Hach sensION+ EC5 ユーザーマニュアル

タイプ
ユーザーマニュアル

このマニュアルも適しています

DOC022.97.90255
© HACH Company, 2010, All rights reserved. Printed in Spain
Manual
Manual
Manual de instruções
Manuel utilisateur
操作手册
07/2010, Edition 1
07/2010, Edición 1
07/2010, Edição 1
07/2010, Edition 1
2010
7月,第一版
sensION+ EC5 HACH 0710.indd 1sensION+ EC5 HACH 0710.indd 1 7/9/10 12:50:547/9/10 12:50:54
sensION+ EC5 HACH 0710.indd 2sensION+ EC5 HACH 0710.indd 2 7/9/10 12:50:557/9/10 12:50:55
ENGLISHESPAÑOLFRANÇAIS PORTUGUÊSCONTACT
中文
sensION+ EC5 HACH 0710.indd 3sensION+ EC5 HACH 0710.indd 3 7/9/10 12:50:557/9/10 12:50:55
sensION+ EC5 HACH 0710.indd 4sensION+ EC5 HACH 0710.indd 4 7/9/10 12:50:557/9/10 12:50:55
1
Table of contents
1. Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . 3
2. General information . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.1. Safety information . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.1.1. Use of hazard information . . . . . . . . . . . . . . 4
2.1.2. Precautionary labels . . . . . . . . . . . . . . . . . . 4
2.2. General product information . . . . . . . . . . . . . . . . . . 4
3. Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.1. Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.1.1. Component list . . . . . . . . . . . . . . . . . . . . . . 5
3.1.2. Accessories . . . . . . . . . . . . . . . . . . . . . . . . 5
3.2. Battery power . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.2.1. Inserting the batteries . . . . . . . . . . . . . . . . . 6
3.3. Disconnecting/connecting the sensor . . . . . . . . . . . . 7
3.4. Carrying case . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4. Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4.1. General information . . . . . . . . . . . . . . . . . . . . . . . . 8
4.2. User interface and navigation . . . . . . . . . . . . . . . . . 8
4.2.1. Keyboard and display . . . . . . . . . . . . . . . . . 8
5. Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.1. Type of cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.2. Temperature coefficient (TC) . . . . . . . . . . . . . . . . . . 9
5.3. Reference temperature (RT) . . . . . . . . . . . . . . . . . . . 9
5.4. TDS Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
6. Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
6.1. Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
6.1.1. Stability measuring . . . . . . . . . . . . . . . . . . . 10
6.1.2. Continuous measuring . . . . . . . . . . . . . . . . . 10
6.2. Changing measuring units (CE/NaCl/TDS) . . . . . . . . 10
6.3. Calibrating the conductivity cell . . . . . . . . . . . . . . . . 11
6.3.1. Preparation . . . . . . . . . . . . . . . . . . . . . . . . 11
6.3.2. One-point calibration . . . . . . . . . . . . . . . . . 11
6.3.3. Two- and three-point calibration . . . . . . . . . . 12
6.4. Readjusting temperature . . . . . . . . . . . . . . . . . . . . . 13
6.5. Manual calibration . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.5.1. Preparation . . . . . . . . . . . . . . . . . . . . . . . . 14
6.5.2. Calibration . . . . . . . . . . . . . . . . . . . . . . . . 14
6.6. Displaying calibration data . . . . . . . . . . . . . . . . . . . 15
6.7. Restoring factory settings . . . . . . . . . . . . . . . . . . . . . 16
6.8. Disabling auto power off . . . . . . . . . . . . . . . . . . . . 16
6.9. Changing temperature units . . . . . . . . . . . . . . . . . . 16
6.10. Display backlighting . . . . . . . . . . . . . . . . . . . . . . . 17
6.11. Shutting down the instrument . . . . . . . . . . . . . . . . . 17
7. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
7.1. Cleaning the measuring instrument and accessories . . 18
7.2. Storing the instrument . . . . . . . . . . . . . . . . . . . . . . 18
7.3. Changing batteries. . . . . . . . . . . . . . . . . . . . . . . . . 18
8. Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
9. Accessories and spare parts . . . . . . . . . . . . . . . . . . . 20
10. Standard solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
11. Limited warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Contact Information
ENGLISH
sensION+ EC5 HACH 0710.indd ANG:1sensION+ EC5 HACH 0710.indd ANG:1 7/9/10 12:50:557/9/10 12:50:55
sensION+ EC5 HACH 0710.indd ANG:2sensION+ EC5 HACH 0710.indd ANG:2 7/9/10 12:50:557/9/10 12:50:55
3
1. Technical Specifications
Subject to change without notice.
