Graphics Tablet NI 5102

National Instruments Graphics Tablet NI 5102 ユーザーマニュアル

  • こんにちは!このNI高速デジタイザのスタートガイドについて、ご質問にお答えします。このドキュメントには、NI 5102からNI 5922までの様々なモデルのインストール、設定、テスト、そしてNI-SCOPEソフトウェアを使ったプログラミング方法が詳しく記載されています。ハードウェアの接続方法やトラブルシューティングについても触れられていますので、お気軽にご質問ください。
  • ソフトウェアをインストールする前にハードウェアをインストールしても大丈夫ですか?
    NI-SCOPEソフトウェアはどこで入手できますか?
    デバイスの仕様はどこで確認できますか?
    SMCベースのデバイスとは何ですか?
March 2009
371133M
GETTING STARTED GUIDE
NI High-Speed Digitizers
This document contains English and Japanese language instructions.
This document explains how to install, configure, and test NI high-speed digitizers and accessories,
and how to begin programming them using the NI-SCOPE instrument driver software. This document
applies to the following digitizers and accessories: NI 5102, NI 5105, NI 5112, NI 5114, NI 5122,
NI 5124, NI 5132, NI 5133, NI 5142, NI 5152, NI 5153, NI 5154, NI 5620, NI 5621, NI 5622, NI 5900,
NI 5911, and NI 5922.
For more information on features and programming, refer to the NI High-Speed Digitizers Help.
For device specifications, refer to the specifications document included with your device. Both
documents are available at Start»All Programs»National Instruments»NI-SCOPE»
Documentation.
For the most current versions of documentation, visit
ni.com/manuals. For the latest version of
NI-SCOPE, visit
ni.com/idnet.
Contents
Conventions ......................................................................................................................................... 2
1. Verifying System Requirements...................................................................................................... 3
2. Unpacking ........................................................................................................................................ 3
3. Verifying the Kit Contents............................................................................................................... 3
Other Required Items................................................................................................................... 4
4. Installing the Software ..................................................................................................................... 4
5. Installing the Hardware.................................................................................................................... 5
PXI Modules ................................................................................................................................ 5
PXI Express Modules .................................................................................................................. 6
PCI Devices ................................................................................................................................. 7
USB Devices................................................................................................................................ 8
USB Cable Strain Relief ...................................................................................................... 9
Mounting the USB-513x.............................................................................................................. 10
Desktop Use......................................................................................................................... 10
DIN Rail Mounting.............................................................................................................. 10
Panel Mounting.................................................................................................................... 11
Windows Device Recognition ..................................................................................................... 11
6. Configuring and Testing in MAX.................................................................................................... 12
7. Programming the Device ................................................................................................................. 14
Acquiring Data Interactively ....................................................................................................... 14
Acquiring Data Programmatically ............................................................................................... 14
NI-SCOPE Examples................................................................................................................... 14
NI Example Finder...............................................................................................................14
8. Making Your First Measurement..................................................................................................... 14
NI High-Speed Digitizers Getting Started Guide 2 ni.com
Appendix A: Front Panels for SMC-Based Devices and USB Devices .............................................. 15
NI 5105 Front Panels ................................................................................................................... 15
NI 5114 Front Panels ................................................................................................................... 16
NI 5122/5124/5142/5922 Front Panels........................................................................................ 17
NI 5132/5133 Front Panels.......................................................................................................... 19
NI 5152/5153/5154 Front Panels................................................................................................. 20
NI 5622 Front Panels ................................................................................................................... 21
Appendix B: Front Panels for Traditional NI-DAQ (Legacy) Devices............................................... 22
NI 5102 Front Panels ................................................................................................................... 22
NI 5112 Front Panels ................................................................................................................... 25
NI 5620/5621 Front Panels.......................................................................................................... 26
NI 5911 Front Panel............................................................................................................
