminations on the pH
probe.
Table 1 Cleaning agents for the pH probe
Contamination Cleaning agent
Proteins Pepsin cleaning solution
Grease, oils, fats Electrode cleaning solution
Limescale 0.1 N HCl solution
Replace the magnetic stirrer
If
the magnetic stirrer does not start, follow the numbered steps to replace
the magnetic stirrer.
English 9
1 2
10 English
Troubleshooting
Refer
to the following table for common problem messages or symptoms,
possible causes and corrective actions.
Table 2 Calibration warnings and errors
Error/Warning Solution
Asymmetry > 58
mV
Slope out of range (accepted values ± 58 mV).
Repeat with .
Connect a new probe.
Buffer or electrode
in poor conditions.
Repeat with
.
Examine the probe: Clean the probe (refer to Clean the
probe
on page 9 for more information); make sure that there
are no air bubbles in the membrane. Shake the probe like
a thermometer; Connect a different probe to verify if
problem is with probe or meter.
Examine the buffer solution: Make sure that the buffer used
matches the buffer specified in configuration; make sure of
the temperature specification in configuration; use new
buffer solution.
Sens. (a) < 70%
BUFFER NOT
RECOGNIZED
UNSTABLE
READING
Time t> 100 s
Repeat with
.
Examine the probe: Clean the probe (refer to Clean the
probe
on page 9 for more information); make sure that there
are no air bubbles in the membrane. Shake the probe like
a thermometer; connect a different probe to verify if problem
is with probe or meter.
Make sure that the membrane and the diaphragm are
properly immersed in the sample.
Table 2 Calibration warnings and errors (continued)
Error/Warning Solution
Electrode in poor
conditions.
Examine the probe: Clean the probe (refer to Clean the
probe on
page 9 for more information); make sure that there
are no air bubbles in the membrane. Shake the probe like a
thermometer; connect a different probe to verify if problem is
with probe or meter.
Check the
electrode
SAME BUFFERS
Repeat with
.
Examine the probe: Clean the probe (refer to Clean the
probe
on page 9 for more information); make sure that there
are no air bubbles in the membrane. Shake the probe like
a thermometer; connect a different probe to verify if problem
is with probe or meter.
Examine the buffer solution: Use new buffer solution.
Table 3 Measurement warnings and errors
Error/Warning Solution
pH 12.78 19°C
Repeat with
.
pH out of range Examine the probe: Clean the probe (refer to Clean the
probe on page 9 for more information); make sure that there
are no air bubbles in the membrane. Shake the probe like a
thermometer; connect a different probe to verify if problem is
with probe or meter.
Temp out of range
°C
Examine the temperature sensor.
Connect a different probe to verify if problem is with probe or
meter.
Time > 150 s Make sure that the membrane and the diaphragm are properly
immersed in the sample.
Examine the temperature.
Examine the probe: Clean the probe (refer to Clean the
probe on page 9 for more information); make sure that there
are no air bubbles in the membrane. Shake the probe like a
thermometer; connect a different probe to verify if problem is
with probe or meter.
English 11
Replacement parts and accessories
Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.
Replacement parts
Description Item no.
sensION+ PH3 Lab pH-meter with accessories, without
probe
LPV2000.97.0002
sensION+ PH31 Lab pH-meter, GLP, with accessories,
without probe
LPV2100.97.0002
sensION+ MM340 Lab pH & Ion-meter, GLP, 2 channels,
with accessories, without probe
LPV2200.97.0002
sensION+ EC7 Lab conductivity meter, with accessories,
without probe
LPV3010.97.0002
sensION+ EC71 Lab conductivity meter, GLP, with
accessories, without probe
LPV3110.97.0002
sensION+ MM374, 2 channel Lab meter, GLP, accessories,
without probes
LPV4110.97.0002
Consumables
Description Item no.
pH buffer solution 4.01, with certificate, 125 mL LZW9460.99
pH buffer solution 7.00, with certificate, 125 mL LZW9461.97
pH buffer solution 10.00, with certificate, 125 mL LZW9470.99
pH buffer solution 4.01, with certificate, 250 mL LZW9463.99
pH buffer solution 7.00, with certificate, 250 mL LZW9464.97
pH buffer solution 10.00, with certificate, 250 mL LZW9471.99
pH buffer solution 4.01, with certificate, 1000 mL LZW9466.99
pH buffer solution 7.00, with certificate, 1000 mL LZW9467.97
Replacement parts and accessories (continued)
Description Item no.
pH buffer solution 10.00, with certificate, 1000 mL LZW9472.99
Electrolytic solution (KCl 3M), 125 mL LZW9510.99
Electrolytic solution (KCl 3M), 250 mL LZW9500.99
Electrolytic solution (KCl 3M), 50 mL LZW9509.99
Electrolytic solution 0.1 M, 125 mL LZW9901.99
Enzyme solution 2964349
Pepsin Cleaning Solution 2964349
Electrode cleaning solution 2965249
0.1 N HCl solution 1481253
Ethanol, 95% (Grease, oils, fats) 2378900
Accessories
Description Item no.