Specifications
Display Liquid crystal, backlit, with pictograms
Keypad Membrane, 5 keys
Measuring ranges
(resolution depends on range)
Conductivity 0.01 µS/cm to 500 mS/cm
Salinity 0.0 to 1999 mg/L NaCl
2.0 to 50.0 g/L NaCl
TDS 0 mg/L to 500 g/L
Temperature –20.0 to 150.0 °C
–4.0 to 302.0 °F
Measuring error (± 1 digit) Conductivity 0.5 %
Salinity 0.5 %
TDS 0.5 %
Temperature 0.2 °C (0.4 °F)
Reproducibility (± 1 digit) Conductivity ± 0.1 %
Salinity ± 0.1 %
TDS ± 0.1 %
Temperature ± 0.1 °C (0.1 °F)
Temperature compensation With built-in Pt1000 temperature sensor
Reference temperature (RT) 20 or 25 °C (68 or 77 °F). Factory settings: 25 °C (77 °F)
Temperature coefficient (TC) 0.00 to 5.00%/°C (factory settings: 2%/ °C.)
TDS conversion factor 0.00 to 4.44 (factory settings: 0.64)
Calibration standards 147 µS/cm, 1413 µS/cm, 12.88 mS/cm and 111.8 mS/cm (at 25 °C/77 °F)
Accepted cell constant Between 0.05 and 50 cm
-1
(factory settings: 1.0 cm
-1
)
Dimensions of the measuring instrument and ambient conditions
Operating temperature 0 °C to 50 °C (32 °F to 122 °F)
Storage temperature –15 °C to 65 °C (5 °F to 149 °F)
Ambient conditions 80% relative humidity (non-condensing)
Weight 300 g
Size 186 x 73 x 38 mm
Additional technical data
Energy management Automatic power off after 5 minutes of inactivity
Type of protection IP 67
Power supply (batteries) Three 1.5V AA batteries. Autonomy of over 400 hours
Connector MP-5, five-contact multipin
Warranty
Warranty 3 years
ENGLISH
sensION+ EC5 HACH 0710.indd ANG:3sensION+ EC5 HACH 0710.indd ANG:3 7/9/10 12:50:557/9/10 12:50:55
4
2. General information
As a result of constant improvements to our products sometimes differences may exist between this manual
and the instructions supplied with the instrument.
2.1. Safety information
Please, read carefully this information before installing and using the instrument !
Pay attention to all danger and caution statements.
2.1.1. Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation that, if not avoided, will result in death
or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation that, if not avoided, may result in death
or serious injury.
CAUTION
Indicates a potentially hazardous situation that, if not avoided, may result in minor or moderate
injury.
Important note: Indicates a situation that, if not avoided, could lead to damage to the instrument.
Important information that requires special emphasis.
Note: Information that supplements points in the main text.
2.1.2. Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could
occur if not observed.
This symbol references the instruction manual for operation or safety information.
Electrical equipment marked with this symbol may not be disposed of in European public disposal
systems after 12 August of 2005. In conformity with European local and national regulations (EU
Directive 2002/96/EC), European electrical equipment users must now return old or end-of life
equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on
how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items
for proper disposal.