......... 27
Appendix C: Accessory Front Panels .................................................................................................. 28
NI 5900 Front Panel..................................................................................................................... 28
Appendix D: Troubleshooting ............................................................................................................. 29
Front Panel ACCESS LED on PXI/PXI Express Module is Off when PXI/PXI Express
Chassis is On ............................................................................................................................ 29
Back Panel LED on USB Module is Off when Device is Plugged In ......................................... 29
Device Does Not Appear in MAX............................................................................................... 29
Device Failed the Self-Test.......................................................................................................... 30
Thermal Shutdown Error ............................................................................................................. 30
Performance Issues Using MXI Connections.............................................................................. 30
Setting Up SMC-Based Devices for Synchronization................................................................. 30
NI 5112 Programming Practices.................................................................................................. 31
Appendix E: Where to Go for Support ................................................................................................ 31
Conventions
The following conventions are used in this manual:
» The » symbol leads you through nested menu items and dialog box options to a final action.
The sequence File» Page Setup»Options directs you to pull down the File menu, select the
Page Setup item, and select Options from the last dialog box.
This icon denotes a tip, which alerts you to advisory information.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to avoid injury, data loss,
or a system crash.
bold Bold text denotes items that you must select or click in the software, such as menu items and
dialog box options.
italic Italic text denotes emphasis or a cross-reference.
monospace Text in this font denotes text or characters that you should enter from the keyboard, sections
of code, programming examples, and syntax examples. This font is also used for disk drives,
paths, directories, filenames, and extensions.
monospace Bold text in this font denotes the messages and responses that the computer automatically
bold prints to the screen. This font also emphasizes lines of code that are different from the other
examples.
© National Instruments Corporation 3 NI High-Speed Digitizers Getting Started Guide
monospace
Italic text in this font denotes text that is a placeholder for a word or value that you must
italic
supply.
Platform Text in this font denotes a specific platform and indicates that the text following it applies only
to that platform.
PXIe PXI Express
1. Verifying System Requirements
Your system must meet certain requirements to use NI high-speed digitizers with NI-SCOPE. For more
information on minimum system, recommended system, and supported application development
environments (ADEs), refer to the NI-SCOPE Readme, which is available on the NI-SCOPE CD.
Note After you install NI-SCOPE, you can access the NI-SCOPE Readme at Start»All Programs»
National Instruments»NI-SCOPE»Documentation.
2. Unpacking
NI high-speed digitizers are shipped in an antistatic bag to protect them from electrostatic discharge
(ESD). Prior to removing the digitizer from the bag, touch the antistatic bag to a metal part of the chassis
to remove any built up static charge.
Caution When handling the digitizer, make sure that you are grounded with a grounding strap or you
are touching a grounded metal object. Handle the digitizer carefully to avoid touching any exposed
pins or electronic circuitry.
Remove the digitizer from the package, remove the packing foam and rubber screw covers (PXI/PXIe
devices only), and inspect the digitizer for loose components or signs of damage. Notify NI if the
digitizer appears damaged in any way. Do not install a damaged digitizer into your computer or chassis.
Store the digitizer in the antistatic bag when not in use.
3. Verifying the Kit Contents
The following items are included in the NI high-speed digitizer kit:
The NI high-speed digitizer or accessory you ordered
NI-SCOPE instrument driver DVD-size case, which contains the NI-SCOPE driver software CD
Other included items:
The specifications document for the device
Read Me First: Safety and Electromagnetic Compatibility
NI High-Speed Digitizers Getting Started Guide (this document)
Maintain Forced-Air Cooling Note to Users (SMC-based devices only)
NI Spectral Measurements Toolkit CD (SMC-based devices with memory options higher than
8 MB only)
USB cable (USB devices only)
NI LabVIEW SignalExpress CD (USB devices only)
Note SMC-based devices are based on the National Instruments Synchronization and Memory Core
architecture. These include the NI 5105, NI 5114, NI 5122, NI 5124, NI 5142, NI 5152, NI 5153,
NI 5154, NI 5622, and NI 5922. For more information, refer to the NI High-Speed Digitizers Help.
