7
2-4. Short press the “Hz/DUTY SELECT” key to convert DCA or ACA,
the DCA range is already set by default after the meter is turned on.
Short press “RANGE” key to convert to manual range, you can
choose 400uA, 4000uA at the uA range, and 40mA, 400mA are
optional at the mA range, 4A, 10A are optional at the A range. The
auto range is already set by default after the meter is turned on.
Note:
(1) If the tested current range is unsure beforehand, please start from
the highest range then find the proper range according to the
reading. Do not measure the voltage in the current jack.
(2) If the reading displays “OL”,the value is beyond the present range.
Now you need to set the knob to the higher range.
(3) Be careful when measuring 10A range. Continuous measurement of
high current will heat the circuit, even affect the accuracy and
damage the meter.
(4) When measuring the large current above 10A, it’s necessary to wear
personal protective equipment (such as approved rubber gloves,
face masks, and flame-retardant clothing etc.) to prevent the injury
from electric shock and arc.
3. Resistance measurement
3-1. Set the knob to resistance range, then connecting the test leads to
the tested resistance.
3-2. Insert the black test lead to “COM” jack, the red one to
“ ”jack.
3-3. You can read the measurement results from the display.
3-4. Auto range method is already set by default after the meter turned
on, short press “RANGE” key to convert to the manual range, The
range 400Ω, 4KΩ, 40KΩ, 400KΩ, 4MΩ, 40MΩ are optional.
Note:
(1) If "OL" displays on LCD, it means over range, now you need to set
the range knob to the higher range. When measuring resistance is