NDS Radiance G2 42" 取扱説明書

  • こんにちは!NDS Radiance G2ユーザーマニュアルの内容を理解しています。この医療用画像システムに関するご質問にお答えします。高解像度ディスプレイ、マルチモダリティイメージング、様々なビデオ入力など、この機器の機能についてご質問ください。
  • 電源の入れ方は?
    接続パネルはどこにありますか?
    画像調整はどのように行いますか?
    機器の清掃方法は?
High-Definition, Multi-Modality Imaging
USER MANUAL
 ENGLISH
Asia Pacic Oces
1st Fl., RISE Building, 5-13,
Asakusabashi 5-Chome,
Taito-Ku, Tokyo, Japan, 111-0053
Tel: + 81 3 6420 3031
42" RADIANCE G2
© 2017 NDS Surgical Imaging, LLC. All rights reserved.
Information in this document has been carefully checked for accuracy; however, no guarantee is given to
the correctness of the contents. This document is subject to change without notice. NDS provides this
information as reference only. Reference to products from other vendors does not imply any
recommendation or endorsement.
This document contains proprietary information protected by copyright. No part of this manual may be
reproduced by any mechanical, electronic, or other means, in any form, without prior written permission of
NDS.
All trademarks are the property of their respective owners.
i
Table of Contents
Tab 1
Warnings and Cautions ----------------------------------------------------------------------------- ii
Recycling ------------------------------------------------------------------------------------------------- ii
Declarations of Conformity ------------------------------------------------------------------------ iii
Legal Statement -------------------------------------------------------------------------------------- iii
Tab 2
About This Manual ------------------------------------------------------------------------------------ 1
Intended Use and Contraindications ------------------------------------------------------------ 1
Image Retention Notice ----------------------------------------------------------------------------- 1
Quick Startup ------------------------------------------------------------------------------------------- 1
Powering On The Unit ----------------------------------------------------------------------------- 1
First Time Users and Initial Test ----------------------------------------------------------------- 1
Tab 3
Connector Panels -------------------------------------------------------------------------------------- 2
Electrical Symbols ------------------------------------------------------------------------------------- 3
Tab 4
Control ---------------------------------------------------------------------------------------------------- 4
Quick Select --------------------------------------------------------------------------------------------- 4
Image Adjustments ----------------------------------------------------------------------------------- 5
Brightness--------------------------------------------------------------------------------------------- 5
Contrast ----------------------------------------------------------------------------------------------- 5
Backlight ---------------------------------------------------------------------------------------------- 5
Tab 5
Menu Systems Overview ---------------------------------------------------------------------------- 6
Video Source -------------------------------------------------------------------------------------------- 7
PIP and Swap ------------------------------------------------------------------------------------------- 8
GPIO ------------------------------------------------------------------------------------------------------- 9
Display Setup ----------------------------------------------------------------------------------------- 10
HD-SDI Picture Menu ---------------------------------------------------------------------------- 10
SD-SDI, S-Video and Composite Picture Menu ------------------------------------------- 10
VGA, SOG, RGBS and YPbPr Picture Menu ------------------------------------------------- 11
DVI-Digital Picture Menu ----------------------------------------------------------------------- 12
Color Menu ----------------------------------------------------------------------------------------- 13
Setup Menu ---------------------------------------------------------------------------------------- 14
Defaults Menu ------------------------------------------------------------------------------------- 15
Tab 6
Troubleshooting ------------------------------------------------------------------------------------- 16
Tab 7
Drawing and Dimensions ------------------------------------------------------------------------- 17
Tab 8
Data Connectors and Pin Outs------------------------------------------------------------------- 18
Cable Bend Radius ---------------------------------------------------------------------------------- 18
Control Connectors and Pin Outs --------------------------------------------------------------- 19
Power Connector and Pin Out ------------------------------------------------------------------- 20
Fuse Replacement ----------------------------------------------------------------------------------- 20
Tab 9
Specications ----------------------------------------------------------------------------------------- 21
Video Inputs
---------------------------------------------------------------------------------------- 22
Video Formats ------------------------------------------------------------------------------------- 22
Cleaning and Disinfecting Instructions -------------------------------------------------------- 23
Electromagnetic Compatibility (EMC) Tables ----------------------------------------------- 24
Contact ---------------------------------------------------------------------------------------------- Back
1
ii
Warnings and Cautions
It is the responsibility of the installer to test the product’s earth ground to verify that it complies with the
hospital, local and national impedance requirements.