NI High-Speed Digitizers Getting Started Guide 4 ni.com
Other Required Items
In addition to the items contained in the kit, you need the following items:
1/8 in. flat-head screwdriver
One of the following configurations:
(PXI Devices) A PXI chassis, a PXI/SCXI combination chassis, or a PXI/CompactPCI chassis
with a controller and the chassis documentation
(PXI Express Devices) A PXI Express chassis with a controller and the chassis documentation
(USB Devices) A desktop or laptop computer and its documentation
(PCI Devices) A desktop computer and its documentation
Note If your application uses NI-TClk synchronization for PCI Devices, you must use a RTSI cable
to connect the PCI Device
S. For more information, refer to NI High-Speed Digitizers Help»
Programming»Reference»NI-TClk Synchronization Help.
4. Installing the Software
Caution Make sure you install the software before you install the hardware.
To install the software, complete the following steps:
1. (Optional) If you are using an application development environment (ADE) such as LabVIEW or
if you are using a third-party tool, install it now. You must install the ADE before installing the
NI-SCOPE instrument driver.
2. Install NI-SCOPE by inserting the NI-SCOPE CD into the CD drive. Click Install Software in the
displayed window.
Note If the installation window does not appear automatically, navigate to the CD drive and
double-click setup.exe.
3. Follow the instructions in the installation prompts.
(Windows Vista) You may see access and security messages during installation. Accept the prompts
to complete the installation.
4. When the installer completes, a dialog box prompts you to Restart, Shut Down, or Restart Later.
Select Restart.
5. If you are using a system running the LabVIEW Real-Time Module, download NI-SCOPE to the
target using Measurement &Automation Explorer (MAX). For more information, refer to the
MAX Remote Systems Help by selecting Help»Help Topics»Remote Systems in MAX.
© National Instruments Corporation 5 NI High-Speed Digitizers Getting Started Guide
5. Installing the Hardware
This section describes how to install hardware for PXI/PXIe, PCI, and USB platforms.
Note You must install the software before you install the hardware.
To prevent damage to the device caused by ESD or contamination, handle the device using the edges or
the metal bracket. For more information, refer to the Read Me First: Safety and Electromagnetic
Compatibility document.
Caution Unless you are using a USB device, you must power off and unplug your PC or chassis
before installing the hardware.
PXI Modules
NI PXI modules are sensitive instruments that should be handled carefully. Do not expose the module
to temperatures or humidity beyond the rated maximums. Keep the module free of dust by cleaning with
compressed air only. Do not clean the device with any solvents or liquids.
You can install PXI modules in any PXI slot marked with a peripheral slot compatibility glyph (a circle
containing the slot number).
To install a PXI module, complete the following steps:
1. Power off and unplug the PXI chassis before installing a PXI module.
2. If the PXI chassis has multiple fan speed settings, make sure that the fans are set to high.
3. Position the PXI chassis so that inlet and outlet vents are not obstructed. For more information,
refer to the chassis documentation.
4. Make sure that the ejector handle is in the unlatched (downward) position.
5. Holding the module by the ejector handle, slide it into an empty slot as shown in Figure 1.
Make sure that the base engages with the guides in the chassis.
6. Slide the module completely into the chassis and latch it by pulling up on the ejector handle.
7. Tighten the captive screws at the top and bottom of the module front panel. Performance may suffer
if both screws are not tightened properly.
8. Verify that the PXI chassis fans are operable and free of dust and other contaminants that may
restrict airflow.
9. Before operating the module, cover all empty PXI slots using PXI filler panels or slot blockers,
which you can purchase at
ni.com.
10. Plug in and power on the PXI chassis.
NI High-Speed Digitizers Getting Started Guide 6 ni.com
Figure 1. PXI Module Installation
PXI Express Modules
Follow the installation instructions in the PXI Modules section and install the module in a PXI Express
slot of the chassis. Refer to the chassis documentation for information about the markings that denote
PXI Express slots.
