GW Instek PPH-1503D クイックスタートガイド

カテゴリー
測定、テスト
タイプ
クイックスタートガイド
Programmable High Precision Dual
Output DC Power Supply
PPH-1503D
Quick Start Guide
GW INSTEK PART NO. 82PH31503DM01
ISO-9001 CERTIFIED MANUFACTURER
Copyright Statement
This manual contains proprietary information, which is protected by
copyright. All rights are reserved. No part of this manual may be
photocopied, reproduced or translated to another language without
prior written consent of Good Will company.
The information in this manual was correct at the time of printing.
However, Good Will continues to improve products and reserves the
rights to change specification, equipment, and maintenance
procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
Table of Contents
1
Table of Contents
OVERVIEW ...................................................................... 2
Front Panel ....................................................... 3
Rear Panel ........................................................ 6
Load and DVM Connection ............................... 7
BASIC OPERATION .......................................................... 9
Source Function ............................................... 9
DVM ............................................................... 16
Pulse Current Measurement ........................... 18
Long integration ............................................. 22
Current Sink Function..................................... 26
External Relay Control .................................... 27
Sequence Function ......................................... 30
Main Specifications ........................................ 31
PPH-1503D Quick Start Guide
2
OVERVIEW
This quick start guide is for those users who are
not very familiar with the usage of this instrument.
It provides a way of quickly understanding this
device. For details about functions such as system
settings, saving & recalling features and remote
control instructions, please refer to the PPH-1503D
user manual.
This quick start guide contains the following content:
Introduction
Introduces the instrument panels and main
features.
Operation
Describes the display interface and the functions
of the interface keys.
Overview
The PPH-1503D is a high-precision, compact, dual
output, multifunction, programmable DC power
supply with flexible operating configurations. In
addition to the basic power supply functionality,
it is also able to measure pulse current, the
average current over long periods of time and
other functions such as battery simulation.
The PPH-1503D is designed for testing the power
consumption of battery powered wireless
communication devices (e.g. cell phones). Such
devices often have large load variations within a
short time span. The high precision power supply
has excellent voltage stability during pulsed loads
and is capable of simultaneously measuring the
pulse current, even for very short pulses. In
OVERVIEW
3
Front Panel
LCD display
Numberpad and
secondary function keys
Arrow keys
DVM inputs
Front panel output
terminals
Function keys
Power button
Display Interface
LCD Display
Voltmeter
Indicators
Displays the output voltage with up to 5 digits of
resolution. The default units are Volts (V).
Ammeter
Indicator
Displays the output current with up to 5 digits of
resolution, depending on the current range
(CH1:5A/500mA/5mA; CH2:1.5A/5mA/ AUTO).
The current range is selectable between A and mA.
Setting Display
Displays the voltage and current settings.
Parameter
Settings Display
Displays the parameter settings. For details on
setting parameters, see page 9.
PPH-1503D Quick Start Guide
4
Status Display
Display the current status of the instrument.
Function Display
Displays the unit functions. There four functions:
Basic power supply function (V AND I);
Pulse current meter function (PULSE);
Long integration current measurement function
(LONG INT);
Digital Voltmeter function (DVM).
Function Keys
Menu key
Menu key to enter or exit from
system settings.
Voltage and
Current Setting
key
Voltage and Current setting
toggle switch. See page 12 for
operation details.
CH1/CH2 Toggle
Switch
CH1 and CH2 setting toggle
switch.
CH1 Front and
Rear output
toggle key
Front and rear output toggle
switch. The key will be lit when
the output is set to the rear
outputs.
Output key
The Output key turns the output
on or off. The Output key will
light up when the output is on. It
has no affect when DVM is
activated.
OVERVIEW
5
Tab /LOCK key
The Tab key is used to toggle
between various parameters. The
Lock key is used to disable all the
panel keys except for the Output
key. Pressing the Lock key for at
least 2 seconds will turn the panel
lock on or off. The Lock key can
also be used to exit from remote
control mode. When the panel
lock is active the Lock key will
light up.