A ground post, located on the back of the product, may be used for the purpose of grounding the unit’s
chassis. Any such ground must be installed in accordance with applicable electrical codes. The ground
post is shown on the mechanical drawing found on page 2.
AC Power Input: 100-240Vac, 50/60Hz; 5-2.5A
Power Cord: Use a hospital grade power cord with the correct plug for your power source.
Disconnect the power cord from the AC mains. The power cord is the only recognized disconnect device.
The MEDICAL EQUIPMENT should be positioned so that its disconnect device is readily accessible.
The product should be powered from a center tapped circuit when used in the US at voltages over 120
volts. Product is intended for continuous operation.
Recycling:
Follow local governing ordinances and recycling plans regarding the recycling or disposal of this
equipment.
This symbol alerts the user that important information regarding the installation and / or
operation of this equipment follows. Information preceded by this symbol should be read
carefully in order to avoid damaging the equipment.
This symbol warns user that un-insulated voltage within the unit may have sufcient magnitude
to cause electrical shock. Therefore, it is dangerous to make contact with any part inside the unit.
To reduce the risk of electric shock, DO NOT remove cover (or back). There are no user serviceable
parts inside.
Refer servicing to qualied service personnel.
This symbol cautions the user that important information regarding the operation and / or
maintenance of this equipment has been included. Information preceded by this symbol should
be read carefully to avoid damage to the equipment.
This symbol denotes the manufacturer.
This symbol denotes the manufacturer’s European Community representative.
To prevent re or shock hazards, do not expose this unit to rain or moisture. Also, do not use this unit's
polarized plug with an extension cord receptacle or other outlets unless the prongs can be fully inserted.
The product is designed to meet the medical safety requirements for a patient vicinity device.
This product is a Class I medical device. No modications are allowed.
This equipment/system is intended for use by healthcare professionals only.
Safety Compliance:
This product is T.U.V. approved with respect to electric shock, re and mechanical hazards
only in accordance with CAN/CSA C22.2 No. 60601-1 and ANSI/AAMI ES60601-1.
Safety Compliance:
This product meets the requirements of EN 60601-1 so as to conform to the Medical Device
Directive 93/42/EEC and 2007/47/EC (general safety information).
iii
FCC and Council Directives of European Standards:
This device complies with Part 15 of FCC rules and 93/42/EEC and 2007/47/EC of the Council Directives of European Standards. Operation
is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesirable results.
1. Use the attached specied cables with the color monitor so as not to interfere with radio and television reception. Use of other cable
and adapters may cause interference with other electronic equipment.
2. This equipment has been tested and found to comply with the limits pursuant to FCC part 15 and CISPR 11. This equipment
generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause
harmful interference to radio communications.
IEC:
This equipment has been tested and found to comply with the limits for medical devices to the IEC 60601-1-2. These limits are designed to
provide reasonable protection against harmful interference in a typical medical installation. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other
devices in the vicinity.
FCC, Council Directives of European Standards and IEC:
There is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio
or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
 Reorient or relocate the receiving antenna.
 Increase the separation between the equipment and receiver.
 Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
 Consult your dealer or an experienced radio/TV technician for help.
Declarations of Conformity
Legal Statement
Accessory equipment connected to this product must be certied according to the respective IEC Standards (i.e., IEC 60950-1) for data
processing equipment and IEC 60601-1 for medical equipment. Furthermore, all congurations shall comply with the system standard, IEC
60601-1-1. Anyone who connects additional equipment to the signal input part or signal output part congures a medical system, and is
therefore responsible that the system complies with the requirements of system standard IEC 60601-1-1. Whoever is responsible for
securing the unit to a system needs to insure that the mounting equipment used with this product complies to IEC standard 60601-1. If in
doubt, consult the technical services department or your local representative.