2.2. General information
The
sensION
+ EC 5
is a portable conductivity meter. To be able to use the instrument, you will need
to connect a sensor (Fig. 3, pg. 7).
sensION+ EC5 HACH 0710.indd ANG:4sensION+ EC5 HACH 0710.indd ANG:4 7/9/10 12:50:557/9/10 12:50:55
5
3. Installation
3.1.1. Component list
Code Instrument Cell Accessories Batteries Manual
LPV3500.97.0002
sensION
+ EC 5
–– ––
✓✓
LPV3560.97.0002
sensION
+ EC 5
LZW5060.97.0002
✓✓✓
LPV3562.97.0002
sensION
+ EC 5
LZW5062.97.0002
✓✓✓
ENGLISH
WARNING
Risk of falling. Measurements are made directly in situ with this instrument. To prevent the risk of
falling, ensure compliance with all local safety regulations on securing with ropes and on the use of
adequate protective clothing and footwear.
Risk of injury. Do not attach the measuring instrument to your body. The measuring sensor could
be accidentally caught in a motor or stirrer and could endanger the user through the connected
sensor cable.
3.1. Content of
sensION
+ EC 5
EC5
sensION
After removing the instrument, check each part for possible damage. All components on the attached list
must be present. If there are any missing or damaged parts, contact the manufacturer or distributor.
3.1.2. Accessories
• Carrying case.
Standard solutions, 125 mL, 147 µS/cm,
1413 µS/cm and 12.88 mS/cm (25 °C).
Calibration tubes 147 µS/cm,
1413 µS/cm and 12.88 mS/cm.
• Two 90-mL bottles for sample and sensor
cleaning.
Fig. 1. Equipment composition
sensION+ EC5 HACH 0710.indd ANG:5sensION+ EC5 HACH 0710.indd ANG:5 7/9/10 12:50:557/9/10 12:50:55
6
3. Installation
3.2. Battery power
WARNING
Risk of fire or explosion. Use only AA alkaline batteries for the measuring instrument and ensure that
the batteries are correctly inserted in their compartment. Inserting batteries incorrectly can damage
the instrument and cause fires or explosions.
The measuring instrument works with three batteries (1.5 V, AA).
The batteries supplied have a useful life of approximately 400 hours.
Important note: never use different types of batteries together.
Important note: if the measuring instrument is not going to be used for a long period of time, remove
the batteries to prevent leakage in the instrument.
3.2.1. Inserting the batteries
1. Open the lid of the battery compartment by pressing and pulling on the tab.
2. Insert the batteries supplied (1.5 V AA). Note the polarity markings inside the battery compartment.
3. Close the lid of the battery compartment.
Fig. 2. Inserting the batteries
1
Tab
2
Battery compartment lid
3
Batteries
4
Battery compartment
+
+
+
2
1
3
4
sensION+ EC5 HACH 0710.indd ANG:6sensION+ EC5 HACH 0710.indd ANG:6 7/9/10 12:50:567/9/10 12:50:56
7
3. Installation
ENGLISH
3.3. Disconnecting/connecting the sensor
The
sensION
+ EC 5
kits are supplied with the sensor attached to the instrument.
Fig. 3. Disconnecting/connecting the sensor to the measuring instrument
Sensor disconnection
Sensor connection
Important note: do not unscrew the connector.
3.4. Carrying case
The carrying case can be used as a support for both, instrument and sensor, see figure.
Electrode position when
it is not in use.
sensION+ EC5 HACH 0710.indd ANG:7sensION+ EC5 HACH 0710.indd ANG:7 7/9/10 12:50:577/9/10 12:50:57
8
4. Start-up
4.1. General information
Important note: before start-up, make sure that the sensor is connected to the measuring instrument.
Fig. 4. Keypad descriptiont
Display
1
Measured conductivity/salinity/
TDS values.