Figure 2. PXI Express Module Installation
1 PXI Chassis
2 Ejector Handle in Downward Position
3 Screws
4 NI PXI Module
1 PXI Express Chassis
2 Ejector Handle
3 Screws
4 NI PXI Express Module
5 Chassis Slot Markings
3
2
4
NI PXI-1042
1
3
4
2
1
NI PXIe-1062Q
5
© National Instruments Corporation 7 NI High-Speed Digitizers Getting Started Guide
PCI Devices
To install a PCI device, complete the following steps:
1. Power off and unplug the PC.
2. Remove the PC cover.
3. Insert the device into an open PCI slot as shown in Figure 3.
Figure 3. PCI Installation
Tip To maximize airflow and extend the life of the PCI device, leave any adjacent PCI slots empty.
(NI PCI-5911) When you install the NI PCI-5911, verify that the BNC connector is centered in the
metal frame of the chassis, as shown in Figure 4. Signal degradation can occur if the outer shell of
the BNC connector touches the computer chassis. To minimize noise, do not allow the shell of the
BNC connector to touch or lie near the metal on the PC.
Figure 4. NI PCI-5911 Installation
Multiple NI 5911 devices in the same computer can raise operating temperatures beyond
specification and produce imprecise data. NI strongly recommends leaving an empty PCI slot
between multiple NI 5911 devices or adding a fan.
1 NI PCI Device 2PCI Slot 3PC
1
2
3
NI High-Speed Digitizers Getting Started Guide 8 ni.com
4. Secure the device to the PCI chassis with a screw.
Caution It is important to completely screw the device front panel into the PCI slot, both for
mechanical stability and to create a solid ground connection. Improperly secured devices may affect
the accuracy of the device.
(SMC-Based Devices) Some computer manufacturers use a securing lever made of plastic to secure
PCI devices; such a lever is unacceptable and must be removed. Use the screw provided in the kit
to screw down the digitizer. Otherwise, you must use a different computer chassis.
5. Replace the PC cover.
6. Plug in and power on the PC.
7.
(SMC-Based Devices) Verify that spread-spectrum clocking is enabled in the PC BIOS.
For information about how to verify this setting, refer to the PC user documentation.
Note For SMC-based devices, spread-spectrum clocking varies the clock signal to spread the timing
clock signal over a small frequency range. Disabling spread-spectrum clocking may affect the
accuracy of device specifications.
USB Devices
To install a USB device, connect the USB cable to the PC and the digitizer, as shown in Figure 5.
Figure 5. USB Installation
1 Laptop Computer 2 NI USB High-Speed Digitizer 3USB Cable
2
1
3
© National Instruments Corporation 9 NI High-Speed Digitizers Getting Started Guide
USB Cable Strain Relief
The two strain relief options for your USB cable are as follows:
Groove Method—Press the USB cable into one of the two grooves on the underside of the USB
device. Choose the USB cable groove that matches your USB cable size, as shown in Figure 6.
Zip Tie Method—Thread a zip tie through the zip tie bar on the underside of the USB device and
tighten around the USB cable, as shown in Figure 6.
Figure 6. USB Cable Strain Relief Options
1 Groove Method
2 Zip Tie Method
3 USB Cable Strain Relief Groove (Large)
4 USB Cable Strain Relief Groove (Small)
5 USB Cable
6Zip Tie
7Zip Tie Bar
4
5
5
7
6
or
1
2
3
NI High-Speed Digitizers Getting Started Guide 10 ni.com
Mounting the USB-513x
You can use the USB-513x on a desktop or mount it to a standard DIN rail or a panel.
Desktop Use
The USB-513x has plastic guides on the underside that allow it to be stacked on top of other USB-513x
devices.
For secure desktop use, you can adhere the supplied rubber non-skid feet to the underside of the device,
as shown in Figure 7.
