Niigata seiki WDI-*** ユーザーマニュアル

タイプ
ユーザーマニュアル
10
20
30
80
70
40
90
60
50
0
5
0
25
20 10
15
0-30mm
JEWELED
SHOCK-PRO OF
WDI-3 058SP
0.01mm










 

PP
-(:(/('
6+2&.3522)
:',63
PP
Minimize
10
20
30
80
70
40
90
60
50
0
5
0
25
20 10
15
0-30mm
JEWELED
SHOCK-PR OOF
WDI-3 058SP
0.01mm
In this manual, indicates RISK OF
PERSONAL INJURY OR PROPERTY
DAMAGE if not followed. The symbol
indicates something which is PROHIBITED,
and the symbol Indicates REQUIRED
step or necessary condition.
For marking acceptance
range for Pass/Fail testing.
Shows measurement on
Main Scale. rotate Bezel ( )
to turn.
Pointer indicates Main Scale
x100
Each division is Main Scale
x100
For holding and mounting
Gauge.
Limit Markers ········
Pointer (Needle) ···
Main Scale ············
Revolution Counter
··
Rev. Counter Scale
··
Stem ·····················
Shaft moves up and down
with measurement.
Probe which contacts
workpiece.
Rotate to turn Main Scale
( )
Standard back cover.
Alternate mounting point.
For mounting Gauge.
Accessory back cover for
mounting Gauge using Lug.
Spindle ·····················
Contact Point ···········
Bezel ························
Flat Rear Plate ·········
Lug ···························
Mounting Hole ·········
Rear Plate with Lug
··
LONG STROKE DIAL GAUGE
Model No.
INSTRUCTION MANUAL
Thank you for purchasing the Niigata Seiki Long Stroke Dial Gauge.
Used with a Magnetic Base or Indicator Stand, this gauge will show the difference in height
or position relative to a zero point set at a reference position.
Comparing to a master part during inspection.
Measuring machine tool positioning accuracy.
Measuring runout for rotary shafts.
Checking vise parallelism on milling machines.
Measuring flatness of surfaces and assemblies.
Confirming machine tool feed distance.
For safe and proper use of this product, please read this
instruction manual before use and follow the procedures
described. Please keep manual where it is accessible to
user for future reference.
Keep this manual with the instrument if transferred or
leased to a third party.
For inquiries about this product, please contact dealer or
Niigata Seiki at the address listed on the following page.
SAFETY NOTIFICATIONS
PROFESSIONAL
PARTS IDENTIFICATION & FUNCTION
APPLICATIONS
Contact Point
(Probe)
Spindle Stem
Pointer
(Needle)Bezel
Gauge Holder
Support
Arm
Stem
Model WDI-3058SP shown
Lug
Mounting Hole
(6.5mm
Rear Plate
With Lug
Flat Rear plate
Limit
Markers
Main Scale
Rev. Counter
Scale
Revolution
Counter
WDI-3058SP
WDI-5058SP
WDI-8058SP
WDI-
10078SP
Replacement Contact Point Part No.: DI-CP
WDI-3058SP
WDI-5058SP
WDI-8058SP
WDI-10078SP
7
9
11
15
3.0
3.5
4.0
4.5
1
30
50
80
100
0
-
100
Model No. Scale
g
g
g
g
210
270
290
530
Weight
(For Flat
Rear Plate)
Gradu-
ation
(mm)
Measurement
Range
(mm)
Hysteresis
Error
(m)
Measuring
Force
(N)
1
Revolution
(mm)
14
15
35
40
50
5
6
1/10 Rev
Indication
Error
(m)
Indication Error
over the Entire
Measuring
Range
(m)
SPECIFICATIONS
TROUBLESHOOTING
Remove any dust or dirt after use. Do not lubricate.
Wipe any contamination from Spindle sliding surface using a dry cloth, or cloth moistened with
alcohol.
To clean other surfaces, wipe with a soft dry cloth, or a cloth moistened with a mild cleaner.
Check for wear of Contact Point.
Measurement accuracy will be affected by worn Contact Point. Regularly check for wear and
replace Contact Point if worn.
Store in provided case in a cool, dark, and dry location.
During storage, make sure there is no force on the Spindle (such as pushed in, or lateral force.)
Keep away from moisture and direct sunlight, and secure from unauthorized personnel.
Origin position shifts during measurement.
Temperature changes during measurement can cause repeatability error.
Please try the following solutions:Use in location with constant temperature.
When taking measurements, periodically adjust zero point using a Master reference to correct for
temperature induced drift.
Measurement is not stable, or measurement accuracy is poor.
Contact Point may be worn.
Worn Contact Point will affect accuracy. Periodically check for wear, and replace if wear is affecting
measurement accuracy.
AFTER USE CARE, STORAGE
To Use the Rear Plate with Lug,
unscrew the standard Flat Rear
Plate screws, replace, and use
screws to secure.
[Accessories]
SAFETY PRECAUTIONS
PREPARATION - Mounting
Please Observe
Always follow the procedures specified below in order to prevent harm to yourself or others, and to
prevent damage to property.
HOW TO USE - Preparation
Dial gauge must be securely mounted such as on a comparator stand or magnetic base.
Please follow these guidelines.
Read the manual and follow all instructions.
Use of product other than as described in the
manual may cause accident.
Use only as Indicator Gauge.
