Elcometer 138 ユーザーマニュアル

タイプ
ユーザーマニュアル
User Guide
Elcometer 138
Bresle Kit & Patches
www.elcometer.com
R
Section
Page
1
Overview
en-2
2
Box Contents
en-2
3
Test Procedure: ISO 8502-6 / ISO 8502-9
en-3
4
Test Procedure: US Navy PPI 63101-000
en-5
5
Using the Conductivity Meter
en-7
6
Care & Maintenance
en-12
7
Technical Specification
en-13
8
Spares & Accessories
en-14
9
Legal Notices & Regulatory Information
en-15
www.elcometer.com
For the avoidance of doubt, please refer to the original English language version.
Kit Dimensions: 346 x 292 x 84mm (13.6 x 11.5 x 3.3")
Kit Weight: 1.1kg (2lb 7oz)
Material Safety Data Sheets for the Elcometer 138 Standard Calibration Solutions are available to download via
our website:
http://www.elcometer.com/images/stories/MSDS/elcometer_138_b771_calibration_solution.pdf
http://www.elcometer.com/images/stories/MSDS/Calibration_Solution_T13827352-1_to_T13827352-3_EU.pdf
http://www.elcometer.com/images/stories/MSDS/Calibration_Solution_T13827352-1_to_T13827352-3_Americas.pdf
© Elcometer Limited 2009-2016. All rights reserved. No part of this document may be reproduced, transmitted,
transcribed, stored (in a retrieval system or otherwise) or translated into any language, in any form or by any
means (electronic, mechanical, magnetic, optical, manual or otherwise) without the prior written permission of
Elcometer Limited.
CONTENTS
R
en-1
The Elcometer 138 Bresle Kit and Patches provides all the materials
and equipment required to determine the surface chloride
contamination level.
Chloride salts are extracted from the surface using the Bresle Patch
method and the chloride content of the test solution is measured using
a conductivity meter.
These instructions incorporate two test methods:
ISO 8502-6 / ISO8502-9
US NAVY PPI 63101-000 (Rev 27)
The Elcometer 138 Bresle Kit and Patches can also be used in
accordance with ISO 8502-11; AS 3894.6-A and SSPC Guide 15.
a
For IMO PSPC , the surface salts should be measured and recorded.
The Elcometer 138 Bresle Kit and Patches can be used for this.
1 OVERVIEW
www.elcometer.com
en-2
2 BOX CONTENTS
Elcometer 135B Bresle Patch, Pack of 25 or;
Elcometer 135C Bresle Test Patch, Pack of 25
(depending on kit ordered)
Elcometer 138 Conductivity Meter & Sensor
Standard Calibration Solution 1.41 mS/cm; 14ml (0.47 fl oz)
Moistening Solution; 14ml (0.47 fl oz)
Bottle of Pure Distilled Water; 250ml (8.5 fl oz)
Syringes, 5ml (0.17 fl oz); x3
Needles (Blunt); x3
Plastic Beaker; 30ml (1 fl oz)
CR2032 Lithium Batteries; x2
Transit Case
User Guide
R
a
International Maritime Organisation, Performance Standard for Protective Coatings.
Note: The Elcometer 138 Conductivity Meter included in the test kit measures the
conductivity of aqueous solutions. The meter is NOT designed to measure
solids, organic solvents, surfactant, oil, adhesive, alcohol, strong acids
(pH: 0 to 2) or strong alkalis (pH: 12 to 14). The life of the sensor will be
extremely short if these substances are measured.
3.1 BEFORE YOU START
1 Calibrate the conductivity meter, see Section .5 on page . 5 en-10
2 As the test is extremely sensitive, clean latex or nitrile gloves
should be worn during the extraction of soluble salts to prevent
contamination of the surface.
3.2 TEST PROCEDURE
1 Remove the printed protective backing
and foam centre from the Bresle
patch.