Number pad
a. The number pad is used to
enter various parameters and
values. The Clear key can be used
to clear set parameters. Pressing
the C/Pict key for at least 2
seconds will take a screenshot.
b. H/L/A Pulse current
measurement shortcut keys.
These short cut keys only work in
the Pulse current measurement
main menu.
H: High measurement mode
L: Low measurement mode
A: Average measurement mode
Directional keys
and Enter key
The directional keys are used for
parameter and menu selection as
well for fine adjustment of the
current/voltage settings.
The Enter key is used to confirm
the selection of any settings or
parameters and to exit after a
setting is complete.
PPH-1503D Quick Start Guide
6
Power Button
Turns the power on or off.
On:
Off:
Rear Panel
AC power socket
and fuse
External
control port
Lan port
USB port
GPIB port
Heat sink fan
Terminals
AC input
socket
(contains the
line fuse)
The AC input accepts 100 to
240±10%VAC. The frequency
is 50Hz/60Hz;
Fuse: 2A slow-blow type
USB port
USB device port for remote
Rear panel interface
OVERVIEW
7
GPIB port
GPIB slave port for remote
control. Abides to IEEE488.2
(SCPI) protocol.
LAN & Host
port
LAN and USB Host port for
remote control.
CH1/CH2
rear panel
output
interface
A total of 5 ports: 1 positive
output terminals, 1 negative
output terminals, a Sense+
terminal, a Sense- terminal and
a ground terminal.
Relay control
interface
A total of 5 ports: A +5V input
terminal, a ground terminal, a
logic level input terminal and 2
CH1/CH2 terminals for relay
control.
Load and DVM Connection
Recommended
Cables
Model
Specification
Usage
GTL-117
1kV
Front panel DVM
input
GTL-204A
10A
Front panel Source
terminal
GTL-203A
3A
Front panel Sense
terminal
Front panel wiring
Use the GTL-204A cables for the
front panel source connections.
PPH-1503D Quick Start Guide
8
Use the GTL-203A cables for the
sense connections.
Use the GTL-117 cables for the
DVM connections.
Rear panel
connections
Insert the wires into the
appropriate terminal according
to the labels printed under the
terminals.
Note
For safety considerations, please keep in mind that
the wiring must be equivalent to the wiring on the
front terminals.
BASIC OPERATION
9
BASIC OPERATION
Source Function
V/I Read back
display area
CH1 function display (V and I, Pulse, Long)
CH2 function display
(V and I, DVM, Pulse,
Long Int)
Status bar
V/I Setting
area
Note
1. Under the Source interface:
Press the Tab key to toggle between "Function
Setting", "VSet" and "ISet". CH1 has three
functions: V and I / Pulse / Long Int. CH2 has 4
functions: V and I / DVM / Pulse / Long Int.
Press the Enter key to enter the corresponding
channel’s parameter settings. The arrow keys
can be used to switch to the secondary
parameters. Press the Menu key to exit the
parameter settings.
2. When "Function Setting" is selected (displayed
font is black), press the arrow keys to switch
between the different "Function Setting".
PPH-1503D Quick Start Guide
10
Parameter
Description
CH1 and CH2 operate as a basic power supply
with the ability to simultaneously display V/I
settings and readback values. The output from
CH1 can be toggled between the front and rear
outputs using the Rear key. When the Rear key is
lit, it indicates that the rear panel output is
activated and that the front panel output is off.
Both outputs cant be activated at the same time.
IntRate
The data sampling period derived
from the number of power line
cycles.
The setting range is:
0.1PLC to 10.00PLC (power line
cycles)
1PLC = 16.7ms(60Hz)/20ms(50Hz).
*PLC stands for power line cycles.
AverRead
Readback refresh rate. This will
display the average number count.
CurrRange
The current range selection. Ch1 has
three settings: 5A, 500mA and 5mA.