NDS sells its products through other medical device manufacturers, distributors and resellers and therefore, purchasers of this NDS
product should consult with the entity through which this product was originally purchased regarding the terms of any applicable
product warranties provided by such entity, if any.
NDS neither assumes nor authorizes any person to assume for it any other liabilities in conjunction with and/or related to the sale and/or
use of its products. To ensure proper use, handling and care of NDS products, customers should consult the product specic literature,
instruction manual, and/or labeling included with the product or otherwise available.
Customers are cautioned that system conguration, software, the application, customer data and operator control of the system, among
other factors, affect the product’s performance. While NDS products are considered to be compatible with many systems, specic
functional implementation by customers may vary. Therefore, suitability of a product for a specic purpose or application must be
determined by the consumer and is not warranted by NDS.
NDS SPECIFICALLY DISCLAIMS ALL WARRANTIES OF ANY KIND, WHETHER EXPRESS, IMPLIED AND/OR STATUTORY, INCLUDING, BUT NOT
LIMITED TO WARRANTIES OF MERCHANTABILITY, FITNESS AND/OR OF SUITABILITY FOR A PARTICULAR PURPOSE, AND NON-
INFRINGEMENT WITH RESPECT TO ALL NDS PRODUCTS OR SERVICES. ANY AND ALL OTHER WARRANTIES, REPRESENTATIONS AND/OR
GUARANTEES, OF ANY TYPE, NATURE OR EXTENT, BE IT IMPLIED, EXPRESS AND/OR WHETHER ARISING UNDER OR AS A RESULT OF ANY
STATUTE, LAW, COMMERCIAL USAGE, CUSTOM, TRADE OR OTHERWISE, ARE HEREBY EXPRESSLY EXCLUDED AND DISCLAIMED.
NDS, its suppliers and/or distributors are not liable, directly or by way of indemnity for any special, incidental, consequential, punitive,
exemplary or indirect damages, including but not limited to alleged damages for delayed shipment, non-delivery, product failure, product
design or production, inability to use such products or services, loss of future business (lost prots), or from any other cause, whatsoever,
in connection with or arising from the purchase, sale, lease, rental, installation or use of such NDS products, these terms and conditions, or
with respect to any the terms of any agreement which incorporates these terms and conditions.
SOME JURISDICTIONS DO NOT ALLOW EXCLUSIONS AND DISCLAIMERS OF CERTAIN WARRANTIES OR LIMITATIONS OF LIABILITY, SO THE
LIMITATIONS AND/OR EXCLUSIONS, SET FORTH HEREIN, MAY NOT APPLY. IN THAT EVENT LIABILITY WILL BE LIMITED TO THE GREATEST
EXTENT PERMITTED BY LAW IN THE SUBJECT JURISDICTION.
The information provided in this document, including all designs and related materials, is the valuable property of NDS and / or its
licensors and, as appropriate, they reserve all patent, copyright, and other proprietary rights to this document, including all design,
manufacturing reproduction, use, and sales rights thereto, except to the extent said rights are expressly granted to others.
2
1
Powering On The Unit:
Connect the power supply to the display via the power plug. Plug in the AC adapter. Connect a video
source to the display. Apply power to the peripheral device, then to the display. The NDS logo is displayed,
followed shortly by video.
The electronics, designed by NDS, incorporates proprietary SmartSync™ technology which at initialization,
examines the incoming signal and automatically displays the video image in its proper format. This
eliminates adjustments for most video sources. To ne tune the image, please refer to “Image
Adjustments” on page 5.
First time users and initial test:
Visually, Flat-Panel (LCD) images will look crisper than those of a traditional CRT. For the same reason, live
video may appear a little blocky. Users not familiar with the image differences should familiarize
themselves before utilization in a critical application and determine its usability. It is recommended that
rst time users view the display next to a CRT to familiarize themselves with any subtle differences in
viewing quality.
Quick Startup
Warning: Leaving a xed (constant) image on the monitor for a long period of time can result in
image retention ("burn in"). Avoid leaving a xed image on the monitor’s screen or turn the monitor
off when it is not in use.