2 Apply the patch to the surface
pressing firmly around the perimeter of
the patch to ensure a complete seal. If
using the Elcometer 135C Bresle Test
Patch, remove the clear protective film
cover using the orange tab.
3 Fill a syringe with 3ml of pure water.
4 Insert the syringe into the patch
through the spongy foam perimeter at
an angle of approximately 30° to the
test surface so that it passes through
the foam into the compartment formed
by the elastomer film and the test
surface. If the patch is positioned in a
difficult position bend the needle as
required.
5 Inject the pure water into the patch.
Do not remove the needle.
If necessary, any air can be evacuated
into the syringe and allowed to stay above
the water in the syringe. Take care not to
re-insert the air during steps 6 and 7.
b
6 During a suitable period of time ,
without removing the needle, suck and
re-inject the solution at least four
c
times .
3 TEST PROCEDURE: ISO 8502-6 / ISO 8502-9
www.elcometer.com
en-3
R
b
On un-pitted blast-cleaned areas, a period of 10 minutes has been found to be satisfactory, though this
time should be agreed by the interested parties.
c
During steps 6 and 7, it is essential that no solution is lost. If any solution is lost, the test shall be rejected.
3 TEST PROCEDURE: ISO 8502-6 / ISO 8502-9 (continued)
7 At the end of the period, extract as
much solution as possible and remove
c
the syringe from the patch .
8 Measure the conductivity of the
solution using the Elcometer 138
Conductivity Meter, see Section 5.6 on
page en-11. Inject the sample directly
into the sensor cell. Rinse the sensor
cell several times with the solution to
be measured before taking the
reading.
3.3 AFTER TEST
1 Record the temperature of the solution.
2 Remove the patch from the surface and clean the surface. If
required, any adhesive residue from the patch left on the test
surface can be removed by wiping with a cloth moistened with a
suitable solvent. Ensure that the solvent will not damage the
surface before use.
3 Rinse all components of the test kit, other than the patch, in
fresh, pure water. The components can then be used again.
3.4 TO CALCULATE THE SURFACE DENSITY OF SALTS
Multiply the reading by one of the following factors:
www.elcometer.com
en-4
R
d
Surface Density of Salts: Factors
ISO Salt Mix
Reading
2
mg/m
µg/cm
2
2
mg/m
µg/cm
2
µS/cm
x1.2
x0.12
x1.1
x0.11
c
During steps 6 and 7, it is essential that no solution is lost. If any solution is lost, the test shall be rejected.
d 2
Based on an area of 12.5cm and a volume of 3ml.
1
2
3
4
5
Note: ISO 8502-9 allows for the measurement of the pure water before testing and
then the subtraction of that value from that obtained in section 3.2 step 8 above.
4.1 BEFORE YOU START
1 Calibrate the conductivity meter, see Section on page .5.5 en-10
2 As the test is extremely sensitive, clean latex or nitrile gloves
should be worn during the extraction of soluble salts to prevent
contamination of the surface.
4.2 TEST PROCEDURE
Measurements are to be made randomly over the prepared surface.
2 2
Five measurements should be taken every 90m (1000ft ). Five
2 2
measurements should be taken for areas less than 90m (1000ft ).
1 Remove the printed protective
backing and foam centre from the
Bresle patch.
2 Apply the patch to the surface
pressing firmly around the perimeter
of the patch to ensure a complete
seal. If using the Elcometer 135C
Bresle Test Patch, remove the clear
protective film cover using the
orange tab.
3 Fill a syringe with 3ml of pure water.
4 Insert the syringe into the patch
through the spongy foam perimeter
and inject 1.5ml of pure water into
the patch. Do not remove the
syringe.
5 With the syringe still in the patch,
reposition the needle and evacuate
any air in the patch.
6 Once the air has been removed,
inject the remaining 1.5ml of pure
water.
7 Remove the syringe from the patch.
www.elcometer.com
en-5
R
4 TEST PROCEDURE: US NAVY PPI 63101-000
www.elcometer.com
en-6
R
8 Rub the surface of the patch gently for
10 to 15 seconds to allow the water to
dissolve surface contaminants.