CH2 has three settings: 5A, 5mA and
Auto.
LimMode
Current limiting mode. There are 4
settings for the current limiting
mode: Limit, Trip, LimitRelay and Trip
Relay.
The Limit settings will limit the
current. When the current reaches
the setting value, the current remains
constant, as in CC mode.
The Trip setting will turn the output
BASIC OPERATION
11
off when the current limit has been
reached.
See page 27 for details on the Limit
Relay and Trip Relay settings.
RelayControl
The relay control settings have 2
configurations: Zero/One.
The Zero setting means that if the
output from the Relay control
interface OUT port is low, the
external relays will energize.
The One setting is just the opposite
of the Zero setting.
See page 27 for further details.
Resistance
The Setting range for battery
resistance simulation is 0.000Ω ~
1.000Ω and the setting resolution is
0.001Ω.
Note. This feature is only for CH1.
O.V.P
The overvoltage settings have a
setting range of 1.00~16.00V (CH1)/
1.00~13.00V (CH2), OFF or Auto.
RecallSetup
There are 6 sets of save/recall
memories. Rst/ SAV0 to SAV4
Output Range
Voltage
CH10.000V~15.000V
CH20.000V~12.000V
Current
CH10.0000A~3.0000A (0V~15V)
0.0000A~5.0000A (0V~9V)
CH20.0000A~1.5000A
PPH-1503D Quick Start Guide
12
Parameter
Settings
Voltage
Press the Vol/Cur key and the
voltage setting on the LCD is
activated. The
corresponding number will
turn black on white
background.
(a) Use the number pad (keys: 0~9, . ,
Clear) to set the voltage value.
Press the Enter key to confirm.
(b) Step Setting:
Press the left and right arrow
keys ( ) to fine tune the
voltage setting at the digit level
(The corresponding number will
turn black on white background).
Press the up and down arrow
keys ( ) to adjust the selected
digit. Press the Enter key to
complete the setting.
Current
Press the Vol/Cur key and the
current setting on the LCD is
activated. The corresponding
number will turn black in
white background.
The input method is similar
to inputting the voltage.
BASIC OPERATION
13
IntRate
Press the Enter key to bring up the
CH1 parameter setting bar. The
default setting is Intrate. Press the
Enter key to enter the desired
parameter. Press the Enter key to
complete the setting. The parameter
range is from 0.01 to 10.00PCL. Press
the arrow keys to set the other
parameters for the setting.
AverRead
Press the arrow keys to toggle to
AverRead item. Press the Enter key to
enter the desired parameter. Press
the Enter key to complete the setting.
The parameter range is from 01 to 10
samples. Press the arrow keys to
select the other parameters for the
setting.
CurrRange
Press the arrow keys to toggle to
CurrRange item. Press the Enter key
and then press the up and down
arrow keys to select the desired
current range. Press the Enter key to
complete the setting. Press the arrow
keys to select the other parameters
for the setting.
LimMode
Press the arrow keys to toggle to
LimitMode. Press the Enter key and
then press the up and down arrow
keys to set the current limited mode.
Press the Enter key to complete the
setting. Press the arrow keys to select
the other parameters for the setting.
PPH-1503D Quick Start Guide
14
RelayControl
Press the arrow keys to toggle to
RelayControl. Press the Enter key and
then press the up and down arrow
keys to set the desired initial state of
relay control. Press the Enter key to
complete the setting. Press the arrow
keys to select the other parameters
for the setting.
Resistance
Press the Arrow keys to toggle to
Resistance. Press the Enter key and
then press numeric keys to enter
parameters (Range: 0.000 to 1.000Ω).
Press the Enter key to complete the
setting. Press the arrow keys to select
the other parameters for the setting.
This feature is only for CH1.
O.V.P
Press the arrow keys to select O.V.P.
Press Enter key and down arrow key
to select the desired OVP State.