Note Do not apply the rubber feet if you are panel mounting the USB-513x or stacking it on top of
another USB-513x device.
Figure 7. Applying Rubber Feet to the USB-513x
DIN Rail Mounting
The DIN rail mounting kit (part number 779689-01, not included in your USB-513x kit) is an accessory
you can use to mount the USB-513x family of products to a standard DIN rail.
Note Apply strain relief, as described in the USB Cable Strain Relief section, before mounting the
USB-513x to a DIN rail.
1 NI USB High-Speed Digitizer 2 Plastic Guides 3 Rubber Feet
3
2
2
1
© National Instruments Corporation 11 NI High-Speed Digitizers Getting Started Guide
Panel Mounting
To mount the USB-513x to a board or panel, complete the following steps while referring to Figure 8.
Figure 8. Mounting the USB-513x on a Panel
Note Do not apply the rubber feet to the USB-513x when panel mounting the device.
Note Apply strain relief, as described in the USB Cable Strain Relief section, before panel mounting
the USB-513x.
1. Download and print the panel mounting template PDF attached in the KnowledgeBase document,
USB-4065/5132/5133/6509 Panel Mounting Template. Go to
ni.com/info and enter the info
code
rd3233 to locate the KnowledgeBase document.
2. Using the template, mark the bottom point and top point on the panel. The points will be 162 mm
(6.375 in.) apart.
3. Remove the USB cable from the connector on the USB-513x.
4. Screw a #8 or M4 screw into the bottom point on the panel.
5. Set the USB-513x on the screw by fitting it into the bottom screw notch on the underside of the
USB-513x.
6. Screw a #8 or M4 screw through the USB-513x top screw hole into the panel.
Windows Device Recognition
Windows recognizes any newly installed device the first time the computer reboots after hardware is
installed. On some Windows systems, the Found New Hardware wizard opens with a dialog box for
every NI device installed. Install the software automatically (Recommended) is selected by default.
Click Next or Yes to install the software for each device.
NI High-Speed Digitizers Getting Started Guide 12 ni.com
Note (USB devices) When you first install an NI USB-5132/5133, Windows will recognize a new
device. Click Next on any dialog boxes that appear to complete the installation.
After Windows recognizes newly installed device, a dialog box prompts you to select from the following
options, which may vary depending on the devices and software installed on your system:
Begin a Measurement with This Device Using NI LabVIEW SignalExpress opens LabVIEW
SignalExpress.
Use This Device Interactively launches the NI-SCOPE Soft Front Panel (SFP).
Begin an Application with This Device launches LabVIEW.
Configure and Test This Device opens MAX to your device so you can configure settings.
Take No Action leaves your device in the system but does not launch an application.
6. Configuring and Testing in MAX
1. Launch MAX by double-clicking the Measurement & Automation icon on the desktop.
Note When you configure your device in MAX, remember that all SMC-based devices and USB
devices are configured under NI-DAQmx, and are referred to in MAX as NI-DAQmx Devices.
Legacy devices are configured in MAX under Traditional NI-DAQ (Legacy) Devices. However,
after you configure these devices in MAX, you use NI-SCOPE to program them.
Figure 9 shows the MAX Configuration Tree, which lists both NI-DAQmx Devices and Traditional
NI-DAQ (Legacy) Devices.
Figure 9. MAX Configuration Tree
2. Expand Devices and Interfaces to see the list of installed devices. If you are using a digitizer with
the LabVIEW Real-Time Module, expand Remote Systems. Find your target IP address or name,
expand it, and then expand Devices and Interfaces.
3. If your device is not listed, press <F5> to refresh. If the device is still not listed, repeat the steps in
section 5. Installing the Hardware. For more information about using MAX, refer to the help files
available within MAX.
© National Instruments Corporation 13 NI High-Speed Digitizers Getting Started Guide
Note Windows Vista does not support some Traditional NI-DAQ (Legacy) devices. Refer to the
NI-SCOPE Readme to determine which operating system is compatible with your digitizer.