2 Reference temperature.
3 Temperature of the sample.
4 Measuring time (hh:mm:ss).
5 Units of measurement.
6 Battery indicator.
Keypad
- Instrument on/off.
- Display light on.
- Start of measuring.
- Start of calibration.
- Selection of parameter and stan-
dard to calibrate.
- Selection of temperature units.
- Start of set-up.
1
2
3
4
@25ºC
@20ºC
μS/cm
mS/cm
mg/l
g/l
ºF
ºC
Bat
EC
TDS
Sal
T
t
6
5
4.2 User interface and navigation
4.2.1 Keyboard and display
Figure 4 indicates the instrument keys and display.
1. Connect the sensor (if necessary) to the
measuring instrument (Fig. 3, pg. 7).
2. Turn on the instrument.
3. Perform calibration (see pg. 11-12).
4. Take measurement (see pg.10).
Note: With the instrument switched off,
- Press the
key and the instrument
will begin measuring.
- Press the
key and the instrument
will begin calibration.
Software version. Instrument
serial number.
Platinum configured
cell (see pg. 9)
Standby screen.
sensION+ EC5 HACH 0710.indd ANG:8sensION+ EC5 HACH 0710.indd ANG:8 7/9/10 12:50:587/9/10 12:50:58
9
5. Set-up
ENGLISH
The
sensION
+ EC 5
is supplied configured with the following parameters:
- Cell: platinum (50 60).
- Temperature coefficient: 2.00%/°C.
- Reference temperature: 25°C.
- TDS calculation factor: 0.64.
To display and/or modify these parame-
ters, follow the diagram below.
5.1. Cell
5060, Platinum cell: Pt.
5062, Titanium cell: tI.
Important note:
You must configure the correct cell to en-
sure correct measuring.
5.2. Temperature
coefficient (TC)
Enter the relevant coefficient value (0.00
to 9.99%/°C).
Recommended value for aqueous sam-
ples: 2.00%/°C.
5.2. Reference
temperature (RT)
Select a value between 20 and 25°C.
5.3. TDS factor (F)
Enter the relevant value
(0.00 to 4.44).
2 s
2 s
2 s
2 s
2 s
2 s
2 s
2 s
2 s
2 s
2 s
2 s
sensION+ EC5 HACH 0710.indd ANG:9sensION+ EC5 HACH 0710.indd ANG:9 7/9/10 12:50:597/9/10 12:50:59
10
6. Operation
6.1. Measurement
Note: the instrument must be calibrated before reading
the first conductivity value.
1. Press
to turn on the instrument.
2. Remove the sensor protector.
3. Clean the sensor with deionized water.
4. Stir the sensor uniformly in the sample, holding it by
the top.
6.1.1. Stability measuring
Follow the diagram.
The reading is not stabilized on the screen until the
sensor signal varies less than ± 1 digit for 6 sec.
Once the reading has stabilized, the screen lights up
for 2 sec.
6.1.2. Continuous measuring
Once stability measuring has begun, press the
key
twice.
The instrument displays the measured value directly on
screen at all times.
The continuous reading should always be started after
beginning a stability reading (see diagram).
To end a continuous reading, press
.
The display will light up for 2 sec.
Note: If the measuring instrument is not used for 5 minu-
tes, it will shut off automatically.
6.2. Changing measuring units
(EC/NaCl/TDS)
To change the units of measurement from con-
ductivity (EC) to salinity (Sal/NaCl) or to TDS,
or vice versa.