9 Insert the syringe into the patch
through the spongy foam perimeter
and extract the solution from the
patch.
10 Measure the conductivity of the
solution using the Elcometer 138
Conductivity Meter, see Section on 5.6
page . Inject the sample directly en-11
into the sensor cell. Rinse the sensor
cell several times with the solution to
be measured before taking the
reading.
4.3 AFTER TEST
1 Record the temperature of the solution.
2 Remove the patch from the surface and clean the surface. If
required, any adhesive residue from the patch left on the test
surface can be removed by wiping with a cloth moistened with a
suitable solvent. Ensure that the solvent will not damage the
surface before use.
3 Rinse all components of the test kit, other than the patch, in fresh,
pure water. The components can then be used again.
4.4 PASS / FAIL CRITERIA
For immersed applications, conductivity due to soluble salts (total
ionic) shall not exceed 30µS/cm.
For non-immersed applications, conductivity due to soluble salts shall
not exceed 70µS/cm.
Note: The charts produced by the US Navy for the calculation of chloride levels are not
required for this test method. Please contact Elcometer or your local Elcometer
supplier if you require a copy of these charts.
4 (continued)TEST PROCEDURE: US NAVY PPI 63101-000
1
2
3
4
5
5.1 CAUTION
Do not drop the conductivity meter.
Never apply undue force when opening the meter (to change the
batteries or sensor).
Do not exert undue force on the sensor.
Do not allow utensils (tweezers, pipette, etc) to touch sensor cell.
Do not measure samples hotter than 40°C (105°F)
Do not allow contact with solvents.
Do not subject the conductivity meter to high temperature or
humidity.
Although the product is waterproof, avoid immersing it
completely. If the meter is accidentally dropped in water, take it
out and remove the moisture.
5.2 FITTING THE BATTERIES
The Elcometer 138 Conductivity Meter uses dry cell batteries only.
Two CR2032 lithium batteries are supplied in the kit.
To fit or replace the batteries:
1 Place batteries in battery clips ensuring
correct polarity.
2 To reassemble the meter, slide the sensor
onto the body of the meter and push the
body and sensor together gently until
sensor retaining clip engages.
When the battery voltage becomes low, the low
battery warning indicator will flash. Replace
both batteries immediately.
Note: Lithium batteries must be disposed of carefully to
avoid environmental contamination. Please consult
your local Environmental Authority for information on
disposal in your region. Do not dispose of any
batteries in fire.
www.elcometer.com
en-7
R
5 USING THE CONDUCTIVITY METER
Battery
Battery
Clip
en-8
R
5 USING THE CONDUCTIVITY METER (continued)
5.3 CONTROLS AND DISPLAY
The Elcometer 138 Conductivity Meter is operated using 3 buttons and
displays readings and other information on the LCD screen.
1
2
3
4
5
6
7
8
9
10
11 12
13
14
15
GUIDE TO CONDUCTIVITY METER & DISPLAY INDICATORS
1
Measurement Cell
Place a liquid sample in this cell to measure it
with the electrode located on the bottom of
the cell.
2
Protection Cover
Protects the measurement cell and at sensor
in storage.
3
Lithium Batteries
CR2032 x 2
4
MEAS Button
Switches the calibration mode to the
measurement mode, activates / deactivates
the reading locking function in the
measurement mode and starts / applies
settings in the special setting mode.
5
ON/OFF Button
Turns the meter On / Off.
6
Strap Eyelet
A strap can be attached here.
7
CAL Button
Starts calibration and switches items /
settings in the special setting mode.
8
Waterproof Gasket
Makes the meter waterproof.
www.elcometer.com
en-9
5 USING THE CONDUCTIVITY METER (continued)
GUIDE TO CONDUCTIVITY METER & DISPLAY INDICATORS
9
MEAS Icon
Flashes until the measured value is stabilised
and illuminates steadily when the measured
value is settled, while the reading locking
function is active.