About This Manual
This manual is designed to assist the user with proper installation, setup and operation of the LCD display.
Depending on the model and options that were purchased, some of the features and options in this
manual may not apply to the display you are using.
A numbered tab on the side of the page denotes the beginning of a section.
The functional descriptions in this manual are representative of:
Part Numbers: 90R0070
Firmware BIOS: 58J0105 Version A and later.
Intended Use and Contraindications
Intended Use:
This monitor is intended for use in a medical environment to display high quality video and graphic
images .
Contraindications:
The monitor may not be used in the presence of ammable anesthetics mixture with air, oxygen or nitrous
oxide.
No part of this product may come in contact with a patient. Never touch the product and a patient at the
same time.
This product is capable of displaying Radiology (PACS) images for reference, not diagnostic, purposes only.
For mission critical applications, we strongly recommend that a replacement unit be immediately
available.
Image Retention Notice
2
Connector Panels
3
Notes
1. An S-Video signal may be applied via 2 BNC terminated cables to the Y and C labeled BNC connectors
or a DIN 4 terminated cable, but not both.
2. DVI 1 IN 1920 x 1200 maximum resolution. DVI 2 IN 1920 x 1080 maximum resolution.
3. 3G-SDI 1 and 3G-SDI 2 accept 3G-SDI signals. The provided accessory cable is not 3G-SDI compliant.
4. An RGBS, YPbPr or VGA signal may be applied via the DVI 2 connector or the R / Pr , G / Y / SOG 2, B /Pb
and HS / C-SYNC BNC connectors. The VS BNC is used when the applied signal has both H sync and V
sync.
5. Set the REDRIVE switch to Off when the RGBS and YPbPr signals will not be daisy chained to another
display. The REDRIVE switch is set to On when the RGBS and YPbPr signals will be daisy chained to a
second display.
6. DVI 1, 3G-SDI 1 and S-VIDEO re-drives are always active when the display is powered on.
7. FIBER OPTIC (90R0053 only) and DVI 1 inputs may be connected simultaneously, but may not be
displayed simultaneously.
8. The ND-OS connector is used for installing BIOS upgrades. The ND-OS connector may also be used for
controlling the display. See Communication Port on page 14.
POWER SWITCH
AC INPUT
3
Electrical Symbols
Open (Off) Switch:
This symbol appears on the open, or off, side of the display’s rocker switch.
Equipotentiality:
This symbol appears next to the display’s potential equalization connector. (ground post)
Closed (On) Switch:
This symbol appears on the closed, or on, side of the display’s rocker switch.
Quick Select
4
Control
A 6 button keypad, located bottom center on the front of the display, allows the user to make adjustments
to various display parameters using the On Screen Menu (OSM) system. To display the standard input
menu, press the INPUT button twice.
4
The QUICK SELECT menu, shown below, is displayed when the keypad’s INPUT button is pressed once. If
the current input is one of the ve shown in the QUICK SELECT menu, it will be illuminated. Select one of
these inputs by pressing the keypad button under the QUICK SELECT key that points to it. For example: To
select DVI-1, press the MENU button under the DVI-1 label. If a selection is not made the QUICK SELECT
menu closes after 10 seconds. The QUICK SELECT menu closes 30 seconds after the last selection is made.
Quick Select changes only the Primary input. If a Secondary input is being displayed and the input chosen
with Quick Select is the same as the Secondary input, then the PIP image will be cleared and the chosen
input will become the Primary input.
Accessing the full input menu is described on page 7.
5
Image Adjustments
Press this keypad button to adjust the Brightness, Contrast and / or the Backlight settings.
Press the Brightness / Contrast button three times, ve times when a PIP
image is visible, to display the Backlight control. Press the or button
to set the backlighting.
Note: Lowering the backlight level will increase the backlight lifetime.
Adjust Backlight
Press the Brightness / Contrast button to display the Brightness control.
Press the or button to adjust Primary brightness. When a PIP image is
visible, press the Brightness / Contrast button again to access the
Secondary brightness control.