Magnetic stirrer with sensor holder, for sensION+ MM benchtop LZW9319.99
3x50 mL printed beakers for benchtop pH calibration LZW9110.97
Three-sensor holder, for sensION+ benchtop instruments LZW9321.99
Holder and clamp for three sensors LZW9155.99
Pyrex glass chamber, continuous flow measurements LZW9118.99
PP protector, electrode storage LZW9161.99
Standard solutions
Technical buffer solutions (DIN 19267)
Refer
to Table 4 pH and ORP (mV) values of specific buffer sets at varying
temperatures.
12 English
Table 4 pH, ORP (mV) and temperature values
Temperature pH mV
°C °F
0 32 2.01 4.01 7.12 9.52 10.30
10 50 2.01 4.00 7.06 9.38 10.17 245
20 68 2.00 4.00 7.02 9.26 10.06 228
25 77 2.00 4.01 7.00 9.21 10.01 220
30 86 2.00 4.01 6.99 9.16 9.96 212
40 104 2.00 4.03 6.97 9.06 9.88 195
50 122 2.00 4.06 6.97 8.99 9.82 178
60 140 2.00 4.10 6.98 8.93 9.76 160
70 158 2.01 4.16 7.00 8.88
80 176 2.01 4.22 7.04 8.83
90 194 2.01 4.30 7.09 8.79
English 13
Caractéristiques techniques
Les caractéristiques techniques peuvent être modifiées sans préavis.
Caractéristiques Détails
Dimensions 35 x 20 x 11 cm (13,78 x 7,87 x 4,33 po)
Poids 1 100 g (2,43 lb)
Boîtier de l'appareil IP42
Alimentation (externe) 100–240 V, 0,4 A, 47-63 Hz
Classe de protection de
l'instrument
Classe II
Température de stockage –15 à +65 °C (5 à +149 °F)
Température de
fonctionnement
0 à 40 °C (41 à 104 °F)
Humidité de fonctionnement < 80 % (sans condensation)
Résolution pH : 0,1/0,01/0,001, ORP : 0,1/1 mV,
température : 0,1 ºC (0,18 ºF)
Erreur de mesure (± 1 chiffre) pH : ≤
0,002, ORP : ≤ 0,2 mV, température : ≤ 0,2
ºC (≤ 0,36 ºF)
Reproductibilité (± 1 chiffre) pH : ± 0,001, ORP : ± 0,1 mV, température : ± 0,1
ºC (± 0,18 ºF)
Connexions Sonde d'indicateur ou combinée : connecteur
BNC (imp. > 10
12
Ω) ; électrode de référence :
connecteur banane ; type A.T.C. Pt 1000 :
connecteur banane ou téléphonique ; agitateur
magnétique : connecteur RCA
Correction de température Manuelle, sonde de température Pt 1000 (A.T.C.),
sonde NTC 10 kΩ
Verrouillage d'affichage de
mesure
Mesure continue, par stabilité
Afficheur Cristal liquide, rétroéclairé, 128 x 64 points
Caractéristiques Détails
Clavier PET avec traitement de protection
Certification CE
Généralités
Les éditions révisées se trouvent sur le site Internet du fabricant.
Consignes de sécurité
Veuillez
lire l'ensemble du manuel avant le déballage, la configuration ou
la mise en fonctionnement de cet appareil. Respectez toutes les
déclarations de prudence et d'attention. Le non-respect de cette
procédure peut conduire à des blessures graves de l'opérateur ou à des
dégâts sur le matériel.
Assurez-vous que la protection fournie avec cet appareil ne soit pas
compromise, n'utilisez pas ou n'installez pas cet appareil d'une autre façon
que celle décrite dans ce manuel.
Interprétation des indications de risques
D A N G E R
Indique une situation de danger potentiel ou imminent qui, si elle n'est pas évitée,
entraîne des blessures graves, voire mortelles.