There are three states: Off / On /
Auto. If the On state is selected, you
will need to enter the OVP value.
Press the Enter key again to complete
the setting. The input parameter
range is from 1.00 to 16.00V (CH1)/
1.00 to 13.00V (CH2). There is no
need to set the OVP value for both
the Off and Auto states. When Auto
is selected, the OVP function will
activate if the output value is higher
than the setting value by 0.8V.
BASIC OPERATION
15
Operation
REAR /
FRONT
Press the Rear key will toggle the
output between the front and rear
terminals for CH1. When the Rear
key is lit, it indicates that the rear
panel output is activated for CH1.
When the Rear key is not lit, it
indicates that the front panel output
for CH1 is activated.
Output
Press the Output key to turn the
output on. When the output is on,
the Output key will light up. When
the output is off, the Output key will
not be lit.
Status
Description
CV/CC
CV appears in green (CH1)
or in blue (CH2)
CH1CH2
CC appears in red
O.V.P
OVP will appear in green
(CH1) or in blue (CH2) when
the OVP has not been
tripped.
CH1 CH2
When the OVP has tripped, the
output will be turned off and a small
prompt window appears.
When the OVP protection
has not been activated, it will
be greyed-out.
RST
Displays the power-on state
setting
PPH-1503D Quick Start Guide
16
DVM
Description
The PPH-1503D has a separate digital voltmeter
with a measurement range of 0~+20VDC.
Note: DVM and CH2 have a common ground
design. So when using the DVM- terminal, it cant be
shorted with the negative output of CH2. In addition,
when using the voltage meter, the power supply must
be properly grounded.
Parameter
Description
Intrate
Sets the reading rate of DVM
measurements based on the number
of PLCs. The setting range is:
0.1PLC to 10.
1PLC=16.7ms(60Hz) /20ms(50Hz).
*PLC stands for Power Line Cycle
AverRead
The number of samples used to
calculate the average.
Normally the unit will display
measurement results onto the screen
as soon as they are captured.
However when more stable results
are needed, averaging can be used.
The AverRead function collects
several samples of data and then
performs an averaging operation on
the data before displaying the
averaged result on the screen.
BASIC OPERATION
17
Parameter
Setttings
IntRate
Press Enter key to bring up the CH1
parameter setting column. The
default setting is IntRate. Press the
Enter key and then press numeric
keys to enter parameters (Range: 0.01
to 10.00). Press the Enter key to
complete the setting. Press the Tab
key to select the other parameter for
settings.
AverRead
Press the arrow keys to select
AverRead item. Press the Enter key
and then press numeric keys to enter
parameters (Range: 1 to 10). Press the
Enter key to complete the setting.
Press the Tab key to select the other
parameter for settings.
Operation
Press CH1 / CH2 key to switch to the CH2
setting. Press the Tab key to switch to the function
selection mode (V and I is changed from
to ). Press the right arrow key to go to
DVM mode ( ). After switching to the
DVM mode, the device is synchronized to start
measuring. When voltage is measured, it doesn’t
affect the operation of the power supply. The
output can be turned on or off by pressing the
Output button.
PPH-1503D Quick Start Guide
18
Pulse Current Measurement
Description
Changes in the load current allow us to measure
the pulse current.
There are three ways that pulse current can be
measured:
1. Measuring the peak current over a single cycle
(High Measurement).
2. Measuring the trough current over a single
cycle (Low Measurement).
3. Measuring the average current over a single
cycle (Average Measurement).
The high and average measurements are
triggered by the rising edge of the pulse current
are performed for the time specified for the
measurement.
Low measurement is triggered by the falling edge
of the pulse current.
Note: Pulse current measurement is only valid
up to 5A (CH1) and 1.5A (CH2).
Parameter
Description
IntTime
Integration Time.
The integration measurement
time can be set to automatic or to
one of the manual settings (High
Time, Low Time and Aver Time).
When the integration
measurement time is set to
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60