Setting the brightness too high or too low will decrease the amount of
visible grayscales.
Adjust Brightness
Press the Brightness / Contrast button twice, three times when a PIP image
is visible, to display the Contrast control. Press the or button to
adjust Primary contrast. When a PIP image is visible, press the Brightness /
Contrast button again to access the Secondary contrast control.
Setting the contrast too high or too low causes loss of some grayscales.
Color saturation may appear incorrect.
Adjust Contrast
6
Press the MENU button once to open the On Screen Menu System (OSM). The primary and secondary
inputs are shown at the top of the menu.
The Picture menu is displayed when the OSM is opened.
Press the or button to select the menu you want to work with, then press the SCROLL button to
select the parameter. Press the or button to set the parameter to the desired value. Press the MENU
button to save your changes and close the Menu System.
Notes:
1. All parameter names change to the language selected in the Setup menu.
2. Grayed out parameters are not currently accessible.
Menu System Overview
5
Languages: English, Deutsch, Français, Italiano, Svenska, Español, Nederlands, Pycckий
Video Source
7
Input Secondary
DVI 1 DVI Fiber SDI 1 SDI 2 DVI 2 RGBS YPbPr SOG 2 S-Video Composite SOG 1 VGA
DVI 1
DVI Fiber
SDI 1
P SDI 2
r DVI 2
i RGBS
m YPbPr
SOG 2
a S-Video X
r Composite X
y SOG 1 X
VGA X
Inputs Menu
Press the INPUT button twice to open the input menu. The Input menu shows a: P for primary input in the
left most column and an: S for secondary input in the column to the right of the cursor. When a secondary
input is selected, its information is shown at the top of the input menu to the right of the primary input
information . Press the SCROLL button to highlight the desired input. Finally, press button to make it
the primary input or press to make it the secondary input. The secondary input may be cleared by
highlighting it using the SCROLL button and pressing the button. Selecting a secondary input is
optional. Note: Grayed out inputs are not accessible.
The Input menu will automatically turn off 30 seconds after the most recent button press. It may also be
turned off by pressing the Input button.
The table below shows which inputs may be secondary when a given input is primary. Inputs marked with
a in the secondary column may be designated as the secondary input. Inputs marked with an X in the
secondary column may not be designated as the secondary input. The gray cells denote those inputs that
share a common connector or common electronics and thus may not used simultaneously.
DVI 2, RGBS, YPbPr and SOG 2 share the same input. Thus if one is selected as the primary input, the others
may not be selected as secondary. The same is true for Composite, SOG 1 and VGA.
8
PIP and Swap
Selecting a secondary input will display a small image of the secondary source in the upper right corner of
the display.
Secondary Image Size Control
The size of the secondary image is controlled by pressing or buttons. Pressing the button will cycle
through:
Small PIP = Width of secondary image is 25% of total screen width.
Large PIP = Width of secondary image is 40% of total screen width.
Split-Screen = Primary and Secondary have equal width side by side (half of screen width).
Split-Screen Overscan = Primary and Secondary have equal width with overscan applied
Full Screen Primary = No secondary image displayed.
The images at the bottom of the page illustrate the above sequence.
Pressing the button will cycle through the above sizes in reverse order.
Pressing the SCROLL / SWAP button will exchange the primary and secondary inputs, and exchange their
respective locations on the display. Pressing the button a second time will restore the inputs to their
original primary/secondary status. It is not necessary for both images to be displayed in order to swap
primary and secondary images.
Secondary image size and image swapping may also be controlled via the GPIO port. GPIO details are on
page 9 and the GPIO connector pin out is described on page 19.
Small PIP Large PIP Split Screen
Split Screen Overscan Full Screen Primary
9
General Purpose Input Output (GPIO):
The GPIO control allows the user to step through the Secondary image sizes as described on page 8, swap
the Primary and Secondary images or display a Record indicator in the upper left corner of the display.
Refer to the Input Menu shown below when setting up the Primary and Secondary inputs.