A V E R T I S S E M E N T
Indique une situation de danger potentiel ou imminent qui, si elle n'est pas évitée,
peut entraîner des blessures graves, voire mortelles.
A T T E N T I O N
Indique une situation de danger potentiel qui peut entraîner des blessures
mineures ou légères.
A V I S
Indique une situation qui, si elle n'est pas évitée, peut occasionner
l'endommagement du matériel. Informations nécessitant une mise en avant
particulière.
14 Français
Etiquettes de mise en garde
Lire
toutes les informations et toutes les étiquettes apposées sur l’appareil.
Des personnes peuvent se blesser et le matériel peut être endommagé si
ces instructions ne sont pas respectées. Si un symbole ‘danger’ ou
‘attention’ se trouve sur l’instrument, une explication est indiquée dans le
manuel.
Lorsque ce symbole est présent sur l’instrument, reportez-vous au
manuel d’instructions pour obtenir des informations relatives au
fonctionnement et/ou à la sécurité.
L’équipement électrique portant ce symbole ne peut être mis au rebut
dans
les systèmes de mise au rebut publics européens après le 12 août
2005. Conformément aux règlements nationaux et européens
(Directive 2002/98/EC), les appareils électriques doivent, depuis le 12
août 2005, ne pas être mis au rebut dans les décharges traditionnelles,
mais être, à la fin de leur service, renvoyés par les utilisateurs
européens au fabricant, qui se chargera de les éliminer à ses frais.
Remarque : Pour le retour à des fins de recyclage, veuillez contactez le fabricant
ou le fournisseur d'équipement pour obtenir les instructions sur la façon de
renvoyer l'équipement usé, les accessoires électriques fournis par le fabricant, et
tous les articles auxiliaires pour mise au rebut appropriée.
Présentation du produit
Les appareils de mesure de la série ION
+ s'utilisent avec des sondes
pour mesurer différents paramètres dans l'eau.
L'appareil de mesure sensION
+ PH3 mesure le pH, l'ORP (mV) et la
température.
Composants du produit
Consultez la Figure 1 pour vous assurer que tous les éléments ont bien
été reçus. Si des éléments manquent ou sont endommagés, contactez
immédiatement le fabricant ou un représentant commercial.
Figure 1 Composants de l'appareil
1 Electrolyte pour la sonde 5 Alimentation
2 Solutions
tampons (pH 4,01, pH 7,00
et pH 10,00)
6 Tige avec joint torique
3 Béchers d'étalonnage (avec barre
magnétique)
7 Sonde (comprise uniquement dans
les kits)
4 Support de sonde 8 Appareil de mesure
Installation
Monter le support de sonde
Suivez les étapes numérotées afin de monter le support de sonde et de
brancher l'agitateur magnétique.
Français 15
1 2
3 4
16 Français
Branchement sur alimentation CA
D A N G E R
Risque d'électrocution Si cet équipement est utilisé à l'extérieur ou dans
des lieux potentiellement humides, un disjoncteur de fuite à la terre
(GFCI/GFI) doit être utilisé pour le branchement de l'équipement à sa
source d'alimentation secteur.
L'appareil de mesure peut être alimenté sur secteur à l'aide d'un
adaptateur d'alimentation universel.
1. Sélectionnez
la fiche d'adaptateur appropriée à la prise électrique dans
le kit adaptateur.
2. Branchez l'adaptateur d'alimentation universel à l'appareil (Figure 2).
3. Branchez l'adaptateur d'alimentation universel à une prise CA
(Figure 3).
4. Allumez l'appareil.
Figure 2 Panneau des connecteurs
1 Connecteur de l'agitateur
magnétique
4 Connecteur de la sonde de
température
2 Connecteur de l'électrode (ou
indicateur) pH combinée
5 Connecteur de la sonde de
température séparée
3 Connecteur de l'électrode de
référence (électrodes séparées)
6 Alimentation
Figure 3 Branchement sur alimentation CA
Français 17
Interface utilisateur et navigation
Interface utilisateur
Description du clavier
1 Touche CALIBRATION
(ETALONNAGE) : permet de lancer
l'étalonnage, d'afficher les données
d'étalonnage et de modifier la
fréquence d'étalonnage
5 Touche
BAS : permet de faire défiler
les autres options ou de modifier une
valeur
2 Touche TEMPERATURE/STIRRER
(TEMPRATURE/AGITATEUR) :
permet de modifier la température
(°C, °F) ou la vitesse de brassage
6 MARCHE/ARRET : allumage ou
extinction de l'appareil
3 Touche HAUT : permet de faire
défiler les autres options ou de
modifier une valeur
7 Touche de mesure mV : permet de
lancer une mesure ORP (mV)
4 Touche de mesure du pH : permet de
lancer une mesure de pH
Description de l'affichage
L'affichage
de l'appareil de mesure indique la concentration, les unités, la
température, l'état d'étalonnage, la date et l'heure.