Use for any purpose other than measuring may
damage or wear the instrument. Improper use
may also cause accident.
Use in an environment which meets the
following conditions:
Temperature within range of 0~40 C, humidity
30~70% (non-condensing.)
Location with minimal dust, oil, oil mist, and
protected from direct sunlight.
Location protected from use by children and
unauthorized people.
Use in location contrary to the above may cause
poor accuracy, damage to the product, or may
result in accident or injury.
Handle With Care.
Do not drop or subject to shock, do not place
under heavy objects. Damage may cause failure
or poor accuracy.
Content marked as follows indicates risk of injury or damage if not followed.
These symbols mark content that must be observed.
Caution
Indicates risk of personal injury or property damage if not followed.
WARNING
Denotes a prohibition – You MUST NOT doDenotes a requirement – You MUST do
Do not disassemble or modify.
It may damage Gauge and cause
poor accuracy
Do not shock Spindle.
Rapid motion, or lateral force may
damage Gauge and cause poor
accuracy.
Make sure Gauge Holder is rigid.
Holder must be sufficiently secure to prevent deflection from
the weight of the Gauge.
Holder must be rigid enough to hold Gauge and not to lift
from measurement force.
Holder support arm should be as short as possible to prevent
deflection.
Deflection or lifting will cause measurement error such as
origin position error and inaccuracies in measured reading.
Dial Gauge must only be attached by Stem or Rear Lug.
Mounting of gauge by other than Stem or Lug will cause
inaccuracy and product damage.
During installation, do not over-tighten the Stem.
Excessive force on the Stem may cause Spindle to bind.
No Rapid
Motion
No Lateral
Force
CALIBRATION
1. Confirm that the Contact Point and Rear Cover are tightly fastened.
If loose, tighten Contact Point and Rear Plate Screws.
2. Attach to the Gauge Holder using the Stem or Rear Lug.
Mounting Gauge by other than Stem or Lug will cause inaccuracy and product damage.
3. Confirm that Pointer and Revolution Counter movement is smooth.
Using fingertip, gently press on Contact Point to move the Spindle up and down. Motion of Pointers
should be smooth. If it is not smooth make sure Stem is not clamped too tight, and adjust. Also make
sure Pointer is stable at set position.
4. Make sure Spindle axis is perpendicular to
measured surface.
If Spindle (Contact Point) is not at a right angle to
surface, Gauge will not operate properly and
measurement will be inaccurate. Always keep the
Spindle axis perpendicular.When used to check parallelism of Milling Machine
vise, use a Magnetic Base to mount the Gauge
perpendicular to surface, and move it out of the way
during operation to prevent interference.
HOW TO USE - Comparison Measurements
1. Set Up Reference Part.
Carefully lift Spindle with fingertip, and, taking care
not to hit Spindle from the side, insert the Reference
Part or Master under Contact Point.
2. Set the Origin.
Adjust the gauge mount or rotate the Bezel to set the
Gauge to "0"
3.
Remove Reference, and begin measurements.
Remove Reference or Master, careful not to
shock Spindle. Insert part to be measured and
read the measurement off the Scale.
Setting the Limit Markers
Limit Markers can be moved to show acceptance
range for measurements.
HOW TO USE - Parallelism, Flatness, Runout, etc.
1. Position Contact Point on surface.
Carefully lifting Spindle with fingertip, and taking care not to hit Spindle from the side, position
the surface to be measured under the Contact Point.
2. Set the Origin.
Adjust the gauge mount, or rotate the Bezel to set the Gauge to "0"
3. Read the scale as the measured part is moved.
Slowly move the part while monitoring the Pointer and reading the measurement.
Right
angle
Mounting Example
Pointer
Limit Markers Ex.) Pass Range:
-
0.05 to +0.15mm
WDI-3058SP WDI-5058SP
WDI-8058SP
8
WDI-10078SP
10mm
Measured Object Measured Object
0.01
Niigata Seiki Co., Ltd.
5-3-14, Tsukanome, Sanjo, Niigata, Japan, 955-0055
Tel. : +81-256-33-5522 Fax. : +81-256-33-5518
MAIL intl.sales@niigataseiki.co.jp
URL http://www.niigataseiki.co.jp
In order to maintain instrument accuracy, it is recommended
that accuracy is confirmed through calibration on a periodic
basis.
Wear of measuring surfaces from repeated use may affect
accuracy and periodic accuracy checks should be performed.
We provide calibration services. Please contact
agent in country of purchase to make arrangements.
Repeat-
ability
(m)










 
PP
-(:(/('
6+2&.3522)
:',63
PP










 

PP
-(:(/('
6+2&.3522)
:',63
PP










 
PP
-(:(/('
6+2&.3522)
:',63
PP
ȭ
J
J
J
J
ȭ
ȭ
URL http://www.niigataseiki.co.jp
  • Page 1 1
  • Page 2 2

Niigata seiki WDI-*** ユーザーマニュアル

タイプ
ユーザーマニュアル

他の言語で