4. Record the device number or device name assigned by MAX. You need this number when you
program your device.
(Traditional NI-DAQ (Legacy) Devices) Select the device to see its properties in the configuration
view. The device number appears in the Value column. By default, the resource name for the
device is
DAQ::
n
, where
n
is the device number MAX assigned to your device. NI-SCOPE
devices that are configurable under Traditional NI-DAQ are not supported by the LabVIEW
Real-Time Module.
(NI-DAQmx Devices) The assigned device name is appended to the device in its configuration
tree label. For example, after you install the digitizer, the device configuration tree label may
appear as
NI PXIe-5122: "
dev1
", where
dev1
is the device name. When you develop
your application, the resource name is the device name assigned by MAX. Only the
NI-SCOPE devices that appear under NI-DAQmx Devices are supported by the
LabVIEW Real-Time Module.
Note To avoid modifying existing applications that use a Traditional NI-DAQ (Legacy) device
number, change the assigned NI-DAQmx device name by right-clicking on the device, selecting
Rename, and entering the Traditional NI-DAQ (Legacy) device number used in your application.
For more information about device naming conventions, refer to the
niScope_init function or the
niScope Initialize VI at NI High-Speed Digitizers Help»Programming»Reference.
5. Perform a self-test on the device to verify installation.
(Traditional NI-DAQ (Legacy) Devices) Right-click the device, select Properties, and click
Test Resources.
(NI-DAQmx) Right-click the device and select Self-Test.
A dialog box indicates whether the device has passed the test.
Note If the device does not pass the self-test, repeat the instructions in section 5. Installing the
Hardware. If the device still does not pass, visit NI Technical Support at
ni.com/support.
6. Run the test panels on the device to verify the signal.
a. Connect a signal to the digitizer and select the appropriate device parameters for this signal
such as range, input limits, sample rate, and sample mode.
b. Access the test panel.
(Traditional NI-DAQ (Legacy) Devices) Click Run Test Panels in the Properties window.
(NI-DAQmx) Right-click the device and select Test Panels.
Note For both types of devices, you can enable triggering by clicking the Advanced button on the
test panel.
All NI digitizers have self-calibration capabilities. You can access this feature for all NI digitizers
programmatically with NI-SCOPE and your ADE, or you can use the NI-SCOPE Soft Front Panel
(SFP). However, only NI-DAQmx devices can be self-calibrated using MAX.
7. Exit MAX when you have finished configuring and testing the digitizer.
NI High-Speed Digitizers Getting Started Guide 14 ni.com
7. Programming the Device
You can acquire data interactively using the NI-SCOPE SFP or programmatically using the NI-SCOPE
instrument driver in your application. You can also run the NI-SCOPE examples to demonstrate the
functionality of the digitizer.
Acquiring Data Interactively
Launch the NI-SCOPE SFP from Start»All Programs»National Instruments»NI-SCOPE»SCOPE
Soft Front Panel. The NI-SCOPE SFP provides context-sensitive help for its controls.
Acquiring Data Programmatically
You can use NI-SCOPE to begin programming the digitizer in your ADE. Refer to the Programming
section of the NI High-Speed Digitizers Help at Start»All Programs»National Instruments»
NI-SCOPE»Documentation.
NI-SCOPE Examples
Programming examples for using NI-SCOPE with LabVIEW, LabWindows
/CVI
, and C are included
on the NI-SCOPE CD. For a complete list of examples and their installed locations, refer to the
NI-SCOPE Readme at Start»All Programs»National Instruments»NI-SCOPE»Documentation.
NI Example Finder
LabVIEW 7.1 or later and CVI 7.1 or later users can use the NI Example Finder to search or browse
examples. NI-SCOPE examples are classified by keyword, so you can search for a particular device or
measurement function.