10
CAL Icon
Flashes during calibration and illuminates
steadily when calibration is nished.
11
Battery Alarm Icon
Illuminates when the batteries are low and
need to be replaced.
12
Temperature Alarm Icon
Flashes when the measuring environment
temperature does not meet the specied
operating temperature of 5°C to 40°C
(41°F to 105°F).
13
Stability Icon
Illuminates when measured value is
stabilised.
14
Measurement Unit Display
A unit symbol illuminates corresponding to the
value displayed on the measured value
display (15). The default setting is ‘S/cm’.
15
Measured Value Display
Displays a measured, set and status value.
5.4 SETTING THE MEASUREMENT UNITS
1 Press and hold the MEAS button for
over 3 seconds in the measurement
mode to enter the special setting mode.
All items appear on the LCD then the
display changes as shown.
2 Press the MEAS button for 0.5 seconds.
The current setting is displayed.
3 Press the CAL button for 0.5 seconds to
change the setting. Pressing the CAL
button continuously changes the
settings sequentially (see diagram on
page en-10).
4 Press the MEAS button to apply the
setting. The instrument returns to
measurement mode.
R
5 USING THE CONDUCTIVITY METER (continued)
www.elcometer.com
en-10
R
5.5 ONE-POINT CALIBRATION
1 To set ‘One-point calibration’, switch the
meter on, press and hold the MEAS button
for 3 seconds. Press the CAL button until
appears.
2 Press the MEAS button again for 0.5
seconds and the calibration type will be
displayed (1 or 2). Press the CAL button
to change the setting to 1 as required and
press MEAS again to apply the setting.
3 Open the protection cover and place some
drops of the 1.41mS/cm standard solution
into the measurement cell. Washing the
sensor with standard solution beforehand
may provide more accurate calibration.
CAL
Conductivity S/cm Display
Conductivity S/m Display
TDS ppm Display
Salt Concentration % Display
Press the CAL button for
0.5 seconds
Press the CAL button for
0.5 seconds
Press the CAL button
for 0.5 seconds
Press the CAL button
for 0.5 seconds
MEAS
Button
CAL
Button
Protection
Cover
5 USING THE CONDUCTIVITY METER (continued)
4 Close the protection cover and press the
CAL button for over 2 seconds. and
ash and the calibration value is
displayed. After the calibration is
completed, and stop ashing and
illuminate steadily.
5 Clean the sensor with tap water and
remove moisture.
6 Press the MEAS button for 0.5 seconds to
enter the measurement mode and prepare
for measurement.
Note: The Elcometer 138 Conductivity Meter has a two-point calibration mode. For
details, see the full Elcometer 138 operating instructions which is available to
download via our website. The two-point calibration mode requires two standard
solutions.
5.6 TAKING A READING
1 Press the ON/OFF button to switch the
meter on.
2 Open the protection cover and place drops
of the sample on to the sensor.
Put an appropriate amount of the test sample
into the measurement cell avoiding the
inclusion of bubbles. Bubbles in the solution
may cause the conductivity measurement to be
inaccurate.
3 Close the protection cover.
Ambient air may cause the measurement values to fluctuate. To reduce
environmental interference, close the protection cover.
Without using the reading locking function:
4 Read the value displayed when appears.
CAL
CAL
www.elcometer.com
en-11
Cell
Electrode
Appropriate
Too Little
CAL
MEAS
Button
CAL
Button
R
e
When using the reading locking function :
4 After appears, press the MEAS button
for 0.5 seconds. The reading locking
function is activated. ashes until
the measured value is stabilised. When
the measured value is stable, stops
ashing and the displayed value is locked
with and illuminated steadily.
5 Read the displayed value.
6 Press the MEAS button for 0.5 seconds.
The reading locking function is deactivated
and disappears.
Note: If a measurement result is out of the specied measurement range, the
displayed measured value ashes.