GPIO
GPIO Primary / Secondary Source Setup:
1. Press the INPUT button twice to open the Input Menu.
2. Press the SCROLL button to highlight the input that will be designated as Primary.
3. Select it by pressing the button a P appears in the column to the left of the cursor.
4. After the Primary input is chosen you may designate a Secondary input.
5. Press the SCROLL button to highlight the input that will be designated as Secondary.
6. Select it by pressing the button an S appears in the column to the right of the cursor. The selected
Secondary input appears as a small PIP in the upper right corner of the display.
7. The Secondary input may be cleared by highlighting it using the SCROLL button and pressing the
button again.
Using GPIO Control Fixture:
1. Connect an appropriately wired xture to the GPIO connector.
2. Press the xture’s PIP Size button.
3. The display will cycle through the PIP sizes. See page 8 for the sequence.
4. Press the xture’s Swap button. The Primary and Secondary images swap locations
5. Press the xture’s Record button, the Record indicator is displayed until the Record button is released.
Note: GPIO connector pin out is on page 19.
Setup:
By default the Priority Input Select feature is enabled with the Primary (DVI) and Secondary (SDI) inputs
preselected.
If the Priority Input Select feature has been disabled, then a secondary input (pages 7 and 8) must be
selected before the feature may be re-enabled. After selecting the secondary (PIP) input, press and hold
the INPUT button for 3 seconds. When the Priority Input Select feature becomes active, a “Priority Input
Select On” message appears in the middle of the screen for about 2 seconds, the PIP image is placed
behind the primary, the PIP + and - ( and ) buttons are disabled, and the SWAP button remains
enabled. When Priority Input Select is enabled, holding the INPUT button down for 3 seconds disables it.
Operation:
When Priority Input Select is enabled and the primary signal is lost, the secondary input becomes the
displayed image. If the primary signal is restored, its image becomes the displayed image again, and the
secondary image is placed behind the primary image. If both the primary and secondary signals are lost,
the monitor alternately scans the primary and secondary inputs until a signal is detected on one of them.
When a signal is detected on the primary or secondary input, its image is displayed on the monitor.
Priority Input Select
10
Display Setup
HD-SDI Picture Menu
SD-SDI, S-Video and Composite
Picture Menu
Horizontal Position (SD-SDI, S-Video and Composite only):
This control is disabled when Overscan is set to 0.
Moves the image to the left or right. Press or to horizontally center the image.
Vertical Position (SD-SDI, S-Video and Composite only):
This control is disabled when Overscan is set to 0.
Moves the image up or down. Press or to vertically center the image.
Sharpness:
Press or to adjust the sharpness (focus) of the displayed image.
Overscan:
0 = The image is displayed at a size that lls the screen without losing any video information. The image
presented to the display may include black bars top and bottom or left and right.
1, 2, 3, 4, 5 or 6 = The image is linearly enlarged, while remaining centered, in incremental steps. As the
image becomes larger video information will be lost from the top and bottom and / or left and right.
Select using or buttons.
Aspect (SD-SDI, S-Video and Composite only):
Pressing the or button changes the aspect ratio of the image from 4:3 to 16:9 or from 16:9 to 4:3.
Note: Displaying a 4:3 image in 16:9 causes its elements stretched horizontally. Likewise, displaying a 16:9
image in 4:3 causes its elements to be compressed horizontally.
Mirroring:
Pressing the or button presents the displayed image swapped from right to left and displays a
mirroring indicator in the lower right corner. Pressing the or button again returns the image to its
normal left to right orientation and removes the mirroring indicator.
11
VGA,SOG, RGBS and YPbPr
Picture Menu
Horizontal Position:
Moves the image to the left or right. Press or to horizontally center the image.
Vertical Position:
Moves the image up or down. Press or to vertically center the image.
Sharpness:
Press or to adjust the sharpness (focus) of the displayed image.
Phase:
Press or to adjust the phase of the display’s pixel clock.
Frequency:
Adjusts the frequency of the display’s pixel clock. With Scaling set to Fill adjust until image just lls the
screen horizontally. Press or to adjust the frequency of the display’s pixel clock.
Overscan (Video):
This parameter is enabled when the input is 16:9 or 480P or 576P or interlaced.