Figure 4 Affichage sur écran unique
1 Unité
de mesure et valeur (pH, ORP
(mV))
3 Température d'échantillon (°C ou °F)
2 Mode de mesure ou date et heure 4 Minuteur visuel de mesure
Navigation
Utilisez la touche d'étalonnage
pour étalonner la sonde. Utilisez la
touche de paramétrage pour mesurer un échantillon. Utilisez les touches
TEMPERATURE
et STIRRER (TEMPERATURE et AGITATEUR)
pour
modifier la température (°C, °F) ou la vitesse de brassage. Utilisez les
touches fléchées
pour accéder à d'autres options ou modifier une
valeur. Veillez à observer l'écran pendant les tâches car les écrans
changent rapidement. Consultez les instructions spécifiques à chaque
tâche.
Démarrage
Allumage et extinction de l'appareil de mesure
A V I S
Veillez à ce que la sonde soit connectée à l'appareil de mesure avant d'allumer
l'appareil de mesure.
Appuyez sur la touche pour allumer ou éteindre l'appareil de mesure.
Si l'appareil ne s'allume pas, vérifiez que l'alimentation CA est bien
branchée à une prise électrique.
Modifier la langue
La langue de l'affichage est sélectionnée au premier allumage de
l'appareil.
Utilisez
les touches
ou pour sélectionner une langue dans la liste. Le
retour à l'écran de mesure se fait automatiquement après 3 secondes.
18 Français
Remarque : Pour modifier la langue à partir du menu principal, appuyez sur .
Opérations standard
Etalonnage
A V E R T I S S E M E N T
Risque d'exposition chimique. Respectez les procédures de sécurité du laboratoire
et portez tous les équipements de protection personnelle adaptés aux produits
chimiques que vous manipulez. Reportez-vous aux fiches techniques de sécurité
des matériaux (MSDS) pour connaître les protocoles de sécurité.
Procédure d'étalonnage
Cette procédure s'utilise généralement avec des solutions d'étalonnage
liquides. Reportez-vous aux documents inclus avec chaque sonde pour
plus d'informations.
Remarque : Les solutions doivent être brassées lors de l'étalonnage. Pour plus
d'informations sur les paramètres de brassage, reportez-vous à Modifier les
paramètres de brassage à la page 20.
1. Versez les solutions tampons et d'étalonnage dans les béchers
d'étalonnage étiquetés.
2. Dans
le menu principal, appuyez sur
pour sélectionner l'étalonnage
pH ou mV.
3. Rincez la sonde à l'eau déminéralisée et placez-la dans le premier
bécher d'étalonnage. Vérifiez qu'aucune bulle d'air ne s'est formée
dans la membrane.
4. Appuyez sur pour mesurer la première solution d'étalonnage.
La solution d'étalonnage suivante apparaît.
5. Rincez la sonde à l'eau déminéralisée et placez-la dans le premier
bécher d'étalonnage. Vérifiez qu'aucune bulle d'air ne s'est formée
dans la membrane.
6. Appuyez sur pour mesurer la seconde solution d'étalonnage.
La solution d'étalonnage suivante apparaît.
7. Rincez la sonde à l'eau déminéralisée et placez-la dans le premier
bécher d'étalonnage. Vérifiez qu'aucune bulle d'air ne s'est formée
dans la membrane.
8. Appuyez sur pour mesurer la troisième solution d'étalonnage.
Une fois l'étalonnage correct, l'écran affiche brièvement 3rd Buffer OK
(3e tampon OK), puis repasse au menu principal.
Afficher les données d'étalonnage
Les données d'étalonnage les plus récentes peuvent être affichées.
1. Sur le menu principal, appuyez sur .
2. Utilisez pour consulter les dernières données d'étalonnage. Le
retour
à l'écran de mesure se fait automatiquement après 3 secondes.
Régler le rappel d'étalonnage
Le rappel d'étalonnage peut être réglé entre 0 et 23 heures, ou entre 1 et
7 jours. L'écran indique le temps restant avant le prochain étalonnage.