5.7 AFTER MEASUREMENT
1 Press the ON/OFF button to switch the meter off.
2 Wash the sensor with tap water and wipe away any residual
water using a clean tissue.
3 Replace the sensor protection cap.
Note: If the meter is to remain unused for a long period of time, use pure water instead
of tap water to wash the sensor.
5 USING THE CONDUCTIVITY METER (continued)
R
MEAS
MEAS
MEAS
www.elcometer.com
en-12
MEAS
e
When using the reading locking function, deactivate the function before starting every measurement.
6 CARE & MAINTENANCE
The Elcometer 138 Conductivity Meter is designed to give many years
reliable service under normal operating conditions.
Prolonged periods of non-use may cause the sensor to dry out.
This can result in malfunction or unstable readings. Pour
moistening solution into the sensor cell and leave for a few
minutes to allow the sensor to become saturated. Wash the
sensor with water prior to use.
If the measuring surface of the sensor is contaminated or if air
bubbles are regularly present in the sample, clean the sensor
using a diluted neutral detergent (diluted 100 times).
MEAS
6 CARE & MAINTENANCE (continued)
The Elcometer 138 Conductivity Meter incorporates a Liquid
Crystal Display. If the display is heated above 50°C (120°F) it
may be damaged. This can happen if the conductivity meter is
left in a car parked in strong sunlight.
Always store the components of the Elcometer 138 Bresle Kit and
Patches in the carrying case when the kit is not being used.
The Elcometer 138 Conductivity Meter does not contain any user-
serviceable components. In the unlikely event of a fault, the meter
should be returned to your local Elcometer supplier or directly to
Elcometer Limited - contact details can be found on our website,
www.elcometer.com. The warranty will be invalidated if the instrument
has been opened.
www.elcometer.com
en-13
7 TECHNICAL SPECIFICATION
7.1 BRESLE PATCH
Patch Size
5cm x 5cm
Test Area
2
12.5cm
Sample Volume
3ml
7.2 CONDUCTIVITY METER
Measurement Principle
2 AC Bipolar Method
Measurement Mode
Conductivity
Conductivity Range &
Repeatability
0 - 199µS/cm
±5µS/cm
0.2 - 1.99mS/cm
±0.05mS/cm
2 - 19.9mS/cm
±0.5mS/cm
Display
Liquid Crystal Display (LCD); 2½ digits
Measurement
Temperature
5°C to 40°C (41°F to 105°F)
Dimensions
164 x 29 x 20mm (6.5 x 1.1 x 0.79")
Weight
47g (1.7 oz) - including batteries
Battery Type
2 x CR2032 lithium
R
8 SPARES & ACCESSORIES
R
The Elcometer 138 Bresle Kit is complete with all the items required to
get started and take measurements however, over the life of the kit,
replacements may be required. The following items are available from
Elcometer or your local Elcometer supplier.
8.1 ELCOMETER 138 CALIBRATION SOLUTIONS
Description Part Number
Standard 1.41 mS/cm (1410 µS/cm) Calibration T13823926
Solution; 6 x 14ml (0.47 oz) Bottles
Standard 12.9 mS/cm (12900 µS/cm) Calibration T13824404
Solution; 6 x 14ml (0.47 oz) Bottles
Standard 447 µS/cm (0.447 mS/cm) Calibration T13827352-1
Solution; 4 x 20ml (0.74 oz) Single Use Pouches
Standard 1413 µS/cm (1.413 mS/cm) Calibration T13827352-2
Solution; 4 x 20ml (0.74 oz) Single Use Pouches
Standard 15000 µS/cm (15 mS/cm) Calibration T13827352-3
Solution; 4 x 20ml (0.74 oz) Single Use Pouches
8.2 BRESLE PATCHES
Description Part Number
Elcometer 135B Bresle Patch, Pack of 25 E135----B
Elcometer 135C Bresle Test Patch, Pack of 25 E135----C25
Elcometer 135C Bresle Test Patch, Pack of 100 E135----C100
8.3 MISCELLANEOUS ACCESSORIES
Description Part Number
Elcometer 138 Conductivity Meter T13823925
Replacement Sensor for Conductivity Meter T13823928
Bottle of Pure Distilled Water; 250ml (8.5  oz) T13827259
Syringes, 5ml (0.17  oz); x3 T13818517
Needles (Blunt); x3 T13818518
Plastic Beaker; 30ml (1  oz) T13818519
www.elcometer.com
en-14
R
TMA-0571 Issue 05 - Text with cover 23942
9 LEGAL NOTICES & REGULATORY INFORMATION
This product meets the Electromagnetic Compatibility Directive.
