16
2.3.2 Method 2: Tube Insertion
To enhance visibility, the Folding Tool can be removed once the prosthesis has been
pushed into the Loading Tube.
Note:
1. Remove the Folding Tool
Remove the Folding Tool by unlocking and disconnecting it from the Loading
Tube (Fig. 8).
2. Enter the TE-puncture
resistance, dilatation and/or lubrication can ease the insertion.
3. Insert the voice prosthesis
4. Release the voice prosthesis
and pull the Loading Tube straight out from the puncture. This will release the
Note:
rotate the voiceprosthesis into place.
5. Finalize the procedure
2.3.3 Method 3: Overshoot Insertion
1. Remove the Folding Tool (Optional)
After the prosthesis has been pushed into the Loading Tube, remove the Folding
Tool by unlocking and disconnecting it from the Loading Tube (Fig. 8).
2. Enter the TE-puncture
resistance, dilatation and/or lubrication can ease the insertion.
3. Insert the voice prosthesis
prosthesis is fully deployed in the esophagus (Fig.11).
is fully deployed in the esophagus (Fig. 12).
4. Release the voice prosthesis
Pull the Loading Tube straight out from the puncture. The prosthesis remains in
5. Finalize the procedure
2.3.4 Method 4: Capsule Insertion
Intended use
Description of the device
WAR NINGS
• Always ensure that the voice prosthesis is folded correctly and that it is properly
attachment of the Capsule.
prosthesis after insertion.
PRECAUTIONS
•
may damage the voice prosthesis.
the voice prosthesis.
procedure the Loading Tube is not used.
1. Insert the prosthesis into Provox Capsule