2 s
+
@25ºC
μS/cm
ºC
EC
T
t
@25ºC
μS/cm
ºC
EC
T
t
@25ºC
μS/cm
ºC
EC
T
t
@25ºC
μS/cm
ºC
EC
T
t
@25ºC
μS/cm
ºC
EC
T
t
@25ºC
μS/cm
ºC
EC
T
t
@25ºC
μS/cm
ºC
EC
T
t
@25ºC
μS/cm
ºC
EC
T
t
@25ºC
μS/cm
ºC
EC
T
t
Stability
measuring
Continuous
measuring
2 s 2 s
TDS
mg/l
Sal
mg/l
@25ºC
μS/cm
EC
Stability measuring
Hold down the
key for 2 s.
sensION+ EC5 HACH 0710.indd ANG:10sensION+ EC5 HACH 0710.indd ANG:10 7/9/10 12:50:597/9/10 12:50:59
11
6. Operation
ENGLISH
6.3.2. One-point calibration
Dip the sensor into the tube several times to homogenize the liquid. Do not
hold it by the tube to avoid heating of the standard solution.
Follow the diagram on the next page.
This form of calibration is acceptable when conductivity values around that
of the standard solution used are being measured.
Note: if the measuring instrument is not used for 5 minutes, it shuts off
automatically.
If errors occur during calibration, an error message will appear on screen
(see pg. 19).
Fig. 6. Preparation
6.3. Calibrating the conductivity sensor
Important note: before turning on the instrument, check that the sensor is connected.
The instrument must be calibrated before taking the first conductivity reading.
The calibration can be performed with 1, 2 or 3 points.
6.3.1. Preparation
1. Fill the tubes with the standard solutions (note the fill level).
2. Unscrew the protector and rinse the sensor with plenty of deionized water.
3. Screw the sensor on to the tube containing the first standard.
50 60
ºC
50 60
0.2... 200.000 μS/cm -30... 80ºC
50 60
0.2... 200.000 μS/cm -30... 80ºC
50 60
50 60
0 60
6
ºC
ºC
0
0
50 60
50 60
0 60
6
0 2 200000 μS/c
0.2... 200.000 μS/c
m -30 8
m -30... 8
0
0
ºC
C
50 60
0.2... 200.000 μS/cm -30... 80ºC
50 60
0.2... 200.000 μS/cm -30... 80ºC
1 2 3
sensION+ EC5 HACH 0710.indd ANG:11sensION+ EC5 HACH 0710.indd ANG:11 7/9/10 12:50:597/9/10 12:50:59
12
6. Operation
6.3.3. Two- and three-point calibration
Follow the diagram below.
The equipment will repeat the process described for the first stan-
dard solution with the second and third.
Dip the sensor into the tube several times to homogenize the li-
quid. Do not hold it by the tube to avoid heating of the standard
solution.
Note: if the measuring instrument is not used for 5 minutes, it shuts
off automatically.
If errors occur during calibration, an error message will appear
on screen (see pg. 19).
When you switch standards, rinse the cell with deionized water.
μS/cm
μS/cm
ºC
T
t
μS/cm
EC
ºC
T
t
μS/cm
EC
μS/cm
ºC
T
t
μS/cm
EC
ºC
T
t
μS/cm
EC
@25ºC
μS/cm
ºC
EC
T
t
50 60
0.2... 200.000 μS/cm -30... 80ºC
50 60
0.2... 200.000 μS/cm -30... 80ºC
50 60
0.2... 200.000 μS/cm -30... 80ºC
50 60
0 60
6
0 2 200000 S/c
0.2... 200.000 μS/c
m308
m -30... 8
0
0
ºC
ºC
0
0
0
50 60
0 60
6
0 2 200000 S/
0.2... 200.000 μS/c
30 8
m -30... 8
0
0
ºC
ºC
50 60
50 60
0.2... 200.000 μS/cm -30... 80ºC
50 60
50 60
0 60
6
0
0
0
50 60
0 60
6
Screw the
sensor
to the 1st tube
Screw the
sensor to the
2nd tube
To exit calibration
Two-point calibration
To exit calibration.
One-point calibration
sensION+ EC5 HACH 0710.indd ANG:12sensION+ EC5 HACH 0710.indd ANG:12 7/9/10 12:51:007/9/10 12:51:00
13
6. Operation
ENGLISH
6.4. Readjusting temperature
The instrument can correct the deviation of a temperature
probe (built into conductivity cells) at 25°C and 85°C (77°F
and 185°F).