This product is Class B, Group 1 ISM equipment according to CISPR 11.
Group 1 ISM product: A product in which there are intentionally generated and/or used conductively
coupled radio-frequency energy which is necessary for the internal functioning of the equipment itself.
Class B product are suitable for use in domestic establishments and in establishments directly
connected to a low voltage power supply network which supplies buildings used for domestic
purposes.
is a registered trademark of Elcometer Limited, Edge Lane, Manchester, M43 6BU.
United Kingdom
All other trademarks acknowledged.
The Elcometer 138 Bresle Kit and Patches is packed in a cardboard package. Please ensure that this
packaging is disposed of in an environmentally sensitive manner. Consult your local Environmental
Authority for further guidance.
CAUTION
The needles supplied for use with this kit are blunt, but care must be exercised when using
and disposing of these needles to prevent accidental needle stick injuries. It is
recommended that used needles be disposed of as special waste, and not in landfill.
If the standard solution used for calibration of the meter comes into contact with the skin,
wash the skin with fresh water. If the standard solution comes into contact with eyes,
immediately flush the eye with large amounts of fresh water and seek medical advice.
R
Guide d'utilisation
Elcometer 138
Kit & Patchs de Bresle
www.elcometer.com
R
www.elcometer.com
En cas de doute, merci de vous référer à la version originale anglaise de ce manuel.
Dimensions du Kit : 346 x 292 x 84 mm (13.6 x 11.5 x 3.3")
Poids du Kit : 1.1 Kg (2lb 7oz)
Les fiches techniques sur la sécurite des matériaux pour les solutions de calibration standard de l'Elcometer 138
son téléchargeable sur notre site Internet :
http://www.elcometer.com/images/stories/MSDS/elcometer_138_b771_calibration_solution.pdf
http://www.elcometer.com/images/stories/MSDS/Calibration_Solution_T13827352-1_to_T13827352-3_EU.pdf
http://www.elcometer.com/images/stories/MSDS/Calibration_Solution_T13827352-1_to_T13827352-3_Americas.pdf
© Elcometer Limited 2009-2016. Tous droits résers. Aucune partie de ce document ne peut être reproduite,
transmise, transcrite, stockée (dans un système documentaire ou autre) ou traduite dans quelque langue que ce
soit, sous quelque forme que ce soit ou par n'importe quel moyen lectronique, mécanique, magnétique, optique,
manuel ou autre) sans la permission écrite préalable d’Elcometer Limited.
SOMMAIRE
R
fr-1
Section
Page
1
Présentation
fr-2
2
Colisage
fr-2
3
Procédure de test : ISO 8502-6 / ISO 8502-9
fr-3
4
Procédure de test : US Navy PPI 63101-000
fr-5
5
Utilisation du conductivimètre
fr-7
6
Entretien & Maintenance
fr-12
7
Caractéristiques techniques
fr-13
8
Pièces de rechange et accessoires
fr-14
9
Informations légales et règlementaires
fr-15
Le Kit de Patchs de Bresle Elcometer 138 contient tous les élements
et équipements nécessaires pour déterminer le niveau de
contamination d'une surface par les chlorures.
Les sels de chlorure sont prélevés sur la surface grâce à la méthode
des Patchs de Bresle, et la teneur en chlorures de la solution de test
est mesurée à l'aide d'un conductivimètre.