To browse for NI-SCOPE examples available in LabVIEW or LabWindows/CVI, launch the
application, select Help»Find Examples, then navigate to Hardware Input and Output»Modular
Instruments»NI-SCOPE (High-Speed Digitizers).
8. Making Your First Measurement
To begin making measurements with an NI high-speed digitizer, complete the following steps:
1. Launch your ADE.
2. Navigate to Start»All Programs»National Instruments»NI-SCOPE»Examples and open one of
the examples. If you are not sure which example to run, begin with the Getting Started example.
3. Enter the correct string into the resource name control or parameter. This string varies depending
on whether the digitizer is a Traditional NI-DAQ device or an NI-DAQmx device. For more
information about device names, refer to any of the following resources.
Section 6. Configuring and Testing in MAX
NI High-Speed Digitizers Help»Programming»Reference»NI-SCOPE LabVIEW
Reference»VIs»niScope Initialize
NI High-Speed Digitizers Help»Programming»Reference»NI-SCOPE Function
Reference Help»Functions»niScope_init
4. Adjust the parameters, if necessary, to capture the input signal that you want to acquire.
5. Connect the signal that you want to acquire to one of the input channels of the digitizer. For
information about the appropriate connections, refer to Appendix A: Front Panels for SMC-Based
Devices and USB Devices or to Appendix B: Front Panels for Traditional NI-DAQ (Legacy)
Devices.
6. Run the example program.
© National Instruments Corporation 15 NI High-Speed Digitizers Getting Started Guide
Appendix A: Front Panels for SMC-Based Devices and USB Devices
This appendix describes digitizer front panels and signal connections for SMC-based devices and USB
devices: NI 5105, NI 5114, NI 5122, NI 5124, NI 5132, NI 5133, NI 5142, NI 5152, NI 5153, NI 5154,
NI 5622, and NI 5922.
NI 5105 Front Panels
Figure 10 shows the NI PXI-5105 and NI PCI-5105 front panels.
Figure 10. NI 5105 Front Panels
Table 1 describes the signal connections for the NI 5105.
Table 1. NI 5105 Front Panel Signal Connections
Connector Description Function
CH 0 to CH 7 Standard SMB jack connector Analog input connection; digitizes data and triggers
acquisitions.
PFI 1 Standard SMB jack connector Multipurpose PFI line for digital trigger in/out,
external clock in, reference clock in/out, and
timebase out.
NI PXI-5105
12-Bit 60 MS/s Digitizer
PFI 1
CH 0
CH 1
CH 2
CH 3
CH 4
CH 5
CH 6
CH 7
PFI 1
CH 0
CH 1
CH 2
CH 3
CH 4
CH 5
CH 6
CH 7
NI PCI-5105
NI High-Speed Digitizers Getting Started Guide 16 ni.com
NI 5114 Front Panels
Figure 11 shows the NI PXI-5114 and NI PCI-5114 front panels. For the AUX connector pinout
information, refer to Figure 13.
Figure 11. NI 5114 Front Panels
Table 2 describes the signal connections for the NI 5114.
Table 2. NI 5114 Front Panel Signal Connections
Connector Description Function
CH 0,
CH 1
Standard BNC female
connector
Analog input connection; digitizes data and triggers
acquisitions.
TRIG Standard BNC female
connector
External analog trigger connection; signals on the TRIG
connector cannot be digitized.
CLK IN Standard SMB jack
connector
Input for an external reference clock or sample clock to the
digitizer.
AUX I/O 9-pin mini-circular DIN
connector
Provides access to the multipurpose digital timing and
triggering lines, PFI 0, and PFI 1 (with optional cable).
For pinout information, refer to Figure 13.
TRIG
CH 0
CH 1
AUX I/O
+5 V
MAX
CLK IN
NI PCI-5114
CH 0
CH 1
CLK IN
AUX I/O
TRIG
NI PXI-5114
© National Instruments Corporation 17 NI High-Speed Digitizers Getting Started Guide
NI 5122/5124/5142/5922 Front Panels
Figure 12 shows the NI PXI-5122/5124/5142/5922 and NI PCI-5122/5124/5142/5922 front panels.