1. Dans le menu principal, appuyez sur
puis sur jusqu'à ce que
l'écran affiche Cal. frequency (Fréquence d'étal.).
2. Utilisez les touches ou pour modifier la valeur.
Le retour à l'écran de mesure se fait automatiquement après 3
secondes.
Mesures d'échantillon
Chaque sonde a des étapes de préparation et procédures spécifiques
pour
prendre les mesures d'échantillons. Pour des instructions pas à pas,
se reporter aux documents inclus avec la sonde.
Remarque : Les solutions doivent être brassées lors de la prise de mesures. Pour
plus d'informations sur les paramètres de brassage, reportez-vous à Modifier les
paramètres de brassage à la page 20.
Remarque : Lors d'une mesure REDOX (mV), la température ne s'affiche pas si la
sonde de température n'est pas connectée.
Appuyez sur pH ou REDOX (mV) pour réaliser une mesure d'échantillon.
Lors des mesures, le paramètre clignote et le minuteur affiche le temps
de stabilisation. Pour modifier le paramètre de mesure (le cas échéant),
appuyez sur pH ou REDOX (mV) et maintenez la touche enfoncée.
Pour mesurer l'échantillon de façon continue, appuyez sur pH ou REDOX
(mV) pour lancer la mesure et appuyez à nouveau sur pH ou REDOX (mV)
durant la phase de stabilisation. Le paramètre clignote pour indiquer le
mode de mesure continue.
Français 19
Opérations avancées
Modifier la date et l'heure
La date et l'heure sont modifiables depuis le menu Date and Time (Date
et heure).
1. Utilisez la touche pour accéder au menu Date and Time (Date et
heure). La date et l'heure sont au format jj-mm-aaaa 24 h.
L'attribut
à modifier est mis automatiquement en surbrillance et passe
à l'attribut suivant si les touches
ou ne sont pas activées.
2. Pour modifier un attribut, appuyez sur les touches ou lorsque
l'attribut est en surbrillance.
La date et l'heure en cours apparaissent sur l'écran.
Régler le contraste écran
1. Appuyez simultanément sur les touches et afin d'accéder au
menu Display contrast (Contraste écran).
2. Utilisez les touches ou pour régler le contraste de l'écran. Le
retour
à l'écran de mesure se fait automatiquement après 3 secondes.
Modifier les paramètres de brassage
La vitesse de brassage peut être modifiée lors de l'étalonnage ou au cours
d'une mesure.
1. Appuyez sur
lors de l'étalonnage ou au cours d'une mesure pour
accéder au menu Stirring (Brassage).
2. Utilisez les touches ou pour modifier la vitesse de brassage en
%.
Modifier les unités de température
Les unités de température peuvent être indiquées en Celsius ou en
Fahrenheit.
1. Sur l'écran principal, appuyez sur .
2. Appuyez sur pour sélectionner Celsius ou Fahrenheit.
Maintenance
A V E R T I S S E M E N T
Dangers multiples. Ne démontez pas l'appareil pour l'entretien. Si les composants
internes doivent être nettoyés ou réparés, contactez le fabricant.
A T T E N T I O N
Risque de blessures corporelles Seul le personnel qualifié est autorisé à
entreprendre les opérations décrites dans cette section du manuel.
Nettoyage de l'appareil
A V I S
N'utilisez jamais d'agents de nettoyage tels que térébenthine, acétone ou autres
produits similaires pour nettoyer l'appareil, ni son écran et ses accessoires.
Nettoyez l'extérieur de l'appareil avec un chiffon humide et une solution
de détergent doux.
Nettoyer la sonde
Nettoyez la sonde selon les besoins. Reportez-vous à la section
Dépannage à la page 22 pour plus d'informations sur le nettoyage.
Reportez-vous à la documentation de la sonde pour plus d'informations
sur son entretien.
Utilisez les agents nettoyants indiqués Tableau 1 dans les cas de
contamination de la sonde pH.
Tableau 1 Agents nettoyants pour la sonde pH
Contamination Agent nettoyant
Protéines Solution de nettoyage de pepsine
Graisse, huile, corps gras Solution de nettoyage d'électrode
Entartrage Solution de HCl 0,1 N
Remplacer l'agitateur magnétique
Si
l'agitateur magnétique ne démarre pas, remplacez-le en respectant les
étapes dans l'ordre.
20 Français
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92

Hach sensION+ pH3 ユーザーマニュアル

タイプ
ユーザーマニュアル