Les présentes instructions contiennent deux méthodes de test :
ISO 8502-6 / ISO8502-9
US NAVY PPI 63101-000 (Rev 27)
Les Kit & Patchs de Bresle peuvent également être utilis conformément
aux normes ISO 8502-11; AS 3894.6-A et SSPC Guide 15.
a
Pour IMO PSPC , les sels de surface doivent être mesurés et les
valeurs enregistrées. Les Kit & Patchs de Bresle peuvent être utilisés
dans ce cadre.
1 PSENTATION
www.elcometer.com
fr-2
2 COLISAGE
Elcometer 135B Patchs de Bresle, Lot de 25 ou;
Elcometer 135C Test Patch de Bresle, Lot de 25
(selon le kit commandé)
Elcometer 138 Conductivimètre & Capteur
Solution de Calibration Standard 1.41 mS/cm; 14 ml (0.47 fl oz)
Solution de mouillage; 14 ml (0.47 fl oz)
Flacon d'eau pure; 250 ml (8.5 fl oz)
Seringues, 5 ml (0.17 fl oz); x 3
Aiguilles (pointe émoussée); x 3
Bécher Plastique; 30 ml (1 fl oz)
Piles Lithium CR2032; x 2
Valise de transport
Guide d'utilisation
R
a
Organisation Maritime Internationale, Norme de Performance des Retements de Protection.
Note : Le conductivimètre Elcometer 138 inclus dans le test mesure la conductivité de
solutions acqueuses. Il n'est PAS conçu pour mesurer des solides, des solvants
organiques, des tensio-actifs, de l'huile, des colles, de l'alcool, des acides forts
(pH : 0 à 2) ou des alcalis forts (pH : 12 à 14). La durée de vie du capteur se
trouverait extrêmement réduite par la mesure de telles substances.
3.1 AVANT DE COMMENCER
1 Calibrez le conductivimètre - Voir Section 5.5 en page fr-10.
2 Le test étant extrêmement sensible, il est recommandé de porter
des gants en latex ou en nitrile pendant l'extraction des sels
solubles pour éviter toute contamination de la surface.
3.2 PROCEDURE DE TEST
1 Retirez la protection imprimée et la
partie centrale de la mousse au dos
du patch de Bresle.
2 Appliquez le Patch sur la surface et
appuyez fermement sur les contours
du Patch pour garantir une parfaite
étanchéité. Si vous utilisez le Test de
Patch de Bresle Elcometer 135C,
retirez le film de protection en utilisant
la languette orange.
3 Remplissez une seringue de 3 ml
d'eau pure.
4 Insérez la seringue à l'intérieur du
patch à travers la bordure en mousse
à un angle d'environ 30° de la surface
de test; l'aiguille doit traverser la
bordure en mousse et pénétrer dans
le compartiment formé par le film
élastomère et la surface de test.
5 Injectez l'eau pure dans le patch. Ne
retirez pas l'aiguille.
Si nécessaire, l'air emprisonné peut être
retiré à l'aide de la seringue (il peut rester
au-dessus du niveau d'eau dans la
seringue). Faites attention à ne pas ré-
injecter cet air lors des étapes 6 et 7.
b
6 Sur une durée raisonnable , sans
retirer l'aiguille, aspirez et ré-injectez
c
la solution au moins quatre fois .
3 PRODURE DE TEST : ISO 8502-6 / ISO 8502-9
www.elcometer.com
fr-3
R
b
Sur une surface sabe non piquetée, l'expérience montre qu'une durée de 10 minutes est satisfaisante.
Cependant, cette due doit être convenue entre les parties intéressées.
c
Durant les étapes 6 et 7, il est primordial de ne pas perdre de solution. Si une quantité de solution est
perdue, le test doit être reje.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112

Elcometer 138 ユーザーマニュアル

タイプ
ユーザーマニュアル