The NI PXI-5122 and the NI PXIe-5122 front panels are identical.
Figure 12. NI 5122/5124/5142/5922 Front Panels
Table 3 describes the signal connections for the NI 5122/5124/5142/5922.
Table 3. NI 5122/5124/5142/5922 Front Panel Signal Connections
Connector Description Function
CH 0,
CH 1
Standard BNC female
connector
Analog input connection; digitizes data and triggers
acquisitions.
TRIG Standard BNC female
connector
External analog trigger connection; signals on the TRIG
connector cannot be digitized.
CLK IN Standard SMB jack
connector
NI 5122/5124/5142: Input for an external reference clock or
sample clock to the digitizer.
NI 5922: Input for an external reference clock to the digitizer.
NI High-Speed Digitizers Getting Started Guide 18 ni.com
Figure 13 shows the pin assignments for the 9-pin DIN connector.
Figure 13. 9-Pin DIN Connector Pin Assignments for NI 5114/5122/5124/5142/5922
CLK OUT Standard SMB jack
connector
NI 5122/5124/5142: Output for the reference clock or sample
clock.
NI 5922: Output for the reference clock.
AUX I/O 9-pin mini-circular DIN
connector
Provides access to the multipurpose digital timing and
triggering lines, PFI 0, and PFI 1 (with optional cable).
For pinout information, refer to Figure 13.
1 +5 V (Fused)
2GND
3 Reserved
4 Reserved
5 Reserved
6PFI 1
7 Reserved
8 Reserved
9PFI 0
Table 3. NI 5122/5124/5142/5922 Front Panel Signal Connections (Continued)
Connector Description Function
4
3
5
6
7
8
9
1
2
© National Instruments Corporation 19 NI High-Speed Digitizers Getting Started Guide
NI 5132/5133 Front Panels
Figure 14 shows the NI USB-5132 and NI USB-5133 front panels.
Figure 14. NI 5132/5133 Front Panels
Figure 15 shows the NI 5132/5133 back panel.
Figure 15. NI 5132/5133 Back Panel
Table 4 describes the signal connections for the NI 5132/5133.
1 Recessed USB port 2LED 3 Ground
Table 4. NI 5132/5133 Front Panel Signal Connections
Connector Description Function
CH 0,
CH 1
Standard BNC connector Analog input connection; digitizes data and triggers
acquisitions.
PFI 1 Standard BNC connector Multipurpose PFI line for sample clock in, digital
trigger in/out, and probe compensation.
CH 0
NI USB-5133
CH 1
PFI 1
8-bit, 100 MS/s Digitizer
CH 0
NI USB-5132
CH 1
PFI 1
8-bit, 50 MS/s Digitizer
1
2
3
NI High-Speed Digitizers Getting Started Guide 20 ni.com
NI 5152/5153/5154 Front Panels
Figure 16 shows the NI PXI-5152/5153/5154 and NI PCI-5152/5153/5154 front panels.
Figure 16. NI 5152/5153/5154 Front Panels
Table 5 describes the signal connections for the NI 5152/5153/5154.
Table 5. NI 5152/5153/5154 Front Panel Signal Connections
Connector Description Function
CH 0,
CH 1
Standard BNC female
connector
Analog input connection; digitizes data and triggers
acquisitions.
TRIG Standard BNC female
connector
External analog trigger connection; signals on the
TRIG connector cannot be digitized.
PFI 0 Standard SMB jack
connector
Multipurpose PFI line for reference clock in,
sample clock in, and digital trigger in/out.
PFI 1 Standard SMB jack
connector
Multipurpose PFI line for reference clock out,
probe compensation, and digital trigger in/out.
CH 0
CH 1
TRIG
PFI 1PFI 0
ACCESS ACTIVE
.
CH0
CH1
TRIG
PFI0
PFI1
/