This allows the instrument to be used as a precision thermo-
meter.
Follow the diagram.
Note: the process must be completed in order to store the
readjustment data.
2 s
μS/cm
@25ºC
μS/cm
ºC
EC
T
t
Enter the correction of
the temperature sensor at
25°C (77 °F).
To exit adjustment wi-
thout saving the data
Enter the correction of
the temperature sensor at
85°C (185 °F).
sensION+ EC5 HACH 0710.indd ANG:13sensION+ EC5 HACH 0710.indd ANG:13 7/9/10 12:51:027/9/10 12:51:02
14
6. Operation
6.5. Manual calibration
Important note: before turning on the instrument, check that the sensor is con-
nected.
Manual readjustment of conductivity (salinity or TDS) measured at any value within
the range.
When readjusting, the instrument behaves as though it were calibrated with a
single standard solution.
6.5.1. Preparation
1. Unscrew the sensor protector.
2. Rinse the sensor with plenty of deionized water.
3. Dip the sensor into the solution in which manual calibration is to be perfor-
med.
Note: Before beginning manual calibration, you must take a measurement in the
standard to be used in the calibration.
6.5.2. Calibration
Stir gently holding the sensor by the handle.
Follow the diagram.
(1) Hold down
and then press
.
Note: if the measuring instrument is not used for 5 minutes, it shuts off automa-
tically.
If errors occur during calibration, an error message will appear on screen (see
pg. 19).
Before beginning manual calibration, you must take a measurement in the stan-
dard to be used in the calibration.
Manual calibration may be performed in conductivity, salinity or TDS. Start the
process after selecting the correct units of measurement.
Important note: To exit calibration without saving the adjustment, press
.
@25ºC
μS/cm
ºC
EC
T
t
@25ºC
μS/cm
ºC
EC
T
t
@25ºC
μS/cm
EC
@25ºC
μS/cm
EC
(1)
sensION+ EC5 HACH 0710.indd ANG:14sensION+ EC5 HACH 0710.indd ANG:14 7/9/10 12:51:027/9/10 12:51:02
15
6. Operation
ENGLISH
6.6. Displaying calibration data
To view the data obtained after calibration on screen.
Note: during calibration, the instrument adjusts a constant for each standard.
Hence, a single constant will be displayed if one-point calibration was performed,
2 constants in the case of two-point calibration and 3 constants for three-point
calibration.
(1) Press the
keys simultaneously.
μS/cm
@25ºC
μS/cm
ºC
EC
T
t
Constant obtained from calibra-
tion with the 1st standard
Constant obtained from calibra-
tion with the 2nd standard
Calibration temperature
Calibration temperature
(1)
sensION+ EC5 HACH 0710.indd ANG:15sensION+ EC5 HACH 0710.indd ANG:15 7/9/10 12:51:027/9/10 12:51:02
16
6. Operation
6.7. Restoring factory settings
2 s
@25ºC
μS/cm
ºC
EC
T
t
6.9. Changing temperature units
The instrument is shipped ready to measure temperature and express the values in °C.
Follow the diagram to change the units to °F.
6.8. Disabling auto power off
if the measuring instrument is not used for 5 minutes, it shuts off automa-
tically. This automatic power off feature can be disabled.
2 s
etc.
Automatic power
off is disabled
Stable measurement
@25ºC
μS/cm
ºC
EC
T
t
@25ºC
μS/cm
EC
T
t
ºF
The current calibration can be erased and restored to the factory
settings.
(1)
(1) Press the
keys simultaneously.
sensION+ EC5 HACH 0710.indd ANG:16sensION+ EC5 HACH 0710.indd ANG:16 7/9/10 12:51:027/9/10 12:51:02
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118

Hach sensION+ EC5 ユーザーマニュアル

タイプ
ユーザーマニュアル
このマニュアルも適しています