0 = The image is displayed at a size that lls the screen without losing any video information. The image
presented to the display may include black bars top and bottom or left and right.
1, 2, 3, 4, 5 or 6 = The image is linearly enlarged, while remaining centered, in incremental steps. As the
image becomes larger video information will be lost from the top and bottom and / or left and right.
Select using or buttons.
Scaling (Graphics):
This parameter is enabled when the input signal is not 16:9, not interlaced and not 480P or 576P.
Fill = Expands the video image to ll the entire screen. The aspect ratio may not be accurately displayed.
Aspect = Expands the video image until its largest dimension lls the screen. Image may be displayed with
black bars on the top and bottom or the left and right. 1:1 = Displays the video data in its native size and
aspect ratio. Image may be displayed with black bars on the top and bottom and on the left and right.
Select using or buttons.
SmartSync™ / Alternative Modes
On initialization NDS’ proprietary SmartSync technology examines the incoming signal and automatically
displays the video image in its proper format. To run SmartSync highlight the SmartSync / Alternative
Modes parameter and press the button.
To select an alternate mode (format) highlight the SmartSync / Alternative Modes parameter and press
the button. The mode increments each time the button is pressed until the selected mode equals
the maximum available, the next time pressed the rst mode is restored. Alternative Modes are used to
manually distinguish between modes (resolutions) whose timing characteristics are very close.
Mirroring:
Pressing the or button presents the displayed image swapped from right to left and displays a
mirroring indicator in the lower right corner. Pressing the or button again returns the image to its
normal left to right orientation and removes the mirroring indicator.
12
DVI Digital Picture Menu
Overscan (Video)
This parameter is enabled when the input is 16:9 or 480P or 576P or interlaced.
0 = The image is displayed at a size that lls the screen without losing any video information. The image
presented to the display may include black bars top and bottom or left and right.
1, 2, 3, 4, 5 or 6 = The image is linearly enlarged, while remaining centered, in incremental steps. As the
image becomes larger video information will be lost from the top and bottom and / or left and right.
Select using or buttons.
Scaling (Graphics)
This parameter is enabled when the input signal is not 16:9, not interlaced and not 480P or 576P.
Fill = Expands the video image to ll the entire screen. The aspect ratio may not be accurately displayed.
Aspect = Expands the video image until its largest dimension lls the screen. Image may be displayed with
black bars on the top and bottom or the left and right. 1:1 = Displays the video data in its native size and
aspect ratio. Image may be displayed with black bars on the top and bottom and on the left and right.
Select using or buttons.
Mirroring:
Pressing the or button presents the displayed image swapped from right to left and displays a
mirroring indicator in the lower right corner. Pressing the or button again returns the image to its
normal left to right orientation and removes the mirroring indicator.
13
Color Menu
Gamma:
Press or to select a preset Gamma (1.8, 2.0, 2.2, 2.4 or 2.6), Video or PACS
Notes:
1. Video is a color corrected Look Up Table (LUT).
2. PACS is a DICOM-like LUT. Note: This product is capable of displaying Radiology (PACS) images for
reference, not diagnostic, purposes only.
Color Temperature:
Press the or button to select a preset color temperature. Note: When a preset is selected and any
parameter is changed, the current values are copied to the User presets and User is selected.
Red, Green, Blue:
Press the or button to increase or decrease the intensity of the selected color.
Saturation:
Press or to set the saturation (color intensity) of the image.
Hue:
Press or to set the hue (color tint) of the image.
Brightness:
Press the or button to adjust brightness. This is a duplicate of the Brightness control described on
page 5.
Contrast:
Press the or button to adjust contrast. This is a duplicate of the Contrast control described on page 5.
Video Level:
Normal: Provides a dynamic range equivalent to the incoming signal with no change.
Expanded: Expands the signal level such that an input signal with range 16 (black) to 235 (white) will be
expanded to a range of 0 (black) to 255 (white). Note: If a signal is already full range (0 to 255), changing
the setting to Expanded will cause the signal to “clip” or “saturate”.
Press the or button to select.
/